The Standard Clamp, Disposable is indicated for use in laparoscopic procedures to grasp, clamp, and manipulate soft tissues. The Clamp is indicated for use in bariatric procedures such as sleeve gastrectomy as a guide, to clamp and manipulate flat tissue and organs, such as the stomach.
Device Story
Disposable, non-energized, hand-held laparoscopic surgical instrument; used to grasp, clamp, and manipulate soft tissue (e.g., stomach) during surgery. Comprised of pistol-grip handle, shaft (fits 12mm trocar), and end effector (upper/lower jaws). Surgeon manually operates trigger/release to open/close jaws; used to guide endoscopic staplers during resection. Used in hospital setting by surgeons. Provides mechanical tissue stabilization; benefits patient by facilitating tissue manipulation during minimally invasive procedures.
Clinical Evidence
Bench testing only. Data demonstrates the 22 cm jaw length meets the same clamping and slip force specifications as the previously cleared 25 cm predicate device.
Technological Characteristics
Disposable, non-energized, manual laparoscopic instrument. Materials: aluminum end effector, stainless steel shaft, polycarbonate handle. Dimensions: 22 cm jaw length, fits 12 mm trocar. Sterilization: Gamma. No software or energy source.
Indications for Use
Indicated for use in laparoscopic procedures to grasp, clamp, and manipulate soft tissues, including bariatric procedures such as sleeve gastrectomy to guide, clamp, and manipulate flat tissue and organs like the stomach.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:
1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.
Predicate Devices
Standard Bariatrics Standard Clamp, Disposable (K181608)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for Standard Bariatrics. The logo consists of a stylized red letter "S" enclosed in a rounded square on the left, followed by the words "STANDARD BARIATRICS" in black, sans-serif font on the right. The trademark symbol is located to the right of the word "BARIATRICS".
## 510(K) Summary
## I. SUBMITTER
Standard Bariatrics 4362 Glendale Milford Road Cincinnati, OH 45242 USA
Phone: 513-304-7971 Email: alison(@standardbariatrics.com
Contact Person: Alison Sathe Date Prepared: March 15, 2019
II. DEVICE
Name of Device: Standard Clamp, Disposable Common or Usual Name: Laparoscopic Surgical Instrument Classification Name: Endoscope and Accessories (21 CFR 876.1500) Regulatory Class: II Product Code: GCJ
III. PREDICATE DEVICE
Standard Bariatrics Standard Clamp, Disposable K181608
### IV. DEVICE DESCRIPTION
The Standard Clamp, Disposable is a disposable non-energized, hand held surgical instrument designed for use in laparoscopic procedures to clamp long planes of soft tissue such as the stomach. The Standard Clamp, Disposable is provided sterile (gamma). There are no accessories provided with the device. The instrument is comprised of three main sections: handle, shaft, and end effector. The handle is a pistol grip with a trigger and release button which are manually activated in order to open and close the end effector. The shaft is sized to fit a standard 12 mm trocar. The clamp is comp is comprised of an upper and lower jaw that close to grasp and hold tissue. The environment of use is in a hospital during laparoscopic surgical procedures. The current Standard Clamp, Disposable has a jaw length of 25 cm. This Special 510(k) is being filed to add a new shorter jaw length (22 cm) to accommodate patients with smaller abdomens.
The Standard Clamp, Disposable is intended to be used by a surgeon on a single patient during a single laparoscopic surgery under normal operating conditions. The Standard Clamp, Disposable is intended to be inserted into the peritoneal cavity through a 12mm trocar and used in up to six cycles of opening and closure to grasp, clamp and manipulate long, 1.2 to 4.5mm thick, planes of soft tissues such as the stomach. The Standard Clamp, Disposable is not intended for direct contact with the cardiovascular system, lymphatic system or cerebrospinal fluid. The Standard Clamp, Disposable can be used while a surgeon uses devices such as a stapler adjacent to tissue that is clamped with the Standard Clamp, Disposable, e.g., to help guide endoscopic staplers during resection of tissue such as in sleeve gastrectomy.
### V. INDICATIONS FOR USE
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image is a logo for Standard Bariatrics. The logo features a red square with a stylized white "S" inside of it. To the right of the square is the text "STANDARD BARIATRICS" in black, bold letters. The letters are all capitalized, and there is a trademark symbol after the word "BARIATRICS".
The Standard Clamp, Disposable is indicated for use in laparoscopic procedures to grasp, clamp, and manipulate soft tissues. The Clamp is indicated for use in bariatric procedures such as sleeve gastrectomy as a guide, to clamp and manipulate flat tissue and organs, such as the stomach.
## COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE VI. DEVICE
The Standard Clamp, Disposable has the same technological characteristics as the predicate device with the exception of jaw length.
| | Table 1: Technological Characteristics |
|--|----------------------------------------|
| | |
| Product | Predicate Standard Clamp, Disposable<br>(product code SCD25)<br>(K181608) | Standard Clamp, Disposable<br>(product code SCD22)<br>(Subject) |
|------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------|
| Indications for Use | The Standard Clamp, Disposable is<br>indicated for use in laparoscopic<br>procedures to grasp, clamp, and<br>manipulate soft tissues. The Clamp is<br>indicated for use in bariatric procedures<br>such as sleeve gastrectomy as a guide, to<br>clamp and manipulate flat tissue and<br>organs, such as the stomach. | Same |
| Disposable/Reusable | Disposable | Same |
| Materials | End Effector: aluminum<br>Shaft: stainless steel<br>Handle: polycarbonate | Same |
| Overall Length | 26.3 inches (66.8 cm) | 25.1 inches (63.8 cm) |
| End Effector Length | 10.6 inches (27 cm) | 9.4 inches (24 cm) |
| End Effector Clamping Length | 9.8 inches (25 cm) | 8.9 inches (22 cm) |
| End Effector Width | 0.47 inches (12 mm) | Same |
| Shaft Width | 0.39 inches (10 mm) | Same |
| Weight | 0.5 lbs. | 0.5 lbs. |
| Articulation | None | Same |
| End effector design | Aluminum jaws with non-piercing<br>ridges to grasp tissue | Same |
| End effector closure | Distal to proximal closure | Same |
| Handle | Pistol grip | Same |
{2}------------------------------------------------
| Sterilization Method | Gamma Sterilization | Same |
|-------------------------------|---------------------------------------------------------------------|-------------------------------------|
| Anatomical Site Used | Various soft tissues accessible during<br>laparoscopic procedures | Same |
| Types of procedures | Laparoscopic procedures | Same |
| Energy Delivered? | No | Same |
| Compatible with 12 mm trocar? | Yes | Same |
| Biocompatibility | Biocompatible for blood/bone/tissue<br>contact for limited duration | Same |
| Packaging | Tyvek pouch in a cardboard box | Tyvek covered tray in cardboard box |
## VII. PERFORMANCE DATA
Standard Clamp, Disposable performance bench testing was provided in K170379. Testing for the Standard Clamp, Disposable 22 cm is provided in Section 018 - Performance Testing - Bench of this submission to the same specifications as the Standard Clamp, Disposable 25 cm previously provided in K 170379. The data demonstrates that the Standard Clamp, Disposable 22 cm meets the same specifications for clamping and slip force as the Standard Clamp, Disposable 25 cm. The packaging change has been validated and data is provided in Section 014 - Sterilization and Shelf Life.
## VIII. CONCLUSIONS
The Standard Clamp, Disposable has the same intended use and principles of operation as its predicate device. The only difference in technological characteristics is the jaw being offered in a shorter length in order to accommodate varying anatomy. The Standard Clamp, Disposable 22 cm meets the same specifications as the previously cleared Standard Clamp, Disposable 25 cm. There are no new issues of safety or effectiveness. Standard Clamp, Disposable is substantially equivalent to the predicate device.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym and name on the right. The Department of Health & Human Services logo is a stylized depiction of a human figure, while the FDA acronym and name are written in blue, with the acronym in a square and the name written out as "U.S. Food & Drug Administration".
Standard Bariatrics Alison Sathe, Regulatory 4362 Glendale Milford Rd. Cincinnati. Ohio 45242
April 18, 2019
Re: K190788
Trade/Device Name: Disposable Standard Clamp Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: Class II Product Code: GCJ Dated: March 21, 2019 Received: March 27, 2019
Dear Alison Sathe:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
{4}------------------------------------------------
https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Please wait...
If this message is not eventually replaced by the proper contents of the document, your PDF viewer may not be able to display this type of document.
You can upgrade to the latest version of Adobe Reader for Windows®, Mac, or Linux® by visiting http://www.adobe.com/go/reader_download.
For more assistance with Adobe Reader visit http://www.adobe.com/go/acrreader.
Windows is either a registered trademark of Microsoft Corporation in the United States and/or other countries. Mac is a trademark of Apple Inc., registered in the United States and other countries. Linux is the registered trademark of Linus Torvalds in the U.S. and other countries.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.