SPEED-CUT
Device Facts
| Record ID | K961407 |
|---|---|
| Device Name | SPEED-CUT |
| Applicant | Gallini Intl., Inc. |
| Product Code | KNW · Gastroenterology, Urology |
| Decision Date | Nov 21, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 876.1075 |
| Device Class | Class 2 |
Indications for Use
An automatic biopsy system designed to take core samples from a variety of soft tissues, including (but not limited to) liver, prostate, kidney.
Device Story
Speed-Cut® is an automatic, spring-loaded, depth-programmable, disposable biopsy instrument. Used by surgeons to obtain soft tissue core samples (liver, prostate, kidney). Operates via spring-loaded mechanism to capture tissue. Small form factor and extended needle lengths allow use in laparoscopic procedures, guided by fiber optics or ultrasound. Output is a tissue core sample for diagnostic evaluation. Benefits include precise, automated tissue acquisition with safety margin established by firing-to-failure testing.
Clinical Evidence
Bench testing only. Testing included firing tests (mean 50 fires before failure), intracutaneous reactivity, acute systemic activity, pyrogen testing, and cytotoxicity. All tests passed.
Technological Characteristics
Automatic spring-loaded biopsy instrument; disposable; depth-programmable; various sizes available for laparoscopic use; compatible with fiber optic or ultrasound guidance.
Indications for Use
Indicated for patients requiring soft tissue core biopsies, including liver, prostate, and kidney; suitable for use in laparoscopic procedures.
Regulatory Classification
Identification
A gastroenterology-urology biopsy instrument is a device used to remove, by cutting or aspiration, a specimen of tissue for microscopic examination. This generic type of device includes the biopsy punch, gastrointestinal mechanical biopsy instrument, suction biopsy instrument, gastro-urology biopsy needle and needle set, and nonelectric biopsy forceps. This section does not apply to biopsy instruments that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Predicate Devices
- ACECUT Automatic Biopsy System (Meadox Medical, Inc.)
Related Devices
- K013069 — BIOSPEED · H.S. Hospital Services S.P.A. · Dec 12, 2001
- K024120 — MODIFICATION TO TEMNO BIOPSY NEEDLES · Allegiance Healthcare Corp. · Jan 15, 2003
- K981301 — SPEED-CUT AUTOMATIC BIOPSY SYSTEM · Gallini U.S., LLC · Jul 8, 1998
- K962905 — TX BIOPSY NEEDLE · Gallini S.R.L. · Oct 11, 1996
- K203141 — Uramix CuraWay Biopsy Needle · Uramix, Inc. · Sep 17, 2021