K022362 · Given Imaging , Ltd. · NEZ · Aug 9, 2002 · Gastroenterology, Urology
Device Facts
Record ID
K022362
Device Name
GIVEN DIAGNOSTICS SYSTEM
Applicant
Given Imaging , Ltd.
Product Code
NEZ · Gastroenterology, Urology
Decision Date
Aug 9, 2002
Decision
SESE
Submission Type
Special
Regulation
21 CFR 876.1300
Device Class
Class 2
Intended Use
The Given® Diagnostic System is intended for visualization of the small bowel mucosa. It may be used as an adjunctive tool in the detection of abnormalities of the small bowel
Device Story
System consists of M2A™ disposable wireless capsule, data recorder set, and RAPID® workstation. Capsule travels via peristalsis through digestive tract; transmits digital images to external receiving antennas worn on patient belt. Data recorder stores images; downloaded to workstation post-test. Physician reviews images on workstation to identify small bowel abnormalities. Adjunctive tool for clinical decision-making; aids in visualization of mucosa. Benefits include non-invasive small bowel assessment.
Clinical Evidence
No clinical data provided; substantial equivalence based on verification activities and compliance with FDA special controls guidance.
Technological Characteristics
System includes ingestible wireless capsule, external antenna array, data recorder, and workstation. Capsule utilizes wireless transmission of digital image data. Operates via peristaltic movement. Complies with Class II Special Controls Guidance Document for Ingestible Telemetric Gastrointestinal Capsule Imaging Systems.
Indications for Use
Indicated for visualization of small bowel mucosa as an adjunctive tool in the detection of small bowel abnormalities.
Regulatory Classification
Identification
An ingestible telemetric gastrointestinal capsule imaging system is used for visualization of the small bowel mucosa as an adjunctive tool in the detection of abnormalities of the small bowel. The device captures images of the small bowel with a wireless camera contained in a capsule. This device includes an ingestible capsule (containing a light source, camera, transmitter, and battery), an antenna array, a receiving/recording unit, a data storage device, computer software to process the images, and accessories.
Special Controls
*Classification.* Class II (special controls). The special control is FDA's guidance, “Class II Special Controls Guidance Document: Ingestible Telemetric Gastrointestinal Capsule Imaging Systems; Final Guidance for Industry and FDA.”
K040248 — MODIFICATION TO GIVEN DIAGNOSTIC SYSTEM · Given Imaging , Ltd. · Feb 17, 2004
K020341 — GIVEN DIAGNOSTIC SYSTEM · Given Imaging , Ltd. · May 2, 2002
K031033 — GIVEN DIAGNOSTIC SYSTEM · Given Imaging , Ltd. · Jul 1, 2003
K032405 — GIVEN DIAGNOSTIC SYSTEM · Given Imaging , Ltd. · Oct 29, 2003
K022980 — MODIFICATION TO GIVEN DIAGNOSTIC SYSTEM · Given Imaging , Ltd. · Feb 27, 2003
Submission Summary (Full Text)
{0}------------------------------------------------
# 510(K) SUMMARY Given® Diagnostic System 510(k) Number K 022362
### Applicant's Name:
Given Imaging Ltd. 2 Carmel St. P.O. Box 258 New Industrial Zone Yokneam 20692, Israel Tel.: 011-972-4-9097730 Fax: 011-972-4-9592466
## Contact Person:
Shoshana Friedman, RAC V.P. Regulatory & Medical Affairs Tel: 011-972-4- 9097784 Fax: 011-972-4-9592466 Email: shosh@givenimaging.com
# Trade Name:
Given® Diagnostic System
# Classification Name:
Ingestible Telemetric Gastrointestinal Capsule Imaging System
# Classification:
FDA has classified Ingestible Telemetric Gastrointestinal Capsule Imaging System as class II devices (product code 78NZE) and they are reviewed by the Gastroenterology Panel.
{1}------------------------------------------------
#### Predicate Device:
- · Given® Diagnostic Imaging System (K010312)
- · Given® Diagnostic System (K020341)
### Performance Standards and Special Controls:
The Given® Diagnostic System complies with the requirements presented in "Class II Special Controls Guidance Document; Ingestible Telemetric Gastrointestinal Capsule Imaging System; Final Guidance for Industry and FDA" issued on November 28,2001
### Intended Use:
The Given® Diagnostic System is intended for visualization of the small bowel mucosa. It may be used as an adjunctive tool in the detection of abnormalities of the small bowel
### Device Description:
The Given® Diagnostic System is comprised of three subsystems: M2A™ Capsule, Data Recorder Set, and RAPID® Workstation.
The M2A® Capsule is a wireless, disposable capsule designed to glide smoothly through the digestive system by the peristaltic activity of the intestinal muscles and is excreted naturally. During its passage, the capsule transmits digital data that is captured by receiving antennas attached to the patient's body. The images are stored in the DataRecorder that is connected to the receiving antennas and is worn on a belt around the waist of the patient. When the test is over, the antennas and recorder are removed from the patient's body. The images from the recorder are downloaded to the RAPID® Workstation for processing and viewing by the physician.
### Substantial Equivalence:
The proposed modified Given® Diagnostic System is substantially equivalent to the Given® Diagnostic System cleared under K010312 and K020341. The verification activities demonstrated that the proposed modifications do not raise any new safety and/or effectiveness issue.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image is a black and white circular seal. The seal contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. In the center of the seal is a stylized image of an eagle with its wings spread, facing to the left. The eagle's body is formed by three curved lines.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
AUG 0 9 2002
Shoshana (Shosh) Friedman, RAC V.P. Regulatory & Medical Affairs Given Imaging Limited New Industrial Park PO Box 258 Yogneam 20692 ISRAEL
#### Re: K022362
Trade/Device Name: Given Diagnostic System Regulation Number: 21 CFR 876.1300 Regulation Name: Ingestible telemetric gastrointestinal capsule imaging system Regulatory Class: II Product Code: 78 NEZ
Dated: July 18, 2002 Received: July 22, 2002
Dear Ms. Friedman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 8xx.1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
#### INDICATIONS FOR USE STATEMENT
0223402 510(k) Number (if known):_
Device Name:
Given® Diagnostic System
## Indications for Use:
The Given® Diagnostic System is intended for visualization of the small bowel mucosa. It may be used as an adjunctive tool in the detection of abnormalities of the small bowel
(PLEASE DO NOT WRITE BELOW THIS LINE -CONTINUE ON ANOTHER PAGE IF NEEDED)
510(k) Number _
L Prescription Use (Per 21 CFR 801.109)
OR
Over the Counter Use__________________________________________________________________________________________________________________________________________________________
Nancy C Brandon
Division Sign Off
(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number 6-6 K022362
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.