← Product Code [OCW](/submissions/GU/subpart-b%E2%80%94diagnostic-devices/OCW) · K181141

# OverStitch Endoscopic Suturing System and Accessories (K181141)

_Apollo Endosurgery · OCW · Jun 27, 2018 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/GU/subpart-b%E2%80%94diagnostic-devices/OCW/K181141

## Device Facts

- **Applicant:** Apollo Endosurgery
- **Product Code:** [OCW](/submissions/GU/subpart-b%E2%80%94diagnostic-devices/OCW.md)
- **Decision Date:** Jun 27, 2018
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Intended Use

The OverStitch Endoscopic Suturing System is intended for endoscopic placement of suture(s) and approximation of soft tissue.

## Device Story

The OverStitch Endoscopic Suturing System is a single-use, sterile endoscopic accessory designed for tissue approximation within the gastrointestinal tract. It is compatible with dual-channel or single-channel endoscopes. The system includes a Needle Driver Assembly, Anchor Exchange Device, Tissue Helix, Suture Cinch, and Suture-Anchor Assembly. In operation, the Needle Driver Assembly attaches to the endoscope; the physician uses the handle to actuate the needle body and exchange the Suture-Anchor Assembly to perform stitching. The Tissue Helix gathers target tissue, pulling it toward the needle. The Suture Cinch uses a PEEK component to secure the suture in situ. The device is used in clinical settings by physicians. It enables endoscopic suturing, providing a minimally invasive alternative to traditional surgical tissue approximation, potentially reducing patient trauma and recovery time.

## Clinical Evidence

No clinical data was required or provided. Substantial equivalence was demonstrated through bench testing, including needle passing reliability, pull-off strength, Suture Cinch deployment, Tissue Helix acquisition, and packaging integrity (ASTM F2096-11, ASTM D4169-16, ASTM F88/F88M-15). Biocompatibility was verified per ISO 10993-1.

## Technological Characteristics

Materials include stainless steel, ABS, silicone, nitinol, PTFE, and PEEK (VESTAKEEP). System is terminally sterilized via EO (SAL 10^-6). Components include Needle Driver Assembly (117-130cm length), Anchor Exchange, Suture Cinch, and Tissue Helix. Actuation is mechanical via handle manipulation. Compatible with dual-channel or single-channel endoscopes (8.8-9.8mm diameter).

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

## Predicate Devices

- OverStitch™ Endoscopic Suturing System ([K171886](/device/K171886.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services seal on the left and the FDA acronym and name on the right. The FDA acronym and name are in blue, with the acronym in a square and the name in a sans-serif font. The logo is simple and professional, and it is easily recognizable.

June 27, 2018

Apollo Endosurgery % Ms. Maritza Ward Manager of Regulatory Affairs 1120 S. Capital of Texas Hwy Ste. 300 Austin, Texas 78746

# Re: K181141

Trade/Device Name: OverStitch™ Endoscopic Suturing System and Accessories Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: Class II Product Code: OCW Dated: April 27, 2018 Received: May 1, 2018

Dear Ms. Ward:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

K181141

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# David Krause -S

for Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K181141

#### Device Name

Overstitch Endoscopic Suturing System

Indications for Use (Describe)

The OverStitch Endoscopic Suturing System is intended for endoscopic placement of suture(s) and approximation of soft tissue.

| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

# CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary

| Owner's Name & Address: | Apollo Endosurgery<br>1120 S. Capital of Texas Hwy.<br>Building 1, Suite 300<br>Austin, TX 78746                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
|-------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person:         | Maritza Ward<br>Manager, Regulatory Affairs<br>Phone: (512)-279-5114<br>Email: Maritza.Ward@apolloendo.com                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Date:                   | June 22, 2018                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Trade Name:             | OverStitch™ Endoscopic Suturing System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Common Name:            | Endoscopic Tissue Approximation Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Product Code:           | OCW                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Classification:         | Class II (21 CFR 876.1500)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Classification Name     | Endoscope and Accessories                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Predicate Devices:      | K171886 – OverStitch™ Endoscopic Suturing System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Device Description      | The OverStitch™ Endoscopic Suturing System and accessories<br>are intended for endoscopic placement of sutures and<br>approximation of soft tissue within the gastrointestinal tract<br>utilizing either a dual channel or single-channel endoscope. The<br>system is comprised of the Needle Driver Assembly and Anchor<br>Exchange Device (collectively referred to as ESS), and<br>accessories such as the Tissue Helix, Suture Cinch and Suture-<br>Anchor Assembly devices. All devices are sterile packaged and<br>designed for single use and are manufactured from various<br>thermoplastic, silicone, stainless steel and other medical grade<br>materials. |
|                         | OverStitch ESS<br>The OverStitch ESS is designed for compatibility with dual<br>channel endoscopes. The endcap of the Needle Driver Assembly<br>attaches to the distal end of the endoscope, and the handle<br>attaches to the proximal end. The handle of the Needle Driver<br>Assembly is squeezed to actuate the needle body and exchange<br>the Suture-Anchor Assembly with the Anchor Exchange to<br>perform stitching operations.                                                                                                                                                                                                                              |

{4}------------------------------------------------

## OverStitch Sx ESS

The OverStitch Sx ESS is designed for compatibility with single channel endoscopes. The ESS is mounted onto the endoscope using silicone straps distributed along the length of the catheter and endcap assembly. The external catheter sheath has two working channels through which the Anchor Exchange and OverStitch accessories can operate, independent of the endoscope channel. The handle of the Needle Driver Assembly is squeezed to actuate the needle body and exchange the proprietary Suture-Anchor Assembly with the Anchor Exchange to perform stitching operations.

## OverStitch Tissue Helix

The OverStitch Helix is an optional device that is designed to acquire tissue by rotating the device handle into targeted area until tissue is gathered into the exposed helix coil. The acquired tissue is then pulled into proximity of the needle body to complete the stitching operation.

#### OverStitch Suture-Anchor Assembly

The OverStitch Suture-Anchor Assembly is comprised of a 510(k) cleared suture product, manufactured from either polydioxanone or polypropylene materials attached to an implantable anchor manufactured from cobalt chrome and stainless steel. The Anchor component is intended to function with the OverStitch ESS devices to perform stitching operations and serves as an anchor to secure suture placement.

#### OverStitch Suture Cinch

The Overstitch Suture Cinch device is comprised of thermoplastic and stainless steel materials and includes an implantable PEEK Cinch component designed to secure the placement of suture as a final step of an OverStitch procedure. The device functions by squeezing the handle and deploying the PEEK components that press-fit onto the tail end of the suture to maintain suture position in situ.

Indications for Use: The OverStitch Endoscopic Suturing System is intended for endoscopic placement of suture(s) and approximation of soft tissue.

{5}------------------------------------------------

Technological Characteristics:

| NEEDLE DRIVER ASSEMBLY | Gen 2 Needle Driver Assembly                                                                                                                                                           | Sx Needle Driver Assembly                                                                     |                                                        |
|------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|--------------------------------------------------------|
| 510(k) Clearance       | K171886                                                                                                                                                                                |                                                                                               |                                                        |
| Single Use             | Yes                                                                                                                                                                                    |                                                                                               |                                                        |
| Shelf-Life Claim       | 3-years                                                                                                                                                                                | 1-year                                                                                        |                                                        |
| General Materials      | Stainless Steel ABS                                                                                                                                                                    | Stainless Steel ABS Silicone                                                                  |                                                        |
| Dimensions             | Length: 130cm<br>O.D.: 2mm (distal end)                                                                                                                                                | Length: 117cm<br>O.D.: 2mm (distal end)                                                       |                                                        |
| Suture Mechanism       | The handle of the Needle Driver Assembly is squeezed to actuate the needle body and exchange the proprietary suture assembly with the Anchor Exchange to perform stitching operations. |                                                                                               |                                                        |
| Compatible Scopes      | Olympus  GIF-2TH180 GIF-2T160                                                                                                                                                          | Single Channel Endoscopes that are 8.8 - 9.8 mm in diameter and up to 110cm in working length |                                                        |
| Endcap + Scope         | Position                                                                                                                                                                               | Seats on distal scope face                                                                    | Side mount to distal end in any rotational orientation |
|                        | Attachment                                                                                                                                                                             | Interference fit                                                                              | Two straps                                             |
| Sterilization          | Terminally sterilized to SAL 10-6 using a validated EO method                                                                                                                          |                                                                                               |                                                        |

| ANCHOR EXCHANGE                           | Gen 2 Anchor Exchange                                                                                                                                     |  |
|-------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| 510(k) Clearance                          | K171886                                                                                                                                                   |  |
| Single Use                                | Yes                                                                                                                                                       |  |
| Shelf-Life Claim                          | 3-years (when packaged as ESS-G02-160)<br>1-year (when packaged as ESS-G02-Sx1)                                                                           |  |
| General Materials                         | • Stainless Steel<br>• ABS                                                                                                                                |  |
| Dimensions                                | O.D.: 3.2mm (distal end)                                                                                                                                  |  |
| Compatible Scopes                         | Dual Channel<br>Olympus<br>• GIF-2TH180<br>• GIF-2T160<br>Single Channel<br>Endoscopes that are 8.8 - 9.8mm in diameter and up to 110cm in working length |  |
| Method of Actuation                       | Button on top of handle                                                                                                                                   |  |
| Integrated With Needle<br>Driver Assembly | No                                                                                                                                                        |  |
| Sterilization                             | Terminally sterilized to SAL 10-6 using a validated EO method                                                                                             |  |

{6}------------------------------------------------

| SUTURE CINCH        | Gen 2 Suture Cinch                                                                                                                                         |
|---------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Clearance    | K171886                                                                                                                                                    |
| Single Use          | Yes                                                                                                                                                        |
| Shelf-Life Claim    | 3-years                                                                                                                                                    |
| Materials           | Stainless Steel ABS VESTAKEEP PEEK                                                                                                                         |
| Cinch design        | PEEK Plug and collar components designed to press-fit together onto suture and hold construct in place.                                                    |
| Dimensions          | Suture Cinch Assembly:<br>Length: 152cm<br>O.D.: 2.4mm (distal end)<br><br>Deployed Cinch:<br>Length: 8mm<br>O.D. 2.4mm                                    |
| User Sequence       | Helix, Device Handle, Anchor Exchange, Cinch per DFU                                                                                                       |
| Compatible Scopes   | Dual Channel<br>Olympus  GIF-2TH180 GIF-2T160              Single Channel<br>Endoscopes that are 8.8 - 9.8mm in diameter and up to 110cm in working length |
| Method of Actuation | Squeezing of Handle                                                                                                                                        |
| Sterilization       | Terminally sterilized to SAL $10^{-6}$ using a validated EO method                                                                                         |

| TISSUE HELIX                    | Gen 2 Tissue Helix                                                               |            |
|---------------------------------|----------------------------------------------------------------------------------|------------|
| 510(k) Clearance                | K171886                                                                          |            |
| Single Use                      | Yes                                                                              |            |
| Shelf-Life Claim                | 3-years                                                                          |            |
| Materials                       | Stainless Steel                                                                  |            |
|                                 | Nitinol                                                                          |            |
|                                 | PTFE                                                                             |            |
|                                 | ABS                                                                              |            |
| Dimensions                      | Length: 165cm                                                                    |            |
|                                 | O.D.: 2.4mm (distal end)                                                         |            |
| Compatible Scopes               | Dual Channel                                                                     |            |
|                                 | Olympus                                                                          |            |
|                                 | •                                                                                | GIF-2TH180 |
|                                 | •                                                                                | GIF-2T160  |
|                                 | Single Channel                                                                   |            |
|                                 | Endoscopes that are 8.8 - 9.8mm in diameter and up to 110cm in<br>working length |            |
| Outer Sheath<br>Protecting Coil | Yes                                                                              |            |
| Sterilization                   | Terminally sterilized to SAL 10-6 using a validated EO method                    |            |

{7}------------------------------------------------

| SUTURE-ANCHOR ASSEMBLY | Gen 2 Anchor                                                                                                                                                                                                                                                                                                                |        |        |        |         |        |        |       |        |        |
|------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|--------|--------|---------|--------|--------|-------|--------|--------|
| 510(k) Clearance       | K171886                                                                                                                                                                                                                                                                                                                     |        |        |        |         |        |        |       |        |        |
| Single Use             | Yes                                                                                                                                                                                                                                                                                                                         |        |        |        |         |        |        |       |        |        |
| Shelf-Life Claim       | 3-years                                                                                                                                                                                                                                                                                                                     |        |        |        |         |        |        |       |        |        |
| Materials              | <ul><li>Stainless Steel 316L</li><li>Cobalt Chromium MP35</li></ul>                                                                                                                                                                                                                                                         |        |        |        |         |        |        |       |        |        |
| Dimensions             | <table><tr><th></th><th>Anchor</th><th>Suture</th></tr><tr><td>Length:</td><td>6.6 mm</td><td>185 cm</td></tr><tr><td>O.D.:</td><td>1.0 mm</td><td>2.0 mm</td></tr></table>                                                                                                                                                 |        | Anchor | Suture | Length: | 6.6 mm | 185 cm | O.D.: | 1.0 mm | 2.0 mm |
|                        | Anchor                                                                                                                                                                                                                                                                                                                      | Suture |        |        |         |        |        |       |        |        |
| Length:                | 6.6 mm                                                                                                                                                                                                                                                                                                                      | 185 cm |        |        |         |        |        |       |        |        |
| O.D.:                  | 1.0 mm                                                                                                                                                                                                                                                                                                                      | 2.0 mm |        |        |         |        |        |       |        |        |
| Suture Anchor Design   | Suture-Anchor has a recessed feature on one end designed to engage with the Anchor Exchange, and a latch feature on the opposite end designed to engage with the Needle Body Assembly. The suture is swaged onto the center of the needle body to allow for the needle body to double as an anchor for the suture construct |        |        |        |         |        |        |       |        |        |
| Sterilization          | Terminally sterilized to SAL 10-6 using a validated EO method                                                                                                                                                                                                                                                               |        |        |        |         |        |        |       |        |        |

Non-Clinical Performance Data: Appropriate product testing was performed on all subject devices evaluate conformance to product specifications and to equivalence to the predicate designs. Verification and validation for the proposed system of devices was conducted in accordance with protocol. All devices were evaluated to their individual functional and reliability requirements, as well as system compatibility.

#### Performance Testing

Bench testing included needle passing reliability, needle pull off strength, Suture Cinch deployment and pull-off strength, Tissue Helix acquisition reliability and bond strength.

#### Packaging Integrity

Packaging Integrity was confirmed by repeating testing in accordance with ASTM F2096-11, ASTM DI69-16 and ASTM F88/F88M-15.

The results of all studies confirmed equivalency between the subject and predicate devices, and that no new issues of safety or efficacy were raised.

- Biocompatibility: Biocompatibility testing and toxicological assessments were performed on subject OverStitch devices in accordance with their risk category requirements, as defined in ISO 10933-1. Testing included cytotoxicity and material mediated pyrogenicity.
Clinical Performance Data: Clinical testing was not required to demonstrate substantial equivalence.

- Basis of Substantial Equivalence: Based on a comparison of indications for use and technological characteristics, the proposed devices have demonstrated substantial equivalence to the predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/GU/subpart-b%E2%80%94diagnostic-devices/OCW/K181141](https://fda.innolitics.com/submissions/GU/subpart-b%E2%80%94diagnostic-devices/OCW/K181141)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
