SINGLE-SITE MEDIUM-LARGE CLIP APPLIER, SINGLE-SITE CADIERE GRASPER, SINGLE-SITE FUNDUS GRASPER, SINGLE-SITE CROCODILE

K120215 · Intuitive Surgical, Inc. · NAY · Apr 30, 2012 · Gastroenterology, Urology

Device Facts

Record IDK120215
Device NameSINGLE-SITE MEDIUM-LARGE CLIP APPLIER, SINGLE-SITE CADIERE GRASPER, SINGLE-SITE FUNDUS GRASPER, SINGLE-SITE CROCODILE
ApplicantIntuitive Surgical, Inc.
Product CodeNAY · Gastroenterology, Urology
Decision DateApr 30, 2012
DecisionSESU
Submission TypeSpecial
Regulation21 CFR 876.1500
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Intuitive Surgical da Vinci® Single-Site™ Instruments and Accessories used with the da Vinci Si Surgical System (IS3000) are indicated for use by trained physicians in an operating room environment for endoscopic manipulation of tissue, grasping, cutting, blunt and sharp dissection, approximation, clip-ligation and electrocautery during single incision laparoscopic cholecystectomy with the da Vinci Single Site Instruments and Accessories, including graspers, dissectors, scissors, suction irrigators, monopolar cautery, 5mm curved cannulae, 5mm and 10mm straight cannulae, flexible blunt obturators, and the 5mm Single-Site Port.

Device Story

Device set enables single-incision laparoscopic cholecystectomy using da Vinci Si Surgical System (IS3000). Components include non-wristed, semi-rigid shaft instruments (graspers, dissectors, scissors, suction irrigators, monopolar cautery), fixed-shape curved cannulae, accessory cannula for manual instruments, semi-rigid blunt obturator, and single fascial port. Operated by trained physicians in OR. Instruments provide mechanical tissue manipulation and electrocautery. Output is direct surgical action on patient tissue. System facilitates minimally invasive access through single incision. Benefits include reduced incision count for cholecystectomy.

Clinical Evidence

No clinical data. Evidence consists of bench testing (dimensional measurements, mechanical/functional verification, electrical safety) and simulated use in an animal model.

Technological Characteristics

Non-wristed, semi-rigid shaft instruments; fixed-shape curved cannulae; single fascial port. Materials transitioned from machined to molded components. Energy source: electrocautery. Connectivity: integrated with da Vinci Si Surgical System (IS3000).

Indications for Use

Indicated for trained physicians in an operating room for endoscopic tissue manipulation, grasping, cutting, dissection, approximation, clip-ligation, and electrocautery during single incision laparoscopic cholecystectomy.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ APR 3 0 2012 K120215 SURGICAL Taking surgery beyond the limits of the human hand™ # 510(K) SUMMARY (per 21 CFR 807.92) Submitter: Intuitive Surgical, Inc. 1266 Kifer Road Sunnyvale, CA 94086 Ph: (408) 523-2100 Fax: (408) 523-1390 Contact: Melissa S. Gonzalez Sr. Regulatory Affairs Specialist Date Summary Prepared: March 9, 2012 Device Name: | Trade Name: | Intuitive Surgical da Vinci® Single-Site™ Instruments and Accessories | |----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Common Name: | system, surgical, computer controlled instrument | | Classification Name: | Endoscope and Accessories (21 CFR 876.1500, Product Code NAY and GCJ) | | Predicate Device: | The Intuitive Surgical da Vinci® Single-Site™ Instruments and Accessories were originally cleared on December 8, 2011 under K112208, and are currently marketed by Intuitive Surgical, Inc. (Sunnyvale, CA). | | Device Description: | The da Vinci Single-Site Instruments and Accessories are a set of devices developed by Intuitive Surgical to enable single incision cholecystectomy using the IS3000 da Vinci Si Surgical System.<br>The da Vinci Single-Site Instruments and Accessories consist of non-wristed, semi-rigid shaft instruments, two fixed-shape curved cannulae, an accessory cannula for insertion of manual laparoscopic instruments, a semi-rigid blunt obturator, and a single fascial port (with insufflation tubing and stopcock) for the placement and insertion of multiple cannulae/instruments through a single incision.<br>The da Vinci Single-Site Instruments and Accessories include instruments to provide grasping, cautery, cutting, clip ligation and suction/irrigation functions. | K120215, Single-Site Instruments & Accessories Page 1 of 2 11 {1}------------------------------------------------ #### Taking surgery beyond the limits of the human hand™ N T II I T I V F · SURGICAL® # Indications For Use: # Technological Characteristics: ### Performance Data: Summarv: The Intuitive Surgical da Vinci® Single-Site™ Instruments and Accessories used with the da Vinci Si Surgical System (IS3000) are indicated for use by trained physicians in an operating room environment for endoscopic manipulation of tissue, grasping, cutting, blunt and sharp dissection, approximation, clip-ligation and electrocautery during single incision laparoscopic cholecvstectomy with the da Vinci Single Site Instruments and Accessories, including graspers, dissectors, scissors, suction irrigators, monopolar cautery, 5mm curved cannulae, 5mm and 10mm straight cannulae, flexible blunt obturators, and the 5mm Single-Site Port. The modified Intuitive Surgical da Vinci Single-Site Instruments and Accessories are equivalent to the predicate devices in terms of their intended use, indications for use and technological characteristics. These modifications include simplifying the back-end drive mechanism, minor design and dimensional changes to the distal grip and moving from machined to molded components. Performance testing (bench and animal) was conducted to demonstrate that the modified devices are substantially. equivalent to the predicate devices, and that the design outputs meet the design inputs. The results of the testing did not raise any new safety or effectiveness questions. The tests conducted consisted of dimensional measurements, mechanical and functional verification, electrical safety, and simulated use in an animal model. Based on the Intended Use, indications for use, technological characteristics and performance data, the modified Intuitive Surgical da Vinci Single-Site Instruments and Accessories are equivalent to the predicate devices in terms of safety and effectiveness. K120215, Single-Site Instruments & Accessories {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure with outstretched wings. The bird is positioned above the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA", which is arranged in a circular fashion around the bird. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 Intuitive Surgical, Inc. % Ms. Melissa Gonzalez Sr. Regulatory Affairs Specialist 1266 Kifer Road Sunnyvale, California 94086 APR 3 0 2012 Re: K120215 Trade/Device Name: Intuitive Surgical da Vinci® Single-Site™ Instruments and Accessories Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: Class II Product Code: NAY, GCJ Dated: April 10, 2012 Received: April 11, 2012 Dear Ms. Gonzalez: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act and the limitations described below. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. The Office of Device Evaluation has determined that there is a reasonable likelihood that this device will be used for an intended use not identified in the proposed labeling and that such use could cause harm. Therefore, in accordance with Section 513(i)(1)(E) of the Act, the following limitation must appear in the Precautions/Warnings/Contraindications section of the device's labeling: The safety and effectiveness of this device for use in the performance of general laparoscopic abdominal surgery procedures have not been established. This device is only intended to be used for single incision laparoscopic cholecystectomy with the da Vinci Single Site Instruments and the da Vinci Si Surgical System (IS3000). Please note that the above labeling limitations are required by Section 513(i)(1)(E) of the Act. Therefore, a new 510(k) is required before these limitations are modified in any way or removed from the device's labeling. {3}------------------------------------------------ ## Page 2 - Ms. Melissa Gonzalez The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and permits your device to proceed to the market. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification if the limitation statement described above is added to your labeline. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, Chiots fo Christy Foreman Director Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ Intuitive Surgical Inc. #### K120215 510(k) Number if known: Device Name: da Vinci® Single-Site™ Instruments and Accessories # INDICATIONS FOR USE: The Intuitive Surgical da Vinci® Single-Site™ Instruments and Accessories used with the da Vinci Si Surgical System (IS3000) are indicated for use by trained physicians in an operating room environment for endoscopic manipulation of tissue, grasping, cutting, blunt and sharp dissection, approximation, clip-ligation and electrocautery during single incision laparoscopic cholecystectomy with the da Vinci Single Site Instruments and Accessories, including graspers, dissectors, scissors, suction irrigators, monopolar cautery, 5mm curved cannulae, 5mm and 10mm straight cannulae, flexible blunt obturators, and the 5mm Single-Site Port. Prescription Use X (Per 21 CFR 801 Subpart D) C) AND/OR Over-the-Counter Use (Per 21 CFR 807 Subpart (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Neil R.P. Doyle for mkm (Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices 510(k) Number K120215
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...