XENON LIGHT SOURCE AUTO LIP 5123, MODEL 2123.011

K983628 · Richard Wolf Medical Instruments Corp. · GCT · Dec 11, 1998 · Gastroenterology, Urology

Device Facts

Record IDK983628
Device NameXENON LIGHT SOURCE AUTO LIP 5123, MODEL 2123.011
ApplicantRichard Wolf Medical Instruments Corp.
Product CodeGCT · Gastroenterology, Urology
Decision DateDec 11, 1998
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2

Indications for Use

Light Projector AUTO LP 5123 provides light for examination, diagnostic and therapeutic applications, in particular, in endoscopy. The Xenon Light Source is classified as Cardiac Floating (CF) which permits its use for cardiac procedures when used in conjunction with the proper instrumentation for entry into the cardiac system.

Device Story

Auto LP 5123 Xenon Light Projector provides illumination for endoscopes. Device utilizes 180W xenon arc lamp (approx. 5600K color temperature) to generate light, transmitted via light guide to endoscope. Illumination level is automatically controlled by video input. Used in clinical settings (e.g., OR, endoscopy suites) by physicians/clinicians. Device classified as Cardiac Floating (CF) for cardiac system entry. Output provides necessary visualization for diagnostic/therapeutic interventions, aiding clinical decision-making during endoscopic procedures.

Clinical Evidence

No clinical tests performed. Bench testing conducted per IEC601-1 (amendment 1 and 2), IEC601-1-1, and IEC601-2-18 standards.

Technological Characteristics

180 W xenon arc lamp; color temperature ~5600 K; service life ~500 hours. Type CF applied part per IEC601-1. Automatic illumination control via video input. Light output via light guide port.

Indications for Use

Indicated for patients undergoing endoscopic examination, diagnostic, or therapeutic procedures, including cardiac procedures when used with appropriate instrumentation.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ 문 Pg 1 of 2 353 Corporate Woods Parkway Vernon Hills, Illinois 60061 Phone: 847.913.1113 Fax: 847.913.1488 RICHARD MEDICAL INSTRUMENTS C # 510(k) Summary of Safety and Effectiveness | Submitter | | |--------------------------------------|----------------------------------------| | Company / Institution name: | Richard Wolf Medical Instruments Corp. | | Division name (if applicable): | N.A. | | Street address: | 353 Corporate Woods Parkway | | City: | Vernon Hills | | State/Province: | Illinois | | Country: | USA | | ZIP/Postal Code: | 60061 | | Contact name: | Mr. Robert L. Casarsa | | Contact title: | Quality Assurance Manager | | Date of Preparation | October 13, 1998 | | FDA establishment regulation number: | 14 184 79 | | Phone number (include area code): | (847) 913-1113 | | FAX number (include area code): | (847) 913-0924 | | Product Information: | | | Trade name: | Auto LP 5123 Xenon Light Projector | | Model number: | 5123.011 | | Common name: | Xenon Light Source | | Classification Name: | Xenon Light Source | Information on devices to which substantial equivalence is claimed: | 510(k) Number | Trade or proprietary or model name | Manufacturer | |---------------|------------------------------------|-----------------------| | 1 K952696 | 1 Xenon Light Source | 1 Richard Wolf M.I.C. | | 2 | 2 | 2 | | 3 | 3 | 3 | ### 1.0 Description Light Projector AUTO LP 5123 uses a xenon arc lamp to provide illumination for an endoscope via a light guide which is connected to the projector's light port. The illumination level is automatically controlled by video. 8 - 1 {1}------------------------------------------------ K 983628 pg 2 of 2 #### 2.0 Intended Use Light Projector AUTO LP 5123 provides light for examination, diagnostic and therapeutic applications, in particular, in endoscopy. The Xenon Light Source is classified as Cardiac Floating (CF) which permits its use for cardiac procedures when used in conjunction with the proper instrumentation for entry into the cardiac system. #### Technological Characteristics 3.0 The color temperature of the 180 W xenon arc lamp is about 5600 K. The service life is approximately 500 hours and will decline with time due to the number of starts. The light projector is a type CF applied part per IEC601-1. #### 4.0 Substantial Equivalence The submitted devices pose the same type of questions about safety and effectiveness as existing devices. The new technological characteristics have not diminished safety or effectiveness. The submitted devices are substantially equivalent to existing 510(k) devices sold by Richard Wolf. #### 5.0 Performance Data Independent laboratories tested the Xenon Light Source 5123 according to specified standards IEC601-1 amendment 1 and 2, IEC601-1-1, and IEC601-2-18. The certifications for UL and CE compliance are pending. #### 6.0 Clinical Tests No clinical tests performed. #### 7.0 Conclusions Drawn These devices are designed and tested to guarantee the safety and effectiveness when used according to the instruction manual. By: Robert J. Casassa Robert L. Casarsa Quality Assurance Manager Date: Oct 12 98 {2}------------------------------------------------ Image /page/2/Picture/0 description: The image shows a logo with a circular border containing the words "U.S. DEPARTMENT OF HEALTH & HUMAN SERVICES" in a sans-serif font. Inside the circle is a stylized symbol consisting of three overlapping human profiles or figures, arranged in a way that suggests movement or progression. The symbol is rendered in black, contrasting with the likely white or light background of the logo. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 DEC 1 1 1998 Mr. Robert L. Casarsa Quality Assurance Manager Richard Wolf Medical Instruments Corp. 353 Corporate Woods Parkway Vernon Hills, Illinois 60061 Re: K983628 Auto LP 5123 Xenon Light Projector Dated: October 13, 1998 Received: October 15, 1998 Regulatory Class: II 21 CFR 876.1500/Procode: 78 GCT Dear Mr. Casarsa: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address "http://www.fda.gov/cdrf/dsmaldsmamain.html". Sincerely yours Lillian Yin, Ph.D. Director, Division of Reproductive Director, Division of Reproductive, Abdominal, Ear, Nose and Throat and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ ndication 510(k) Number (if known):_____________________________________________________________________________________________________________________________________________________ Device Name:__Light Projector AUTO LP 5123 ## Intended Use: Light Projector AUTO LP 5123 provides light for examination, diagnostic and therapeutic applications, in particular, in endoscopy. The Xenon Light Source is classified as Cardiac Floating (CF) which permits its use for cardiac procedures when used in conjunction with the proper instrumentation for entry into the cardiac svstem. (PI.E.ASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concursence of CDRH Office of Device Evaluation (ODE) Voriel a. Seymon (Division Sign-Off) Division of Reproductive, Abdominal, ENT, and Radiological Devices 510(k) Number _
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...