← Product Code [NAY](/submissions/GU/subpart-b%E2%80%94diagnostic-devices/NAY) · K050005

# INTUITIVE SURGICAL MONOPOLAR CURVED SCISSORS, MODEL 400179; TIP COVER ACCESSORY, MODEL 400180 (K050005)

_Intuitive Surgical, Inc. · NAY · Jan 25, 2005 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/GU/subpart-b%E2%80%94diagnostic-devices/NAY/K050005

## Device Facts

- **Applicant:** Intuitive Surgical, Inc.
- **Product Code:** [NAY](/submissions/GU/subpart-b%E2%80%94diagnostic-devices/NAY.md)
- **Decision Date:** Jan 25, 2005
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

The Intuitive Surgical® Endoscopic Instrument Control System is intended to assist in the accurate control of Intuitive Surgical® Endoscopic Instruments including rigid endoscopes, blunt and sharp endoscopic dissectors, scissors, scalpels, ultrasonic shears, forceps/pickups, needle holders, endoscopic retractors, stabilizers, electrocautery and accessories for endoscopic manipulation of tissue, including grasping, dissection, approximation, ligation. and sharp cutting, blunt electrocautery, suturing and delivery and placement of microwave ablation probes and accessories during general laparoscopic surgical non-cardiovascular thoracoscopic surgical procedures. general procedures, and thoracoscopically assisted cardiotomy procedures. The system can also be employed, with adjunctive mediastinotomy to perform coronary anastomosis during cardiac revascularization. It is intended for use by trained physicians in an operating room environment in accordance with the representative specific procedures set forth in the Professional Instructions for Use.

## Device Story

The da Vinci Surgical System is a computer-controlled endoscopic instrument system. It consists of a Surgeon Console, Patient Side Cart, and Insite Vision System. The surgeon operates from the console using Master Tool Manipulators to control endoscopic instruments and a stereo endoscope. Commands are relayed via cables to Patient Side Manipulators on the cart, which hold instruments in a fixed position relative to the patient. The vision system fuses two independent images into a 3-D view for the surgeon. This modification adds monopolar electrosurgical capability to the curved scissor instrument. The system assists in tissue manipulation, including grasping, dissection, and suturing. It is used by trained physicians in an operating room to enhance surgical precision during minimally invasive procedures.

## Clinical Evidence

Bench testing only. No clinical data provided. Design analysis and bench testing were conducted to confirm that the modified instrument's functional characteristics are substantially equivalent to the predicate devices and that design outputs meet design input requirements.

## Technological Characteristics

Computer-controlled surgical instrument system. Features a Surgeon Console with Master Tool Manipulators and a Patient Side Cart with Patient Side Manipulators. Includes a stereo endoscopic vision system (Insite Vision System) providing 3-D visualization. The subject modification incorporates monopolar electrosurgical capability into a curved scissor instrument. Connectivity is via cable-based relay from console to patient-side cart. System is intended for use in an operating room environment.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

## Predicate Devices

- Intuitive Surgical® da Vinci® Surgical System and Endoscopic Instruments ([K990144](/device/K990144.md))
- Intuitive Surgical® da Vinci® Surgical System and Endoscopic Instruments ([K002489](/device/K002489.md))
- Intuitive Surgical® da Vinci® Surgical System and Endoscopic Instruments ([K011002](/device/K011002.md))
- Intuitive Surgical® da Vinci® Surgical System and Endoscopic Instruments ([K013416](/device/K013416.md))
- Intuitive Surgical® da Vinci® Surgical System and Endoscopic Instruments ([K021036](/device/K021036.md))
- Intuitive Surgical® da Vinci® Surgical System and Endoscopic Instruments ([K022574](/device/K022574.md))
- Intuitive Surgical® da Vinci® Surgical System and Endoscopic Instruments ([K040237](/device/K040237.md))
- Intuitive Surgical® da Vinci® Surgical System and Endoscopic Instruments ([K043153](/device/K043153.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K05 0005 1/3 007

## JAN 2 5 2005 Section II

### 510(k) SUMMARY

This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

### 510(k) Number:

| Date                  | December 30, 2004                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
|-----------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter             | Intuitive Surgical, Inc.<br>950 Kifer Road<br>Sunnyvale, CA 94086                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| ER Number             | 2955842                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Contact               | Mike Yramategui<br>Director, Regulatory Affairs<br>Telephone: (408) 523 - 2145<br>Fax: (408) 523 - 1390<br>e-mail: mike.yramategui@intusurg.com                                                                                                                                                                                                                                                                                                                                                                                                            |
| Subject<br>Device     | Name: Intuitive Surgical® da Vinci® Surgical System and Endoscopic<br>Instruments<br>Classification Name: System, Surgical, Computer Controlled Instrument<br>(21 CFR 876.1500)<br>Common Name: Endoscopic Instrument, Monopolar Curved Scissors                                                                                                                                                                                                                                                                                                           |
| Predicate<br>Devices  | Intuitive Surgical® da Vinci® Surgical System and Endoscopic<br>Instruments (legally marketed under K990144 / K002489 / K011002 /<br>K013416 / K021036 / K022574 / K040237 / K043153)                                                                                                                                                                                                                                                                                                                                                                      |
| Device<br>Description | This special 510(k) is being submitted for a modification to the<br>endoscopic instruments to incorporate monopolar electrosurgical<br>capability on an endoscopic curved scissor instrument. This modification<br>affects the instrument only and there are no changes in the design,<br>technology, materials, manufacturing, performance, specifications, and<br>method of use for the da Vinci® Surgical System.<br>The endoscopic instruments are used with the da Vinci® Surgical System<br>which consists of two integrated sub-systems as follows: |

{1}------------------------------------------------

K050005 2/3 008

#### Device Description (continued)

Intuitive Surgical® Endoscopic Instrument Control System: This subsystem is comprised of the Surgeon Console and Patient Side Cart. While seated at the Surgeon Console, the surgeon controls critical aspects of the procedure, including movement of the endoscopic instruments and endoscope, within the operative field. Endoscopic instrument and camera movements are controlled by the surgeon through use of the Master Tool Manipulators (MTM), two hand operated mechanisms residing within the Surgeon Console. The endoscopic instruments are held in a fixed position (with respect to the patient) by either two (or optionally three) unique arms known as Patient Side Manipulators (PSM), which are located on the Patient Side Cart (PSC). The endoscope is also held in a fixed position (with respect to the patient) by another arm, similar to the PSM, known as the Endoscope Camera Manipulator (ECM) and also located on the PSC. Commands from the Surgeon Console are relayed to the PSC, which is located immediately adjacent to the patient, via cables. Instrument and endoscope changes are performed by another provider positioned adjacent to the PSC.

Intuitive Surgical® Stereo View Endoscopic System: The endoscopic vision system used with the da Vinci® Surgical System, also known as Intuitive Surgical Insite™ Vision System, consists of a stereo endoscope, endoscopic camera, and various accessories, including a light source and light guides. The Insite™ Vision System provides two independent images that are relayed to the viewer located in the Surgeon Console, where they are fused to form a 3-D (or alternatively a 2-D image) image of the surgical field.

#### The Intuitive Surgical Endoscopic Instrument Control System is Intended Use intended to assist in the accurate control of Intuitive Surgical® Endoscopic Instruments including rigid endoscopes, blunt and sharp endoscopic dissectors, scissors, scalpels, ultrasonic shears, forceps/pickups, needle holders, endoscopic retractors, stabilizers, electrocautery and accessories for endoscopic manipulation of tissue, including grasping, dissection, approximation, ligation. and sharp cutting, blunt electrocautery, suturing and delivery and placement of microwave ablation probes and accessories during general laparoscopic surgical non-cardiovascular thoracoscopic surgical procedures. general procedures, and thoracoscopically assisted cardiotomy procedures. The system can also be employed, with adjunctive mediastinotomy to perform coronary anastomosis during cardiac revascularization. It is intended for use by trained physicians in an operating room environment in accordance with the representative specific procedures set forth in the Professional Instructions for Use.

{2}------------------------------------------------

Kosovoos 3/3 009

| Comparison to<br>Predicate<br>Device | The design, technology, materials, manufacturing methods, performance,<br>specifications and methods of use are essentially the same for the<br>Monopolar Curved Scissors as for other Intuitive Surgical® Endoscopic<br>Instruments with the primary difference being that monopolar<br>electrosurgical capability is incorporated into a curved scissor instrument.<br>There are no changes in the design, technology, materials,<br>manufacturing, performance, specifications, and method of use for the da<br>Vinci® Surgical System |
|--------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Technological<br>Characteristics     | The technological characteristics of the subject devices are essentially the<br>same as for the predicate devices.                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Performance<br>Data                  | Design analysis and comparison, as well as bench testing, has been<br>conducted to confirm that basic functional characteristics are<br>substantially equivalent to the predicate devices cited, and that design<br>output meets the design input requirements.                                                                                                                                                                                                                                                                           |
| Conclusion                           | Based upon the technical information, intended use, and performance<br>information provided in this pre-market notification, the da Vinci®<br>Surgical System and Monopolar Curved Scissors described herein has<br>been shown to be substantially equivalent to current legally marketed<br>predicate devices, and the results of the design control process confirm<br>that the design output meets the design input requirements.                                                                                                      |

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/12 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the circle is an abstract symbol that resembles an eagle or other bird-like figure.

Public Health Service

JAN 2 5 2005

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Mr. Michael Yramategui Director, Regulatory Affairs Intuitive Surgical, Inc. 950 Kifer Road Sunnyvale, California 94086

Re: K050005

Trade/Device Name: Intuitive Surgical® Endoscopic Instrument Control System Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessoreis Regulatory Class: II Product Code: NAY Dated: December 30, 2004 Received: January 3, 2005

Dear Mr. Yramategui :

We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your becalent of the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for asso stated in the encreations of the enactment date of the Medical Device Amendments, or to conniner of the to rias if its accordance with the provisions of the Federal Food, Drug, de nees that have been route approval of a premarket approval application (PMA). and Cosmetic Free (110) that the device, subject to the general controls provisions of the Act. The r ou may, areres, include requirements for annual reguirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set Of It Fart 877, its systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{4}------------------------------------------------

Page 2 -- Mr. Michael Yramategui

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Miriam C. Provost

Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

# Indications for Use

510(k) Number (if known):

Intuitive Surgical® Endoscopic Instrument Control System Device Name:

Indications For Use:

The Intuitive Surgical® Endoscopic Instrument Control System is intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments including rigid endoscopes, blunt and sharp endoscopic dissectors, scissors, scalpels, ultrasonic shears, forceps/pickups, needle holders, endoscopic retractors, stabilizers, electrocautery and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, and delivery and placement of microwave ablation probes and accessories during general laparoscopic surgical procedures, general non-cardiovascular thoracoscopic surgical procedures, and thoracoscopically assisted cardiotomy procedures. The system can also be employed, with adjunctive mediastinotomy to perform coronary anastomosis during cardiac revascularization. It is intended for use by trained physicians in an operating room environment in accordance with the representative specific procedures set forth in the Professional Instructions for Use.

Prescription Use × (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Miriam C. Provost

(Division Sign-Off) Division of General. Restorative. and Neurological Devices

510(k) Number_4650005

Page 1 of 1

---

**Source:** [https://fda.innolitics.com/submissions/GU/subpart-b%E2%80%94diagnostic-devices/NAY/K050005](https://fda.innolitics.com/submissions/GU/subpart-b%E2%80%94diagnostic-devices/NAY/K050005)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
