# VISERA CYSTO-NEPHRO VIDEOSCOPE OLYMPUS CYF TYPE V2, VISERA CYSTO-NEPHRO VIDEOSCOPE OLYMPUS CYF TYPE VA2, VISERA CYSTO-NE (K133538)

_Olympus Corporation of the Americas · NWB · Aug 7, 2014 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/GU/subpart-b%E2%80%94diagnostic-devices/NWB/K133538

## Device Facts

- **Applicant:** Olympus Corporation of the Americas
- **Product Code:** [NWB](/submissions/GU/subpart-b%E2%80%94diagnostic-devices/NWB.md)
- **Decision Date:** Aug 7, 2014
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

VISERA CYSTO-NEPHRO VIDEOSCOPE OLYMPUS CYF TYPE V2, CYF TYPE VA2, CYF TYPE V2R : These instruments have been designed to be used with an Olympus video system center, light source, documentation equipment, monitor, EndoTherapy accessories, and other ancillary equipment for endoscopic diagnosis and treatment within the bladder, urethra, ureter, and kidney. EVIS EXERA II VIDEO SYSTEM CENTER OLYMPUS CV-180 : This video system center has been designed to be used with Olympus camera heads, endoscopes, light source, monitors, endo-therapy accessories and other ancillary equipment for endoscopic diagnosis, treatment and video observation. EVIS EXERA II XENON LIGHT SOURCE OLYMPUS CLV-180 : This light source has been designed to be used with Olympus endoscopes, video system center, and other ancillary equipment for endoscopic diagnosis, treatment and video observation.

## Device Story

EVIS EXERA II 180 SYSTEM is an endoscopic video imaging platform for urological procedures. System components include CYF series cysto-nephro videoscopes, CV-180 video system center, and CLV-180 xenon light source. Videoscopes capture internal anatomical images of bladder, urethra, ureter, and kidney; video system center processes signals for display on monitors; light source provides illumination. Used in clinical settings by physicians for diagnosis, treatment, and video observation. System integrates with ancillary equipment and EndoTherapy accessories. Output allows real-time visualization to guide clinical decision-making and therapeutic interventions, benefiting patients through minimally invasive diagnostic and treatment capabilities.

## Clinical Evidence

Bench testing only. Risk analysis and design verification tests were conducted based on ISO 14971:2007 to confirm that modifications do not affect safety or effectiveness.

## Technological Characteristics

System comprises flexible cysto-nephro videoscopes, video processor (CV-180), and xenon light source (CLV-180). Utilizes standard endoscopic imaging technology. Materials changed for specific components; design meets applicable safety standards. Connectivity via standard video/ancillary interfaces. No new technological features compared to predicate.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

## Predicate Devices

- EVIS EXERA II 180 SYSTEM ([K062049](/device/K062049.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features a stylized image of three faces in profile, stacked on top of each other. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the image.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

August 7, 2014

OLYMPUS MEDICAL SYSTEMS CORP. % Sheri L. Musgnung Manager, Regulatory Affairs Olympus Corporation of the Americas 3500 Corporate Parkway, PO Box 610 Center Valley, PA 18034-0610

Re: K133538

> Trade/Device Name: EVIS EXERA II 180 SYSTEM Regulation Number: 21 CFR§ 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: NWB, FAJ Dated: June 30, 2014 Received: July 1, 2014

Dear Sheri L. Musgnung,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

# Herbert P. Lerner -S

for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### Indications for Use

510(k) Number (if known) K133538

Device Name EVIS EXERA II 180 SYSTEM

#### Indications for Use (Describe)

VISERA CYSTO-NEPHRO VIDEOSCOPE OLYMPUS CYF TYPE V2, CYF TYPE VA2, CYF TYPE V2R : These instruments have been designed to be used with an Olympus video system center, light source, documentation equipment, monitor, EndoTherapy accessories, and other ancillary equipment for endoscopic diagnosis and treatment within the bladder, urethra, ureter, and kidney.

#### EVIS EXERA II VIDEO SYSTEM CENTER OLYMPUS CV-180 :

This video system center has been designed to be used with Olympus camera heads, endoscopes, light source, monitors, endo-therapy accessories and other ancillary equipment for endoscopic diagnosis, treatment and video observation.

#### EVIS EXERA II XENON LIGHT SOURCE OLYMPUS CLV-180 :

This light source has been designed to be used with Olympus endoscopes, video system center, and other ancillary equipment for endoscopic diagnosis, treatment and video observation.

Type of Use (Select one or both, as applicable)

> Prescription Use (Part 21 CFR 801 Subpart D)

- Over-The-Counter Use (21 CFR 801 Subpart C)

#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

#### FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Olympus. The word "OLYMPUS" is written in large, bold, blue letters. Below the word is a thin, yellow line. Below the line, the words "Your Vision, Our Future" are written in a smaller font.

# 510(k) Summary

## 1. General Information

. Applicant: OLYMPUS MEDICAL SYSTEMS CORP. 2951 Ishikawa-cho, Hachioji-shi, Tokyo, 192-8507, Japan Establishment Registration No.: 8010047

- Sheri L. Musgnung . Official Correspondent: Manager, Regulatory Affairs Olympus Corporation of the Americas 3500 Corporate Parkway PO Box 610 Center Valley, PA 18034-0610, USA Phone: 484-896-3147 FAX: 484-896-7128
- Prepared Date: November 15, 2013

### 2. Device Identification

| ■ | Device Trade Name:    | EVIS EXERA II 180 SYSTEM                                                                            |
|---|-----------------------|-----------------------------------------------------------------------------------------------------|
| ■ | Common Name:          | ENDOSCOPIC VIDEO IMAGING SYSTEM                                                                     |
| ■ | Regulation Number:    | 876.1500                                                                                            |
| ■ | Regulation Name:      | Endoscope and Accessories                                                                           |
| ■ | Regulatory Class:     | II                                                                                                  |
| ■ | Classification Panel: | Gastroenterology and urology                                                                        |
| ■ | ■ Product Code:       | NWB (endoscope, accessories, narrow band spectrum) FAJ (cystoscope and accessories, flexible/rigid) |

### 3. Predicate Devices

EVIS EXERA II 180 SYSTEM (K062049)

> Section6 510(k) Summary Page 2 of 4

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Olympus. The word "OLYMPUS" is written in large, bold, dark blue letters. Underneath the word is a thin, horizontal, yellow line. Below the line, the words "Your Vision, Our Future" are written in a smaller font.

# 4. Device Description

The EVIS EXERA II 180 SYSTEM consists of Olympus camera heads, endoscopes, video system center, light source, monitors, EndoTherapy accessories and other ancillary equipment. This system with the compatible cystscopes is intended for endoscopic diagnosis, treatment and video observation of the bladder, urethra, ureter, and kidney.

The primary components of the subject system are:

- VISERA CYSTO-NEPHRO VIDEOSCOPE OLYMPUS CYF TYPE V2, ◆
- CYF TYPE VA2, CYF TYPE V2R (CYF-V2, CYF-VA2, CYF-V2R) EVIS EXERA II VIDEO SYSTEM CENTER OLYMPUS CV-180
- EVIS EXERA II XENON LIGHT SOURCE OLYMPUS CLV-180

### 5. Indications for Use

### VISERA CYSTO-NEPHRO VIDEOSCOPE OLYMPUS CYF TYPE V2, CYF TYPE VA2, CYF TYPE V2R :

These instruments have been designed to be used with an Olympus video system center, light source, documentation equipment, monitor, EndoTherapy accessories, and other ancillary equipment for endoscopic diagnosis and treatment within the bladder, urethra. ureter, and kidney.

### EVIS EXERA II VIDEO SYSTEM CENTER OLYMPUS CV-180 :

This video system center has been designed to be used with Olympus camera heads, endoscopes, light source, monitors, endo-therapy accessories and other ancillary equipment for endoscopic diagnosis. treatment and video observation.

### EVIS EXERA II XENON LIGHT SOURCE OLYMPUS CLV-180 :

This light source has been designed to be used with Olympus endoscopes, video system center, and other ancillary equipment for endoscopic diagnosis, treatment and video observation.

# 6. Technological Characteristics

The purpose of this notification is to add additional urology-related marketing claims to this system, add an additional Cysto-nephro videoscope, add additional compatible items, and add a change to a type of material. There are no new technological features incorporated in this system. The proposed video system center and xenon light source and the subject endoscopes have identical technological features to the predicate device. The subject system has been designed to meet the applicable safety standards.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Olympus. The word "OLYMPUS" is written in large, bold, blue letters. Below the word is a thin, yellow line. Underneath the line, the words "Your Vision, Our Future" are written in a smaller, lighter font.

## 7. Summary of Non-clinical Testing

Risk analysis was carried out in accordance with established in-house acceptance criteria based on ISO 14971:2007. The design verifications tests and their acceptance criteria were identified and performed as a result of this risk analysis assessment.

### 8. Conclusion

When compared to the predicate devices, the EVIS EXERA II 180 SYSTEM does not incorporate any significant changes in intended use, methods of operation, materials, or design that could affect the safety or effectiveness. Based on the design control activities, Olympus believes the EVIS EXERA II 180 SYSTEM and the predicate device selected are substantially equivalent and do not change fundamental scientific technology and intended use of the market-cleared device.

---

**Source:** [https://fda.innolitics.com/submissions/GU/subpart-b%E2%80%94diagnostic-devices/NWB/K133538](https://fda.innolitics.com/submissions/GU/subpart-b%E2%80%94diagnostic-devices/NWB/K133538)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).