← Product Code [GCJ](/submissions/GU/subpart-b%E2%80%94diagnostic-devices/GCJ) · K221781

# Image Processing Unit, Model: OPTO-CAM104K, Camera Head, Model: OPTO-CHD104KH (K221781)

_Guangdong Optomedic Technologies, Inc. · GCJ · Feb 2, 2023 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/GU/subpart-b%E2%80%94diagnostic-devices/GCJ/K221781

## Device Facts

- **Applicant:** Guangdong Optomedic Technologies, Inc.
- **Product Code:** [GCJ](/submissions/GU/subpart-b%E2%80%94diagnostic-devices/GCJ.md)
- **Decision Date:** Feb 2, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

Image Processing Unit is intended to be used with compatible laparoscopes, camera heads, light sources, monitors, and other ancillary equipment for visualization, image recording and documentation during laparoscopic procedures. Endoscopic Camera Head is intended to be used with compatible laparoscopes, image processing unit, and other ancillary equipment for visualization, image recording and documentation during laparoscopic procedures.

## Device Story

Endoscopic camera system consisting of Image Processing Unit (IPU) and Camera Head; used in laparoscopic procedures. Camera head captures optical images via laparoscope; converts to electronic signals using internal image sensor. IPU acts as control center; processes signals; outputs real-time video to 4K monitors via DVI, HDMI, or SDI cables. System used in surgical settings by physicians; provides real-time visualization to assist surgical decision-making. Benefits include high-resolution imaging for improved anatomical visualization during minimally invasive surgery. Device is reusable; requires high-level disinfection between uses.

## Clinical Evidence

Bench testing only. No clinical data provided. Testing included electrical safety (ANSI/AAMI ES 60601-1), electromagnetic compatibility (IEC 60601-1-2), endoscopic equipment safety (IEC 60601-2-18), shipping container performance (ASTM D4169-16), and reprocessing validation (AAMI TIR 30:2016, AAMI TIR 12:2020).

## Technological Characteristics

System comprises IPU and Camera Head. Connectivity via LEMO interface; video output via DVI, HDMI, 12G-SDI, 3G-SDI. Power: 110-240V AC, 50/60Hz. Reusable, non-sterile; requires high-level disinfection. Software validated as Moderate Level of Concern. Standards: ANSI/AAMI ES 60601-1, IEC 60601-1-2, IEC 60601-2-18, ASTM D4169-16, AAMI TIR 30, AAMI TIR 12.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

## Predicate Devices

- Imagel SPIES ([K131953](/device/K131953.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Guangdong OptoMedic Technologies, Inc. Jane Guo Regulatory Affairs Manager Suite 503, Building A. Golden Valley Intellicreation Community. No. 2 Yonganbei Street Foshan, Guangdong 528200 China

February 2, 2023

Re: K221781

Trade/Device Name: Image Processing Unit, Model: OPTO-CAM104K; Camera Head, Model: OPTO-CHD104KH Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: Class II Product Code: GCJ Dated: January 4, 2023 Received: January 5, 2023

Dear Jane Guo:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Colin K. Chen -S

for

Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K221781

Device Name Image Processing Unit: OPTO-CAM104K; Camera Head: OPTO-CHD104KH

#### Indications for Use (Describe)

Image Processing Unit is intended to be used with compatible laparoscopes, camera heads, light sources, monitors, and other ancillary equipment for visualization, image recording and documentation during laparoscopic procedures.

Endoscopic Camera Head is intended to be used with compatible laparoscopes, image processing unit, and other ancillary equipment for visualization, image recording and documentation during laparoscopic procedures.

Type of Use (Select one or both, as applicable)

| <span style="font-size: 10pt;"> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)   </span> |
|--------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-size: 10pt;"> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)   </span>                 |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for OptoMedic. The logo consists of a green and yellow swirl on the left, followed by the word "OptoMedic" in black, bold font. The logo is simple and modern, and it is likely used to represent a company that specializes in eye care.

K221781

# 510(k) Summary

# Prepared in accordance with the requirements of 21 CFR Part 807.92

Date Prepared: July 13, 2022

## I. General Information

| 510(k) Submitter/Owner:            | Guangdong OptoMedic Technologies, Inc.                          |
|------------------------------------|-----------------------------------------------------------------|
|                                    | Suite 503, Building A, Golden Valley Intellicreation Community, |
|                                    | No. 2 Yonganbei Street, Daxu, Guicheng, Nanhai, Foshan,         |
|                                    | Guangdong, 528200, P.R. China                                   |
| Establishment Registration Number: | Not yet registered                                              |
| Primary Contact Person:            | Jane Guo                                                        |
|                                    | Regulatory Affairs Engineer                                     |
|                                    | Tel: +86 (757) 8670 2920                                        |
|                                    | Email: guoweijuan@optomedic.com                                 |
| Secondary Contact Person:          | Alice Lau                                                       |
|                                    | Regulatory Affairs Engineer                                     |
|                                    | Tel: +86 (757) 8670 2920                                        |
|                                    | Email: liuli@optomedic.com                                      |

## II. Device Identification

| Device Trade Name:    | Image Processing Unit; Camera Head                                |
|-----------------------|-------------------------------------------------------------------|
| Common or Usual Name: | Endoscopic Camera System                                          |
| Model:                | Image Processing Unit: OPTO-CAM104K<br>Camera Head: OPTO-CHD104KH |
| Regulation Name:      | Endoscopes and accessories                                        |
| Regulation Number:    | 21 CFR 876.1500                                                   |
| Regulatory Class:     | Class II                                                          |
| Product Code:         | GCJ                                                               |

# III. Predicate Device

510(k) Number: K131953

006_6.0_510(k)_Summary

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for OptoMedic. The logo consists of a green and yellow circular graphic on the left, followed by the text "OptoMedic" in a bold, sans-serif font. The "Opto" portion of the text is in a lighter shade of gray, while the "Medic" portion is in a darker, almost black shade of gray.

Product Name: Imagel SPIES

The predicate device has not been subject to a design-related recall.

### IV. Device Description

The Endoscopic Camera System comprises the Image Processing Unit, the Camera Head (including the objective lens), power supply cord and video cables. The Camera Head with integrated cable can be connected to the Image Processing Unit through the LEMO interface.

During surgical procedures, the Endoscopic Camera System is used to provide real-time visible imaging similar to that provided by conventional imaging system used in surgical endoscopy. The area of interest is illuminated with visible light from the light source and the resulting reflected light is collected by a variety of laparoscopes that are attached to the camera head via equipment coupler. The camera head receives optical image from the connected laparoscope and converts optical image into electronic signal using the incorporated image sensor. The Image Processing Unit serves as the control center of the system and should be connected to a compatible light source in order to provide visualization. It can be connected to a 4K video monitor via one of the video cables provided to display image in real time: DVI cable, HDMI cable, 12G-SDI cable, and 3G-SDI cable.

The proposed system is a reusable device and provided non-sterile. The Camera Head should be cleaned and high-level disinfected prior to the first use and after every subsequent use. The recommended compatible Light Source is OPTO-LED104K (510k exempted, product code: NTN).

#### V. Indications for Use

Image Processing Unit (hereafter referred to as "IPU") is intended to be used with compatible laparoscopes, camera heads, light sources, monitors, and other ancillary equipment for visualization, image recording and documentation during laparoscopic procedures.

Endoscopic Camera Head is intended to be used with compatible laparoscopes, image processing unit, and other ancillary equipment for visualization, image recording and documentation during laparoscopic procedures.

| Description                              | Subject Device                                                                                                                                                                                                                                                                                                        | Predicate Device (K131953)                                                                                                                                                                                              |
|------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Regulation Number                        | 21 CER 876.1500                                                                                                                                                                                                                                                                                                       | 21 CER 876.1500                                                                                                                                                                                                         |
| Product Code                             | GCJ                                                                                                                                                                                                                                                                                                                   | FET                                                                                                                                                                                                                     |
| Device class                             | Class II                                                                                                                                                                                                                                                                                                              | Class II                                                                                                                                                                                                                |
| Indication for use                       | Image Processing Unit<br>(hereafter referred to as “IPU”)<br>is intended to be used with<br>compatible laparoscopes,<br>camera heads, light sources,<br>monitors, and other ancillary<br>equipment for visualization,                                                                                                 | The Imagel SPIES is a camera control<br>unit (CCU) for use with camera heads or<br>video endoscopes for visualization,<br>image recording and documentation<br>during general endoscopic and<br>microscopic procedures. |
|                                          |                                                                                                                                                                                                                                                                                                                       |                                                                                                                                                                                                                         |
|                                          | image recording and<br>documentation during<br>laparoscopic procedures.<br>Endoscopic Camera Head is<br>intended to be used with<br>compatible laparoscopes,<br>image processing unit, and<br>other ancillary equipment for<br>visualization, image recording<br>and documentation during<br>laparoscopic procedures. |                                                                                                                                                                                                                         |
| Prescription/<br>Over-the-counter<br>use | Prescription                                                                                                                                                                                                                                                                                                          | Prescription                                                                                                                                                                                                            |
| Testing                                  | ANSI/AAMI ES 60601-1:<br>2005+A2 (R2012) +A1<br>IEC 60601-1-2:2014<br>IEC 60601-2-18:2009<br>ASTM D4169-16<br>AAMI TIR 30:2016<br>AAMI TIR 12:2020                                                                                                                                                                    | IEC 60601-1<br>IEC 60601-1-2                                                                                                                                                                                            |
| Dimensions                               | IPU:<br>422mm(L)×370mm(W)×149<br>mm(H)<br>Camera Head:<br>92mm(L)×39mm(W)×50mm(<br>H)                                                                                                                                                                                                                                 | Image1 S Connect TC200 :<br>30.5cm×5.4cm×32.0cm (W×H×D)<br>TH 100&TH 101:<br>100mm×48mm×38mm(L×H×W)<br>TH 103:<br>88mm×47mm×35mm(L×H×W)<br>TH 104:<br>100mm×49mm×39mm(L×H×W)                                            |
| Weight                                   | IPU: 9Kg±15%<br>Camera Head: 400(g)±20                                                                                                                                                                                                                                                                                | Image1 S Connect TC200: 2.1kg<br>Image1 S H3-Link TC300: 1.86kg<br>Image1 S X-Link TC301: 1.86kg<br>TH100: 270g<br>TH101: 270g<br>TH103: 227g<br>TH104: 299g                                                            |
| Power supply                             | 110-240V AC, 50/60Hz                                                                                                                                                                                                                                                                                                  | 100-120VAC, 50/60Hz<br>200-240VAC, 50/60Hz                                                                                                                                                                              |
| Video Output                             | 3G-SDI×4<br>HDMI ×1<br>DVI ×1                                                                                                                                                                                                                                                                                         | 2×DVI-D<br>1×3G-SDI                                                                                                                                                                                                     |
|                                          |                                                                                                                                                                                                                                                                                                                       |                                                                                                                                                                                                                         |
|                                          |                                                                                                                                                                                                                                                                                                                       | 12G-SDI ×1                                                                                                                                                                                                              |
| Signal<br>format                         | output                                                                                                                                                                                                                                                                                                                | 4096×2160p, 3840×2160p, 1920×1080p, 50/60Hz<br>1920×1080p,50Hz/60Hz                                                                                                                                                     |
| Sensor                                   |                                                                                                                                                                                                                                                                                                                       | 4K CMOS sensor assembly<br>3×1/3"                                                                                                                                                                                       |

# VI. Comparison of Technological Characteristics with The Predicate Device

Table 1 General Comparison

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the word "OptoMedic" in a bold, sans-serif font. The "O" in "Opto" is stylized with a green and yellow gradient circle that is open on the right side. The rest of the letters are black.

.

1

..............................................................................................................................................................................

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for OptoMedic. The logo consists of a green circular design on the left, followed by the text "OptoMedic" in black. The circular design has a gradient of green, with the top being a lighter shade and the bottom being a darker shade. The text is in a bold, sans-serif font.

The differences technological characteristics do not raise different questions of safety and effectiveness.

### VII. Performance data

Non clinical tests were conducted to verify that the proposed device met all design specifications as is Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:

- ANSI/AAMI ES 60601-1: 2005+A2 (R2012) +A1 Medical Electrical Equipment- Part 1: . General requirements for basic safety and essential performance.
- IEC 60601-1-2:2014, Medical electrical equipment- Part 1-2: General requirements for basic . safety and essential performance- Collateral standard: Electromagnetic compatibility-Requirements and tests.
- IEC 60601-2-18:2009 Medical electrical equipment Part 2-18: Particular requirements for . the basic safety and essential performance of endoscopic equipment
- ASTM D4169-16 Standard Practice for Performance Testing of Shipping Containers and . Systems.
- AAMI TIR 30:2016 A compendium of processes, materials, test methods, and acceptance . criteria for cleaning reusable medical devices
- AAMI TIR 12:2020 Designing, testing, and labeling reusable medical devices for reprocessing . in health care facilities: A guide for medical device manufacturers

The software of the proposed device was validated as Moderate level of concern (LoC) in accordance with the following guidance documents: Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices - Guidance for Industry and FDA Staff

Performance testing were also conducted on the subject device and demonstrate that the proposed system performs according to specifications and functions as intended.

#### VIII. Conclusions

The performance testing summarized above supports a substantial equivalence determination. The performance testing demonstrate that the subject device is as safe and as effective as the legally marketed predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/GU/subpart-b%E2%80%94diagnostic-devices/GCJ/K221781](https://fda.innolitics.com/submissions/GU/subpart-b%E2%80%94diagnostic-devices/GCJ/K221781)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
