DA VINCI FLUORESCENCE IMAGING VISION SYSTEM, MODEL FF 100
K101077 · Intuitive Surgical, Inc. · GCJ · Feb 4, 2011 · Gastroenterology, Urology
Device Facts
Record ID
K101077
Device Name
DA VINCI FLUORESCENCE IMAGING VISION SYSTEM, MODEL FF 100
Applicant
Intuitive Surgical, Inc.
Product Code
GCJ · Gastroenterology, Urology
Decision Date
Feb 4, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.1500
Device Class
Class 2
Indications for Use
The da Vinci® Fluorescence Imaging Vision System is intended to provide real-time endoscopic visible and near-infrared fluorescence imaging. The da Vinci® Fluorescence Imaging Vision System enables surgeons to perform minimally invasive surgery using standard endoscopic visible light as well as visual assessment of vessels, blood flow and related tissue perfusion using near infra-red imaging.
Device Story
Endoscopic imaging system for minimally invasive surgery; provides real-time visible light and near-infrared (NIR) fluorescence imaging. Input: optical signals captured via 12mm endoscopes and 3D HD stereoscopic camera head. Transformation: system processes optical inputs to display standard visible light and NIR fluorescence overlays; utilizes IndoCyanine Green (ICG) fluorescence imaging agent for contrast. Output: visual assessment of vessels, blood flow, and tissue perfusion displayed to surgeon. Used in OR by surgeons to assist in intra-operative tissue assessment. Benefits: enables real-time visualization of perfusion, potentially aiding surgical decision-making during minimally invasive procedures.
Clinical Evidence
Bench testing only. Verification and validation testing performed to evaluate device requirements and specifications against the da Vinci Si Surgical System and SPY Scope Intra-operative Imaging System. Results demonstrated that all components met design and performance criteria.
Technological Characteristics
Endoscopic imaging system; 12mm endoscopes (0 or 30 degree); 3D HD stereoscopic camera head; NIR-optimized illuminator; ICG fluorescence imaging kit. Connectivity: integrates with existing video processor unit. Energy: standard endoscopic illumination. Sterilization: not specified.
Indications for Use
Indicated for patients undergoing minimally invasive surgery requiring endoscopic visible light and near-infrared fluorescence imaging for visual assessment of vessels, blood flow, and tissue perfusion.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
Predicate Devices
Intuitive Surgical da Vinci SI Surgical System: Model IS3000 (K081137)
Novadaq Spy Scope Intra-operative Operating System (K091515)
Submission Summary (Full Text)
{0}------------------------------------------------
KI017 page (1) of (2)
FEB - 4 2011 INTUITIVE
falling surgery beyond the limits of the numan hund "
SURGICAL
## 510(k) Summary (As Required by 21 CFR 807.92(c))
April 16, 2010
Submitter: Intuitive Surgical, Inc. 1266 Kifer Road Sunnyvale, CA 94086
- Official Contact: Karen Uyesuai Vice President, Clinical and Regulatory Affalrs Ph: 408-523-8918 Fax: 408-523-1390 karen.uyesugl@Intusurg.com
Trade Name: Intuitive Surgical da Vinci® Fluorescence Imaging Vislon System
Common Name: system, surgical, computer controlled Instrument
Product Code: NAY (GCJ/IZI)
Classification: endoscope and accessories, 21 CFR 876. 1500
Predicate Devices: Intuitive Surgical da Vinci� SI Surgical System: Model IS3000 (K081137) Novadaq Spy Scope Intra-operative Operating System (KO91515)
Device Description: The Intultive Surgical da Vinci® Fluorescence Imaging Vision System is an endoscopic imaging system for high definition (HD) visible ilght and near-Infrared fluorescence Imaging during minimally invasive surgery. The Intuittve Surgical da Vincl® Fluorescence Imaging Vision System is a variation of the Intultive Surgical da Vincle SI Surgical System: Model IS3000 (K081137) with the following modified/additional components:
- 12mm endoscopes (0 degree or 30 degree) optimized for NIR . fluorescence Imaging In addition to VIS Imaging,
- A 3D High Definition stereoscopic camera head optimized for NIR . fluorescence imaging In addition to VIS Imaging,
- A new illuminator for use with the existing video processor unit, and ◆
- A Fluorescence Imaging Klt (IndoCyanine Green (ICG) fluorescence ● Imaging agent, aqueous solvent, and syringe trays)
Intuitive Surgical, Inc.
Confidential
page 5-2 of 55
{1}------------------------------------------------
K151077
ge (2) of (2)
falling surgery be rand the lines of the human hand "
N T II I T I V F SURGICAL
Intended Use/Indications for Use: The da Vincle Fiuorescence Imaging Vision System is Intended to provide real-time endoscopic visible and near-infrared fluorescence imaging. The da Vincl® Fluorescence Imaging Vision System enables surgeons to perform minimally invasive surgery using standard endoscopic visible light as well as visual assessment of vessels, blood flow and related tissue perfusion using near infra-red imaging.
Technological Characteristics: The da Vinci® Fluorescence Imaging Vision System is equivalent to the predicate devices in terms of its indications for use, design, technology and performance specifications.
Performance Data: Verification and validation were performed on the da Vincle Fluorescence Imaging Vision System to evaluate the device requirements/specifications and to demonstrate that they are substantially equivalent to the predicate devices (da Vinct®Si Surgical System: Model IS3000 and SPY Scope Intra-operative Imaging System).
The results of verification and validation tests demonstrate that the da Vincle Fluorescence imaging Vision System components met their design and performance criteria and are substantlally equivalent to the predicate devices.
Summary: Based on the Indications for use, design, technological characteristics and performance data the da Vincl® Fluorescence Imaging Vision System Is substantially equivalent to the predicate devices in terms of safety, efficacy and performance.
Confidential
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure with three curved lines representing its body and wings. The logo is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" in a circular arrangement.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Intuitive Surgical, Inc. % Ms. Karen Uyesugi Vice President, Clinical and Regulatory Affairs 1266 Kifer Road Sunnyvale, California 94086
FEB - 4 201
Re: K101077
Trade/Device Name: da Vinci® Fluorescence Imaging Vision System Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: Class II Product Code: GCJ, NAY, IZI Dated: January 10, 2011 Received: January 11, 2011
Dear Ms. Uyesugi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{3}------------------------------------------------
Page 2 - Ms. Karen Uyesugi
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Mark N. Melkerson Director Division of Surgical, Orthopedic And Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## 510(k) Number if known: K101077
Device Name: da Vinci® Fluorescence Imaging Vision System
## INDICATION FOR USE:
The da Vinci® Fluorescence Imaging Vision System is intended to provide realtime endoscopic visible and near-infrared fluorescence imaging. The da Vinci® Fluorescence Imaging Vision System enables surgeons to perform minimally invasive surgery using standard endoscopic visible light as well as visual assessment of vessels, blood flow and related tissue perfusion using near infra-red imaging.
Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Per 21 CFR 801 Subpart D) Subpart C)
AND/OR
Over-the-Counter Use (Per 21 CFR 807
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Niel R. Dyke
Signature
Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K101077