ARTEMIS LIGHT ENGINE

K141164 · Quest Medical Imaging · NTN · Mar 16, 2015 · Gastroenterology, Urology

Device Facts

Record IDK141164
Device NameARTEMIS LIGHT ENGINE
ApplicantQuest Medical Imaging
Product CodeNTN · Gastroenterology, Urology
Decision DateMar 16, 2015
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2

Indications for Use

The Artemis Light Engine is a primary LED light source when used in conjunction with laparoscopes or surgical cameras to illuminated surgical sites that allow observation or manipulation of body cavities, hollow organs, and canals.

Device Story

Artemis Light Engine is an LED light source for clinical and laboratory use; provides illumination for surgical sites during endoscopic procedures. Device connects to laparoscopes or surgical cameras. Intensity is controlled via PC or directly through a USB interface (RS232-via-USB). Output power is calibrated for linear response over intensity and lifetime. Used in clinical settings by physicians/surgeons to visualize body cavities, hollow organs, and canals. Provides consistent illumination to assist in surgical observation and manipulation; benefits patient by enabling clear visualization during minimally invasive procedures.

Clinical Evidence

Bench testing only. Performance testing verified light output and linear response. Software verification testing was successfully completed. Device complies with IEC 60601-1-2 (EMC) and ISO 14971 (Risk Management) standards.

Technological Characteristics

LED light source; intensity controlled via PC or USB (RS232-via-USB) interface. Calibrated for linear output response. Complies with IEC 60601-1-2 and ISO 14971 standards. Connectivity via USB/serial port.

Indications for Use

Indicated for use as a primary LED light source in conjunction with laparoscopes or surgical cameras to illuminate surgical sites for observation or manipulation of body cavities, hollow organs, and canals in patients undergoing endoscopic procedures.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized design of three human profiles facing right, layered on top of each other. The profiles are connected by a curved line that forms the shape of a bird. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the logo. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 March 16. 2015 Ouest Medical Imaging Mr. Martin Heuvelmans Quality Assurance Manager Industrieweg 41 1775 PW Middenmeer Netherlands Re: K141164 Trade/Device Name: Artemis Light Engine Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: Class II Product Code: NTN Dated: February 13, 2015 Received: February 13, 2015 Dear Mr. Heuvelmans: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical {1}------------------------------------------------ device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, # Jennifer R. Stevenson -S For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K141164 Device Name Artemis Light Engine Indications for Use (Describe) The Artemis Light Engine is a primary LED light source when used in conjunction with laparoscopes or surgical cameras to illuminated surgical sites that allow observation or manipulation of body cavities, hollow organs, and canals. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------------------------------------------------------------------------------------------------|--| | <div style="display:flex; align-items:center;"><span style="margin-right:5px;">☑</span>Prescription Use (Part 21 CFR 801 Subpart D)</div> | | | <div style="display:flex; align-items:center;"><span style="margin-right:5px;">☐</span>Over-The-Counter Use (21 CFR 801 Subpart C)</div> | | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the logo for Quest Medical Imaging. The logo features a stylized letter "Q" in gray and blue, with the word "Quest" in bold black letters next to it. To the right of "Quest" is a vertical blue line, followed by the words "medical imaging" in a smaller, italicized font. 510(k) Summary (21 CFR 807.92(c)) # Administrative information | Manufacturer Name: | Quest Medical Imaging<br>Industrieweg 41<br>1775 PW Middenmeer<br>Netherlands | |-------------------------|------------------------------------------------------------------------------------------------------------------------------------| | Contact Person: | Martin Heuvelmans, QA Manager<br>Tel: +31 227 604046<br>Fax: +31 227 604053<br>E-mail: martin.heuvelmans@quest-<br>innovations.com | | Official Correspondent: | Quest Medical Imaging<br>Martin Heuvelmans, QA Manager<br>E-mail: martin.heuvelmans@quest-<br>innovations.com | ### FDA Establishment Registration Number: 3010590422 | Date prepared: | March 3, 2014 | |----------------------|----------------------| | Name of Device: | | | Trade Name: | Artemis Light Engine | | Common Name: | LED light source | | Classification Name: | 21 CFR 876.1500 | Identification of Predicate Device(s): Product Code: | 510(k)<br>Number | Device | Manufacturer | |------------------|-------------------------|------------------------------| | K130819 | CLLV1, LED Light Source | Olympus Winter & Ibe<br>GmbH | | K123956 | Power LED 175 | Karl Storz GmbH & Co. KG | NTN ### Intended Use Statement The Artemis Light Engine is a primary LED light source when used in conjunction with laparoscopes or surgical cameras to illuminated surgical sites that allow observation or manipulation of body cavities, hollow organs, and canals. ### Device description The Light Engine is a family of light engines that are designed for laboratory and clinical use in applications. The Light Engine provides a light source of which the {4}------------------------------------------------ intensity can be controlled via a PC. The output power of the Light Engine is calibrated as to provide a linear response over the full range of intensity and life-time. The Light Engine can be controlled by an external software device, or directly with the serial port provided by the USB interface (RS232-via-USB). ## Performance Testing The performance testing to verify light output of the Artemis Light Engine was done as per the Bench Testing protocol attached. Design Test and Verification demonstrates that the device functions as it is intended and its performance does not raise any issues of safety and effectiveness. Software verification testing was successfully performed and completed. The manufacturer complies with the following voluntary standards: - IEC 60601-1-2 Edition 3: 2007-03 Medical Electrical Equipment Part1-2: ● General Requirements For Basic Safety And Essential Performance -Collateral Standard: Electromagnetic Compatibility - Requirements And Tests. - . ISO 14971 Second edition 2007-03-01, Medical Devices - Application of Risk Management to Medical Devices. # Predicate Device Comparison The Artemis Light Engine is substantially equivalent to the above listed 510(k) cleared devices on the market. Wording differences between indications for use of the Artemis Light Engine and those of the above-listed predicate devices do not constitute a new intended use. The characteristics and technology used are comparable with the predicate devices. ### Conclusions The Artemis Light Engine has similar intended use and technical features as the predicate devices listed above. Therefore, the Artemis Light Engine is substantially equivalent to the predicate devices listed above.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...