Endomina EZFuse system

K252400 · Endo Tools Therapeutics S.A. · OCW · Feb 13, 2026 · Gastroenterology, Urology

Device Facts

Record IDK252400
Device NameEndomina EZFuse system
ApplicantEndo Tools Therapeutics S.A.
Product CodeOCW · Gastroenterology, Urology
Decision DateFeb 13, 2026
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2
AttributesTherapeutic

Intended Use

The endomina® EZFuse system is intended for endoscopic placement of suture(s) and approximation of soft tissue in the gastrointestinal tract.

Device Story

Endoscopic system for soft tissue approximation in GI tract; consists of triangulation platform and suture delivery instrument. Platform attaches to flexible endoscope; provides bendable therapeutic channel for delivery device; enables triangulation while keeping endoscope channel free. Delivery device pierces tissue with needle; releases anchors connected by suture to create interrupted stitches. Used by physicians in surgical endoscopy. System facilitates tissue approximation via endoscopic access; benefits patient by enabling minimally invasive repair/closure of GI tissues.

Clinical Evidence

No clinical data. Substantial equivalence demonstrated via bench testing, including needle welding, piercing, tightening/breaking forces, handle actuation, and implant release functionality. Sterility validated per EN ISO 11135:2014/A1:2019; biocompatibility per ISO 10993-1:2018; packaging per ASTM F1929-23 and ASTM F1886-16.

Technological Characteristics

Single-use, sterile, EO-sterilized (SAL 10^-6). Materials: thermoplastic, stainless steel, biocompatible 3D printing materials. Components: triangulation platform (bracket, tubing, handle) and suture delivery instrument (needle, anchors, non-absorbable braided polyester suture, PEEK). Needle tip: backbevel (3 grinding facettes). Dimensions: 1295mm total length. Shelf-life: 2 years.

Indications for Use

Indicated for endoscopic placement of suture(s) and approximation of soft tissue in the gastrointestinal tract in adult populations.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION February 13, 2026 Endo Tools Therapeutics S.A. Marine Rouyer Regulatory and Quality Director Rue Clément Ader, 10 Gosselies, 6041 Belgium Re: K252400 Trade/Device Name: Endomina EZFuse system Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope And Accessories Regulatory Class: Class II Product Code: OCW Dated: January 12, 2026 Received: January 12, 2026 Dear Marine Rouyer: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K252400 - Marine Rouyer Page 2 (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. {2} K252400 - Marine Rouyer Page 3 For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, ANTHONY LEE -S Anthony C. Lee Ph.D., M.B.A. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity, and Transplant Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. Submission Number (if known) K252400 Device Name Endomina EZFuse system Indications for Use (Describe) The endomina® EZFuse system is intended for endoscopic placement of suture(s) and approximation of soft tissue in the gastrointestinal tract. Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) # CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4} E T EndoTools Therapeutics K252400 - Page 1 of 5 # 510(k) SUMMARY ## Endo Tools Therapeutics' endomina® EZFuse system In accordance with 21 CFR 807.92 of Federal code of Regulations, the following 510(k) summary is submitted for endomina® EZFuse system. The endomina® EZFuse system is composed of endomina® platform and endomina® EZFuse suture. The endomina® platform is a short-term, single-use platform providing additional therapeutic channels for a flexible endoscope to be used in surgical endoscopy in the gastrointestinal tract consisting of a bracket that closes around the endoscope, a tubing system and a handle. The endomina® EZFuse suture is a device composed of two parts: one suture chain and one delivery device. The suture chain (implantable) is loaded into the delivery device (not implantable). endomina® EZFuse system Traditional 510(k) submission {5} E T EndoTools Therapeutics K252400 - Page 2 of 5 Submitter Name & Address Endo Tools Therapeutics S.A. Rue Clément Ader, 10 6041 Gosselies Belgium Contact Person Marine ROUYER Regulatory and Quality Director regulatory@endotools.be Phone: +32 71 91 94 09 Date: January 28, 2026 510(k) Number: K252400 Device identification Trade Device Name: endomina® EZFuse system FDA Product Code: OCW Classification Name: Endoscope and accessories Regulation Number: 21 CFR 876.1500 Regulatory Class: Class II Predicate Device: K211309 – endomina® system (Endo Tools Therapeutics S.A.) Device Description: The endomina® EZFuse system is intended for endoscopic placement of suture(s) and approximation of soft tissue in the gastrointestinal tract utilizing an endoscope. The system is comprised of a triangulation platform (endomina® platform) and an instrument for tissue piercing and approximation (endomina® EZFuse delivery system and suture). The endomina® EZFuse system is sterile packaged and designed for single use and is manufactured from various thermoplastic, stainless steel materials, biocompatible 3D printing materials and other medical grade materials. endomina® platform: The endomina® platform can be assembled on a regular flexible endoscope inside the gastrointestinal tract. It includes a bendable therapeutic channel meant for the endoscopic tool endomina® EZFuse suture. This channel can be deployed perpendicular to the axis of the endoscope, thereby ensuring the triangulation, leaving the channel of the endoscope free for other instruments. endomina® EZFuse system Traditional 510(k) submission {6} E T EndoTools Therapeutics K252400 - Page 3 of 5 endomina® EZFuse suture: Endomina® EZFuse suture is an instrument intended to be used with the endomina® platform and a flexible endoscope for tissue approximation in the gastrointestinal. It is inserted in the bendable arm of endomina®'s platform. It enables piercing and approximation of two, internal tissues with a needle which are then linked by releasing anchors connected with suture. These anchors create a stitch that can then be tightened, generating interrupted stitches in the gastrointestinal tract. Indications for Use: The endomina® EZFuse system is intended for endoscopic placement of suture(s) and approximation of soft tissue in the gastrointestinal tract. Technological Characteristics | Product Characteristics | | endomina® system | endomina® EZFuse system | | --- | --- | --- | --- | | Intended use | | | | | 510(k) Clearance | | K211309 | K252400 | | Area | | Gastrointestinal tract | | | Indications for Use | | The endomina® system, composed of a triangulation platform (endomina® platform) and an instrument for tissue piercing and approximation (TAPES), is intended for endoscopic placement of suture(s) and approximation of soft tissue in the gastrointestinal tract. The system is to be used on an adult population. | The endomina® EZFuse system is intended for endoscopic placement of suture(s) and approximation of soft tissue in the gastrointestinal tract. | | Intended use | | Endoscopic placement of suture(s) and approximation of soft tissue. | | | Technological characteristics | | | | | Single use | | Yes | | | Sterilization | | Terminally sterilized to SAL 10^{-6} using a validated EO method | | | Shelf-Life Claim | | 2 years | | | | Design | Lancet (3 grinding facettes) | Backbevel (3 grinding facettes) | endomina® EZFuse system Traditional 510(k) submission {7} E T EndoTools Therapeutics K252400 - Page 4 of 5 | Product Characteristics | | endomina® system | endomina® EZFuse system | | --- | --- | --- | --- | | Needle tip design | Dimensions | • ID=1,40 + 0,04/-0,01mm • OD=1,80 ± 0,015mm • Length=17 ± 0,3mm • Reduced portion=5 ± 0,2mm • OD Reduced portion=1,60 ± 0,02mm • Main needle grinding angle=15°±1° | | | Device length | | 1375mm | 1295mm | | Device length inserted in the platform and patient body | | 1056 mm | 1056 mm | | Anchors and suture | | Anchors are linked by two braided suture and a knot. | | | Dimensions | • Suture | 2 wires of 0.15 mm in parallel | | | | • Anchors | 1 anchor of 10 mm x 1.1mm OD 1 anchor of 10 mm x 1.1mm OD | 1 anchor of 10 mm x 1.2mm OD 1 anchor of 10 mm x 1.1mm OD | | Raw materials | • Suture • Anchors | • Non-absorbable braided polyester • PEEK (Optima LT2) | | endomina® EZFuse system Traditional 510(k) submission {8} E T EndoTools Therapeutics K252400 - Page 5 of 5 ## Non-clinical Performance Data: Appropriate product testing was performed on endomina® EZFuse system to evaluate conformance with standard requirements and substantial equivalence to the predicate device. Verification and validation for the proposed system was conducted in accordance with protocols. ### Performance testing: Bench tests were performed to demonstrate the substantial equivalence of the needle welding, needle heel, needle piercing, tightening fuse break forces, implant strength/breaking force, the handle actuation force during piercing, the functionality (implant release), the performance of the device and the tissue injury of the device. ### Sterility: Sterility claims (SAL 10⁻⁶) were confirmed by executing a sterilization validation in accordance with EN ISO 11135:2014 /A1:2019. ### Packaging and Shelf-life: Packaging integrity was confirmed by repeating testing in accordance with ASTM F1929-23 and ASTM F1886-16. Shelf-life claims were confirmed by functional testing on aged products. ## Biocompatibility: Biocompatibility testing and toxicological assessments were performed on the predicate and are applicable for all endomina® EZFuse system components in accordance with their risk category requirements, as defined in ISO 10993-1: 2018. Testing included cytotoxicity in accordance with ISO 10993-5: 2009, sensitization in accordance with ISO 10993-10: 2021, irritation in accordance with ISO 10993-10: 2021, acute systemic toxicity in accordance with ISO 10993-11: 2017, material mediated pyrogenicity in accordance with USP <151> and implantation in accordance with ISO 10993-6: 2016. ## Clinical Performance Data: Clinical testing was not required to demonstrate substantial equivalence ## Basis of Substantial Equivalence: The results of all studies confirmed substantial equivalence between the subject and predicate devices and that no new issues of safety nor efficacy were raised. Based on a comparison of indications for use and technological characteristics, the proposed system has demonstrated substantial equivalence to its predicate. endomina® EZFuse system Traditional 510(k) submission
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...