K083840 · Msi Medserv International Deutschland GmbH · GCM · Sep 28, 2009 · Gastroenterology, Urology
Device Facts
Record ID
K083840
Device Name
MSI MEDSERV INTERNATIONAL-RIGID ENDOSCOPE
Applicant
Msi Medserv International Deutschland GmbH
Product Code
GCM · Gastroenterology, Urology
Decision Date
Sep 28, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.1500
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
MSI MedServ International Deutschland GmbH rigid endoscopes and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs and canals.
Device Story
Rigid endoscope; tubular device; provides access, illumination, and visualization of body cavities, hollow organs, and canals. Used with fiberoptic light source and carrier for illumination. Operated by clinicians in clinical settings. Facilitates observation or manipulation of internal anatomy. Benefits include minimally invasive access to body cavities for diagnostic or therapeutic procedures.
Clinical Evidence
No clinical data provided; substantial equivalence is based on design, construction, and material similarity to the predicate device.
Technological Characteristics
Rigid tubular endoscope; fiberoptic illumination system; materials and construction similar to predicate device (K950076).
Indications for Use
Indicated for use by clinicians to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
{0}------------------------------------------------
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
JUL 27 2015
MSI MedServ International Deutschland GmbH Ms. Christine Strobel Managing Director Escad Strasse 3 D-88630 Pfullendorf Germany
Re: K083840
Trade/Device Name: MSI MedServ International - Rigid Endoscope Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: GCM Dated (Date on orig SE ltr): September 11, 2009 Received (Date on orig SE ltr): September 11, 2009
Dear Ms. Strobel,
This letter corrects our substantially equivalent letter of September 28, 2009.
We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be
{1}------------------------------------------------
Page 2 -
found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Benjamin R. Fisher -S
Beniamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
FDA CORRENE
SEP 11 2009
510(k) Number (if known):
K 083840
OUTLINE OF
PROPOSED RESEARCH
Device Name:
MSI MedServ International - Rigid Endoscope
Indications for Use:
Rigid Endoscopes
MSI MedServ International Deutschland GmbH rigid endoscopes and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs and canals.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| | <div style="text-align:left;">for mkm</div> |
|---------------------|---------------------------------------------|
| (Division Sign-Off) | |
Division of Surgical, Orthopedic,
and Restorative Devices
| 510(k) Number | K083840 |
|---------------|---------|
|---------------|---------|
{3}------------------------------------------------
K683840
:・、
...
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
## i ·
## Section 4. 510(k) Summary :
| General Provisions | Submitter's<br>Name and Adress | MSI MedServInternational GmbH<br>Escad Str. 3<br>D-88630 Pfullendorf |
|------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------|
| | Contact Person | Mr. Karl-Heinz Puscher<br>49 7552-936-570<br>khpuscher@medservinternational.de |
| | Classification Name | Endoscopes and accessories |
| | Proprietary Name | MSI MedServInternational<br>Rigid endoscopes |
| Name of Predicate<br>Device | Predicate Device | 510 (k) Reference No.<br>Olympus America Inc.<br>K950076 |
| Device Description | A rigid endoscope is a tubular endoscopic device with any of<br>a group of accessory devices which attach to the endoscope<br>and is intended to provide access, illumination and allow<br>observation or manipulation of body cavities, hollow organs,<br>and canals. It is typically used with a Fiberoptic light source<br>and carrier to provide illumination. | |
| Intended Use | The rigid endoscopes and accessories of MSI MedServ<br>International Deutschland GmbH are devices used to provide<br>access, illumination, and allow observation or manipulation of<br>body cavities, hollow organs and canals. | |
| Summary of<br>Technological<br>Characteristics | The rigid endoscopes of MSI are similar in construction and<br>materials to the previously cleared endoscope K950076 of<br>Olympus, America. | |
| Summary of<br>substantial<br>equivalence | The rigid endoscopes of MSI are considered to be<br>substantially equivalent to the currently marketed Olympus<br>rigid endoscope based on a comparison of the intended<br>uses, designs and results of the testing and evaluations<br>performed. | |