1688 4K Camera Control Unit with Advanced Imaging Modality (1688010000)

K230216 · Stryker · GCJ · Feb 23, 2023 · Gastroenterology, Urology

Device Facts

Record IDK230216
Device Name1688 4K Camera Control Unit with Advanced Imaging Modality (1688010000)
ApplicantStryker
Product CodeGCJ · Gastroenterology, Urology
Decision DateFeb 23, 2023
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 876.1500
Device ClassClass 2
AttributesPediatric

Indications for Use

Endoscopic white light and near-infrared illumination and imaging during endoscopic procedures.

Device Story

1688 4K Camera System captures live surgical video via rigid/flexible endoscopes; utilizes CMOS image sensors to acquire continuous image data in visible and near-infrared (fluorescence/transillumination) spectra. System comprises Camera Control Unit (CCU), camera head, and coupler. CCU processes image data to generate 4K video stream for display. Used in OR by surgeons to visualize surgical field; aids real-time clinical decision-making during minimally invasive procedures. Benefits include enhanced visualization of anatomy and tissue perfusion via advanced imaging modalities.

Clinical Evidence

Bench testing only. No clinical data required. Testing included electrical safety (ANSI/AAMI ES60601-1, IEC 60601-2-18), electromagnetic compatibility (IEC 60601-1-2), CCU timing, video compatibility, and software verification (IEC 62304).

Technological Characteristics

4K (3840 x 2160) resolution, 60 fps digital video output. CMOS image sensor. Modes: White Light, NIR-fluorescence, NIR-transillumination. Components: CCU, camera head, coupler. Standards: IEC 60601-1, IEC 60601-1-6, IEC 60601-2-18, IEC 60601-1-2. Software verified per IEC 62304.

Indications for Use

Indicated for adults and pediatric patients undergoing general laparoscopy, nasopharyngoscopy, ear endoscopy, sinuscopy, neurosurgery, and plastic surgery where endoscopic visualization is required. Common procedures include cholecystectomy, hernia repair, appendectomy, pelvic lymph node detection, hysterectomy, spinal fusion, ACL reconstruction, arthroscopy, lung biopsy, and cardiac valve replacement. Used by general, pediatric, gynecological, cardiac, thoracic, plastic, orthopedic, ENT, neuro, and urological surgeons.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the seal of the Department of Health & Human Services. To the right of the seal is the FDA logo in blue, with the words "U.S. FOOD & DRUG" on one line and "ADMINISTRATION" on the line below. The logo is simple and professional, and it is easily recognizable. February 23, 2023 Stryker Michelle Stephens Senior Staff Regulatory Affairs Specialist 5900 Optical Ct San Jose, California 95138 Re: K230216 Trade/Device Name: 1688 4K Camera Control Unit with Advanced Imaging Modality Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope And Accessories Regulatory Class: Class II Product Code: GCJ, GWG Dated: January 25, 2023 Received: January 26, 2023 Dear Michelle Stephens: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's {1}------------------------------------------------ requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, # Jianting Wang -S Jianting Wang Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ### Indications for Use 510(k) Number (if known) #### K230216 Device Name 1688 4K Camera System with Advanced Imaging Modality (AIM) #### Indications for Use (Describe) The 1688 Video Camera is indicated for use in general laparoscopy, ear endoscopy, ear endoscopy, sinuscopy, neurosurgery and plastic surgery whenever a laparoscope/ arthroscope/ sinuscope/ sinuscope is indicated for use. The 1688 Video Camera is indicated for adults and pediatric patients. A few examples of the more common endoscope surgeries are Laparoscopic cholecystectomy, Laparoscopic hernia repair, Laparoscopic appendectomy, Laparoscopic pelvic lymph node detection, Laparoscopically assisted hysterectomy, Laparoscopic and thorascopic anterior spinal fusion, Anterior cruciate ligament reconstruction, Knee arthroscopy, Small joint arthroscopy, Decompression fixation, Wedge resection, Lung biopsy, Dorsal sympathectomy, Pleurodesis, Internal mammary artery dissection for coronary artery bypass grafting where endoscopic visualization is indicated and Examination of the evacuated cardiac chamber during performance of valve replacement. The users of the 1688 Video Camera are general and pediatric surgeons, gynecologists, cardiac surgeons, thoracic surgeons, plastic surgeons, orthopedic surgeons, neurosurgeons and urologists. | Type of Use (Select one or both, as applicable) | | |---------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------| | <div> <span> </span> <span>Prescription Use (Part 21 CFR 801 Subpart D)</span> </div> | <div> <span> </span> <span>Over-The-Counter Use (21 CFR 801 Subpart C)</span> </div> | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ### 510(k) Summary ### Submitter: | Applicant: | Stryker Endoscopy<br>5900 Optical Court<br>San Jose, CA 95138 | |-----------------|----------------------------------------------------------------------------------------------------------------------------------| | Contact Person: | Michelle Stephens<br>Senior Staff Regulatory Affairs Specialist<br>Email: michelle.stephens@stryker.com<br>Phone: (408) 754-2473 | | Date Prepared: | February 16, 2023 | #### Subject Device: | Name of Device: | 1688 4K Camera System with Advanced Imaging Modality | |-------------------------|-----------------------------------------------------------------------------------------------------------------| | Common or<br>Usual Name | 3-chip Video Camera | | Classification<br>Name: | Laparoscope, General and Plastic Surgery (21 C.F.R. §876.1500)<br>Endoscope, Neurological (21 C.F.R. §882.1480) | | Regulatory Class: | II | | Product Code: | GCJ<br>GWG | | 510(k) Review<br>Panel: | General & Plastic Surgery<br>Neurology | #### Predicate Device(s): | 780 nm 1688 4K Camera System with Advanced Imaging Modality<br>(primary predicate) | K214046 | |------------------------------------------------------------------------------------|---------| | 1688 4K Camera System with Advanced Imaging Modality<br>(reference device) | K222130 | #### Device Description: The 1688 4K Camera System with Advanced Imaging Modality is an endoscopic camera system that produces live video in the surgical field during surgical endoscopic procedures. The system is sensitive in the visible and infrared spectrum. The optical image is transferred from the surgical site to the camera head by a variety of rigid and flexible endoscopes, which are attached to the camera head. The 1688 4K Camera System consists of three main components: (1) a camera control unit (CCU); (2) a camera head with an integral cable that connects to the CCU; and (3) a coupler for attaching an endoscope to the camera head. {4}------------------------------------------------ ### Indications for Use: | Subject Device | Predicate Devices | | | |-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------| | 1688 4K Camera System<br>This Submission | 780 nm 1688 4K<br>Camera System<br>(K214046 –<br>primary predicate) | 1688 4K Camera<br>System with<br>Advanced Imaging<br>Modality<br>(K222130 –<br>reference device) | | | Intended Use: | Intended Use: | Intended Use: | | | Endoscopic white light and near-infrared illumination and imaging during endoscopic procedures. | Same as subject device | Same as subject device | | | Indications for Use: | Indications for Use: | Indications for Use: | | | The 1688 Video Camera is indicated for use in general laparoscopy, nasopharyngoscopy, ear endoscopy, sinuscopy, neurosurgery and plastic surgery whenever a laparoscope/ endoscope/ arthroscope/ sinuscope is indicated for use. The 1688 Video Camera is indicated for adults and pediatric patients.<br><br>A few examples of the more common endoscope surgeries are Laparoscopic cholecystectomy, Laparoscopic hernia repair, Laparoscopic appendectomy, Laparoscopic pelvic lymph node detection, Laparoscopically assisted hysterectomy, Laparoscopic and thorascopic anterior spinal fusion, Anterior cruciate ligament reconstruction, Knee arthroscopy, Small joint arthroscopy, Decompression fixation, Wedge resection, Lung biopsy, Pleural biopsy, Dorsal sympathectomy, Pleurodesis, Internal mammary artery dissection for coronary artery bypass, Coronary artery bypass grafting where endoscopic visualization is indicated and Examination of the evacuated cardiac chamber during performance of valve replacement.<br><br>The users of the 1688 Video Camera are general and pediatric surgeons, gynecologists, cardiac surgeons, thoracic surgeons, plastic surgeons, orthopedic surgeons, ENT surgeons, neurosurgeons and urologists | Same as subject device | Same as subject device | | | Item | Subject Device<br>1688 4K Camera System<br>(This Submission) | Predicate Devices | | | | | 780 nm 1688 4K Camera System<br>(K214046-primary<br>predicate) | 1688 4K Camera System<br>with Advanced Imaging<br>Modality<br>(K222130 - reference device) | | Manufacturer | Stryker | Same as subject device | Same as subject device | | Imaging<br>Modes | White Light<br>Near-infrared - fluorescence<br>Near-infrared –<br>transillumination | Same as subject device | Same as subject device | | Camera<br>System<br>Components | Camera Control Unit<br><br>Camera Head(s) – Standard,<br>Integrated, Inline, Pendulum,<br>Autoclavable<br><br>Coupler(s) - AIM 4K, AIM<br>4K Autoclavable | Camera Control Unit<br><br>Camera Head(s) – Standard,<br>Integrated, Inline, Pendulum<br><br>Coupler(s) - AIM 4K | Camera Control Unit<br><br>Camera Head(s) – Standard,<br>Integrated, Inline, Pendulum,<br>Autoclave<br><br>Coupler(s) - AIM 4K, AIM<br>4K Autoclave | | Principles of<br>Operations | Via an optical endoscope and<br>coupler, light is projected<br>from a light source onto one<br>or more complementary<br>metal oxide semiconductor<br>image sensors which acquire<br>a continuous stream of image<br>data. The image data is<br>processed to provide a video<br>stream that is then sent to a<br>display for viewing. | Same as subject device | Same as subject device | | Safety<br>Standards | IEC 60601-1<br>IEC 60601-1-6<br>IEC 60601-2-18<br>IEC 60601-1-2 | Same as subject device | Same as subject device | | Modes of<br>Operation | Alternate Frame processing<br>Simultaneous Frame<br>processing | Same as subject device | Same as subject device | | Image Sensor | CMOS image sensor | Same as subject device | Same as subject device | | Image<br>Processing/<br>Video Output<br>Resolution | Digital<br><br>4K (up to 3840 x 2160) | Same as subject device | Same as subject device | | Frame Rate | 60 frames per second | Same as subject device | Same as subject device | | Camera<br>Control Unit<br>PCB Digital<br>Board | Revision F<br><br>NOTE: Modified to account<br>for End of Life (EOL) and<br>supply chain shortages. | Revision E | Revision E | {5}------------------------------------------------ ## Comparison of Technological Characteristics with the Predicate Device: {6}------------------------------------------------ ### Performance Data: The following performance data were provided in support of the substantial equivalence determination: | Test | Method | Result | |----------------------------------------|------------------------------------------------------------------------------------------------------------------|--------| | Electrical Safety | In accordance with:<br>ANSI/AAMI ES60601-1:2005 + A1:2012<br>IEC 60601-2-18:2009<br>IEC 60601-1-6:2010 + A1:2013 | Pass | | Electromagnetic<br>Compatibility (EMC) | In accordance with IEC 60601-1-2:2014 | Pass | | Performance - Bench | CCU Timing<br>Video Compatibility | Pass | | Software Verification | In accordance with IEC 62304:2015 | Pass | NOTE: The 1688 4K Camera System is not patient contacting; therefore, biocompatibility testing was not required to support the determination of substantial equivalence. Additionally, the device modifications to the 1688 4K Camera System do not require clinical studies to support the determination of substantial equivalence. #### Conclusions: The 1688 4K Camera System is substantially equivalent in design, intended use, principles of operation, technological characteristics and safety features to the predicate devices. There are no different issues of safety and/or effectiveness introduced by the 1688 4K Camera System.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...