The Ranfac Aspiration Needle with Adjustable Guide is intended for use for aspiration of bone marrow or autologous blood using a standard piston syringe.
Device Story
Device is a manual aspiration needle used for bone marrow or autologous blood collection. Consists of stainless steel cannula and two stainless steel stylets with ABS plastic handles. User inserts needle into bone marrow cavity by applying pressure and twisting handles; stylet is removed and replaced with a blunt stylet; needle is advanced to depth set by adjustable guide. Aspiration performed by attaching standard piston syringe to Luer fitting and applying negative pressure. Used in clinical settings by healthcare professionals. Benefits include controlled depth aspiration and ability to sample multiple areas via partial withdrawal.
Clinical Evidence
No clinical data. Substantial equivalence is supported by bench testing, material compliance with ISO standards, and design similarity to predicate devices.
Technological Characteristics
Stainless steel cannula and stylets (ISO 9626); ABS plastic handles; adjustable depth guide; Luer lock connections (ISO 594/1, ISO 594-2). Sterilization via Ethylene Oxide (ISO 11135-1, ISO 10993-7). Manual operation; no energy source or software.
Indications for Use
Indicated for aspiration of bone marrow or autologous blood in patients requiring such procedures.
Regulatory Classification
Identification
A gastroenterology-urology biopsy instrument is a device used to remove, by cutting or aspiration, a specimen of tissue for microscopic examination. This generic type of device includes the biopsy punch, gastrointestinal mechanical biopsy instrument, suction biopsy instrument, gastro-urology biopsy needle and needle set, and nonelectric biopsy forceps. This section does not apply to biopsy instruments that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
SwannShidi Bone Marrow Aspiration Needle (K121181)
Submission Summary (Full Text)
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol featuring three stylized human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 11, 2014
Ranfac Corporation Mr. Christopher P. Whelan Senior Vice President 30 Doherty Avenue Avon, Massachusetts 02322
Re: K140991
Trade/Device Name: Ranfac Aspirating Needle with Adjustable Guide Regulation Number: 21 CFR 876.1075 Regulation Name: Gastroenterology-urology-biopsy instrument Regulatory Class: Class II Product Code: KNW Dated: September 2, 2014 Received: September 3, 2014
Dear Mr. Whelan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{2}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# David Krause -S
Radiological Health
Binita S. Ashar, M.D., M.B.A., F.A.C.S. for Director Division of Surgical Devices Office of Device Evaluation Center for Devices and
Enclosure
{3}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
### Indications for Use
510(k) Number (if known) K140991
Device Name
Ranfac Aspirating Needle with Adjustable Guide.
Indications for Use (Describe)
The Ranfac Aspiration Needle with Adjustable Guide is intended for use for aspiration of bone marrow or autologous blood using a standard piston syringe.
Type of Use (Select one or both, as applicable)
ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘ
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY CONTR
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
{4}------------------------------------------------
Image /page/3/Picture/0 description: The image shows a logo for a company called Ranfac. The logo is split into two halves, with the left half being black and the right half being white. The word "Ranfac" is written in black letters above the logo. The logo is a semi-circle shape.
# SECTION 5
# K140991 510(k) Summary
The contents of this 510(k) summary on the following pages have been provided in conformance with 21 CFR § 807.92 Content and format of a 510(k) summary.
# 510(k) Summary
Owner's Name and Address:
Official Contact Person:
Date Summary Prepared:
Device Trade Name:
Common Name:
Classification Name:
Ranfac Corp. 30 Doherty Avenue Avon, MA 02322-0635 FDA Registration Number 1211566
Christopher P. Whelan Senior Vice President Telephone: 508-588-4400 extension: 106 Facsimile: 508-584-8588 e-mail: cwhelan@ranfac.com
April 15, 2014
Ranfac Aspiration Needle with Adjustable Guide
Needle, Aspiration and Injection, Disposable
Gastroenterology-urology biopsy instrument (KNW) Subpart B Diagnostic Devices 21 CFR § 876.1075, Gastroenterology-urology biopsy instrument, Class II
# 510(k) Summary
RANFAC CORP. | PO BOX 635 | 30 DOHERTY AVE | AVON, MA 02322 | USA 800-272-6322 | Fax: 508-584-8588 Phone: 508-588-4400 -Email info@ranfac.com
www.ranfac.com
sincc 1888
{5}------------------------------------------------
Image /page/4/Picture/0 description: The image shows a logo with the text "Ranfac" on the top right. There is a horizontal line above the text. Below the text is a semi-circle that is split in half, with the left half being black and the right half being white.
# 510(k) Summary
#### Predicate Device:
| 510(k)<br>Number | Predicate Description | Manufactured By |
|------------------|------------------------------------------|------------------------|
| K131157 | Ranfac Bone Marrow Aspiration Needle | Ranfac Corp. |
| K121181 | SwannShidi Bone Marrow Aspiration Needle | Alliance Partners, LLC |
#### Background
The Ranfac Aspiration Needle with Adjustable Guide is a variation of the Ranfac Bone Marrow Aspiration Needle (K131157). The Ranfac Aspiration Needle with Adjustable Guide is intended for use for aspiration of bone marrow or autologous blood using a standard piston syringe. The Ranfac Aspiration Needle with Adjustable Guide works in a similar manner to the Ranfac Bone Marrow Aspiration Needle in that it enters the bone by the user applying pressure to the needle/stylet while twisting the handles. Once the needle enters the marrow cavity, the Stylet is removed and a Blunt Stylet is inserted and the needle is advanced to the desired location as set by the Adjustable Guide. Aspiration is performed by attaching a syringe to the Luer fitting of the needle and applying negative pressure. The needle can be partially withdrawn to continue to aspirate from different areas prior to complete withdrawal.
#### Device Description:
The Ranfac Aspiration Needle with Adjustable Guide consists of a stainless steel cannula, and two stainless steel stylets. All of the aforementioned have handles molded of ABS plastic. The cannula has an Adjustable Guide, also molded of ABS plastic that can be used as a depth guide.
RANFAC CORP. | PO BOX 635 | 30 DOHERTY AVE | AVON, MA 02322 | USA Fax: 508-584-8588 800-272-6322 || || || || || || || || || || || || || || || || || || || || || || || || || || || || || || || || || || || || || || || || || || || || || || || || || || || || || || Phone: 508-588-4400 Email info@ranfac.com
www.ranfac.com
sincc 1888
{6}------------------------------------------------
Image /page/5/Picture/0 description: The image shows a logo with the text "Ranfac" at the top. Below the text is a semi-circular shape that is divided vertically into two halves. The left half of the semi-circle is filled with a solid black color, while the right half is white with a black outline.
# 510(k) Summary
The Ranfac Aspiration Needle with Adjustable Guide is intended for use for aspiration of bone marrow or autologous blood using a standard piston syringe.
The Ranfac Aspiration Needle with Adjustable Guide is similar in materials and design to the predicate devices. All devices are comprised of stainless steel cannula and stylets with mating plastic handles. All devices have luer fittings enabling the use of standard piston syringes.
The Stainless Steel Cannula and Stylets, contact the patient in a limited use, and is in conformance with ISO 9626 First edition 1991-09-01, Amendment 1 2001-06-01 Stainless steel needle tubing for the manufacture of medical devices. FDA Standards Recognition Number 6-163.
The luer lock connection that is molded as part of the plastic handle is in conformance with ISO 594/1 Conical fittings with a 6% (Luer) taper for syringes, needles and certain other medical equipment - Part 1: General requirements. FDA Standards Recognition Number 6-11.
The luer lock connection that is molded as part of the plastic handle is in conformance with ISO 594-2 Second edition 1998-09-01, Conical fittings with a 6% (Luer) taper for syringes, needles and certain other medical equipment - Part 2: Lock fittings. FDA Standards Recognition Number 6-129.
RANFAC CORP. | PO BOX 635 | 30 DOHERTY AVE | AVON, MA 02322 | USA Phone: 508-588-4400 | 800-272-6322 | Fax: 508-584-8588 Email info@ranfac.com
www.ranfac.com
Intended Use:
## Technological Characteristics:
### Non-clinical Data: Standards
{7}------------------------------------------------
Image /page/6/Picture/0 description: The image shows a logo for a company called Ranfac. The logo is a semi-circular shape that is divided into two halves. The left half is filled with black, and the right half is white. The word "Ranfac" is written in a sans-serif font above the semi-circle.
The following standards apply to the sterilization of the finished device.
ISO 11135-1, Sterilization of health care products -Ethylene Oxide - Part1: Requirements for development, validation and routine control of a sterilization process for medical devices. FDA Recognition Number 14-331.
ISO 10993-7, Biological evaluation of medical devices -Part 7: Ethylene oxide sterilization residuals. FDA Recognition Number 14-278.
AAMI/ANSI/ISO 14161:2009, Sterilization of health care products - Biological indicators - Guidance for the selection, use and interpretation of results. FDA Recognition Number 14-285
ISO 11607-1:2006, Packaging for terminally sterilized medical devices - Part 1: Requirements for materials, sterile barrier systems and packaging systems. FDA Recognition Number 14-355
#### Not applicable
Clinical Data:
Conclusion:
Based on the similarities in materials, design, principles of function, biocompatibility and sterilization between the Ranfac Aspiration Needle with Adjustable Guide, subject of this premarket notification and the predicate devices, the Ranfac Aspiration Needle with Adjustable Guide has been shown to be substantially equivalent to the predicate devices under the Federal Food, Drug and Cosmetic Act.
RANFAC CORP. | PO BOX 635 | 30 DOHERTY AVE | AVON, MA 02322 | USA Phone: 508-588-4400 | 800-272-6322 | Fax: 508-584-8588 | Email info@ranfac.com
www.ranfac.com
since 1888
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.