← Product Code [GCJ](/submissions/GU/subpart-b%E2%80%94diagnostic-devices/GCJ) · K190592

# YelloPort Elite Port Access System (K190592)

_Surgical Innovations Limited · GCJ · Apr 29, 2019 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/GU/subpart-b%E2%80%94diagnostic-devices/GCJ/K190592

## Device Facts

- **Applicant:** Surgical Innovations Limited
- **Product Code:** [GCJ](/submissions/GU/subpart-b%E2%80%94diagnostic-devices/GCJ.md)
- **Decision Date:** Apr 29, 2019
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

The YelloPort Elite port access system is indicated for use in laparoscopic procedures to give access to the abdominal cavity while maintaining pneumoperitoneum. The YelloPort Elite port access system is also indicated for use in thoracoscopic procedures to give access to the thoracic cavity.

## Device Story

YelloPort Elite Port Access System is a reusable surgical access device used in hospital settings by surgeons. It consists of a trocar (blunt, pencil point, pyramidal, or shielded), a cannula, a universal seal, and a Hasson (fascia) adaptor. The trocar is inserted into the cannula to penetrate the abdominal or thoracic wall; once access is achieved, the trocar is removed, leaving the cannula as a channel for endoscopes and surgical instruments. The universal seal maintains pneumoperitoneum during instrument use. The system facilitates minimally invasive surgery by providing a stable port for instrument passage. The device is reusable (except for the single-use universal seal) and sterilized via pressurized steam. It benefits patients by enabling minimally invasive surgical access, potentially reducing trauma compared to open surgery.

## Clinical Evidence

No clinical studies were conducted. Substantial equivalence was established through non-clinical bench testing, including sealing performance, instrument insertion/compatibility, port insertion force, insufflation, seal inversion, mechanical testing, biocompatibility (ISO 10993-1), shelf life, and sterilization validation. A simulated surgical evaluation by a consultant surgeon and usability assessments were performed.

## Technological Characteristics

Materials: Stainless steel (trocar), PEEK plastic (cannula, Hasson adaptor). Components: Trocar, cannula, universal seal, Hasson adaptor. Energy: Mechanical. Connectivity: None. Sterilization: Pressurized steam (reusable components). Form factor: 10mm and 12mm diameters; 75, 105, 150mm working lengths.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

## Predicate Devices

- YelloPort Plus Port Access System ([K070712](/device/K070712.md))
- YelloPort Port Access System ([K070712](/device/K070712.md))
- ENDOPATH XCEL Bladeless Trocar Universal Seal ([K032676](/device/K032676.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains two logos. On the left is the Department of Health & Human Services logo, which features a stylized caduceus. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG" in blue, and "ADMINISTRATION" in a smaller font size below that.

April 29, 2019

Surgical Innovations Limited Ms. Shazia Chaudhry Quality Assurance and Regulatory Manager Clayton Wood House 6 Clayton Wood Bank Leeds, West Yorkshire, LS16 6QZ, United Kingdom

Re: K190592

Trade/Device Name: YelloPort Elite Port Access System Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: Class II Product Code: GCJ Dated: February 28, 2019 Received: March 7, 2019

Dear Ms. Chaudhry:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801and Part 809: medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or post marketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

for

Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### Indications for Use

510(k) Number (if known) K190592

Device Name YelloPort Elite Port Access System

#### Indications for Use (Describe)

The YelloPort Elite port access system is indicated for use in laparoscopic procedures to give access to the abdominal cavity while maintaining pneumoperitoneum.

The YelloPort Elite port access system is also indicated for use in laparoscopic procedures to give access to the thoracic cavity.

| Type of Use (Select one or both, as applicable) |                                                      |
|-------------------------------------------------|------------------------------------------------------|
| Prescription Use (Part 21 CFR 801 Subpart D)    | <div> <input checked="true" type="checkbox"/> </div> |
| Over-The-Counter Use (21 CFR 801 Subpart C)     | <div> <input type="checkbox"/> </div>                |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

For YelloPort Elite Port Access System 510(k) Summary - 510(k)190592

This 510(k) Summary is submitted in accordance with 21 CFR Part 807, Section 807.92.

#### Submitter's Name:

Surgical Innovations Ltd.

#### Submitter's Address:

Clayton Wood House 6 Clayton Wood Bank Leeds LS16 6QZ United Kingdom

Telephone: +44(0)113 230 7597

#### Establishment Registration Number:

9680952

#### Contact Person:

Miss Shazia Chaudhry (BSc. M.Phil. MBA)

Clayton Wood House 6 Clayton Wood Bank Leeds LS16 6QZ United Kingdom

Telephone: +44(0)113 230 7597

E-Mail: shazia.chaudhry@surginno.co.uk

#### Date Prepared:

24 April 2019

{4}------------------------------------------------

For YelloPort Elite Port Access System 510(k) Summary - 510(k)190592

The information below summarises the Device Classification Information regarding the YelloPort Elite Port Access System:

#### Primary Product Code:

| Regulation<br>Number | Device                                    | Device Class | Product<br>Code | Classification<br>Panel      |
|----------------------|-------------------------------------------|--------------|-----------------|------------------------------|
| 876.1500             | Laparoscope, General &<br>Plastic Surgery | Class 2      | GCJ             | General &<br>Plastic Surgery |

#### Device Trade Name:

YelloPort Elite Port Access System

#### Device Common Name:

YelloPort Elite Port Access System

#### Intended/ Indications Use:

The YelloPort Elite port access system is indicated for use in laparoscopic procedures to give access to the abdominal cavity while maintaining pneumoperitoneum.

The YelloPort Elite port access system is also indicated for use in thoracoscopic procedures to give access to the thoracic cavity.

#### Summary of Substantial Equivalence:

The predicate devices which are used to claim equivalence of the YelloPort Elite Port Access System are detailed in the summary table below.

{5}------------------------------------------------

For YelloPort Elite Port Access System 510(k) Summary – 510(k)190592

#### Summary of Comparison between Proposed Device and Predicate Device

| General Information                                      |                                                                                                                                                                                                                                                                                                                                                        |                                                                                                                                                                                                                                                                                                                                                      |                                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                                                                                         |
|----------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Property                                                 | Proposed Device                                                                                                                                                                                                                                                                                                                                        | Primary Predicate                                                                                                                                                                                                                                                                                                                                    | Secondary Predicate                                                                                                                                                                                                                                                                                                                            | Secondary Predicate                                                                                                                                                                                                     |
|                                                          | YelloPort Elite Port<br>Access System                                                                                                                                                                                                                                                                                                                  | YelloPort Plus Port<br>Access System                                                                                                                                                                                                                                                                                                                 | YelloPort Port<br>Access System                                                                                                                                                                                                                                                                                                                | ENDOPATH XCEL<br>Bladeless Trocar<br>Universal Seal                                                                                                                                                                     |
| Common<br>Name                                           | Laparoscope,<br>General & Plastic<br>Surgery                                                                                                                                                                                                                                                                                                           | Laparoscope,<br>General & Plastic<br>Surgery                                                                                                                                                                                                                                                                                                         | Laparoscope,<br>General & Plastic<br>Surgery                                                                                                                                                                                                                                                                                                   | Laparoscope, General<br>& Plastic Surgery                                                                                                                                                                               |
| Device<br>Manufacturer                                   | Surgical Innovations                                                                                                                                                                                                                                                                                                                                   | Surgical Innovations                                                                                                                                                                                                                                                                                                                                 | Surgical Innovations                                                                                                                                                                                                                                                                                                                           | Ethicon Endo-Surgery,<br>LLC                                                                                                                                                                                            |
| Device<br>Classification                                 | II                                                                                                                                                                                                                                                                                                                                                     | II                                                                                                                                                                                                                                                                                                                                                   | II                                                                                                                                                                                                                                                                                                                                             | II                                                                                                                                                                                                                      |
| Primary<br>Product Code                                  | GCJ                                                                                                                                                                                                                                                                                                                                                    | GCJ                                                                                                                                                                                                                                                                                                                                                  | GCJ                                                                                                                                                                                                                                                                                                                                            | GCJ                                                                                                                                                                                                                     |
| 510(k)<br>Number                                         | N/A                                                                                                                                                                                                                                                                                                                                                    | K070712                                                                                                                                                                                                                                                                                                                                              | K070712                                                                                                                                                                                                                                                                                                                                        | K032676                                                                                                                                                                                                                 |
| Environment                                              | Hospital                                                                                                                                                                                                                                                                                                                                               | Hospital                                                                                                                                                                                                                                                                                                                                             | Hospital                                                                                                                                                                                                                                                                                                                                       | Hospital                                                                                                                                                                                                                |
| Intended Use/<br>Indication for<br>Use                   | The YelloPort Elite<br>Port Access System is<br>indicated for use in<br>laparoscopic<br>procedures to give<br>access to the<br>abdominal cavity while<br>maintaining<br>pneumoperitoneum.<br><br>The YelloPort Elite<br>Port access system is<br>also indicated for use<br>in thoracoscopic<br>procedures to give<br>access to the thoracic<br>cavity. | The YelloPort Plus<br>Port Access System is<br>indicated for use in<br>laparoscopic<br>procedures to give<br>access to the<br>abdominal cavity<br>whilst maintaining<br>pneumoperitoneum.<br><br>The YelloPort Plus<br>Port Access System is<br>also indicated for use<br>in laparoscopic<br>procedures to give<br>access to the thoracic<br>cavity. | The YelloPort Port<br>Access System is<br>indicated for use in<br>laparoscopic<br>procedures to give<br>access to the<br>abdominal cavity<br>whilst maintaining the<br>pneumoperitoneum.<br><br>The YelloPort Port<br>Access System is also<br>indicated for use<br>in laparoscopic<br>procedures to give<br>access to the thoracic<br>cavity. | The ENDOPATH XCEL<br>Bladeless Trocar is<br>intended for use in<br>abdominal, thoracic and<br>gynaecologic minimally<br>invasive surgical<br>procedures, to establish<br>a path of entry for<br>endoscopic instruments. |
| Trocar Types                                             | Blunt<br>Pencil point,<br>Pyramidal<br>Shielded (locking)                                                                                                                                                                                                                                                                                              | Blunt<br>Pencil Point,<br>Pyramidal<br>Quill<br>Shielded (locking and<br>non-locking).                                                                                                                                                                                                                                                               | Blunt<br>Pencil Point,<br>Pyramidal<br>Shielded (locking and<br>non-locking).                                                                                                                                                                                                                                                                  | N/A                                                                                                                                                                                                                     |
| Trocar /<br>Cannula<br>Diameter and<br>Lengths           | Available in 10 and 12<br>mm diameter; and in<br>75, 105 and 150 mm<br>working lengths.                                                                                                                                                                                                                                                                | Available in 3, 5, 10,<br>12 and 16 mm<br>diameter; and in 55,<br>70, 75, 95, 105 and<br>150 mm working<br>lengths.                                                                                                                                                                                                                                  | Available in 3.5, 5.5,<br>7.5, 10.5 and 12.5mm<br>diameter; and in 55,<br>70, 95, 105 and 150<br>mm working lengths.                                                                                                                                                                                                                           | N/A                                                                                                                                                                                                                     |
| Comparison<br>of Use:                                    | Allows instruments<br>from 5mm -12mm in<br>diameter to be used.                                                                                                                                                                                                                                                                                        | Within the range,<br>allows instruments<br>from 5-12mm diameter<br>to be used.                                                                                                                                                                                                                                                                       | Within the range,<br>allows instruments<br>from 5-12mm diameter<br>to be used.                                                                                                                                                                                                                                                                 | Allows instruments from<br>5mm-12mm in diameter<br>to be used.                                                                                                                                                          |
| Trocar<br>Patient<br>Contacting<br>Material              | Stainless Steel                                                                                                                                                                                                                                                                                                                                        | Stainless Steel                                                                                                                                                                                                                                                                                                                                      | Stainless Steel                                                                                                                                                                                                                                                                                                                                | N/A                                                                                                                                                                                                                     |
| Trocar<br>Supplied:                                      | Non-sterile<br>(Reusable)                                                                                                                                                                                                                                                                                                                              | Non-sterile<br>(Reusable)                                                                                                                                                                                                                                                                                                                            | Non-sterile<br>(Reusable)                                                                                                                                                                                                                                                                                                                      | N/A                                                                                                                                                                                                                     |
| Cannula<br>Patient<br>Contacting<br>Materials            | PEEK plastic                                                                                                                                                                                                                                                                                                                                           | PEEK plastic                                                                                                                                                                                                                                                                                                                                         | PEEK plastic                                                                                                                                                                                                                                                                                                                                   | N/A                                                                                                                                                                                                                     |
| Cannula<br>Supplied:                                     | Non-sterile<br>(Reusable)                                                                                                                                                                                                                                                                                                                              | Non-sterile<br>(Reusable)                                                                                                                                                                                                                                                                                                                            | Non- sterile<br>(Reusable)                                                                                                                                                                                                                                                                                                                     | N/A                                                                                                                                                                                                                     |
| Seal                                                     | Consists of:<br>Non-Return Valve &<br>Instrument Seal                                                                                                                                                                                                                                                                                                  | Consists of:<br>Non-Return Valve &<br>Instrument Seal                                                                                                                                                                                                                                                                                                | Consists of :<br>Non-Return Valve &<br>Instrument Seal                                                                                                                                                                                                                                                                                         | Consists of:<br>Non-Return Valve &<br>Instrument Seal                                                                                                                                                                   |
| Seal<br>Supplied:                                        | Sterile                                                                                                                                                                                                                                                                                                                                                | Sterile                                                                                                                                                                                                                                                                                                                                              | Sterile                                                                                                                                                                                                                                                                                                                                        | Sterile                                                                                                                                                                                                                 |
| Hasson<br>(Fascia)<br>Adaptor                            | Available in both<br>10mm and 12mm<br>diameter versions.                                                                                                                                                                                                                                                                                               | Available in both<br>10mm and 12mm<br>diameter versions.                                                                                                                                                                                                                                                                                             | Available in both<br>10mm and 12mm<br>diameter versions.                                                                                                                                                                                                                                                                                       | N/A                                                                                                                                                                                                                     |
| Hasson<br>(Fascia)<br>Adaptor                            | Fully reusable can be<br>sterilised in<br>pressurised steam.                                                                                                                                                                                                                                                                                           | Fully reusable can be<br>sterilised in<br>pressurised steam.                                                                                                                                                                                                                                                                                         | Fully reusable can be<br>sterilised in<br>pressurised steam.                                                                                                                                                                                                                                                                                   | N/A                                                                                                                                                                                                                     |
| Hasson<br>(Fascia<br>Adaptor)<br>Contacting<br>Materials | PEEK plastic                                                                                                                                                                                                                                                                                                                                           | PEEK plastic                                                                                                                                                                                                                                                                                                                                         | PEEK plastic                                                                                                                                                                                                                                                                                                                                   | N/A                                                                                                                                                                                                                     |

{6}------------------------------------------------

For YelloPort Elite Port Access System 510(k) Summary - 510(k)190592

Any technical differences have been justified, both scientifically and using performance testing. These do not affect the safety or effectiveness of the proposed device.

#### Device Description:

The YelloPort Elite Port Access System comprises a trocar, available with a range of tip styles, cannula. Universal Seal and Hasson (fascia) adaptor. The trocar, cannula and Hasson (fascia) adaptor elements of the device are fully reusable and can be sterilised in pressurised steam. The Universal seal is single use only.

In order to obtain access to the surgical site during laparoscopic surgery, the YelloPort Elite Trocar is introduced into the YelloPort Elite Cannula to accomplish cannula penetration of the abdominal wall. The cannula is connected to the YelloPort Elite Universal Seal at its proximal end and once the abdominal/thoracic wall is punctured, the trocar is removed. The cannula acts as a channel for the introduction of the endoscopes and instruments. Generically, trocars and cannulas are available in a range of lengths and diameters.

{7}------------------------------------------------

For YelloPort Elite Port Access System 510(k) Summary - 510(k)190592

#### Technological Characteristics:

A comparative review of the YelloPort Elite Port Access System with the predicate devices found that the technology, mode of operation, and general principles for treatment with this device were substantially equivalent to the predicate devices.

#### Non-Clinical Tests (Performance/Physical Data):

The YelloPort Elite Port Access System was evaluated for its safety and effectiveness based on the following testing:

- Sealing Performance
- Instrument Insertion Force (Universal Seal)
- Seal Contamination Testing
- Instrument Compatibility Testing (Universal Seal)
- Port Insertion Force Testing
- . Insufflation Testing
- Seal Inversion Testing
- . Trocar Knob Pull Testing
- Stopcock Torque Lever Test
- Universal Seal Attachment and Removal Test
- Cannula Mechanical Testing
- Pressure Retention of Universal Seal (With Use of Locking Shielded Trocar) .
- Trocar Peritoneum Insertion Simulation Test (Locking Shielded)
- Torque Force Test (Locking Shielded)
- Tensile Test (Locking Shielded)
- . Biocompatibility Testing in accordance with ISO 10993-1
- . Shelf life and lifecycle studies
- . Transit studies
- . Cleaning and sterilisation validation
- Clinical (Design) validations which include a simulated surgical evaluation by a consultant surgeon
- . Usability assessments.

#### Clinical Studies

No clinical studies were conducted as part of submission to prove substantial equivalence. Non-Clinical Bench testing was sufficient to prove equivalence to the predicate device.

#### Safety and Effectiveness/Conclusion:

Based on the information presented in these 510(k) premarket notifications the YelloPort Elite Port Access System is considered substantially equivalent. The YelloPort Elite Port Access System is as safe and effective as the currently marketed predicate devices.

Based on testing and comparison with the predicate device. the YelloPort Elite Port Access System shows no adverse indications or results. It is our determination that the YelloPort Elite Port Access System is safe, effective and performs within its design specifications and is substantially equivalent to the predicate devices.

---

**Source:** [https://fda.innolitics.com/submissions/GU/subpart-b%E2%80%94diagnostic-devices/GCJ/K190592](https://fda.innolitics.com/submissions/GU/subpart-b%E2%80%94diagnostic-devices/GCJ/K190592)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
