The disposable PillCam® Express™ video capsule delivery device is indicated for the transendoscopic delivery of the PillCam® SB video capsule in patients age 8 and above who are either unable to ingest the PillCam capsule or are known to have slow gastric emptying time.
Device Story
PillCam Express is a single-use, latex-free delivery system for PillCam SB video capsules. System components include a catheter, syringe, and capsule holder. Physician passes catheter through standard endoscope accessory channel; snaps capsule holder onto distal end. Endoscope guides system to proximal duodenum. Capsule released pneumatically via air-filled syringe. Post-deployment, endoscope is withdrawn; capsule holder cut off with scissors; catheter retracted and discarded. Device facilitates capsule placement in patients unable to ingest or with slow gastric emptying, enabling subsequent diagnostic imaging.
Clinical Evidence
Clinical validation study included 25 patients (ages 22-89). Primary endpoint: successful placement of capsule into the duodenum. Results: 100% success rate in capsule placement; no serious adverse events reported.
Technological Characteristics
Single-use, disposable, latex-free delivery system. Components: catheter, syringe, capsule holder. Pneumatic actuation via air-filled syringe. Manual operation via standard endoscope accessory channel.
Indications for Use
Indicated for transendoscopic delivery of PillCam SB video capsule in patients age 8+ unable to ingest capsule or with slow gastric emptying.
Regulatory Classification
Identification
An ingestible telemetric gastrointestinal capsule imaging system is used for visualization of the small bowel mucosa as an adjunctive tool in the detection of abnormalities of the small bowel. The device captures images of the small bowel with a wireless camera contained in a capsule. This device includes an ingestible capsule (containing a light source, camera, transmitter, and battery), an antenna array, a receiving/recording unit, a data storage device, computer software to process the images, and accessories.
Special Controls
*Classification.* Class II (special controls). The special control is FDA's guidance, “Class II Special Controls Guidance Document: Ingestible Telemetric Gastrointestinal Capsule Imaging Systems; Final Guidance for Industry and FDA.”
{0}------------------------------------------------
K101200
Page 1 of 3
GlEN
IMAGING
Given Imaging Limited New Industrial Park PO Box 258, Yoqneam 20692 Israel Voice 972 4 909 7777 Fax 972 4 959 2466
## 510(k) Summary
SEP 30 2010
This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of the Safe Medical Devices Act of 1990 and 21 CFR 807.92.
| Submitter Name and Address: | Given Imaging Ltd.<br>Hermon Building<br>New Industrial Park<br>PO Box 258<br>Yoqneam 20692<br>Israel<br>Tel.: 011-972-4-9097730<br>Fax: 011-972-4-9938060 |
|------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Tim Thomas RAC<br>Vice President<br>Regulatory Affairs & Quality Assurance |
| | Email: tim.thomas@givenimaging.com. |
| Phone Number: | 770-662-0870 ext. 1006 |
| Fax Number: | 770-662-0510 |
| Establishment Registration Number: | 9710107 |
| Date Prepared: | April 27, 2010 |
| Device Trade Name(s): | PillCam® Express™ Video Capsule Delivery Device |
| Device Common Name: | Ingestible telemetric gastrointestinal capsule imaging system<br>& Endoscope and/or accessories |
| Classification: | Regulation No: 21 CFR 876.1300 & 876.1500<br>Class: II<br>Review panel - Gastroenterology/Urology<br>NEZ - System, Imaging, Gastrointestinal, Wireless, Capsule<br>and KOG - Endoscope and/or accessories |
| Predicate Device(s): | Capsule endoscopy delivery device (K040494) |
Given Imaging Ltd. New Industrial Park. PO Box 258, Yoqncam 20692, I-racl
{1}------------------------------------------------
K101200 Page 2 of 3
Image /page/1/Picture/1 description: The image shows the logo for GI-EN Imaging. The logo is in black and white and features the words "GI-EN" in large, bold letters. Below the letters is an image of a medical device. To the right of the image is the word "IMAGING" in smaller letters.
Given Imaging Limited New Industrial Park PO Box 258, Yogneam 20692 Israel Voice 972 4 909 7777 Fax 972 4 959 2466
General Device Description:
The PillCam® Express Capsule Endoscopy Delivery System is comprised of three parts:
- Catheter
- 트 Syringe
- Capsule holder
The catheter is passed through the accessory channel of a standard endoscope and the capsule holder is snapped on to the distal end of the device.
The endoscope is used to guide the Capsule Endoscopy Delivery System to the proximal duodenum. The capsule is then released pneumatically, using an air-filled syringe attached to the proximal end of the catheter. Once the capsule is deployed, the endoscope is withdrawn and the capsule holder is cut off at the distal part of the device with scissors. The catheter is then retracted and discarded.
The PillCam® Express Capsule Endoscopy Delivery System is a single-use, disposable and latex-free product.
Indications for Use: Intended Use: The disposable PillCam® Express™ video capsule delivery device is indicated for the transendoscopic delivery of the PillCam® SB video capsule in patients age 8 and above who are either unable to ingest the PillCam capsule or are known to have slow gastric emptying time The technology characteristics are similar as the predicate Technological device. Characteristics: Performance data demonstrating safety and effectiveness of Performance Data: the PillCam® Express™ video capsule delivery device was obtained from non-clinical bench testing and from a clinical validation study. The non-clinical bench testing included strength testing of the cup to the catheter, force to release the capsule, force to maintain the capsule in the cup, and strength of the catheter. The clinical validation study included twenty five patients in the age range of 22-89 years old. All capsules were successfully placed into the duodenum and no serious adverse events were reported in the clinical study. Based on the technological characteristics and clinical Conclusion: performance of the device, Given Imaging Ltd. believes that
Given Imaging Ltd. New Industrial Park. P() Box 258, Yoqncam 20692, Israel
{2}------------------------------------------------
K101200
Page 3 of 3
### Given Imaging Limited
New Industrial Park .PO Box 258, Yoqneam 20692 İsrael Voice 972 4 909 7777 Fax 972 4 959 2466
Given PillCam® Express™ video capsule delivery device and the predicate capsule delivery device selected are substantially equivalent and do not raise new issues of safety or effectiveness.
Image /page/2/Picture/5 description: The image shows the address of Given Imaging Ltd. The address is New Industrial Park, PO Box 258, Yoqneam 20692, Israel. The text is black and the background is white. The text is centered on the page.
Image /page/2/Picture/6 description: The image shows the logo for Glen Imaging. The logo is in black and white and features the words "Glen Imaging" in a stylized font. The word "Glen" is on the top line, and the word "Imaging" is on the bottom line. There is a graphic of a test tube between the two words.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like symbol on the right side. To the left of the symbol, the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the symbol.
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002
Mr. Tim Thomas VP Regulatory Affairs and Quality Assurance Given Imaging Limited Hermon Building, New Industrial Park P.O. Box 258 Yoqneam 20692 ISRAEL
SEP 3 0 2010
Re: K101200
Trade/Device Name: PillCam® Express™ Video Capsule Delivery Device Regulation Number: 21 CFR §876.1300 Regulation Name: Ingestible telemetric gastrointestinal capsule imaging system Regulatory Class: II Product Code: NEZ Dated: September 20, 2010 Received: September 21, 2010
Dear Mr. Thomas:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related
{4}------------------------------------------------
Page 2
adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
Hubert Lemire MD
Herbert P. Lerner, M.D., Director (Acting) Division of Reproductive. Gastro-Renal and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
#### INDICATIONS FOR USE
510(k) Number (if known):
Device Name: PillCam® Express™ Video Capsule Delivery Device
SEP 3 0 2010
#### Indications for Use:
The disposable PillCam® Express™ video capsule delivery device is indicated for the transendoscopic delivery of the PillCam® SB video capsule in patients age 8 and above who are either unable to ingest the PillCam capsule or are known to have slow gastric emptying time.
| Prescription Use | |
|-----------------------------|--|
| (Part 21 CFR 801 Subpart D) | |
AND/OR
Over-The-Counter Use _ (21 CFR 801 Subpart C)
-------------------------------------------------------------------------------------------
(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Reproductive, Gastro-Renal, and
Urological Devices
510(k) Number K10/200
Given Imaging Ltd. Traditional 510(k) Submission PillCam® Express™ Video Capsule Delivery Device
April 27, 2010
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.