ENDOSCOPIC CLIP APPLIER
Device Facts
| Record ID | K962493 |
|---|---|
| Device Name | ENDOSCOPIC CLIP APPLIER |
| Applicant | Aesculap, Inc. |
| Product Code | OCW · Gastroenterology, Urology |
| Decision Date | Sep 11, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 876.1500 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Indications for Use
The devices presented in this submission are intended for use in endoscopic surgery for ligating and marking vessels and tubular structures.
Device Story
Reusable stainless steel endoscopic clip appliers (10mm diameter, 260-370mm length); used with disposable clip magazine holding 8 titanium clips. Instrument features handle with separate control levers for clip supply and closure. Design allows multiple clip applications without device withdrawal/reinsertion. Disassemblable for cleaning, disinfection, and sterilization. Operated by surgeons in endoscopic surgical settings. Output is mechanical ligation/marking of vessels/tubular structures; facilitates hemostasis or tissue marking.
Clinical Evidence
No clinical data provided. Bench testing only; manufacturing follows ISO and German DIN standards; manufacturer holds ISO 9001 certification.
Technological Characteristics
Reusable stainless steel construction; 10mm diameter; 260-370mm length. Mechanical actuation via handle control levers. Disposable titanium clip magazine. Disassemblable for cleaning/sterilization. Manufactured per ISO and German DIN standards.
Indications for Use
Indicated for use in endoscopic surgery for ligating and marking vessels and tubular structures in patients requiring such procedures.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
Predicate Devices
- Aesculap Endoscopic Clip Appliers (predecessor 510(k)s)
- Baxter Endoscopic Clip Appliers
- Ethicon Endoscopic Clip Appliers
- Horizon Surgical Endoscopic Clip Appliers
- U.S. Surgical Endoscopic Clip Appliers
Related Devices
- K081031 — ENDOSCOPIC CLIP APPLIERS WITH CLIP MAGAZINE · Aesculap, Inc. · May 5, 2008
- K080753 — AESCULAP ENDOSCOPIC CLIP SYTEM · Aesculap, Inc. · Apr 2, 2008
- K102081 — ENDOSCOPIC CLIP SYSTEM · Aesculap, Inc. · Aug 6, 2010
- K221495 — SOLID CLIP Single Use Clip Applier · Medscope Biotech Co., Ltd. · Jul 18, 2024
- K013695 — MICROLINE INC. REUSABLE LAPAROSCOPIC CLIP APPLIER WITH IMPLANTABLE TITANIUM CLIPS, MODEL M/L-10 · Microline Pentax, Inc. · Jan 8, 2002