MIST PANORAMIC LAPAROSCOPE A RIGID LAPAROSCOPE

K971150 · Mist, Inc. · GCJ · Sep 23, 1997 · Gastroenterology, Urology

Device Facts

Record IDK971150
Device NameMIST PANORAMIC LAPAROSCOPE A RIGID LAPAROSCOPE
ApplicantMist, Inc.
Product CodeGCJ · Gastroenterology, Urology
Decision DateSep 23, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2

Indications for Use

The MIST Panoramic Laparoscope is intended to be used by trained physicians for the illumination and direct visualization of body cavities, hollow organs and canals during diagnostic and general surgical procedures.

Device Story

MIST Panoramic Laparoscope is a rigid, fiber-optic endoscope designed for illumination and direct visualization of body cavities, hollow organs, and canals. Device lacks a working channel. Used by trained physicians in clinical or surgical settings to facilitate diagnostic procedures and minimally invasive general surgeries. Provides visual access to internal anatomy, aiding physician assessment and surgical intervention. Sold non-sterile; requires cleaning, sterilization, and reuse by the healthcare facility.

Clinical Evidence

No clinical data provided; substantial equivalence is based on identical design and intended use to the predicate device.

Technological Characteristics

Rigid, fiber-optic laparoscope; no working channel; re-usable; sold non-sterile. Materials are identical to the predicate device.

Indications for Use

Indicated for illumination and direct visualization of body cavities, hollow organs, and canals for diagnostic purposes and minimally invasive general surgeries in patients requiring such procedures.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

Predicate Devices

Submission Summary (Full Text)

{0} SEP 23 1997 K 971150 mdi Exhibit #1 510(k) Summary Page 1 of 2 ## 510(K) Summary ## MIST Panoramic Laparoscope ### Submitter Information: Robert Hefter Vice President of Logistics Minimally Invasive Surgical Technologies (MIST) 3310 US 70 West Smithfield, NC 27577 Telephone Number: (919) 989-6478 ### 510(k) Summary Prepared By: Carolann Kotula MDI Consultants, Inc. 55 Northern Boulevard Great Neck, NY 11021 Telephone Number: (516) 482-9001 Date 510(k) Summary Prepared: March 27, 1997 ### Name of the Device: **Trade or Proprietary Name:** MIST Panoramic Laparoscope, various catalog numbers. **Common Name:** Rigid Laparoscope **Classification Name:** Laparoscope, General and Plastic Surgery, Endoscope and Accessories, including instruments (21 CFR Part 876.1500) Consulting since 1978 {1} mdi Exhibit #1 510(k) Summary Page 2 of 2 ## Identification of Legally Marketed Device to which the Submitter Claims Equivalence: The MIST Panoramic Laparoscope is identical in intended use and materials to the Comeg Endoscopes marketed under K862275. ## Description of the Subject Devices: The MIST Panoramic Laparoscope is a rigid, fiber-optic laparoscope. No working channel is provided. The devices are re-usable, and will be sold non-sterile with instruction for cleaning, sterilization, and re-use. ## Intended Use of the Subject Devices The MIST Panoramic Laparoscope is intended to be used by trained physicians for the illumination and direct visualization of body cavities, hollow organs and canals during diagnostic and general surgical procedures. ## Technological Characteristics of the Subject Devices There are no differences in the characteristics of the MIST Panoramic Laparoscope and the predicate devices. The MIST devices are identical to the predicate. Consulting since 1978 {2} DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20856 SEP 23 1997 MIST c/o Ms. Carolann Kotula Official Correspondent MDI Consultants, Inc. 55 Northern Boulevard, Suite 410 Great Neck, New York 11021 Re: K971150 Trade Name: MIST Panoramic Laparoscope, a rigid laparoscope Regulatory Class: II Product Code: GCJ Dated: July 2, 1997 Received: July 7, 1997 Dear Ms. Kotula: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for {3} Page 2 - Ms. Carolann Kotula devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for *in vitro* diagnostic dcvices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, ![img-0.jpeg](img-0.jpeg) Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4} EXHIBIT #9 510(k) Number (if known): K 971150 Device Name: Mist Panoramic Laparoscope Indications For Use: The MIST Panoramic Laparoscope is a rigid laparoscope intended for the illumination and direct visualization of body cavities, hollow organs and canals for the purpose of diagnosis of diseases and/or minimally invasive general surgeries. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) ![img-1.jpeg](img-1.jpeg) Prescription Use ☑ (Per 21 CFR 801.109) OR Over-The-Counter Use ☐ (Optional Format 1-2-96)
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...