MAMMOTOME REOLVE DUAL VACUUM ASSISTED BIOPSY (VAB) SYSTEM

K123259 · Devicor Medical Products, Inc. · KNW · Nov 15, 2012 · Gastroenterology, Urology

Device Facts

Record IDK123259
Device NameMAMMOTOME REOLVE DUAL VACUUM ASSISTED BIOPSY (VAB) SYSTEM
ApplicantDevicor Medical Products, Inc.
Product CodeKNW · Gastroenterology, Urology
Decision DateNov 15, 2012
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1075
Device ClassClass 2
Attributes3rd-Party Reviewed

Indications for Use

The Mammotome® revolve Dual Vacuum Assisted Biopsy (VAB) System is indicated to provide tissue samples for diagnostic sampling of breast abnormalities. The Mammotome® revolve Dual Vacuum Assisted Biopsy (VAB) System is intended to provide breast tissue for histologic examination with partial or complete removal of the imaged abnormality. The Mammotome® revolve Dual Vacuum Assisted Biopsy (VAB) System is intended to provide breast tissue for histologic examination with partial removal of a palpable abnormality. The extent of a histologic abnormality cannot always be readily determined from palpation or imaged appearance. Therefore, the extent of removal of the palpated or imaged evidence of an abnormality does not predict the extent of removal of a histologic abnormality, e.g., malignancy. When the sampled abnormality is not histologically benign, it is essential that the tissue margins be examined for completeness of removal using standard surgical procedures. In instances when a patient presents with a palpable abnormality that has been classified as benign through clinical and/or radiological criteria (e.g., fibroadenoma, fibrocystic lesion), the Mammotome® revolve Dual Vacuum Assisted Biopsy (VAB) System may also be used to partially remove such palpable lesions. Whenever breast tissue is removed, histological evaluation of the tissue is the standard of care. When the sampled abnormality is not histologically benign, it is essential that the tissue margins be examined for completeness of removal using standard surgical procedures.

Device Story

Mechanical breast biopsy system; utilizes vacuum-assisted technology for tissue acquisition. System components: sterile single-use probe (trocar-tipped needle, rotating cutter, specimen collection chambers), reusable holster (drive motors, gear trains), and reusable control unit (vacuum pump, power supply, touchscreen interface). Operates via vacuum and rotational/translational cutting to extract tissue samples for histologic examination. Used in clinical settings (stereotactic, ultrasound, MRI, molecular imaging) by healthcare providers. Control unit manages vacuum, valve actuators, and cutter positioning. Automated specimen retrieval transports tissue to collection chambers. Benefits include efficient diagnostic sampling of breast abnormalities; provides tissue for histologic evaluation to determine malignancy.

Clinical Evidence

Bench testing only. In vivo porcine model used to compare tissue sample collection performance (sample weight, reliability, quality) against the predicate device. Results confirmed the subject device retrieves tissue samples comparable to the predicate.

Technological Characteristics

System components: sterile single-use probe, reusable holster, reusable control unit. Materials: stainless steel (needle, cutter), plastic (housing), PVC (tubing, DEHP-free). Energy: electrical (control unit), vacuum (pump). Connectivity: touchscreen, remote keypad, footswitch. Sterilization: Cobalt 60 irradiation. Software: closed-loop control for cutter position, automated biopsy functions.

Indications for Use

Indicated for patients requiring diagnostic breast tissue sampling of imaged or palpable abnormalities. Applicable for partial or complete removal of imaged abnormalities and partial removal of palpable abnormalities (e.g., fibroadenoma, fibrocystic lesions).

Regulatory Classification

Identification

A gastroenterology-urology biopsy instrument is a device used to remove, by cutting or aspiration, a specimen of tissue for microscopic examination. This generic type of device includes the biopsy punch, gastrointestinal mechanical biopsy instrument, suction biopsy instrument, gastro-urology biopsy needle and needle set, and nonelectric biopsy forceps. This section does not apply to biopsy instruments that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ 510(k) Premarket Notification Mammotome® revolve Dual Vacuum Assisted Biopsy System Devicor Medical Products, Inc. August 13, 2012 K 123259 ### 510(k) Summary # NOV 1 5 2012 The following information is provided as required by 21 CFR § 807.87 for the Mammotome® revolve Vacuum Assisted Biopsy (VAB) System 510(k) premarket notification. In response to the Safe Medical Devices Act of 1990 the following is a summary of the safety and effectiveness information upon which the substantial equivalence determination is based. ### Company: Devicor® Medical Products, Inc. 300 E-Business Way, Fifth Floor Cincinnati, OH 45241 Establishment Registration Number: 3008492462 ### Contact: Shawna Rose Director, Regulatory Affairs Devicor Medical Products, Inc. 300 E-Business Way, Fifth Floor Cincinnati, OH 45241 Ph: 513-864-9178 Fax: 513-864-9011 E-mail: srose@mammotome.com Date of Submission: August 13, 2012 Proprietary Name: Mammotome® revolve Dual Vacuum Assisted Biopsy (VAB) System Common Name: Biopsy System Regulation: 21 CFR 876.1075 Regulatory Class: II Product Codes: KNW Classification Name: Biopsy Instrument Predicate Device: Mammotome® Biopsy System, K033700 Device Description: The Mammotome® revolve Dual Vacuum Assisted Biopsy (VAB) System is a mechanical breast biopsy device indicated to provide tissue samples for diagnostic sampling of beast abnormalities for histologic examination. {1}------------------------------------------------ K123259 2/6 The Mammotome® revolve Dual VAB System is comprised of three primary subsystems: - 1) a sterile, single-use probe - 2) a reusable holster, and - 3) a reusable control unit. Intended Use: The Mammotome® revolve Dual Vacuum Assisted Biopsy (VAB) System is indicated to provide tissue samples for diagnostic sampling of breast abnormalities. - The Mammotome® revolve Dual Vacuum Assisted Biopsy (VAB) System is intended to provide breast tissue for histologic examination with partial or complete removal of the imaged abnormality. - The Mammotome® revolve Dual Vacuum Assisted Biopsy (VAB) System is intended to � provide breast tissue for histologic examination with partial removal of a palpable abnormality. The extent of a histologic abnormality cannot always be readily determined from palpation or imaged appearance. Therefore, the extent of removal of the palpated or imaged evidence of an abnormality does not predict the extent of removal of a histologic abnormality, e.g., malignancy. When the sampled abnormality is not histologically benign, it is essential that the tissue margins be examined for completeness of removal using standard surgical procedures. In instances when a patient presents with a palpable abnormality that has been classified as benign through clinical and/or radiological criteria (e.g., fibroadenoma, fibrocystic lesion), the Mammotome® revolve Dual Vacuum Assisted Biopsy (VAB) System may also be used to partially remove such palpable lesions. Whenever breast tissue is removed, histological evaluation of the tissue is the standard of care. When the sampled abnormality is not histologically benign, it is essential that the tissue margins be examined for completeness of removal using standard surgical procedures. ### Technological Characteristics: The Mammotome® revolve Dual Vacuum Assisted Biopsy (VAB) System, facilitates the diagnostic removal of tissue through a combination of vacuum and rotational/translational cutting functions. The Mammotome® revolve Dual Vacuum Assisted Biopsy (VAB) System utilizes the same primary subsystems as identified in the predicate device to achieve its intended use: - a sterile single-use probe containing trocar tipped biopsy needle, rotating cutter, 1) specimen collection chambers, and vacuum tubing/valving which enables tissue sample {2}------------------------------------------------ 123259 acquisition under vacuum, tissue cutting, and tissue transport to the specimen collection chambers; 3/6 - 2) a reusable holster containing the drive motors, gear trains and user activation switches, which provides the basic functionality for the procedure including the rotation, advancement and retraction of the cutting mechanism of the probe, as well as the specimen management system rotation; and - 3) a reusable control unit that houses the vacuum pump, power supply, valve actuators, software, user interface touchscreen and control electronics, which provides the primary electrical, electronic, mechanical and vacuum control of the system. In addition, several accessories are available including remote keypad and footswitch controls, a transport cart and probe guides for various stereotactic (ST) tables. The Mammotome® revolve Dual VAB System is an updated version of the currently marketed system, incorporating changes to improve efficiency and ease of use, while maintaining the basic technology, functionality and clinical outcomes of the predicate system. Consistent with the configuration of the currently marketed system, the Mammotome® revolve VAB System will be configured for use in multiple imaging modalities, including Stereotactic, Ultrasound, Magnetic Resonance and Advanced Molecular Imaging. The system relies on software to operate many of its functions including utilization of closed loop control on cutter position. As with the predicate system, the focus of the software is to aid in system set-up and facilitate biopsy functions. The Mammotome® revolve Dual VAB System software has been updated to simplify the touch screen display used to set up the system and create a single button push Biopsy function that transports the tissue sample back to a collection system for retrieval and analysis. A side-by-side comparison of the marketed and proposed devices is provided below. {3}------------------------------------------------ 123259 416 ### Side-by-Side Comparison to Legally Marketed Device | Device Characteristics | Marketed Device Mammotome® ST<br>System | Proposed Device Mammotome®<br>revolve Biopsy System | |-------------------------------------------------------|-------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------| | Indications for Use | Breast Biopsy/complete, partial<br>removal of imaged abnormality/partial<br>removal of palpated abnormality | Breast Biopsy / complete, partial<br>removal of imaged abnormality/partial<br>removal of palpated abnormality | | Needle Assembly | | | | Needle Insertion Method | Fired or Manual | Fired or Manual | | Tip Type | Bladed trocar | Bladed trocar | | Tip Material | Stainless steel | Stainless steel | | Needle Configuration | Dual lumen | Dual lumen | | Needle Material | Stainless steel | Stainless steel | | Sample Aperture<br>Configuration | Lateral aperture in needle | Lateral aperture in needle | | Needle Diameter Sizes | 8G, 11G, and 14G | 8 G & 10 G | | Needle Rotation Method | Manual, Remote Thumbwheel | Manual, Remote Thumbwheel | | Probe Housing | | | | Vacuum Port<br>Attachment (Valve<br>Cartridge) | Yes, tethered to control module | Yes, tethered to Control Module | | Specimen Retrieval /<br>Collection Method | Manual | Automatic | | Housing Material | Plastic | Plastic | | Probe Cutter Subassembly | | | | Cutting Method | Rotation and translation of inner cutter | Rotation and translation of inner cutter | | Tissue Transport Method | Mechanical and Vacuum | Vacuum | | Cutter Material | Stainless steel | Stainless steel | | Cutter Sealing<br>Mechanism | Yes | Yes | | Cutter Sealing Material | Elastomer/Thermoplastic | Elastomer/Thermoplastic | | Knockout Sub Assembly | | | | Knockout Material | Stainless steel | | | Knockout Sealing<br>Mechanism | Yes | No Knockout / Specimen Retrieval | | Knockout Sealing<br>Material | Elastomer/Thermoplastic | Automatic | | Fluid Injection Capability | Yes | | | Packaging | | | | Type | C-film with Tyvek Cover | PETG with Tyvek Cover | | Device Characteristics | Marketed Device Mammotome® ST System | Proposed Device Mammotome® revolve Biopsy System | | Marking Access | Through Needle | Through Device | | Sterilization | | | | Method | Cobalt 60 Irradiation | Cobalt 60 Irradiation | | Tubing Set | | | | Physical Properties | | | | Tubing Material | PVC | PVC with no DEHP | | Features | | | | Attachment to Probe | Luer | Luer | | Attachment to control Module | Cartridge with two parallel vacuum lines leading to probe | Cartridge with two parallel vacuum lines leading to probe + Saline line leading to probe | | Vacuum Control | Solenoid activated stopcocks | Motor activated Rotary Valves | | Sterilization | | | | Method | Cobalt 60 Irradiation | Cobalt 60 Irradiation | | Holster/Cables / Control Module | | | | Features | | | | User interface Mechanisms | Buttons on holster, Pedals on Footswitch or Touchscreen | Buttons on holster, Footswitch Pedals, Touchscreen, Keypad | | Microprocessor and software with Upgrade Capabilities | Yes | Yes | | Display | Yes, LCD display | Yes, LCD display | | Translational Cutter Movement | Automatic or Semi Automatic | Automatic | | Rotational Cutter Movement | Automatic with cutter advancement | Automatic with cutter advancement | | Rotational and Translation Speed control | Yes; Closed-loop control | Yes, Closed loop control | | Automated Initialization of Disposable Probe | Yes | Yes | | Probe firing capability | Yes | Yes | | Motor location | Control Module | Holster | | Drive Train Type | Flexible Mechanical Drive Cable | On board Motor and Gear Train | | Independent Lateral and Axial Vacuum System | Yes | Yes | | Remote Footswitch/Keypad Capability | Yes | Yes | {4}------------------------------------------------ 510(k) Premarket Notification Mammotome® revolve Dual Vacuum Assisted Biopsy System Assisted Biopsy System K123259 5/6 {5}------------------------------------------------ 510(k) Premarket Notification Mammotome® revolve Dual Vacuum Assisted Biopsy System Devicor Medical Products, Inc. August 13, 2012 K123259 6/6 ### Performance testing: To demonstrate substantial equivalence of the proposed device to the predicate device, sideby-side comparison of tissue sample collection, using an in vivo porcine model, was performed. This model has historically been used to evaluate the ability for Mammotome® breast biopsy devices to collect tissue samples. The Mammotome® revolve Dual Vacuum Assisted Biopsy (VAB) System and the predicate device were each used to obtain tissue sample was evaluated against the following criteria: - . Sample weight - Sampling reliability ● - Sample quality . Testing results confirmed that the Mammotome® revolve Vacuum Assisted Biopsy (VAB) System would retrieve a tissue sample comparable to that of the predicate device. Conclusion: The claim of substantial equivalence of the Mammotome® revolve Dual Vacuum Assisted Biopsy (VAB) System to the predicate system is based on the comparison of the intended use, product technical characteristics, and performance characteristics. {6}------------------------------------------------ Image /page/6/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three overlapping wing segments, representing health, human services, and the partnership between the federal government and the states. The eagle is positioned to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA", which is arranged in a circular fashion around the left side of the logo. ### DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002 November 15, 2012 Underwriters Laboratories, Incorporated % Mr. Ned Devine Senior Staff Engineer 333 Pfingsten Road Northbrook Road, Illinois 60062 Re: K123259 Trade/Device Name: Mammotome® revolve Dual Vacuum Assisted Biopsy (VAB) System Regulation Number: 21 CFR 876.1075 Regulation Name: Gastroenterology-urology-urology biopsy instrument Regulatory Class: II Product Code: KNW Dated: October 16, 2012 Received: October 18, 2012 Dear Mr. Devine: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set {7}------------------------------------------------ ### Page 2 - Mr. Ned Devine forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, ## Peter D. Rumm -S Mark N. Melkerson Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {8}------------------------------------------------ ### Indications for Use Statement 510(k) Number: K 123259 Device Name: Mammotome® revolve Dual Vacuum Assisted Biopsy (VAB) System Indications for Use: The Mammotome® revolve Dual Vacuum Assisted Biopsy (VAB) System is indicated to provide tissue samples for diagnostic sampling of breast abnormalities. - The Mammotome® revolve Dual Vacuum Assisted Biopsy (VAB) System is intended to . provide breast tissue for histologic examination with partial or complete removal of the imaged abnormality. - The Mammotome® revolve Dual Vacuum Assisted Biopsy (VAB) System is intended to . provide breast tissue for histologic examination with partial removal of a palpable abnormality. The extent of a histologic abnormality cannot always be readily determined from palpation or imaged appearance. Therefore, the extent of removal of the palpated or imaged evidence of an abnormality does not predict the extent of removal of a histologic abnormality, e.g., malignancy. When the sampled abnormality is not histologically benign, it is essential that the tissue margins be examined for completeness of removal using standard surgical procedures. In instances when a patient presents with a palpable abnormality that has been classified as benign through clinical and/or radiological criteria (e.g., fibroadenoma, fibrocystic lesion), the Mammotome® revolve Dual Vacuum Assisted Biopsy (VAB) System may also be used to partially remove such palpable lesions. Whenever breast tissue is removed, histological evaluation of the tissue is the standard of care. When the sampled abnormality is not histologically benign, it is essential that the tissue margins be examined for completeness of removal using standard surgical procedures. Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) | PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) | | |-------------------------------------------------------------------------|--| | | | oncurrence of CDRH, Office of Device Evaluation (ODE) I, Orthopedic, (Division Sign-Off Division of Surgical and Restorative Devices 510(k) Number 23,255 ين Page 5-2
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...