Osteo-Site Bone Biopsy Needle, Osteo-Site Bone Biopsy Needle Set
Applicant
Cook Incorporated
Product Code
KNW · Gastroenterology, Urology
Decision Date
Feb 28, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.1075
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Osteo-Site® Bone Biopsy Needle and Osteo-Site® Bone Biopsy Needle Set is intended for vertebral body access, biopsy and infusion during a vertebroplasty procedure.
Device Story
Osteo-Site Bone Biopsy Needle and Set are manual surgical instruments used for percutaneous access to the vertebral body. The device consists of a stainless steel cannula, inner stylet, and obturator, with an optional inner biopsy needle. The physician inserts the cannula into the vertebral body; the stylet is removed to allow for biopsy or infusion. The device is used in a clinical setting (e.g., OR or interventional suite) by a physician to facilitate vertebroplasty. The device provides a pathway for bone tissue sampling and subsequent cement infusion. Benefits include minimally invasive access to the vertebral body for diagnostic and therapeutic purposes.
Clinical Evidence
Bench testing only. Performance testing included compatibility and dimension analysis, tensile testing, torque strength, torsional override, compression strength, resistance to breakage, and biopsy performance testing. Biocompatibility testing included cytotoxicity, sensitization, intracutaneous reactivity, acute systemic toxicity, hemolysis, partial thromboplastin time, and complement activation. All tests met predetermined acceptance criteria.
Technological Characteristics
Cannula, stylet, obturator, and inner biopsy needle constructed of stainless steel. Needle holder made of ABS/Polycarbonate. Available in 11, 13, 15 gauge (cannula) and 14, 16 gauge (biopsy needle). Tip configurations include tapered, bevel, Franseen, scoop, trocar, diamond, and trephine. Sterilization via EtO to a 10-6 SAL. Manual, non-powered device.
Indications for Use
Indicated for vertebral body access, biopsy, and infusion during vertebroplasty procedures in patients requiring bone tissue sampling or cement delivery.
Regulatory Classification
Identification
A gastroenterology-urology biopsy instrument is a device used to remove, by cutting or aspiration, a specimen of tissue for microscopic examination. This generic type of device includes the biopsy punch, gastrointestinal mechanical biopsy instrument, suction biopsy instrument, gastro-urology biopsy needle and needle set, and nonelectric biopsy forceps. This section does not apply to biopsy instruments that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Predicate Devices
Stryker Bone and Vertebral Body Biopsy Kit (K032943)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a symbol. The symbol consists of three stylized human profiles facing to the right, with flowing lines beneath them. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 28, 2017
Cook Incorporated Ms. Erum Nasir, MS Regulatory Affairs Team Lead 750 Daniels Way Bloomington, Indiana 47404
Re: K170008
Trade/Device Name: Osteo-Site Bone Biopsy Needle. Osteo-Site Bone Biopsy Needle Set Regulation Number: 21 CFR 876.1075 Regulation Name: Gastroenterology-urology biopsy instrument Regulatory Class: Class II Product Code: KNW Dated: December 30, 2016 Received: January 3, 2017
Dear Ms. Nasir:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and Part 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely Yours,
# Jennifer R. Stevenson -S
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K170008
Device Name
Osteo-Site® Bone Biopsy Needle and the Osteo-Site® Bone Biopsy Needle Set
Indications for Use (Describe)
The Osteo-Site® Bone Biopsy Needle and Osteo-Site® Bone Biopsy Needle Set is intended for vertebral body access, biopsy and infusion during a vertebroplasty procedure.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 2.0 510(k) SUMMARY
# Osteo-Site® Bone Biopsy Needle and the Osteo-Site® Bone Biopsy Needle Set 21 CFR §807.92 Date Prepared: December 30, 2016
#### Submitted By:
| Submission: | Traditional 510(k) Premarket Notification |
|----------------------------|------------------------------------------------------------------------------|
| Applicant: | Cook Incorporated |
| Contact: | Erum Nasir, MS |
| Applicant Address: | Cook Incorporated<br>750 Daniels Way<br>Bloomington, IN 47404 |
| Contact Phone Number: | (812) 335-3575 x102607 |
| Contact Fax Number: | (812) 332-0281 |
| Device Information: | |
| Trade Name: | Osteo-Site® Bone Biopsy Needle and the Osteo<br>Site® Bone Biopsy Needle Set |
| Common Name: | Instrument, Biopsy |
| Classification Name: | Gastroenterology-Urology Biopsy Instrument |
| Classification Regulation: | 21 CFR §876.1075, Product Code KNW |
## Device Class/Classification Panel: Class II, Gastroenterology/Urology
### Predicate Devices:
- Stryker Bone and Vertebral Body Biopsy Kit (K032943) ●
### Device Description:
The Osteo-Site® Bone Biopsy Needle and Osteo-Site® Bone Biopsy Needle Set consists of a cannula with an accompanying inner stylet and an obturator. The device may include an additional inner biopsy needle and obturator. The cannula is available in several sizes (11, 13 or 15 gauge) and lengths (10 or 15 cm) with various cannula/stylet tip configurations. The additional inner biopsy needle is available in two sizes (14 or 16 gauge) and lengths (17.1 or 22.1 cm) with a trephine tip and functions coaxially with the cannula once the stylet is removed. A plastic, non-radiopaque needle holder is also available to aid with needle placement.
{4}------------------------------------------------
# Intended Use:
The Osteo-Site® Bone Biopsy Needle and Osteo-Site® Bone Biopsy Needle Set is intended for vertebral body access, biopsy and infusion during a vertebroplasty procedure.
# Comparison to Predicate Devices:
The Osteo-Site® Bone Biopsy Needle and Osteo-Site® Bone Biopsy Needle Set is substantially equivalent to the predicate device, the Stryker Bone and Vertebral Body Biopsy Kit (K032943), in that these devices have similar indications for use, technological characteristics, design, and method of placement. The differences between the subject devices and the predicate devices are shown in the table below and include the materials, dimensions, cannula/stylet tip configurations, sterilization method, and included components. These minor differences were appropriately assessed and do not raise new questions regarding safety or effectiveness.
| | Stryker Bone and Vertebral<br>Biopsy Kit (K032943) | Osteo-Site® Bone Biopsy Needle<br>and Set -<br>Proposed Device |
|---------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------|
| Regulation | 876.1075 | Identical |
| Product Code | KNW | Identical |
| Classification | II | Identical |
| Intended Use | Stryker's bone and vertebral<br>body biopsy kits can be used<br>as a biopsy tool to remove a<br>sample of bone tissue from a<br>vertebral body for diagnostic<br>purposes using a coring,<br>cutting or aspiration technique. | This device is intended for vertebral<br>body access, biopsy and infusion<br>during a vertebroplasty procedure. |
| Ancillary Device<br>Information | Stryker Introduction<br>Cannula/Stylet is sold<br>separately, but required to<br>access the vertebral body to<br>perform a biopsy with the<br>Stryker Bone and Vertebral<br>Biopsy Kit | N/A<br>The Cannula/Stylet combination is<br>included as a component in the Osteo-<br>Site® Bone Biopsy Needle Set and is<br>subject of this submission. |
| Anatomical Site | Vertebral Body | Identical |
| Method of Placement | Percutaneous | Identical |
| Biopsy technique | Coring/Cutting/Suction | Not specified |
| Biopsy Component | Inner Biopsy Needle | Cannula/Inner Biopsy Needle* |
#### Substantial Equivalence Table
* The proposed set includes the inner biopsy needle and obturator.
{5}------------------------------------------------
| Substantial Equivalence Table (cont'd) | | | |
|----------------------------------------|----------------------|----------------------------------------------------|----------------------------------------------------------------|
| | | Stryker Bone and Vertebral<br>Biopsy Kit (K032943) | Osteo-Site® Bone Biopsy Needle<br>and Set -<br>Proposed Device |
| Cannula** | Material | Stainless Steel | Identical |
| | Gauge | 10, 11, 13 | 11, 13, 15 |
| | Length (cm) | 12.7 | 10, 15 |
| | Tip<br>Configuration | Four-Facet | Tapered, Bevel, Franseen |
| Stylet** | Material | Stainless Steel | Identical |
| | Tip<br>Configuration | Diamond, Bevel | Scoop, Trocar, Bevel, Diamond |
| Inner<br>Biopsy<br>Needle* | Material | Stainless Steel | Identical |
| | Gauge | 10, 11, 13 | 14, 16 |
| | Length (cm) | 12.7, 22.9 | 17.1, 22.1 |
| | Tip<br>Configuration | Four-Facet | Trephine |
| Obturator* | Material | Stainless Steel | Identical |
| | Number | One | One or Two |
| Handle Material | | Unknown | ABS/Polycarbonate |
| Accessories | | Syringe | Needle Holder |
| Obturator Material | | Stainless Steel | Identical |
| Sterilization Method | | Gamma Irradiation,<br>VDmax25 | EtO |
| Sterilization Assurance Level | | 10-6 | 10-6 |
| Packaging | | Peel pouch | Tyvek peel pouch |
#### Substantial Equivalence Table (cont'd)
* The proposed set includes the inner biopsy needle and obturator.
** Stryker Introduction Cannula/Stylet is listed under product code OAR and sold separately, but is required to access the vertebral body to perform a biopsy with the Stryker Bone and Vertebral Biopsy Kit cleared under K032943. Predicate IFU and brochure is provided in Appendix B (Page B-6 and B-9, respectively) for reference purpose only.
# Performance Data:
The following tests were performed to demonstrate that the proposed Osteo-Site® Bone Biopsy Needle and Osteo-Site® Bone Biopsy Needle Set met the applicable design and performance requirements and support a determination of substantial equivalence. Where applicable, testing was done per applicable ISO and other international standards.
- . Compatibility and Dimension Analysis (Cannula, Stylet, Obturator, Inner biopsy needle, Inner biopsy needle obturator) - Testing demonstrated that the devices function as intended.
{6}------------------------------------------------
- . Tensile Testing (Cannula, Stylet, Obturator, Inner biopsy needle, Inner biopsy needle obturator, and Needle holder) - Testing demonstrated that the devices withstand peak load values greater than the predetermined acceptance criterion.
- Torque Strength Testing (Cannula, stylet, Inner biopsy needle) Testing . demonstrated that the devices withstand torque forces greater than the predetermined acceptance criterion.
- Torsional Override Testing (Cannula) Testing demonstrated that the devices ● withstand torsional load greater than the predetermined acceptance criterion.
- Compression Strength Testing (Cannula and stylet assembly, Inner biopsy needle) . – Testing demonstrated that the devices withstand compression forces greater than the predetermined acceptance criterion.
- . Resistance to Breakage Testing (Cannula and stylet assembly)- Testing demonstrated that the devices function as intended.
- Biopsy Testing (Cannula and inner biopsy needle) Testing demonstrated that the . devices function as intended.
- . Biocompatibility Testing - Testing (cytotoxicity, sensitization, intracutaneous reactivity, acute systemic toxicity, hemolysis, partial thromboplastin time, and complement activation) shows that the devices are biocompatible.
# Conclusion:
The results of these tests support a conclusion that the Osteo-Site® Bone Biopsy Needle and Osteo-Site® Bone Biopsy Needle Set met the design input requirements based on the intended use and support the conclusion that this device does not raise new questions of safety or effectiveness and is substantially equivalent to the predicate device, Stryker Bone and Vertebral Body Biopsy Kit (K032943).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.