EndoWrist Vessel Sealer, 8 mm Harmonic ACE Curved Shears, da Vinci Single-Site Instruments and Accessories
Applicant
Intuitive Surgical, Inc.
Product Code
NAY · Gastroenterology, Urology
Decision Date
Apr 21, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.1500
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The EndoWrist Vessel Sealer is a bipolar electrosurgical instrument for use with a compatible da Vinci Surgical System and the ERBE VIO dV electrosurgical generator. It is intended for grasping and blunt dissection of tissue and for bipolar coagulation and mechanical transection of vessels up to 7 mm in diameter and tissue bundles that fit in the jaws of the instrument. The EndoWrist Vessel Sealer has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures, and should not be used for these procedures. The 8 mm Harmonic ACE Curved Shears is intended for soft tissue incisions when bleeding control and minimal thermal injury are desired. It is designed to be used in conjunction with a compatible da Vinci Surgical System and a compatible Ethicon Endo-Surgery Generator and Hand Piece. The Intuitive Surgical da Vinci Single-Site Instruments and Accessories used with a compatible da Vinci Surgical System are indicated for use by trained physicians in an operating room environment for endoscopic manipulation of tissue, grasping, cutting, blunt and sharp dissection, approximation, clip-ligation, electrocautery, suction/irrigation and suturing during single incision laparoscopic cholecystectomy, benign hysterectomy and salpingo-oophorectomy with the da Vinci Single-Site Instruments and Accessories, including graspers, dissectors, needle drivers, scissors, suction irrigators, monopolar cautery, bipolar cautery, clip appliers, 5 mm curved cannulae, 5 mm and 10 mm straight cannulae, 8 mm endoscope cannula, flexible and rigid blunt obturators, cannula seal, and the Single-Site Port.
Device Story
EndoWrist Vessel Sealer, 8 mm Harmonic ACE Curved Shears, and da Vinci Single-Site Instruments are robotic-assisted surgical tools; used with da Vinci Surgical Systems in OR by trained physicians. EndoWrist Vessel Sealer: bipolar electrosurgical instrument; connects to ERBE VIO dV generator; seals/transects vessels up to 7 mm. Harmonic ACE: ultrasonic instrument; connects to Ethicon generator; performs soft tissue incisions. Single-Site Instruments: include graspers, dissectors, needle drivers, scissors, cautery, and port accessories for single-incision procedures. Surgeon controls movement/function via Surgeon Console foot pedals/interface. Devices facilitate precise tissue manipulation, dissection, and hemostasis; improve surgical outcomes through robotic assistance.
Clinical Evidence
No clinical data. Bench testing only; risk analysis performed to evaluate labeling modifications. No changes to design inputs were made.
Technological Characteristics
Sterile, single-use instruments. EndoWrist Vessel Sealer: bipolar electrosurgical, 8 mm, integrated cord. Harmonic ACE: ultrasonic energy delivery, clamp arm and blade. Single-Site: mechanical instruments (graspers, dissectors, etc.) and cannulae. Connectivity: compatible with da Vinci Surgical Systems and specific external generators (ERBE VIO dV or Ethicon).
Indications for Use
Indicated for patients undergoing single incision laparoscopic cholecystectomy, benign hysterectomy, and salpingo-oophorectomy. EndoWrist Vessel Sealer indicated for tissue grasping, blunt dissection, and bipolar coagulation/mechanical transection of vessels up to 7 mm. Harmonic ACE indicated for soft tissue incisions requiring bleeding control and minimal thermal injury. Contraindicated for tubal sterilization/coagulation.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
da Vinci Single-Site Instruments and Accessories (K152448)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 21, 2017
Intuitive Surgical, Inc. Ms. Cheryl Wu Regulatory Affairs Engineer 1266 Kifer Road Sunnyvale, California 94086
Re: K170865
Trade/Device Name: EndoWrist Vessel Sealer, 8 mm Harmonic ACE Curved Shears, da Vinci Single-Site Instruments and Accessories Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: Class II Product Code: NAY Dated: March 21, 2017 Received: March 23, 2017
Dear Ms. Wu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR
{1}------------------------------------------------
Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Jennifer R. Stevenson -S
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### INDICATIONS FOR USE STATEMENT 4.0
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510 (k) Number (if known)
к170865
Device Name EndoWrist® Vessel Sealer
Indications for Use (Describe)
The EndoWrist Vessel Sealer is a bipolar electrosurgical instrument for use with a compatible da Vinci Surgical System and the ERBE VIO dV electrosurgical generator. It is intended for grasping and blunt dissection of tissue and for bipolar coagulation and mechanical transection of vessels up to 7 mm in diameter and tissue bundles that fit in the jaws of the instrument. The EndoWrist Vessel Sealer has not been shown to be effective for tubal coagulation for sterilization procedures, and should not be used for these procedures.
| Type of Use (Select one or both, as applicable) | <span> <span style="padding-right:5px">☒</span>Prescription Use (Part 21 CFR 801 Subpart D) </span> <span> <span style="padding-right:5px">☐</span>Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
|-------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|-------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
[PRAStaff@fda.hhs.gov](mailto:PRAStaff@fda.hhs.gov)
*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."*
FORM FDA 3881 (8/14)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF
Image /page/2/Picture/15 description: The image shows the logo for Intuitive Surgical. The word "INTUITIVE" is in gray, block letters, with a yellow dot above the "I". Below that, the word "SURGICAL" is in smaller, gray letters with a trademark symbol.
{3}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510 (k) Number (if known) K170865
Device Name
8 mm Harmonic ACE® Curved Shears
Indications for Use (Describe)
The 8 mm Harmonic ACE Curved Shears is intended for soft tissue incisions when bleeding control and minimal thermal injury are desired. It is designed to be used in conjunction with a compatible da Vinci Surgical System and a compatible Ethicon Endo-Surgery Generator and Hand Piece.
| Type of Use (Select one or both, as applicable) | <span> ☒ Prescription Use (Part 21 CFR 801 Subpart D) </span> <span> □ Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
|-------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|-------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
[PRAStaff@fda.hhs.gov](mailto:PRAStaff@fda.hhs.gov)
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/14)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF
Image /page/3/Picture/12 description: The image shows the logo for Intuitive Surgical. The word "INTUITIVE" is in all caps and in a light gray color. Below that, the word "SURGICAL" is also in all caps and in a smaller font size and the same light gray color.
{4}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known)
к170865
Device Name
da Vinci® Single-Site® Instruments and Accessories
#### Indications for Use (Describe)
The Intuitive Surgical da Vinci Single-Site Instruments and Accessories used with a compatible da Vinci Surgical System are indicated for use by trained physicians in an operating room environment for endoscopic manipulation of tissue, grasping, cutting, blunt and sharp dissection, approximation, clip-ligation, electrocautery, suction'irrigation and suturing during single incision laparoscopic cholecystectomy, benign hysterectomy and salpingo-oophorectomy with the da Vinci Single-Site Instruments and Accessories, including graspers, dissectors, needle drivers, scissors, suction irrigators, monopolar cautery, bipolar cauters, clip appliers, 5 mm and 10 mm straight cannulae, 8 mm endoscope cannula, flexible and rigid blunt obturators, cannula seal, and the Single-Site Port.
Type of Use (Select one or both, as applicable)
| <span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> | <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
|-----------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------|
|-----------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------|
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
[PRAStaff@fda.hhs.gov](mailto:PRAStaff@fda.hhs.gov)
*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."*
FORM FDA 3881 (8/14)
Page 1 of 1
ing Services (301) 443-6740 EF
Image /page/4/Picture/16 description: The image shows the logo for Intuitive Surgical. The logo consists of the word "INTUITIVE" in a sans-serif font, with a small yellow dot above the "I". Below the word "INTUITIVE" is the word "SURGICAL" in a smaller, sans-serif font. The logo is simple and modern, and it is likely used to represent the company's brand.
{5}------------------------------------------------
# 510(k) Summary
| 510(k) Submitter: | Intuitive Surgical, Inc.<br>1266 Kifer Road<br>Sunnyvale, CA 94086 |
|---------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact: | Cheryl Wu<br>Regulatory Affairs Engineer<br>Phone: 408-523-6401<br>Fax: 408-523-1390<br>Email: cheryl.wu@intusurg.com |
| Date Summary Prepared: | March 21, 2016 |
| Trade Name: | EndoWrist® Vessel Sealer,<br>8 mm Harmonic ACE® Curved Shears,<br>da Vinci® Single-Site® Instruments and Accessories |
| Common Name: | Endoscope and accessories |
| Classification: | Class II<br>21 CFR 876.1500, Endoscope and Accessories |
| Product Codes: | NAY |
| Classification Advisory<br>Committee: | General and Plastic Surgery |
| Predicate Device: | K140189 – EndoWrist Vessel Sealer<br>K143132 – IS4000 8 mm Harmonic ACE Curved Shears<br>K152448 – da Vinci Single-Site Instruments and Accessories |
# Technological Characteristics:
The purpose of this submission is to change the Indications for Use for the EndoWrist Vessel Sealer, 8 mm Harmonic ACE Curved Shears, and da Vinci Single-Site Instruments and Accessories.
Specific references to compatible systems (i.e. "Xi", "Model IS4000") have been removed from the EndoWrist Vessel Sealer, 8 mm Harmonic ACE Curved Shears, and Single-Site Instruments and Accessories Indications for Use statements. The reference to "compatible da Vinci Surgical Systems" in the subject Indications for Use coupled with identification of specific systems compatibility in the Compatibility Information or Device Description section of the User Manual for the subject devices is essentially equivalent to the reference to model "IS4000" in the predicate Indications for Use statements.
{6}------------------------------------------------
The subject devices are identical to their predicate devices in terms of design, technology, and performance specifications. The Indications for Use and Device Description for the subject devices included in this submission are summarized below:
# EndoWrist Vessel Sealer
#### Intended Use/Indications for Use
The EndoWrist Vessel Sealer is a bipolar electrosurgical instrument for use with a compatible da Vinci Surgical System and the ERBE VIO dV electrosurgical generator. It is intended for grasping and blunt dissection of tissue and for bipolar coagulation and mechanical transection of vessels up to 7 mm in diameter and tissue bundles that fit in the jaws of the instrument. The EndoWrist Vessel Sealer has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures, and should not be used for these procedures.
# Device Description
The EndoWrist Vessel Sealer is a sterile, single-use (disposable), 8 mm instrument with an integrated cord that connects to the instrument housing and an Erbe VIO dV generator. The EndoWrist Vessel Sealer device consists of a distal wristed end effector and a proximal housing connected by a tubular shaft. The housing contains mechanisms to actuate the end effector when attached to a compatible da Vinci Surgical System. An integrated cord attached to the housing is connected to a receptacle in the IESU. An electrode sealing surface and a cutting blade within the jaws of the instrument enable sealing of vessels and cutting of sealed vessels and other tissues. The sealing and cutting functions are controlled using the compatible da Vinci Surgical System foot pedals.
# 8 mm Harmonic ACE Curved Shears
#### Intended Use/Indications for Use
The 8 mm Harmonic ACE Curved Shears is intended for soft tissue incisions when bleeding control and minimal thermal injury are desired. It is designed to be used in conjunction with a compatible da Vinci Surgical System and a compatible Ethicon Endo-Surgery Generator and Hand Piece.
#### Device Description
The da Vinci 8 mm Harmonic ACE Curved Shears is a single-use, sterile instrument used to deliver ultrasonic energy to enable transection and coagulation of tissue. The movement and function of the da Vinci 8 mm Harmonic ACE Curved Shears are controlled by the surgeon from the Surgeon Console of a compatible da Vinci Surgical System. This instrument consists of a housing, instrument shaft, and instrument jaws comprised of a clamp arm and blade.
Image /page/6/Picture/12 description: The image shows the logo for Intuitive Surgical. The word "INTUITIVE" is in all caps and in a light gray color. Below that, the word "SURGICAL" is in smaller, light gray letters with a registered trademark symbol to the right.
{7}------------------------------------------------
# da Vinci Single-Site Instruments and Accessories
#### Intended Use/Indications for Use
The Intuitive Surgical da Vinci Single-Site Instruments and Accessories used with a compatible da Vinci Surgical System are indicated for use by trained physicians in an operating room environment for endoscopic manipulation of tissue, grasping, cutting, blunt and sharp dissection, approximation, clip-ligation, electrocautery, suction/irrigation and suturing during single incision laparoscopic cholecystectomy, benign hysterectomy and salpingo-oophorectomy with the da Vinci Single-Site Instruments and Accessories, including graspers, dissectors, needle drivers, scissors, suction irrigators, monopolar cautery, bipolar cautery, clip appliers, 5 mm curved cannulae, 5 mm and 10 mm straight cannulae, 8 mm endoscope cannula, flexible and rigid blunt obturators, cannula seal, and the Single-Site Port.
#### Device Description
The da Vinci Single-Site Instruments and Accessories are a set of devices developed by Intuitive Surgical to enable single incision laparoscopic cholecystectomy, benign hysterectomy and salpingo oophorectomy using the compatible da Vinci Surgical System.
#### Performance Data:
In accordance with the Design Control process, risk analysis was conducted to evaluate the impact of labeling modifications on the predicate devices. Design verification and validation testing were not required since design inputs were not changed.
#### Summary:
Based on the intended use, indications for use, technological characteristics, and performance data, the subject devices are substantially equivalent to currently marketed predicate devices.
Image /page/7/Picture/10 description: The image shows the logo for Intuitive Surgical. The word "INTUITIVE" is written in all caps in a sans-serif font. Below that, the word "SURGICAL" is written in a smaller font, also in all caps. There is a small yellow dot above the word "INTUITIVE".