← Product Code [NAY](/submissions/GU/subpart-b%E2%80%94diagnostic-devices/NAY) · K200049

# Senhance Surgical System (K200049)

_Transenterix, Inc. · NAY · Mar 9, 2020 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/GU/subpart-b%E2%80%94diagnostic-devices/NAY/K200049

## Device Facts

- **Applicant:** Transenterix, Inc.
- **Product Code:** [NAY](/submissions/GU/subpart-b%E2%80%94diagnostic-devices/NAY.md)
- **Decision Date:** Mar 9, 2020
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

The Senhance Surgical System is intended to assist in the accurate control of laparoscopic instruments for visualization and endoscopic manipulation of tissue including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, mobilization, and retraction. The Senhance Surgical System is intended for use in laparoscopic gynecological surgery, colorectal surgery, cholecystectomy, and inquinal hernia repair. The system is indicated for adult use. It is intended for use by trained physicians in an operating room environment in accordance with the instructions for use.

## Device Story

Senhance Surgical System is a robotically-assisted tele-operation platform for laparoscopic surgery. System components include a surgeon cockpit with remote manipulators and video monitor, manipulator arms for instrument/endoscope control, and a communication hub (Node). The Smart Node (Intelligent Surgical Unit) adds enhanced image processing and three new camera control methods. Surgeon operates from the cockpit, receiving visual feedback via monitor and optional tactile force feedback. System translates user hand/eye movements into robotic instrument/endoscope manipulation within the insufflated body cavity. Used in OR by trained physicians to perform tissue manipulation (grasping, cutting, suturing, etc.). Benefits include precise instrument control and enhanced visualization/camera navigation during minimally invasive procedures.

## Clinical Evidence

No human clinical trials. Evidence includes bench testing for compatibility, reliability, and functionality; electrical safety (IEC 60601-1, 60601-1-2, 60601-2-18); software verification/validation; and a single-center, un-blinded, observational simulated-use design validation in a porcine model. Usability testing confirmed users could perform all critical tasks without use errors leading to harm.

## Technological Characteristics

Robotically-assisted tele-operation system. Components: surgeon cockpit, manipulator arms, communication node, and laparoscopic instruments. Features: force feedback, eye tracking, and Smart Node (ISU) for image processing/camera control. Complies with IEC 60601-1, IEC 60601-1-2, and IEC 60601-2-18. Software classified as major level of concern.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

## Predicate Devices

- Senhance Surgical System ([K192877](/device/K192877.md))

## Reference Devices

- AutoLap System ([K152848](/device/K152848.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a seal with an abstract design. To the right of the seal, the text "FDA U.S. FOOD & DRUG ADMINISTRATION" is displayed in blue. The word "FDA" is in a larger, bolder font, and the words "U.S. FOOD & DRUG" are stacked above the word "ADMINISTRATION".

March 9, 2020

TransEnterix, Inc. Kaitlyn Alexander Regulatory Affairs Manager 635 Davis Drive, Suite 300 Morrisville, North Carolina 27650

Re: K200049

Trade/Device Name: Senhance Surgical System Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: Class II Product Code: NAY Dated: January 9, 2020 Received: January 9, 2020

Dear Kaitlyn Alexander:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

### 510(k) Number (if known)

#### K200049

Device Name

TransEnterix® Senhance® Surgical System

#### Indications for Use (Describe)

The Senhance® Surgical System is intended to assist in the accurate control of laparoscopic instruments for visualization and endoscopic manipulation of tissue including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, mobilization. The Senhance Surgical System is intended for use in laparoscopic gynecological surgery, colorectal surgery, cholecystectomy, and inguinal hernia repair. The system is indicated for adult use. It is intended for use by trained physicians in an operating room environment in accordance with the instructions for use.

Type of Use (Select one or both, as applicable)

| <div><span style="font-family: Arial;">☑</span> Prescription Use (Part 21 CER 801 Subpart D)</div> | <div>□ Over-The-Counter Use (21 CER 801 Subpart C)</div> |
|----------------------------------------------------------------------------------------------------|----------------------------------------------------------|
|----------------------------------------------------------------------------------------------------|----------------------------------------------------------|

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

K200049

## 510(K) SUMMARY

[In accordance with 21CFR 807.92]

| 1. Submitter                          |                                                                                            |
|---------------------------------------|--------------------------------------------------------------------------------------------|
| 510(k) Sponsor:                       | TransEnterix, Inc.                                                                         |
| Address:                              | 635 Davis Drive, Suite 300<br>Morrisville, NC 27560                                        |
| Contact Person:                       | Kaitlyn Alexander<br>Regulatory Affairs Manager                                            |
| Contact Information:                  | Email: kalexander@transenterix.com<br>Phone: 919-765-8400 x8505<br>Facsimile: 919.765.8459 |
| Date Summary Prepared:                | 01/09/2020                                                                                 |
| 2. Device                             |                                                                                            |
| Proprietary (Trade) Name:             | Senhance® Surgical System                                                                  |
| Common Name:                          | System, Surgical, Computer Controlled Instrument                                           |
| Classification:                       | Class II                                                                                   |
| Classification Advisory<br>Committee: | General and Plastic Surgery                                                                |
| Regulation Number:                    | 21 CFR 876.1500, Endoscope and Accessories                                                 |
| Product Codes:                        | NAY (System, Surgical, Computer Controlled Instrument)                                     |
| 3. Predicate Device:                  |                                                                                            |
| Predicate Device                      | Senhance® Surgical System (K192877)                                                        |
| Reference Device                      | AutoLap System (K152848)                                                                   |

## 4. Device Description:

The purpose of this submission is to seek clearance for an alternate Node component called the Smart Node (to be marketed as the Intelligent Surgical Unit (ISU)), which introduces enhanced image processing features and augments the endoscope movement capabilities of the

{4}------------------------------------------------

TransEnterix® Senhance® Surgical System. The Smart Node adds three new methods of camera control for the surgeon operating at the Senhance Cockpit.

# 5. Intended Use/ Indications for Use:

The Senhance Surgical System is intended to assist in the accurate control of laparoscopic instruments for visualization and endoscopic manipulation of tissue including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, mobilization, and retraction. The Senhance Surgical System is intended for use in laparoscopic gynecological surgery, colorectal surgery, cholecystectomy, and inquinal hernia repair. The system is indicated for adult use. It is intended for use by trained physicians in an operating room environment in accordance with the instructions for use.

# Comparison with Predicate Device Intended Use/ Indications for Use:

The Senhance Surgical System with Smart Node has the same intended use/ indications for use as previously cleared for the predicate Senhance Surgical System (K192877).

# 6. Summary of Technological Characteristics:

The subject device has the same basic technological characteristics as the predicate Senhance Surgical System (K192877). In sum, the subject and predicate devices involve roboticallyassisted tele-operation as the primary technological principle. It is based on the accurate translation of user input to a robotically assisted output. It involves the use of endoscopic instrumentation for manipulation of tissue and vessels in the insufflated body cavity. The device consists of: a surgeon console (cockpit), which provides remote manipulators or handles to allow the surgeon to maneuver the surgical instruments and a video monitor to display the endoscopic signal; manipulator arms, which hold and maneuver the instruments and endoscope, based on inputs from the surgeon; node, which is the system communication hub, connecting the cockpit and manipulator arms: and instruments, which manipulate the tissue of interest. The Senhance instruments are similar in design and materials to traditional laparoscopic instrumentation.

In addition, force feedback provides an optional tactile sensory input to the surgeon control handles to give a sense of tissue elasticity. An eye tracking feature provides the surgeon an optional method to control the endoscope from the cockpit, other than using their hands. The Senhance Smart Node allows for three new optional methods of camera control, in addition to the optional eye tracking method.

Performance testing demonstrated that the technological differences between the Senhance Surgical System with Smart Node and the predicate device did not raise any different questions of safety or effectiveness.

# 7. Performance Data:

Performance testing of the Senhance system with Smart Node was conducted to support the substantial equivalence to the predicate device. The following performance tests were conducted:

{5}------------------------------------------------

Bench Testing: Bench testing evaluated the performance of the Senhance Smart Node as well as the overall use of the Senhance Surgical System with Smart Node. The following tests confirmed that the Senhance system and the Smart Node perform as intended after tests of compatibility, reliability, functionality, safety, and efficacy.

Electrical Safety and Compatibility: The Senhance Surgical System with Smart Node comply with the current versions of IEC 60601-1 (Basic safety and essential performance), IEC 60601-1-2 (Electromagnetic disturbances), and IEC 60601-2-18 (Endoscopic equipment interactions).

Software Verification and Validation Testing: Software verification and validation were conducted on the Senhance system software modifications to support the subject Smart Node. Documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff "Guidance for the Content of Premarket Submissions for Software Controlled in Medical Devices" and "Guidance for the Content of Premarket Submissions for Management of Cybersecurity in Medical Devices". The software was considered as a "major" level of concern.

Pre-Clinical Design Validation: Design validation of the Senhance system with Smart Node was conducted to ensure that the device performs as intended according to defined user needs and intended use in a simulated use environment. A single-center, un-blinded, observational, simulated use design validation evaluation of the Senhance system was conducted with users who represented the intended primary user population. The design validation was conducted in a live porcine model, which most closely represents the human anatomy for the given procedures. All user level requirements were assessed and found to be met.

Usability Testing: Modifications to the existing Senhance system Usability Engineering file were made based on the new features and functionality of the Smart Node. A confirmatory summative study was performed with final instructions and training materials. In a simulated use environment, the users were able to independently perform all critical tasks without use errors that would lead to harm.

## 8. Conclusions

The Senhance Surgical System with Smart Node is as safe and effective as the predicate Senhance Surgical System (K192877). The subject device has the same intended uses/ indications for use and similar technological characteristics and principles of operation as its predicate device. The minor technological differences between the subject device and predicate device raise no different questions of safety or effectiveness. The performance testing supported the safety and functionality of the device and demonstrate that the device is substantially equivalent to predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/GU/subpart-b%E2%80%94diagnostic-devices/NAY/K200049](https://fda.innolitics.com/submissions/GU/subpart-b%E2%80%94diagnostic-devices/NAY/K200049)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
