The Uramix Curaway Automatic Core Biopsy Instrument is intended for use in obtaining biopsies from soft tissues such as liver, kidneys, prostate, spleen, lymph nodes and various soft tissue tumors. It is not intended for use in bone, or breast.
Device Story
Single-use, automatic core biopsy instrument; consists of inner stylet and cutting cannula with etched tip. Used for soft tissue sampling (liver, kidney, prostate, spleen, lymph nodes, tumors). Operated by clinicians; provides ultrasound/X-ray enhancement for guidance. Features adjustable depth stopper for localization; color-coded handle indicates gauge size. Device functions via mechanical spring-loaded mechanism to capture tissue core. Benefits include accurate, minimally invasive tissue acquisition for diagnostic evaluation. Used in clinical settings.
Clinical Evidence
Bench testing only. Includes biocompatibility (ISO 10993-5, -10, -11), sterilization validation (ISO 11138-2, 11737-2), EO residuals (EN ISO 10993-7), and shelf-life (ISTA2A, ASTM F1980-16). Mechanical testing compared puncture force (2.7N vs 5.3N), connection firmness, and operational resistance against predicate. Ex-vivo porcine tissue studies (muscle, liver) demonstrated equivalent sample length and weight.
Technological Characteristics
Single-use, spring-loaded automatic biopsy instrument. Materials: SUS304/06Cr19Ni10 stainless steel (cannula/stylet), ABS (handle/actuator). Features: etched tip for ultrasound/X-ray visibility, adjustable depth stopper, color-coded hubs. Sterilization: Ethylene Oxide (EO). Sizes: 12G-20G; lengths 80mm-250mm. Mechanical operation; no software or electronic components.
Indications for Use
Indicated for obtaining core biopsy samples from soft tissues including liver, kidney, prostate, spleen, lymph nodes, and soft tissue tumors. Not indicated for use in bone or breast.
Regulatory Classification
Identification
A gastroenterology-urology biopsy instrument is a device used to remove, by cutting or aspiration, a specimen of tissue for microscopic examination. This generic type of device includes the biopsy punch, gastrointestinal mechanical biopsy instrument, suction biopsy instrument, gastro-urology biopsy needle and needle set, and nonelectric biopsy forceps. This section does not apply to biopsy instruments that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). On the left is the seal of the Department of Health & Human Services. To the right of the seal is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in a smaller font size below the words "U.S. FOOD & DRUG".
September 17, 2021
Uramix, Inc. % Michael Mooreville, MD Medical Director Uramix, Inc 272 N. Lansdowne Ave. Lansdowne, Pennsylvania 19050
Re: K203141
Trade/Device Name: URAMIX CuraWay Automatic Core Biopsy Instrument Regulation Number: 21 CFR 876.1075 Regulation Name: Gastroenterology-Urology Biopsy Instrument Regulatory Class: Class II Product Code: KNW Dated: July 21, 2021 Received: July 21, 2021
Dear Dr. Mooreville :
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approv al of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4. Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K203141
#### Device Name
Uramix Curaway Automatic Core Biopsy Instrument
#### Indications for Use (Describe)
The Uramix Curaway Automatic Core Biopsy Instrument is intended for use in obtaining biopsies from soft tissues such as liver, kidneys, prostate, spleen, lymph nodes and various soft tissue tumors. It is not intended for use in bone, or breast.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
__ Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### SUBMITTER
Uramix, Inc 272 N. Lansdowne Ave. Lansdowne, PA 19050
| Contact Person: | Michael Mooreville, MD |
|-----------------|------------------------|
| | Medical Director |
| Telephone: | (610) 304-9070 |
| Email: | uramix@comcast.net |
## DEVICE
Name of the Device: URAMIX CuraWay Automatic Core Biopsy Instrument Regulation Number: 21 CFR 876.1075 Regulation Name: Instrument, Biopsy/Regulatory Class II Regulatory Specialty: Gastroenterology/Urology KNW Product code:
Manufacturer: Zhejiang CuraWay Medical Technology Co., Ltd Company Address: Room 106, Building 1, No.600, 21st Street, Qiantang New Area, Hangzhou, Zhejiang, China
## PREDICATE DEVICE
K133948, BARD® MAX-CORE® Disposable Core Biopsy Primary:
## DEVICE DESCRIPTION
The URAMIX CuraWay Automatic Core Biopsy Instrument is a single use only core biopsy device and is individually packaged and sterilized. The packaging is compatible with the product's EO sterilization method. The sterilization validation confirms the packaging is qualified to maintain the sterilization condition of the device. It is available in several needle gauge sizes and lengths. The side and rear actuator buttons are color coded according to the various gauge sizes. The 18 gauge is the one used for current intended marketing, for prostate and kidney biopsies.
{4}------------------------------------------------
### K203141
### 510(k) Summary
| ABN 1208,<br>ABN 1208A | ABN 1408,<br>ABN 1408A | ABN 1608,<br>ABN 1608A | ABN 1808,<br>ABN 1808A | ABN 2008,<br>ABN 2008A |
|------------------------|-------------------------|--------------------------|------------------------|--------------------------|
| ABN 1210,<br>ABN 1210A | ABN 1410,<br>ABN 1410A | ABN 1610,<br>ABN 1610A | ABN 1810,<br>ABN 1810A | ABN 2010,<br>ABN 2010A |
| ABN 1213,<br>ABN 1213A | ABN 1413,<br>ABN 1413A | ABN 1613,<br>ABN 1613A | ABN 1813,<br>ABN 1813A | ABN 2013,<br>ABN 2013A |
| ABN 1216,<br>ABN 1216A | ABN 1416,<br>ABN 1416A | BN 1616,<br>ABN 1616A | ABN 1816,<br>ABN 1816A | ABN 2016,<br>ABN 2016A |
| ABN 1220,<br>ABN 1220A | ABN 1420,<br>ABN 1420A | ABN 1620,<br>ABN 1620A | ABN 1820,<br>ABN 1820A | ABN 2020,<br>ABN 2020A |
| ABN 1225,<br>ABN 1225A | ABN 1425,<br>ABN 1425A | ABN 1625,<br>ABN 1625A | ABN 1825,<br>ABN 1825A | ABN 2025,<br>ABN 2025A |
| 12G<br>(2.7mm)<br>Blue | 14G<br>(2.1mm)<br>Green | 16G<br>(1.7mm)<br>Violet | 18G<br>(1.2mm)<br>Pink | 20G<br>(0.9mm)<br>Yellow |
According to gauge and length, the following variants are available:
The operating handle is color coded and the Type A is a variant in the position of the needle markings, which provide reference for depth of placement.
### INDUCATIONS FOR USE
The URAMIX CuraWay Automatic Core Biopsy Instrument is intended for use in obtaining core biopsy samples from soft tissue such as kidney, liver, prostrate, spleen, lymph nodes, and various soft tissue tumors. It is not intended for use in bone, or breast.
{5}------------------------------------------------
## COMPARISON TO PREDICATE
| | Predicate Device | Subject Device | |
|-------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------|
| Item | BARD® MAX-CORE® Disposable<br>Core Biopsy Instrument, K133948 | URAMIX CuraWay Automatic Core<br>Biopsy Instrument, K203141 | Outcome |
| Intended<br>Use | Obtaining biopsies from soft tissues<br>such as liver, kidney, thyroid,<br>prostate, spleen, lymph nodes and<br>various soft tissue tumors.<br>It is not intended for use in bone. | The URAMIX Curaway Automatic<br>Core Biopsy Instrument is<br>intended for use in obtaining<br>biopsies from soft tissues such<br>as liver, kidney, prostate, spleen,<br>lymph nodes and various soft<br>tissue tumors. It is not intended<br>for use in bone, or breast. | It was used for the same<br>clinical condition, and same<br>intended purpose, and at the<br>same site. |
| Material | Stainless steel + ABS | The shaft of cutting cannula and inner<br>stylet is made of medical grade stainless<br>steel (SUS304/06Cr19Ni10). The operating<br>handle and actuator button are made of<br>acrylonitrile butadiene styrene (ABS). | Same material used |
| Types/<br>Sizes | 14G, 16G, 18G, 20G | 12G, 14G, 16G, 18G, 20G | 12G is also available from<br>Zhejiang CuraWay, this has<br>no significant influence on<br>clinical use. |
| Design<br>Attributes | Each needle has numerically ordered<br>centimeter markings on the outer<br>cannula to provide reference for depth<br>placement<br>Needles feature an adjustable needle<br>stop which allows the user to restrict<br>forward movement, localizing the<br>needle tip to the biopsy site<br>Color coded stylet hubs indicate<br>gauge size of needles available in a<br>variety of gauge sizes and centimeter<br>lengths<br>Available with an echogenic tip on the<br>outer cannula to promote accurate<br>placement under ultrasound guidance | Automatic Core Biopsy Instrument is a<br>puncture needle consisting of an inner<br>stylet and a cutting cannula with etched tip<br>(s. Figure 1). The cutting cannula has<br>numerically ordered centimeter markings<br>on it to provide reference for depth<br>placement. In the distal area of the cutting<br>cannula, an ultrasound enhancement is<br>available to promote accurate placement<br>under ultrasound or X-ray guidance. An<br>adjustable depth stopper allows the user to<br>restrict forward movement, localizing the<br>needle tip to the biopsy site. The Type A is<br>a variant in position of the depth stopper.<br>The operating handle is color-coded, which<br>indicates gauge size of the needle. The<br>needle is protected in a needle sheath.<br>Automatic Core Biopsy Instrument is<br>available in several needle gauge sizes<br>and lengths. | Very similar design |
| Length<br>(mm) | 100mm, 160mm, 200mm, 250mm | 80mm, 100mm, 130mm, 1600mm, 200mm,<br>250mm | This has no significant<br>influence on clinical use. |
| Slot size | 18mm | 18mm | Same |
| Structure<br>a. Photo<br>(examples) | | | Very similar structure and<br>design concept |
| Packaging | Tyvek 1073B + PE | Tyvek 1073B + PET/PE | Critical component Tyvek is<br>the same.<br>Both packaging systems are<br>common packaging. |
| Biocompati<br>bility | ISO 10093-1, ISO 10993-5 and ISO<br>10993-10 | ISO 10093-1, ISO 10993-5 and ISO 10993-<br>10 | Same. |
| Sterilizatio<br>n method | EO sterilization | EO sterilization | Same. |
| Conclusion | After comparison of the device in question with the predicate device, K133948 regarding<br>clinical, technical and biological characteristics, it can be concluded that both devices have<br>similar design, perform the same function under the same clinical conditions. They use the<br>same materials and have the same durability. Tissue penetration and sample size are the same<br>and the device is in contact with the same human tissues and body fluids. They can be<br>considered as equivalent devices. | | |
{6}------------------------------------------------
## SUMMARY OF NON CLINICAL TESTING
The following testing was conducted to demonstrate the safe and effective use of the URAMIX Curaway Automatic Core Biopsy Instrument:
- Biocompatibility Testing per ISO 10993-5, including in vitro cytotoxicity ●
- ISO 10993-10 tests for irritation and skin sensitization ●
- ISO 10883-11 test for systemic toxicity ●
- Sterilization Testing per ISO 11138-2, and ISO 11737-2 sterility test ●
- Residuals per EN ISO 10993-7 EO residuals = 1.13 mg/device ●
- Packaging and Shelf-life per ISTA2A and ASTM F1980-16 3-year shelf life ●
{7}------------------------------------------------
- Puncture Force Value: Predicate 5.3N; Subject 2.7N ●
- Resistance during operation: Predicate none; Subject none ●
- Abnormal noise during operation: Predicate none: Subject none .
- Connection firmness (needle tube + plastic parts): P 181 to 230N; S 190 to 212N ●
- Samples of various tissues: similar length, weight between Predicate and Subject .
- The ex-vivo tissue studies (porcine muscle, liver) demonstrate equivalent tissue . samples in terms of length and weight.
## CONCLUSION
Based upon the performance data provided in this submission and comparing indicated use, design, materials, principle of operation and overall technological characterisitics, the URAMIX Curaway Automatic Core Biopsy Instrument has been determined to be substantially equivalent to the predicate device, BARD Max-Core, K133948
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.