← Product Code [GCJ](/submissions/GU/subpart-b%E2%80%94diagnostic-devices/GCJ) · K990414

# SMITH & NEPHEW HANDPORT SYSTEM (K990414)

_Smith & Nephew, Inc. · GCJ · Apr 14, 1999 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/GU/subpart-b%E2%80%94diagnostic-devices/GCJ/K990414

## Device Facts

- **Applicant:** Smith & Nephew, Inc.
- **Product Code:** [GCJ](/submissions/GU/subpart-b%E2%80%94diagnostic-devices/GCJ.md)
- **Decision Date:** Apr 14, 1999
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

The Smith & Nephew HandPort System is intended to provide abdominal access to the surgeon's hand while preserving pneumoperitoneum during laparoscopic surgery. The retractor may also be used independently to provide incision retraction and to protect the wound from contamination.

## Device Story

HandPort System provides abdominal access for surgeon's hand during laparoscopic surgery while maintaining pneumoperitoneum; enables tactile feedback and improved manipulation capacity similar to open surgery. Device components include base retractor, sleeve, bracelet, retractor cap (with/without instrument portal), and sterile lubricant. Used in OR by surgeons; facilitates hand-assisted procedures and large specimen extraction. Retractor component also functions independently for incision retraction and wound protection. Benefits include improved surgical access and specimen removal capabilities.

## Clinical Evidence

Evidence includes bench testing confirming pneumoperitoneum maintenance, animal series testing confirming sufficient hand access and pneumoperitoneum maintenance, and a prospective, multicenter clinical trial confirming the device's capability in facilitating hand-assisted surgery across a range of procedures.

## Technological Characteristics

Sterile, single-use laparoscopic access system. Components: base retractor, sleeve, bracelet, retractor cap, and sterile lubricant. Mechanical design maintains pneumoperitoneum while allowing manual access. No electronic or software components.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

## Predicate Devices

- Dexterity PneumoSleeve ([K962147](/device/K962147.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

## K990414

## Section 9- 510k Summary of Safety and Effectiveness

------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

| 9.1<br>Statement                                  | This summary of 510k safety and effectiveness information is being submitted in<br>accordance with the requirements of SMDA 1990 and CFR 807.92                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
|---------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 9.2<br>Submitter                                  | Smith and Nephew Endoscopy, Inc.<br>Endoscopy Division<br>130 Forbes Boulevard<br>Mansfield, Ma. 02048                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| 9.3<br>Company<br>Contact                         | Amy Walters<br>Group Manager, RA/QA/Clinical<br>(508) 261-3776                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| 9.4<br>Device<br>Name                             | Proprietary Name: HandPort System<br>Common Name: Extended Laparoscopic Device and Accessories<br>Classification Name: Endoscopic accessories(78 GCJ)<br>Laparoscopic accessories (85 HET)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| 9.5<br>Predicate<br>Legally<br>Marketed<br>Device | Dexterity PneumoSleeve (K962147), cleared on July 9, 1996                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| 9.6<br>Device<br>Description                      | The HandPort System provides access to the abdominal cavity while maintaining<br>pneumoperitoneum; the laparoscopic surgeon regains the tactile sense and<br>feedback along with the increased hand-eye and instrumentation manipulation<br>capacity of open surgery.<br>The standard HandPort System device consists of the following components:<br>1. Base retractor,<br>2. Sleeve,<br>3. Bracelet,<br>4. Retractor Cap, with or without Instrument Portal, and<br>5. Sterile Lubricant                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| 9.7<br>Intended Use                               | The Smith & Nephew HandPort System is intended to provide abdominal access<br>to the surgeon's hand while preserving pneumoperitoneum during laparoscopic<br>surgery. The retractor may be used independently to provide incision retraction<br>and to protect the wound from contamination.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| 9.8 Device<br>Indications                         | The Smith & Nephew HandPort System is indicated for use in laparoscopic<br>procedures, where entry of the surgeon's hand may facilitate the procedure, and<br>for extraction of large specimens.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| 9.9<br>Substantial<br>Equivalence                 | Predicate Device Comparison:<br>The HandPort System is substantially equivalent to the currently marketed<br>Extended Laparoscopic Device: Dexterity PneumoSleeve. The devices are similar<br>in materials, design, intended use, and indications for use:<br>Both devices are sterile, and intended for single use. Both devices are intended to provide abdominal access to the surgeon's<br>hand while preserving pneumoperitoneum during laparoscopic surgery. With both devices, the retractor may also be used independently to<br>provide incision retraction and to protect the wound from<br>contamination during both laparoscopic and open surgery.          Testing Summary:<br>Bench testing confirmed that the HandPort System reliably maintains<br>pneumoperitoneum under simulated use conditions.<br>Simulated use testing in an animal series confirmed that the HandPort System<br>provides sufficient hand access to anatomical structures while maintaining<br>pneumoperitoneum for the duration of the surgical procedure.<br>A prospective, multicenter clinical trial confirmed the capability of the HandPort<br>System in facilitating hand-assisted surgery in a range of surgical procedures. |

{1}------------------------------------------------

**Applicant** Amy Walters
**Date** 3/29/99

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing right, with flowing lines representing hair or clothing.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

APR 1 4 1939

Ms. Amy L. Walters Group Manager, Regulatory Affairs, Quality Assurance and Clinical Smith & Nephew, Inc. Endoscopy Division 130 Forbes Boulevard Mansfield, Massachusetts 02048

Re: K990414

Trade Name: Smith & Nephew HandPort System Regulatory Class: II Product Code: GCJ Dated: February 5, 1999 Received: February 10, 1999

Dear Ms. Walter:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{3}------------------------------------------------

Page 2 - Ms. Amy L. Walters

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

Celia M. Witten, Ph.D., M.D.

Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

510 (k) Number (If Known): K990414

Device Name: Smith & Nephew HandPort System

Indications For Use:

The Smith & Nephew HandPort System is indicated for use in laparoscopic procedures, where entry of the surgeon's hand may facilitate the procedure, and for extraction of large specimens.

## Intended Use:

The Smith & Nephew HandPort System is intended to provide abdominal access to the surgeon's hand while preserving pneumoperitoneum during laparoscopic surgery. The retractor may also be used independently to provide incision retraction and to protect the wound from contamination.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use (Per 21 CFR 801.109)

Over-The-Counter Use

(Division Division of General Re 510(k) Number

---

**Source:** [https://fda.innolitics.com/submissions/GU/subpart-b%E2%80%94diagnostic-devices/GCJ/K990414](https://fda.innolitics.com/submissions/GU/subpart-b%E2%80%94diagnostic-devices/GCJ/K990414)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).