ENDOSCOPIC APPLICATOR

K120608 · Micromedics, Inc. · GCJ · Mar 15, 2012 · Gastroenterology, Urology

Device Facts

Record IDK120608
Device NameENDOSCOPIC APPLICATOR
ApplicantMicromedics, Inc.
Product CodeGCJ · Gastroenterology, Urology
Decision DateMar 15, 2012
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2
AttributesTherapeutic, 3rd-Party Reviewed

Indications for Use

The Endoscopic Applicator is intended for use in delivering hemostatic agents to bleeding surgical sites through a 5 mm or larger trocar.

Device Story

The Endoscopic Applicator is a sterile, single-use, disposable device used to deliver hemostatic agents to bleeding surgical sites. It is designed for use during endoscopic surgery through a 5 mm or larger trocar. The system consists of a non-reflective cannula and a stylet (obturator). A standard syringe containing a hemostatic agent is attached to the device's luer connector. The clinician inserts the cannula through the trocar and uses the stylet to facilitate the delivery of the agent to the target site. The device provides a mechanical means for precise application of hemostatic materials, assisting the surgeon in controlling bleeding during minimally invasive procedures.

Clinical Evidence

Bench testing only. Performance data included volume testing, flex testing, luer lock testing (ISO 594-2), hemostatic usage testing, leak testing, pull/torque/bend testing, cannula tissue compliance, shipping validation, and shelf-life evaluations. Biocompatibility was assessed per ISO 10993-1 (2009). All tests met pre-defined acceptance criteria.

Technological Characteristics

Device consists of a PolyMed (high strength fiber with ester vinyl resin) cannula and stylet with ABS over-molded luer lock, tip, and handle. Sterile, single-use, disposable. Sterilized via ethylene oxide. Dimensions compatible with 5 mm or larger trocars. Mechanical delivery system; no energy source or software.

Indications for Use

Indicated for delivery of hemostatic agents to bleeding surgical sites in patients requiring endoscopic procedures via a 5 mm or larger trocar.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ 42060.f Micromedics, Inc · Traditional 510{k) Premarket Notification Endoscopic Applicator #### 510(k) Summary 5 MAR 1,5 2012 Date Prepared: December 22, 2011 | 510(k) Submitter | |---------------------------------------| | Micromedics, Inc. | | 1270 Eagan Industrial Road, Suite 120 | | St. Paul, MN 55121-1385 | Contact for Official Correspondence Lise W. Duran - General Manager Tel: 651-452-1977; Fax: 651-452-1787 Email: Lise.Duran@nordsommicromedics.com | General Information | | | | |----------------------------|------------------------------------------------------------------------------------------------------------------|--------------|---------------------------| | Trade Name | Endoscopic Applicator | Common Name | Endoscopic Applicator | | Classification Information | Endoscope and accessories<br>per 21 CFR 876.1500 (Class II) | Product Code | GCJ | | | | Panel | General & Plastic Surgery | | Predicate Devices | Endoscopic Applicator; Ethicon, Inc. (K051732)<br>Endoscopic Applicator; Baxter Healthcare Corporation (K031882) | | | #### Device Description The Endoscopic Applicator is a sterile, single-use, disposable device intended for use in delivering hemostatic agents to bleeding surgical sites through a 5 mm (or larger) trocar. The Endoscopic Applicator is designed with a luer connector, which is used for connection to a syringe containing the hemostatic agent. The Endoscopic Applicator system consists of a cannula and stylet. The non-reflective cannula is a composite sheath constructed of PolyMed" (a high strength fiber with ester vinyl resin) with an Acrylonitrile Butadiene Styrene (ABS) over-molded luer lock. The stylet (obturator) consists of a PolyMed® composite rod, an ABS over-molded tip, and an ABS over-molded handle. The Endoscopic Applicator cannula and stylet are packaged in a double sterile barrier tray configuration and sterilized using ethylene oxide. Six individually sterile packaged applicators are contained in a shelf carton along with instructions for use. ### Intended Use / Indications The Endoscopic Applicator is intended for use in delivering hemostatic agents to bleeding surgical sites through a 5 mm or larger trocar. ## Substantial Equivalence Comparison The Endoscopic Applicator has substantially equivalent intended use/indications, principles of use and functional characteristics as the predicate Ethicon and Baxter endoscopic applicators. All three devices are intended for use in delivering hemostatic agents to bleeding surgical sites through a 5 mm or larger trocar. The subject and predicate device systems consist of a cannula and stylet. The cannula allows a common syringe containing hemostatic agent to be attached and permits delivery of the hemostatic agent to the intended bleeding site with the aid of the stylet. While all three devices allow access through a 5 mm trocar, they differ in length and biomaterials. # Summary of Non-Clinical Performance Data Summaly of Non-Chilitin's crossments through design verification and biocompatibility t its Endosoph reportation of the many of the secordance with ISO 10993 - Biological evaluation of medical devices, Part 1 – Evaluation and tests (2009) show the device is considered safe for of montanian intended biocontact. Non-clinical testing included the below tests and showed the test {1}------------------------------------------------ # Micromedics, Inc KI2.obof --- Traditional 510(k) Premarket Notification Endoscopic Applicator articles met the pre-defined acceptance criteria, therefore demonstrating the mechanical integrity and suitability of the device for its intended use over the labeled shelf life. - Volume test . - . Flex test (stylet and cannula) - Luer Lock tests (ISO 594-2) . - Hemostatic usage test . - Device leak test . - Pull / Torque tests (cannula) o - Pull / Bend tests (stylet) o - Cannula tissue compliance . - Shipping validation - Shelf life evaluations Substantial Equivalence Conclusion The Endoscopic Applicator does not raise new questions of safety or effectiveness when compared to the predicate devices and is, therefore, substantially equivalent. - The remainder of this page intentionally left blank.- {2}------------------------------------------------ # DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus, which is a symbol often associated with medicine and healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the caduceus. The logo is black and white. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 MAR 1 5 2012 Micromedics, Inc. % Intertek Testing Services Ms. Paula Wilkerson 2307 E. aurora Road, Unit B7 Twinsburg, Ohio 44087 Re: K120608 Trade/Device Name: Endoscopic Applicator Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: GCJ Dated: February 28, 2012 Received: February 29, 2012 Dear Ms. Wilkerson: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 {3}------------------------------------------------ Page 2 - Ms. Paula Wilkerson CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours. R Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ K12060f # Indications for Use 510(k) Number (if known): _ Device Name: Endoscopic Applicator Indications for Use: The Endoscopic Applicator is intended for use in delivering hemostatic agents to bleeding surgical sites through a 5 mm or larger trocar. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Neil RP Dyke for mxn Page 1 of 1 (Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices 510(k) Number K120608
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...