K990515 · International Medical Systems · KNW · May 13, 1999 · Gastroenterology, Urology
Device Facts
Record ID
K990515
Device Name
MATHIS VERTEBRAL AND BONE BIOPSY SYSTEM
Applicant
International Medical Systems
Product Code
KNW · Gastroenterology, Urology
Decision Date
May 13, 1999
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.1075
Device Class
Class 2
Indications for Use
The Mathis Vertebral and Bone Biopsy Systems are intended for use by physicians performing bone marrow biopsy procedures.
Device Story
Mathis Vertebral and Bone Biopsy System is a manual bone marrow biopsy needle. Device consists of a cannula with a sharpened distal end and a coaxial stylet with a sharpened distal end. Components are insert-molded to hubs; cannula hub features threadable lug tabs. System includes a removable gripable handle. Operated by physicians in clinical settings to penetrate bone for marrow biopsy. Principle of operation relies on point geometry of cannula and stylet to facilitate controlled bone penetration. Output is a bone marrow sample obtained via manual insertion and extraction.
Clinical Evidence
No clinical data.
Technological Characteristics
Manual bone marrow biopsy needle. Components: cannula and stylet with sharpened distal ends, insert-molded hubs, removable gripable handle. Cannula hub includes threadable lug tabs. No electronic components or software.
Indications for Use
Indicated for physicians performing bone marrow biopsy procedures.
Regulatory Classification
Identification
A gastroenterology-urology biopsy instrument is a device used to remove, by cutting or aspiration, a specimen of tissue for microscopic examination. This generic type of device includes the biopsy punch, gastrointestinal mechanical biopsy instrument, suction biopsy instrument, gastro-urology biopsy needle and needle set, and nonelectric biopsy forceps. This section does not apply to biopsy instruments that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Predicate Devices
Manan Bone Marrow Aspiration / Biopsy Needle (K890925)
Related Devices
K190177 — J-Type Marrow Biopsy Needle,T-Type Marrow Biopsy Needle,Tweezer Bone Marrow Biopsy Needle, · Ranfac Corporation · Jun 12, 2019
K172558 — iVAS 8g Bone Biopsy Kit, iVAS 11g Bone Biopsy Kit, iVAS 10g Bone Biopsy Kit · Stryker Corporation · Oct 5, 2017
K242322 — Bone Marrow Biopsy Needle · Zhejiang Curaway Medical Technology Co., Ltd. · Jan 24, 2025
K013692 — TRAPSYSTEM SET · H.S. Hospital Service S.R.L. · Feb 5, 2002
{0}------------------------------------------------
# IMS
## International Medical Systems
PO Box 4936 Annapolis, Maryland 21403
×990515
#### 510(k) Summary
Submitted by: Charles Simonelli 28 Spring Lane Farmington, CT 06032 Phone: 860-676-0770 Fax:860-676-0871
Date Submitted: February 2, 1999
#### Classification, Common or Usual Name, Device Name:
| Trade Name: | Mathis Vertebral and Bone Biopsy System |
|----------------------|-------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name: | Bone Marrow Biopsy Needle |
| Classification Name: | Needle, Biopsy, Cardiovascular<br>General and Plastic Surgury-878<br>Classification Number-79DWO<br>Regulation Number-878.4800<br>Class I |
### Predicate Device:
Manan Bone Marrow Aspiration / Biopsy Needle (K890925). The device is available in the following sizes:
8 ga x 4" 8 ga x 6" 11 ga x 4" 11 ga x 6" 13 ga x 4" 13 ga x 2-1/2"
#### Device Description:
The Mathis Vertebral and Bone Biopsy System is a standard bone marrow biopsy needle consisting of a cannula with a sharpened distal end and a stylet received coaxially in the cannula and having a sharpened distal end protruding from the distal end of the cannula. The cannula is insert molded to a hub with standard threadable lug tabs. The stylet is insert molded to a standard hub. When the stylet assembly is placed into the cannula assembly, the entire cannula/stylet unit can be threaded into a removable gripable handle.
#### Indications For Use:
The Mathis Vertebral and Bone Biopsy Systems are intended for use by physicians performing bone marrow biopsy procedures.
{1}------------------------------------------------
#### Technological Characteristics:
The key technological characteristic involved in bone marrow biopsy procedures is the effect of the point geometry of the cannula and protruding stylet on the units ability to penetrate the outer layer of the bone in a controllable manner. Stylets and cannula presently in the market have been designed with various sharpened distal ends to facilitate bone penetration. The Mathis Vertebral and Bone Biopsy System will not substantially differ its stylet and cannula point geometry from product currently available in the market.
#### Non-Clinical Data:
Part specification and prints for the Mathis Vertebral and Bone Biopsy System show it to be substantially equivalent to the Manan Bone Marrow Aspiration/Biopsy Needle.
#### Clinical Performance Data:
None available at this time.
#### Conclusions:
It is the conclusion of International Medical Systems that the Mathis Vertebral and Bone Biopsy System is substantially equivalent, relative to its key technical performance characteristics, to product already available in the market.
{2}------------------------------------------------
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAY 1 3 1999
Mr. Charles Simonelli Quality Assurance Manager International Medical Systems P.O. Box 4936 Annapolis, Maryland 21403
Re: K990515 Trade Name: Mathis Vertebral and Bone Biopsy System Regulatory Class: II Product Code: KNW Dated: February 9, 1999 Received: February 18, 1999
Dear Mr. Simonelli:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{3}------------------------------------------------
Page 2 - Mr. Charles Simonelli
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Celia M. Witten, Ph.D., M.D.
Director
Division of General and
Restorative Devices
Office of Device Evaluation
Center for Devices and
Radiological Health
Enclosure
{4}------------------------------------------------
International Medical Systems PO Box 4936 Annapolis, Maryland 21403
Statement For Indications For Use:
K 940215 510(k) Number (if known):_
Device Name: Mathis Vertebral and Bone Biopsy System
Indications For Use:
The Mathis Vertebral and Bone Biopsy Systems are intended for use by physicians performing bone marrow biopsy procedures.
**Prescription Use**
(Per 21 CFR 801.109)
Dooelo
(Division Sign-Off Division of General Restorative Dev 9905 510(k) Number
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.