Cyber Blade

K192499 · SIDAM S.r.l. · GCJ · Jul 31, 2020 · Gastroenterology, Urology

Device Facts

Record IDK192499
Device NameCyber Blade
ApplicantSIDAM S.r.l.
Product CodeGCJ · Gastroenterology, Urology
Decision DateJul 31, 2020
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2
AttributesTherapeutic

Intended Use

Cyber BLADE is intended for use under endoscopic visualization for the morcellation and removal of dissected benign prostatic hyperplasia (BPH) tissue during endoscopic surgical procedures in urology.

Device Story

Cyber Blade is a single-use endoscopic morcellator used by urologists to remove dissected BPH tissue. The device consists of a sterile single-use blade handpiece, a reusable internal motor with a battery pack, and an AC-powered charger. It operates via a cutting window created by the relative movement of internal and external tubes, driven by the motor. Vacuum intensity and cutting speed are adjustable via controls on the handpiece. The device requires an external suction system and a collection canister. By enabling efficient tissue removal under endoscopic visualization, it facilitates minimally invasive urological surgery, potentially reducing procedure time and improving patient outcomes compared to traditional removal methods.

Clinical Evidence

Bench testing only. Performance testing included morcellation efficiency (approx. 17g/min), suction testing, dimensional verification, tensile strength of the blade, battery testing, and comparative testing against the Richard Wolf morcellator. Compliance with IEC 60601-1, IEC 60601-1-2, and various ISO standards for biocompatibility (ISO 10993 series), sterilization (ISO 11135, 11737), and packaging (ISO 11607) was verified.

Technological Characteristics

Materials: Stainless steel. Principle: Mechanical morcellation via relative movement of internal/external tubes. Power: Battery-powered handpiece with AC-powered charger. Dimensions: 38 cm length, 0.47 cm external diameter. Connectivity: None. Sterilization: Ethylene oxide (for single-use morcellator). Software: Embedded firmware for motor control.

Indications for Use

Indicated for morcellation and removal of dissected benign prostatic hyperplasia (BPH) tissue during endoscopic urological surgical procedures.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG ADMINISTRATION" in blue as well. The FDA is a federal agency responsible for regulating and supervising the safety of food, drugs, and other products. July 31, 2020 SIDAM s.r.1. Andrea Tromba Regulatory Affairs Manager Via Statale Sud, 169 Mirandola (MO), 41037 Italy Re: K192499 Trade/Device Name: Cyber Blade Regulation Number: 21 CFR§ 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: II Product Code: GCJ Dated: June 18, 2020 Received: June 22, 2020 Dear Andrea Tromba: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. {1}------------------------------------------------ You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Martha W. Betz, Ph.D. Acting Assistant Director DHT3B: Division of Reproductive. Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K192499 Device Name Cyber Blade Indications for Use (Describe) Cyber BLADE is intended for use under endoscopic visualization and removal of dissected benign prostatic hyperplasia (BPH) tissue during endoscopic surgical procedures in urology. Type of Use (Select one or both, as applicable) 2 Prescription Use (Part 21 CFR 801 Subpart D) __ Over-The-Counter Use (21 CFR 801 Subpart C) #### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED. #### FOR FDA USE ONLY Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below. {3}------------------------------------------------ ### 5. 510(K) SUMMARY | Applicant /<br>Manufacturer<br>Name and Address: | SIDAM s.r.l.<br>Via Statale Sud, 169<br>41037 San Giacomo Roncole<br>MIRANDOLA (MO) - Italy | |-----------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------| | 510(k) Contact Person: | Andrea Tromba<br>Regulatory Affairs Manager<br>SIDAM s.r.l.<br>Email: andrea@sidamit.it<br>Phone: +39 0535.25523<br>Fax:<br>+39 0535.25635 | | Date Prepared: | 31/07/2020 | | Device Name: | Cyber Blade | | Classification: | Class II | | Classification Name: | Endoscopes and accessories | | Common name: | Laparoscope, General and Plastic Surgery | | Regulation Number: | 21 CFR 876.1500 | | Product Code: | GCJ | | Main Predicate device<br>Secondary predicate device | Versacut + Tissue Morcellator (K133272), Lumenis Ltd<br>Morce Scope Set 8970 (K041610), Richard Wolf Medical<br>Instruments Corp. | ### Performance Standards: There are no mandatory performance standards for this device. ### Description of the device: Cyber Blade provides efficient and rapid morcellation under endoscopic visualization. The cutting action is driven by the motor in the handpiece. Cyber Blade comprises: - a single use sterile blade fixed to a handpiece where controls are placed - a re-usable internal motor provided with integrated battery pack - an AC powered charger for the battery pack Cyber blade is also supplied with aspiration tubing and canister for the collection of evacuated tissue. {4}------------------------------------------------ | REF | Description | notes | |-----------|-----------------------------|------------------------| | 20041001 | Cyber Blade Morcellator | sterile single use | | 20042001 | Battery-powered motor | non sterile - reusable | | 20049001 | Charging station | 100-220V, 50/60Hz | | ABM000091 | Suction tube | sterile single use | | 077.0562 | Canister (tissue collector) | | Cyber Blade is comprehensive of the following parts: | Key Specifications, mechanism of action, device function | | |----------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------| | Principle of operation | morcellation through the cutting effect of the cutting window, produced by the relative movement of the internal tube and the external tube. | | Vacuum intensity regulation | Possible through the leverage on the morcellator | | Speed regulation | Possible through the button on the morcellator | | Morcellation velocity 1 | About 17 grams per minute | | Handpiece dimension (external tube) | 38 cm length<br>0.47 cm external diameter | | Materials in contact with the tissue | Stainless steel | | Power supply | The morcellator is battery powered.<br>The battery charger is AC power supplied. | | Supplied Sterile | Yes (only the morcellator) | | Re-usable | No. The morcellator is single use only.<br>Its internal motor is reusable | #### Indications for Use Cyber BLADE is intended for use under endoscopic visualization for the morcellation and removal of dissected benign prostatic hyperplasia (BPH) tissue during endoscopic surgical procedures in urology. #### Comparison of Technological Characteristics: The subject device and the identified predicate devices have comparable general intended use and technological characteristics. Any differences compared to the predicate devices do not present any new types of safety or effectiveness concern. A comparison of the technical characteristics between the subject and predicate devices follows: <sup>1</sup> The morcellated tissue coincides with the removed tissue, as all is aspired with the vacuum system {5}------------------------------------------------ | Specification | Subject device | main predicate device | Secondary predicate device | Comparison to Predicate | |-----------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------| | Device Name<br>(K #) | Cyber Blade | VERSACUT + TISSUE<br>MORCELLATOR (K133272) | MORCE SCOPE SET 8970<br>(K041610) | | | Submitter | SIDAM s.r.l. | LUMENIS LTD. | RICHARD WOLF MEDICAL<br>INSTRUMENTS CORP. | | | Principle of operation | morcellation through the cutting<br>effect of the cutting window,<br>produced by the relative<br>movement of the internal tube<br>and the external tube. | Morcellation through the cutting<br>effect of the cutting window,<br>produced by the relative<br>movement of the internal tube<br>and the external tube. | Morcellation through the cutting<br>effect of the cutting window,<br>produced by the relative<br>movement of the internal tube<br>and the external tube. | Same | | Vacuum intensity<br>regulation | Possible through the leverage on<br>the morcellator | Possible from the vacuum<br>generator | Possible from the vacuum<br>generator | equivalent | | Speed regulation | Possible through the button on<br>the morcellator | Possible from the vacuum<br>generator | Possible from the vacuum<br>generator | equivalent | | Handpiece dimension<br>(external tube) | 38 cm length<br>0.47 cm external diameter | 39.5 cm length<br>0.47 cm external diameter | 35 or 38 cm length<br>0.475 cm external diameter | In the same range | | Materials in contact<br>with the tissue | Stainless steel | Stainless steel | Stainless steel | equivalent | | Power supply | The morcellator is battery<br>powered.<br>The battery charger is AC power<br>supplied. | AC power supplied | AC power supplied | A device battery powered<br>raises less concerns compared<br>to a device AC powered | | Suction pump | Not part of the system, must use<br>an external suction system. | Part of the system | Part of the system | External FDA cleared suction<br>system to be used. | | Supplied Sterile | Yes | No | Yes | A device provided sterile raises<br>less concerns compared to a<br>device intended to be sterilized<br>before use. | | Re-usable | No. The morcellator is single use<br>only.<br>Its internal motor is reusable,<br>but it never comes in contact<br>with the patient | Yes | Available either single use and<br>reusable | A single use device raises less<br>concerns than a re-usable one. | ## Comparison of indications: as detailed in the comparison table below, the indications of the subject device are a sub-set of the indications of its predicate devices. | Specification | Subject device | main predicate device | additional predicate device | Comparison to Predicate | |----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Device Name<br>(K #) | Cyber Blade | VERSACUT + TISSUE<br>MORCELLATOR (K133272) | MORCE SCOPE SET 8970<br>(K041610) | | | Submitter | SIDAM s.r.l. | LUMENIS LTD. | RICHARD WOLF MEDICAL<br>INSTRUMENTS CORP. | | | Intended use | Cyber BLADE is intended for use<br>under endoscopic visualization<br>for the morcellation and<br>removal of dissected benign<br>prostatic hyperplasia (BPH)<br>tissue during endoscopic surgical<br>procedures in urology. | The VersaCut + Tissue<br>Morcellator is intended for use<br>under<br>direct or endoscopic<br>visualization for the<br>morcellation and removal<br>of dissected tissue during<br>pelviscopic, laparascopic,<br>percutaneous<br>and open surgical procedures<br>whenever access to the surgical<br>site is limited. | MORCE Scope Set 8970, in<br>conjunction with a<br>morcellation probe, and with<br>its sheaths and obturators, is<br>used in the cutting<br>(morcellation) and continuous<br>removal of large tissue masses.<br>In combination with the<br>corresponding auxiliary<br>instruments, it can be used as a<br>nephroscope in the<br>disintegration and<br>removal/aspiration of kidney<br>and bladder stones and<br>the removal of tumors via<br>percutaneous (kidney) or<br>transurethral (bladder)<br>passages, in conjunction with | Sub-set of the indications of<br>the predicate devices.<br>Endoscopy is a less invasive<br>access to the body compared<br>to laparoscopy or other<br>invasive techniques. | {6}------------------------------------------------ | Specification | Subject device | main predicate device | additional predicate device | Comparison to Predicate | |----------------------|----------------|--------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------| | Device Name<br>(K #) | Cyber Blade | VERSACUT + TISSUE<br>MORCELLATOR (K133272) | MORCE SCOPE SET 8970<br>(K041610) | | | Submitter | SIDAM s.r.l. | LUMENIS LTD. | RICHARD WOLF MEDICAL<br>INSTRUMENTS CORP. | | | | | | intracorporeal 5-4ithotripters<br>e.g. operated pneumatically,<br>by ultrasound, electro-<br>hydraulically or by laser, under<br>endoscopic control.<br>The POWER CONTROL 2303 in<br>conjunction with POWER STICK<br>M4 serves to drive WOLF<br>morcellators for the<br>continuous removal of ablated<br>tissue in endoscopic<br>operations.<br>The SUCTION PUMP is used for<br>aspirating irrigation fluid in<br>conjunction with a<br>resectoscope or a morcellator<br>following laser TURP. | | # Summary of Non-Clinical Performance Testing : Cyber Blade devices comply with the following recognized consensus standards or guidelines: | STANDARD TITLE or guidelines | standard<br>recognized<br>by FDA? | FDA<br>recognition<br>number | exclusions<br>from the<br>standard | |---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------|------------------------------|------------------------------------| | IEC 60601-1:2012, ed 3.1, Medical Electrical Equipment - Part 1:<br>General Requirements For Basic Safety And Essential Performance | Yes | 19-4 | No | | IEC 60601-1-2 Edition 4: 2014, Medical Electrical Equipment - Part 1-<br>2: General Requirements For Basic Safety And Essential Performance<br>- Collateral Standard: Electromagnetic Compatibility - Requirements<br>And Tests | Yes | 19-8 | No | | EN 556-1:2001/AC:2006 Sterilization of medical devices –<br>Requirements for medical devices to be designated "STERILE" – Part<br>1: requirements for terminally sterilized medical devices | No | N/A | No | | ISO 15223-1 Third Edition 2016-11-01 Medical devices - Symbols to<br>be used with medical device labels, labelling, and information to be<br>supplied - Part 1: General requirements | Yes | 5-117 | No | | EN 1041:2008 Information supplied by the manufacturer of medical<br>devices. | No | N/A | No | | ISO 10993-5 Third edition 2009-06-01 Biological evaluation of<br>medical devices - Part 5: Tests for in vitro cytotoxicity | Yes | 2-245 | No | | ISO 10993-1 Fourth edition 2009-10-15 Biological evaluation of<br>medical devices – Part 1: Evaluation and testing within a risk<br>management process [Including: Technical Corrigendum 1 (2010)] | Yes | 2-220 | No | | STANDARD TITLE or guidelines | standard<br>recognized<br>by FDA? | FDA<br>recognition<br>number | exclusions<br>from the<br>standard | | ISO 10993-7 Second edition 2008-10-15 Biological evaluation of<br>medical devices - Part 7: Ethylene oxide sterilization residuals<br>[Including: Technical Corrigendum 1 (2009)] | Yes | 14-408 | No | | ISO 10993-10 Third Edition 2010-08-01 Biological evaluation of<br>medical devices - Part 10: Tests for irritation and skin sensitization | Yes | 2-174 | No | | ISO 11135 Second edition 2014-07-15 Sterilization of health-<br>care products - Ethylene oxide - Requirements for the development,<br>validation and routine control of a sterilization process for medical<br>devices [Including: Amendment 1 (2018)] | Yes | 14-529 | No | | ISO 11138-1 Third edition 2017-03 Sterilization of health care<br>products - Biological indicators - Part 1: General requirements | Yes | 14-502 | No | | ISO 11607-1 First edition 2006-04-15 Packaging for<br>terminally sterilized medical devices - Part 1: Requirements for<br>materials, sterile barrier systems and packaging systems [Including:<br>Amendment 1 (2014)] | Yes | 14-454 | No | | ISO 11607-2 First edition 2006-04-15 Packaging for<br>terminally sterilized medical devices - Part 2: Validation<br>requirements for forming, sealing and assembly processes<br>[Including: Amendment 1 (2014)] | Yes | 14-455 | No | | ISO 11737-1 Third edition 2018-01 Sterilization of health care<br>products - Microbiological methods - Part 1: Determination of a<br>population of microorganisms on product | Yes | 14-514 | No | | ISO 14971 Second edition 2007-03-01 Medical devices -<br>Application of risk management to medical devices | Yes | 5-40 | No | | ISO 17664 Second edition 2017-10 Processing of health care<br>products - Information to be provided by the medical device<br>manufacturer for the processing of medical devices | Yes | 14-515 | No | | ISO 17664 Second Edition 2017-10 Processing Of Health Care<br>Products - Information To Be Provided By The Medical Device<br>Manufacturer For The Processing Of Medical Devices | Yes | 14-515 | No | | Software Verification and Validation Testing<br>Software verification and validation testing were conducted and<br>documentation was provided as recommended by FDA's Guidance<br>for Industry and FDA Staff, "Guidance for the Content of Premarket<br>Submissions for Software Contained in Medical Devices | - | - | - | {7}------------------------------------------------ The following additional testing were performed: morcellation efficiency, suction test, dimensional specification verification, tensile strength of blade, comparative testing with Richard Wolf morcellator, risk analysis on the performance test, battery testing. {8}------------------------------------------------ # Conclusion As the indications for use are a sub-set of its predicate devices and the principle of operation and specifications are equivalent, Cyber Blade is substantially equivalent to its predicate devices.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...