← Product Code [KNW](/submissions/GU/subpart-b%E2%80%94diagnostic-devices/KNW) · K171531

# Jamshidi Bone Marrow Biopsy/Aspiration Needle, Jamshidi T-Handle Bone Marrow Biopsy/Aspiration Needle, Jamshidi Evolve Bone Marrow Biopsy/Aspiration Needle (K171531)

_Care Fusion · KNW · Sep 11, 2017 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/GU/subpart-b%E2%80%94diagnostic-devices/KNW/K171531

## Device Facts

- **Applicant:** Care Fusion
- **Product Code:** [KNW](/submissions/GU/subpart-b%E2%80%94diagnostic-devices/KNW.md)
- **Decision Date:** Sep 11, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1075
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

Jamshidi Bone Marrow Biopsy/Aspiration Needle – DJ: Intended use for the posterior iliac crest biopsy technique. Jamshidi T-Handle Bone Marrow Biopsy/Aspiration Needle - TJ: Intended use for the posterior iliac crest biopsy technique. Jamshidi Evolve Bone Marrow Biopsy/Aspiration Needle - EJ: Indicated for use in aspirating bone marrow and for use in obtaining core biopsy samples of bone and/or bone marrow.

## Device Story

Manual, sterile, disposable needles for bone marrow aspiration and core biopsy; used in clinical settings by healthcare providers. Device consists of outer cannula with handle, inner stylet, and optional marrow acquisition (MAC) or specimen cradle. Operation involves manual insertion into bone; cradle captures and retrieves core sample during withdrawal. Provides tissue samples for diagnostic evaluation; aids in clinical decision-making for hematological or bone-related conditions. Differences from predicates include expanded needle lengths, centimeter depth markings, updated tip geometry, modified cradle design, and new depth stop accessory.

## Clinical Evidence

No clinical tests were conducted for this submission. Evidence is based on bench-top testing (pull and sharpness testing) and simulated use testing.

## Technological Characteristics

Manual, single-use biopsy/aspiration needles. Components: cannula, handle, inner stylet, specimen/marrow acquisition cradle. Materials comply with ISO 10993 biocompatibility standards. Sterilization via Ethylene Oxide (ISO 11135) or Radiation (ISO 11137). Packaging validated per ISO 11607. Accelerated aging per ASTM F1980.

## Regulatory Identification

A gastroenterology-urology biopsy instrument is a device used to remove, by cutting or aspiration, a specimen of tissue for microscopic examination. This generic type of device includes the biopsy punch, gastrointestinal mechanical biopsy instrument, suction biopsy instrument, gastro-urology biopsy needle and needle set, and nonelectric biopsy forceps. This section does not apply to biopsy instruments that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Predicate Devices

- [K813338](/device/K813338.md)
- [K913306](/device/K913306.md)
- [K003370](/device/K003370.md)
- [K070091](/device/K070091.md)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

September 11, 2017

CareFusion Tamara Brey Regulatory Affairs Manager 75 North Fairway Drive Vernon Hills, Illinois 60061

Re: K171531

Trade/Device Name: Jamshidi Bone Marrow Biopsy/Aspiration Needle, Jamshidi T-Handle Bone Marrow Biopsy/Aspiration Needle, Jamshidi Evolve Bone Marrow Biopsy/Aspiration Needle Regulation Number: 21 CFR 876.1075 Regulation Name: Gastroenterology-Urology Biopsy Instrument Regulatory Class: Class II Product Code: KNW Dated: August 29, 2017 Received: August 31, 2017

Dear Tamara Brey:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Actinclude requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device comples with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply

Image /page/0/Picture/11 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.

{1}------------------------------------------------

with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely. Jennifer R. Stevenson -S3

For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

# Indications for Use

510(k) Number (if known)

K171531

Device Name

Jamshidi Bone Marrow Biopsy/Aspiration Needles

Indications for Use (Describe)

Jamshidi Bone Marrow Biopsy/Aspiration Needle – DJ: Intended use for the posterior iliac crest biopsy technique.

Jamshidi T-Handle Bone Marrow Biopsy/Aspiration Needle - TJ: Intended use for the posterior iliac crest biopsy technique.

Jamshidi Evolve Bone Marrow Biopsy/Aspiration Needle - EJ: Indicated for use in aspirating bone marrow and for use in obtaining core biopsy samples of bone and/or bone marrow.

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|-------------------------------------------------|--|

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Page 1 of 3

# 510(k) SUMMARY – K171531

A summary of 510(k) safety and effectiveness information in accordance with 21 CFR 807.92.

| SUBMITTER INFORMATION                                         |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
|---------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Name                                                          | CareFusion                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Address                                                       | 75 North Fairway Drive, Vernon Hills, IL 60061 USA                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Phone number                                                  | (847) 362-9285                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Fax number                                                    | (312) 949-9245                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Establishment Registration<br>Number                          | 1423507                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Name of contact person                                        | Tamara Brey                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Date prepared                                                 | September 7, 2017                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| DESCRIPTION OF DEVICE                                         |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Trade or proprietary name                                     | Jamshidi® Bone Marrow Biopsy/Aspiration Needle<br>Jamshidi® T-Handle Bone Marrow Biopsy/Aspiration Needle<br>Jamshidi® Evolve Bone Marrow Biopsy/Aspiration Needle                                                                                                                                                                                                                                                                                                                                                                                        |
| Common or usual name                                          | Bone Marrow Biopsy Needle                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Classification name                                           | Gastroenterology-urology biopsy instrument                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Classification panel                                          | Gastroenterology/Urology                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Regulation                                                    | Class II per 21CFR §876.1075                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Product Code                                                  | KNW                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Legally marketed device(s) to<br>which equivalence is claimed | K813338, K913306, K003370, and K070091.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Device description                                            | The Jamshidi® devices are manual, sterile, disposable needles intended<br>to obtain bone marrow aspirate and core biopsy samples from bone<br>and/or bone marrow. The devices are comprised of an outer cannula with<br>a handle and an inner stylet. Some models include a marrow acquisition<br>cradle (MAC) or a specimen cradle. The MAC and specimen cradle fit<br>inside the cannula to capture and retrieve samples by mechanically<br>cutting the core sample. The specimen is contained within the cradle<br>during withdrawal from the cannula. |
| Intended use of the device                                    | Jamshidi Bone Marrow Biopsy/Aspiration Needle - DJ: Intended use for<br>the posterior iliac crest biopsy technique.<br>Jamshidi T-Handle Bone Marrow Biopsy/Aspiration Needle - TJ:<br>Intended use for the posterior iliac crest biopsy technique.<br>Jamshidi Evolve Bone Marrow Biopsy/Aspiration Needle – EJ: Indicated<br>for use in aspirating bone marrow and for use in obtaining core biopsy<br>samples of bone and/or bone marrow.                                                                                                              |

{4}------------------------------------------------

## DEVICE COMPARISON TO PREDICATE DEVICE

The Jamshid® Bone Marrow Biopsy/Aspiration Needles are the predicate devices in the intended use, target population, fundamental scientific technology, operating principles, fundamental mechanical design, performance characteristics, packaging, sterility and biocompatibility.

The technological characteristics of the proposed Jamshidi® Bone Marrow Biopsy/Aspiration Needles utilizes the same basic components as identified in the predicate Jamshidi® devices to achieve its intended use: a cannula, handle, inner stylet and a marrow acquisition cradle (MAC) or a specimen cradle in some models. The proposed and predicate devices are for single use, have the same materials, and operate manually to access bone marrow.

The differences between the proposed and predicate devices include an expanded range of needle lengths, introduction of centimeter depth markings to facilitate placement, a new needle tip geometry, an updated specimen cradle design, and a new depth stop accessory. The general procedures and guidelines in the proposed device's labeling were updated to provide clarity and facilitate biopsy sample collection.

In summary, the proposed changes were appropriately assessed and do not raise any new or significant questions of safety and efficacy. The proposed devices are substantially equivalent to the predicate devices presented.

# PERFORMANCE DATA

#### SUMMARY OF NON-CLINICAL TESTS CONDUCTED FOR DETERMINATION OF SUBSTANTIAL EQUIVALENCE - PERFORMANCE TEST SUMMARY

Bench-top testing (pull testing, sharpness testing) was conducted on EJ devices. Simulated use testing was conducted on DJ, TJ and EJ devices. Comparative testing (sharpness testing, simulated use testing) was conducted on JBC devices. All testing was successfully completed, showing with reasonable assurance that the proposed CareFusion's Jamshidi Bone Marrow Biopsy/Aspiration Needles meet or exceed all performance requirements, and are substantially equivalent to the predicate devices.

| Characteristic   | Standard/Test/FDA Guidance                                                                                                                                                                                           |
|------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Risk Management  | BS EN ISO 14971:2012<br>Medical Devices. Application of Risk Management to Medical Devices                                                                                                                           |
| Packaging        | AAMI / ANSI / ISO 11607-1:2006/(R)2010 [Including: Amendment 1 (2014)] Packaging<br>for Terminally Sterilized Medical Devices - Part 1: Requirements for Materials, Sterile<br>Barrier Systems And Packaging Systems |
| Packaging        | AAMI / ANSI / ISO 11607-2:2006/(R)2010, [including: amendment 1 (2014)]. Packaging<br>for Terminally Sterilized Medical Devices - Part 2: Validation Requirements for Forming,<br>Sealing and Assembly Processes     |
| Biocompatibility | ANSI/AAMI/ISO 10993-1:2009 (R) 2013 Biological Evaluation of Medical Devices - Part<br>1: Evaluation and Testing within a Risk Management Process                                                                    |
| Biocompatibility | ANSI/AAMI/ISO 10993-4:2002 Amendment 1 2006 Biological Evaluation of Medical<br>Devices - Part 4: Selection of Tests for Interactions with Blood                                                                     |
| Biocompatibility | ANSI/AAMI/ISO 10993-5:2009 (R) 2014 Biological Evaluation of Medical Devices -<br>Part 5: Tests for In Vitro Cytotoxicity                                                                                            |
| Biocompatibility | ANSI/AAMI/ISO 10993-7:2008 (R) 2012 Biological Evaluation of Medical Devices - Part<br>7: Ethylene Oxide Sterilization Residuals                                                                                     |
| Biocompatibility | ANSI/AAMI/ISO 10993-10:2010 (R) 2014 Biological Evaluation of Medical Devices -<br>Part 10: Tests for Irritation and Skin Sensitization                                                                              |
| Biocompatibility | ANSI/AAMI/ISO 10993-11:2006 Biological Evaluation of Medical Devices - Part 11:<br>Tests for Systemic Toxicity                                                                                                       |

{5}------------------------------------------------

------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

| Biocompatibility                                                                                                | ANSI/AAMI /ISO 10993-12:2012 Biological Evaluation of Medical Devices – Part 12: Sample Preparation and Reference Materials                                                                                                                |
|-----------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Biocompatibility                                                                                                | ANSI/AAMI/ISO 10993-17:2002 (R) 2012 Biological Evaluation of Medical Devices – Part 17: Establishment of Allowable Limits for Leachable Substances                                                                                        |
| Biocompatibility                                                                                                | ISO 10993-18: 2005 Biological Evaluation of Medical Devices - Part 18: Chemical Characterization of Materials                                                                                                                              |
| Sterilization                                                                                                   | ISO 11135:2014 Sterilization of Health Care Products - Ethylene Oxide Requirements for the Development, Validation and Routine Control of a Sterilization Process for Medical Devices                                                      |
| Sterilization                                                                                                   | ISO 11138-1: 2006/(R) 2010 Sterilization of Health Care Products - Biological Indicators - Part 1: General Requirements                                                                                                                    |
| Sterilization                                                                                                   | ISO 11737-1:Second Edition 2006-04-01 Sterilization of Medical Devices - Microbiological Methods - Part 1                                                                                                                                  |
| Sterilization                                                                                                   | AAMI TIR28:2009/(R)2013 Product Adoption and Process Equivalency for Ethylene Oxide Sterilization                                                                                                                                          |
| Sterilization                                                                                                   | AAMI/ANSI/ISO 11137-1:2006/(R) 2010 Including: Amendment 1 (2013) Sterilization of Health Care Products - Radiation - Part 1: Requirements for Development, Validation, and Routine Control of a Sterilization Process for Medical Devices |
| Stability                                                                                                       | ASTM F1980-07 Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices                                                                                                                                          |
| SUMMARY OF CLINICAL TESTS CONDUCTED FOR DETERMINATION OF SUBSTANTIAL EQUIVALENCE AND/OR OF CLINICAL INFORMATION |                                                                                                                                                                                                                                            |
| N/A - No clinical tests were conducted for this submission                                                      |                                                                                                                                                                                                                                            |
| CONCLUSION OF DEVICE COMPARISON                                                                                 |                                                                                                                                                                                                                                            |
| The Jamshidi Bone Marrow Biopsy/Aspiration Needles are equivalent to the predicate devices in that:             |                                                                                                                                                                                                                                            |
| - The devices have the same intended use.                                                                       |                                                                                                                                                                                                                                            |
| - The devices have substantially equivalent design, function and areas of application.                          |                                                                                                                                                                                                                                            |

| - The devices demonstrate substantially equivalent performance.

---

**Source:** [https://fda.innolitics.com/submissions/GU/subpart-b%E2%80%94diagnostic-devices/KNW/K171531](https://fda.innolitics.com/submissions/GU/subpart-b%E2%80%94diagnostic-devices/KNW/K171531)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
