K033700 · Ethicon Endo-Surgery, Inc. · KNW · Feb 20, 2004 · Gastroenterology, Urology
Device Facts
Record ID
K033700
Device Name
MAMMOTOME EX HAND HELD SYSTEM
Applicant
Ethicon Endo-Surgery, Inc.
Product Code
KNW · Gastroenterology, Urology
Decision Date
Feb 20, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.1075
Device Class
Class 2
Indications for Use
The Mammotome EX Hand Held System is intended to provide breast tissue for histologic examination with partial or complete removal of the imaged abnormality. The Mammotome EX Hand Held System is also intended to provide breast tissue for histologic examination with partial removal of a palpable abnormality.
Device Story
Mechanical breast biopsy device; consists of disposable bladed trocar-tipped probe, reusable holster, and reusable control module. Operates via vacuum and radial cutting functions to extract tissue samples. Used by clinicians for diagnostic sampling of breast abnormalities. Features microprocessor-controlled closed-loop system for cutter translation and rotation; motors relocated to holster for improved ergonomics. Provides standard and rapid operation modes for cutter advancement and specimen retrieval. Independent control of axial and lateral vacuum. Output is tissue sample for histologic examination. Benefits include easier handling and reduced dependence on user interaction or tissue variability during biopsy.
Clinical Evidence
Bench testing only. Preclinical testing demonstrated satisfactory performance in breast tissue biopsy.
Technological Characteristics
Mechanical breast biopsy system. Components: disposable bladed trocar-tipped probe, reusable holster, reusable control module. Vacuum-assisted radial cutting. Microprocessor-based closed-loop control for cutter translation and rotation. Powered by motors in holster. Independent axial and lateral vacuum control.
Indications for Use
Indicated for diagnostic sampling of breast abnormalities, including imaged abnormalities (microcalcifications, masses, spiculated masses, asymmetric densities, multi-focal disease, diffuse tissue) and palpable abnormalities. Used for partial or complete removal of imaged abnormalities and partial removal of palpable lesions. Not for determining extent of histologic abnormality; surgical margins must be examined if findings are not benign.
Regulatory Classification
Identification
A gastroenterology-urology biopsy instrument is a device used to remove, by cutting or aspiration, a specimen of tissue for microscopic examination. This generic type of device includes the biopsy punch, gastrointestinal mechanical biopsy instrument, suction biopsy instrument, gastro-urology biopsy needle and needle set, and nonelectric biopsy forceps. This section does not apply to biopsy instruments that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Predicate Devices
Mammotome® Hand Held System
Mammotome® Hand Held 8 Gauge Probe
Mammotome® Biopsy System
Submission Summary (Full Text)
{0}------------------------------------------------
# FEB 2 0 2004
510(k) Premarket Notification for Mammotome® EX Hand Held System
K033700
loF2
# Mammotome® EX Hand Held System
# 510(k) Summary of Safety and Effectiveness
#### Company
Ethicon Endo-Surgery, Inc. 4545 Creek Rd. Cincinnati, OH 45242
#### Contact
Carol Sprinkle, BSN, RN Associate, Regulatory Affairs
#### Date Prepared:
November 24, 2003
## Name of Device
Trade Name: Mammotome® EX Hand Held System Classification Name: Biopsy Needle
## Predicate Devices:
Mammotome® Hand Held System Mammotome® Hand Held 8 Gauge Probe Mammotome® Biopsy System
#### Device Description
The Mammotome® EX Hand Held System is a mechanical breast biopsy device used in incisional breast biopsy of microcalcifications, masses, spiculated masses, asymmetric densities, multi-focal discase, and diffused tissue.
The Mammotome EX Hand Held System consists of three major components, a disposable bladed trocar tipped needle-like probe, a reusable holster, and a reusable control module. The following accessories are also provided with the system: disposable vacuum tubing set and canister, footswitch, remote keypad, support arm and cart.
#### Intended Use
The Mammotome EX Hand Held System is intended to provide breast tissue for histologic examination with partial or complete removal of the imaged abnormality. The Mammotome EX Hand Held System is also intended to provide breast tissue for histologic examination with partial removal of a palpable abnormality.
{1}------------------------------------------------
#### Ethicon Endo-Surgery, Inc.
510(k) Premarket Notification for Mammotome® FX Hand Held System
K033700
20F-2
### Technological Characteristics
The Mammotome EX Hand Held System is a modification of the currently marketed Mammotome Hand Held System. It represents a refinement in design, software and crgonomic enhancements for ease of use, however, the basic configuration, technology, materials and principles of operation of the proposed and marketed devices are equivalent.
The EX Hand Held biopsy device, used with or without imaging modalities, facilitates the diagnostic removal of tissue with fluid management through a combination of vacuum and radial cutting functions. The proposed and marketed devices contain the same primary components to achieve these functions: a probe, housing/holster, and a control module. The probe needle and cutter, which interface directly with the patient, are similar in both new and marketed devices.
In the proposed device, the control module motors are bypassed and two mechanical cables are eliminated. Two small motors reside in the holster to power the cutter translation and rotation. These physical changes make the probe/holster easier to handle.
Software modifications allow the Mammotome EX Hand Held System to provide standard and rapid modes of operation for cutter advancement and specimen retrieval. A microprocessor provides closed-loop control to reduce dependence of cutter rotation and translation speed on user interaction or tissue variability.
Axial and lateral vacuum can still be controlled independently.
#### Performance Data
Preclinical testing was performed to ensure the device performs as intended. Testing demonstrated satisfactory performance in breast tissue biopsy.
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol resembling three abstract human profiles or a triple-stranded helix, often referred to as the "Human Services Symbol".
Public Health Service
FEB 2 0 2004
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Ms. Carol Sprinkle, BSN, RN Associate, Regulatory Affairs Ethicon Endo-Surgery, Inc. 4545 Creek Road Cincinnati, Ohio 45242
Re: K033700
Trade/Device Name: Mammotome® EX Hand Held System Regulation Number: 21 CFR 876.1075 Regulation Name: Gastroenterology-urology-biopsy instrument Regulatory Class: II Product Code: KNW Dated: November 24, 2003 Received: November 25, 2003
Dear Ms. Sprinkle:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the l'ederal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must or any I cathar all the Act's requirements, including, but not limited to: registration and listing (21 CHR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
# Page 2 - Ms. Carol Sprinkle, BSN, RN
This letter will allow you to begin marketing your device as described in your Section 510(k) I ms icher will anow you to cogin maing of substantial equivalence of your device to a legally prematics notification. The a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific acries in 1984 and 1) 594-4659. Also, please note the regulation entitled, Contact the Office of Commarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small other general informational and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Sincerely yours,
Mark N Milliken
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications for Use Statement
510(k) Number: K 033700
Device Name: Mammotome® EX Hand Held System
Indications for Use:
The Mammotome Biopsy System is indicated to provide tissue samples for diagnostic sampling of breast abnormalities.
- The Mammotome Biopsy System is intended to provide breast tissue for histologic 해 examination with partial or complete removal of the imaged abnormality.
- The Mammotome Biopsy System is intended to provide breast tissue for 보 histologic examination with partial removal of a palpable abnormality.
The extent of a histologic abnormality cannot always be readily determined from palpation or imaged appearance. Therefore, the extent of removal of the palpated or imaged evidence of an abnormality does not predict the extent of removal of a histologic abnormality, c.g., malignancy. When the sampled abnormality is not histologically benign, it is essential that the tissue margins be examined for completeness of removal using standard surgical procedures.
In instances when a patient presents with a palpable abnormality that has been classified as benign through clinical and/or radiological criteria (e.g., fibroadenoma, fibrocystic lesion), the Mammotome Biopsy System may also be used to partially remove such palpable lesions. Whenever breast tissue is removed, histological evaluation of the tissue is the standard of care. When the sampled abnormality is not histologically benign, it is essential that the tissue margins be examined for completeness of removal using standard surgical procedures.
for
Mark N Millerss
(Division Sign-Off)
Division of General, Restorative,
and Neurological Devices
510(k) Number K033700
(Please do not write below this line - continue on another page if necessary)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use: ________
Over-The-Counter-Usc: ________________________________________________________________________________________________________________________________________________________ OR
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.