AUTOLAP SYSTEM

K131375 · Mst - Medical Surgery Technologies , Ltd. · GCJ · Sep 19, 2013 · Gastroenterology, Urology

Device Facts

Record IDK131375
Device NameAUTOLAP SYSTEM
ApplicantMst - Medical Surgery Technologies , Ltd.
Product CodeGCJ · Gastroenterology, Urology
Decision DateSep 19, 2013
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2

Indications for Use

The AutoLap System is indicated for General laparoscopic, Gynecologic and Urologic procedures for the purpose of holding and controlling the movement of a standard laparoscope or rigid endoscope within surgical cavities during minimally invasive laparoscopic surgery.

Device Story

AutoLap System is a laparoscope holder for minimally invasive surgery; holds/controls movement of standard laparoscope or rigid endoscope. System comprises processing unit (electronics/software), base unit (mounted on operating bed), and laparoscopic unit (motors/sensors). Surgeon controls movement via RF-transmitting Command Unit (affixed to hand/instrument) or manual activation button on system. Enables left/right, up/down, and zoom in/out movements. Used in OR by surgeons to stabilize/position camera; reduces manual holding requirements; improves image stability. Bench testing and animal studies (cholecystectomy/Nissen fundoplication) validated movement control, safety features (emergency stop), and performance.

Clinical Evidence

No human clinical data. Evidence includes bench testing (electrical safety per ANSI/AAMI 60601-1, EMC per IEC 60601-1-2, sterilization per ISO 11135-1/10993-7, packaging per ISO 11607-2/ASTM D1649-09) and an animal study (domestic pigs). Animal study evaluated cholecystectomy and Nissen fundoplication; confirmed movement control, safety, and performance with no operative complications.

Technological Characteristics

System includes processing unit, base unit, and laparoscopic unit with motors and sensors. Connectivity via RF signals from Command Unit. Sterilization per ISO 11135-1. Packaging per ISO 11607-2 and ASTM D1649-09. Electrical safety per ANSI/AAMI 60601-1; EMC per IEC 60601-1-2. Software-controlled movement in multiple degrees of freedom.

Indications for Use

Indicated for patients undergoing general laparoscopic, gynecologic, or urologic procedures requiring holding and control of a standard laparoscope or rigid endoscope within surgical cavities.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ### K131375 # 510(k) SUMMARY MST AutoLap System Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared MST - Medical Surgery Technologies Ltd. Kochav Yokneam Building, POB 685, Yokneam Ilit 20692, Israel Phone: +972-73-796-5570 Facsimile: +972-73-796-5571 Contact Person: Tami Harel Date Prepared: August 21, 2013 ## Name of Device and Name/Address of Sponsor - AutoLap System MST - Medical Surgery Technologies Ltd. Kochav Yokneam Building, POB 685, Yokneam Ilit 20692, Israel #### Common or Usual Name SEP 19 2013 Laparoscope Holder #### Classification Name Laparoscope, General and Plastic Surgery 21 C.F.R. 876.1500, Product Code GCJ #### Predicate Devices LapMan Laparoscope Manipulator System and Accessories, Manufactured by MEDSYS, S.A. (K062968) #### Intended Use / Indications for Use The AutoLap System is indicated for General laparoscopic, Gynecologic and Urologic procedures for the purpose of holding and controlling the movement of a standard laparoscope or rigid endoscope within surgical cavities during minimally invasive laparoscopic surgery. ## Technological Characteristics The AutoLap System is designed to hold and control the movement of a laparoscope or rigid endoscope during laparoscopic procedures. The AutoLap System includes both software and hardware. Hardware components include a Processing Unit, which includes the system's electronics and software algorithms, a Base Unit and a Laparoscopic Unit that include motors and sensors. The Base Unit is mounted on the operating bed. An {1}------------------------------------------------ accessory cart that is used for placing the system's components during the procedure and their storage afterwards is also provided. To maneuver the laparoscope to the desired position, the surgeon presses a single button, referred to as Command Unit, which is affixed to his/her hand or to the surgical instrument and transmits RF signals to the system. Additionally, the system enables the surgeon to manually perform larger movements of the laparoscope using the Manual Activation Button affixed to the AutoLap system. Movements can be performed in the left/right, up/down and zoom in/ zoom out directions. ## Performance Data The following testing was conducted to evaluate the device: - Electrical safety was tested in accordance with ANSI/AAMI 60601-1:2005 -. Medical electrical equipment - Part 1: General requirements for basic safety and essential performance. - . Electromagnetic Compatibility (EMC) was tested in accordance with IEC 60601-1-2:2007 - Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests. - Sterilization validation was conducted in accordance with the ISO 11135-1 and . 10993-7 standards. - Packaging integrity and shelf life validation in accordance with ISO 11607-2 and . ASTM D1649-09. - . Bench testing was conducted in order to demonstrate that the AutoLap system performs according to its requirements and specifications. Performance in terms of movement velocities in all DOF, movement range in all DOF, directional movement control and image stability, have all been demonstrated. Safety features such as Emergency Stop Button functionality and warning messages have all been demonstrated to meet their respective acceptance criteria. - . The company performed an animal study evaluating the use of the AutoLap system in laparoscopic cholecystectomy and Nissen fundoplication on domestic pias. Procedures were performed by several surgeons. No operative complications related to the use of the AutoLap were observed and the device met all predefined acceptance criteria. The results demonstrated the system's performance and safety. The system's performance in terms of the movement control was validated. All performance testing demonstrates that the AutoLap System performs according to specifications and functions as intended. {2}------------------------------------------------ ## Substantial Equivalence The AutoLap System is substantially equivalent to the LapMan Laparoscope Manipulator System and Accessories. The AutoLap System has the same intended uses and similar indications, technological characteristics, and principles of operation as its predicate device. The minor technological differences between the AutoLap System and its predicate devices raise no new issues of safety or effectiveness. Performance data demonstrate that the AutoLap System is as safe and effective as the LapMan Laparoscope Manipulator System and Accessories. Thus, the AutoLap System is substantially equivalent. {3}------------------------------------------------ # DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. In the center of the circle is an abstract symbol that resembles an eagle or bird in flight, with three stylized lines forming the body and wings. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 September 19, 2013 Medical Surgery Technologies, Ltd. c/o John Smith, M.D., J.D. Partner Hogan Lovells US LLP 555 Thirteenth Street NW Washington DC 20004 Re: K131375 Trade/Device Name: AUTOLAP Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: Class II Product Code: GCJ Dated: August 21, 2013 Received: August 21, 2013 Dear Dr. Smith: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you: however, that device labeling must be truthful and not misleading, If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must {4}------------------------------------------------ Page 2 - Dr. John Smith comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, # David Krause -S for Mark N. Melkerson Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ ## Indications for Use Statement K131375 510(k) Number (if known): Device Name: AUTOLAP Indications for Use: The AutoLap System is indicated for General laparoscopic, Gynecologic and Urologic procedures for the purpose of holding and controlling the movement of a standard laparoscope or rigid endoscope within surgical cavities during minimally invasive laparoscopic surgery. Prescription Use x (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Page 1 of 1 | Digitally signed by Long H. Chen -A<br>DN: c=US, o=U.S. Government, ou=HHS, | for MXM | |-------------------------------------------------------------------------------------------------------------------------------------|---------| | Long H. Chen -A ou=FDA, ou=People, cn=Long H. Chen - A,<br>0.9.2342.19200300.100.1.1=1300369056<br>Date: 2013.09.18 07:04:48 -04:00 | | (Division Siqn-Off) Division of Surgical Devices 510(k) Number: K131375
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...