← Product Code [GCJ](/submissions/GU/subpart-b%E2%80%94diagnostic-devices/GCJ) · K090275

# GELPORT SINGLE INCISION ACCESS SYSTEM (K090275)

_Applied Medical Resources Corp. · GCJ · Feb 18, 2009 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/GU/subpart-b%E2%80%94diagnostic-devices/GCJ/K090275

## Device Facts

- **Applicant:** Applied Medical Resources Corp.
- **Product Code:** [GCJ](/submissions/GU/subpart-b%E2%80%94diagnostic-devices/GCJ.md)
- **Decision Date:** Feb 18, 2009
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** 3rd-Party Reviewed

## Indications for Use

For use in patients undergoing laparoscopic surgical procedures to obtain abdominal access by instrumentation. Trocars may be placed through the Gel Seal Cap to allow instrument access to the surgical site. The device may be used in procedures such as nephrectomy, colectomy and splenectomy in colorectal, urological and general surgery to access the surgical site.

## Device Story

Single Incision Access System provides abdominal access for laparoscopic surgery; consists of wound retractor (flexible cylinder with semi-rigid polyurethane rings) and Gel-Cap (polycarbonate ring with gel material). Surgeon places retractor in abdominal incision; attaches Gel-Cap via lever. Gel-Cap allows insertion of multiple trocars for instrument access while maintaining pneumoperitoneum. Unlike predicate (K014047), device lacks hand-access slit; optimized for trocar-only use to enable smaller incisions. Used in OR by surgeons; benefits include reduced incision size and maintained peritoneal pressure during instrument exchange. Convenience kits may include trocars with retention flanges to prevent migration.

## Clinical Evidence

Bench testing only. Testing included minimum incision size determination, leak testing (pre/post trocar insertion and after vigorous manipulation), and particulate testing to evaluate resistance to gel degradation.

## Technological Characteristics

Components: wound retractor (polyurethane rings, flexible film), Gel-Cap (polycarbonate ring, gel material). Attachment: lever-based locking mechanism. Form factor: 40-120mm diameter for 15-90mm incisions. Connectivity: None. Sterilization: Not specified. Software: None.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

## Predicate Devices

- Applied GelPort Laparoscopic Hand Access Device ([K014047](/device/K014047.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Page 1 of 4

# K090275

### 510(k) SUMMARY

### FEB 1 8 2009

#### SUBMITTED BY:

Applied Medical Resources Corporation 22872 Avenida Empresa Rancho Santa Margarita, CA-92688 (949) 713-8000 (949) 713-8205 (FAX)

Frans VandenBroek Vice President, Regulatory Affairs fvandenbroek@appliedmedical.com

DATE OF PREPARATION:

The 1

COMMON NAME:

PREDICATE DEVICE:

TRADE NAME:

CONTACT PERSON:

CLASSIFICATION NAME:

To be determined

January 8, 2009

Single Incision Access System

Laparoscope, General & Plastic Surgery (21CFR 876.1500, product code GCJ)

K014047, Applied GelPort Laparoscopic Hand Access Device

DEVICE DESCRIPTION: The subject device is a modified version of the predicate . device. It has a simpler design, specific indications for use and allows a surgeon to choose an alternate method of accessing body cavities. It also improves patient safety by reducing the size of an abdominal incision.

Image /page/0/Figure/17 description: The image shows two medical devices, labeled as "PREDICATE DEVICE K014047" and "SUBJECT DEVICE". Both devices are attached to an abdominal wall and consist of a gel-cap, outer ring, wound retractor film, and inner ring. The predicate device has a hand interacting with it, while the subject device has three gel-caps.

{1}------------------------------------------------

K090275

ﺗ

Page 2 of (4)

The predicate device of K014047 accommodates insertion of a hand during a Hand Assisted Laparascopic (HAL) procedure. It consists of a disk-like Gel-Cap that is positioned over an incision in the patient's abdomen. The Gel-Cap is constructed of a gel material and has a slit in the center. The slit is normally closed, thus allowing insufflation of the peritoneum. When a hand is placed through the slit, the gel material conforms and seals against the hand, thus preventing loss of peritoneal pressure. The Gel-Cap also accepts placement of trocars around the periphery.

The subject device is designed for surgeons who prefer to perform laparascopic surgery without a hand in the peritoneum. There is no slit, thus simplifying the role of the Gel-Cap to a means of accommodating trocars only. The trocars are clustered near the center of the Gel-Cap which allows the abdominal incision to be smaller. The indications for use changes from "hand access plus trocars" to "trocars only". APPLIED refers to the resulting device as a Single Incision Access System. The new system will be available in diameters of 40-120mm for use in incisions ranging from 15-90mm.

INTENDED USE: For use in patients undergoing laparoscopic surgical procedures to obtain abdominal access by instrumentation. Trocars may be placed through the Gel Seal Cap to allow instrument access to the surgical site. The (device) may be used in procedures such as nephrectomy, colectomy and splenectomy in colorectal, urological and general surgery to access the surgical site.

SUMMARY OF TECHNOLOGICAL CHARACTERISTICS: The predicate and subject device consists of a Gel-Cap and wound retractor. The Gel-Cap is constructed of a semirigid polycarbonate ring and a flexible gel-like material. An attachment lever locks the cap onto the wound retractor.

Image /page/1/Figure/6 description: The image shows two medical devices, the GelPort Hand Access System and the GelPort Single Incision System. Both devices have a circular base with a triangular opening in the center, labeled as 'Gel'. An 'X-Slit' is visible near the base of the GelPort Hand Access System. An 'Attachment Lever' is present on both devices, connecting the base to a 'Gel-Cap'.

{2}------------------------------------------------

K090275

Page 3 of (4)

The wound retractor consists of a thin-film flexible cylinder that has semi-rigid polyurethane rings at each end. The inner ring - which is placed in the patient's abdominal cavity - has a tether that aids removal of the wound retractor at the end of the procedure.

ﺔ ﺗ

Image /page/2/Figure/2 description: The image shows two medical devices called GelPort Hand Access System Wound Retractor and GelPort Single Incision System Wound Retractor. Both devices have an outer ring, an inner ring, and a film. The GelPort Single Incision System Wound Retractor also has a removal tether.

The Gel-Cap attaches to the outer ring of the wound retractor via the attachment lever. The graphic below shows the assembled GelPort system. However, in an actual procedure, the Gel-Cap would be attached to the wound retractor only after the retractor has been placed in the patient's abdomen.

Image /page/2/Figure/4 description: The image contains two medical devices, the GelPort Hand Access System Wound Retractor and Gel-Cap and the GelPort Single Incision System Wound Retractor and Gel-Cap. Both devices appear to be circular with a cylindrical extension. The GelPort Single Incision System Wound Retractor and Gel-Cap has a string attached to it with a ring at the end.

APPLIED plans to market the subject device individually and also in a convenience kit. The kit will include trocars from APPLIED's extensive trocar model family. Those trocars will have a flange at the distal end of the cannula to prevent the cannulas from migrating during instrument exchanges (See graphic, next page).

{3}------------------------------------------------

K0 90.275

Page 4 g(4)

Image /page/3/Picture/2 description: The image shows a round, dark-colored object being held in a person's hand. The object has several white, cylindrical components attached to its top and bottom surfaces. The components appear to be evenly spaced around the circumference of the object. The background is a neutral gray color, which helps to highlight the object and the hand holding it.

Bottom view of Gel-Cap with trocars inserted; white disks are retention flanges

DISCUSSION OF NONCLINICAL TESTS SUBMITTED: There are currently no recognized standards that specify performance characteristics of Single Incision Access Systems. Therefore, APPLIED set up testing to confirm safety and efficacy of the subject device relative to the predicate device of K014047. These tests include determination of:

- Minimum Incision Size the minimum incision required to install the inner ring of the . wound retractor.
- Leak Testing -- the ability of the system to maintain pneumoperitoneum before and after . multiple insertions of trocars through the Gel-Cap.
- Particulate Testing the ability of the system to resist gel particulation caused by . inserting trocars through the Gel-Cap.
- Leak testing after vigorous manipulation of trocars. �

A discussion of the test method and results is in Section 15.

CONCLUSIONS DRAWN FROM TESTING: APPLIED's performance and functional testing demonstrated that the GelPort Single Incision Access System is substantially equivalent to the predicate device of K014047 and introduces no new safety and effectiveness issues.

{4}------------------------------------------------

#### DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" are arranged in a circular pattern around the eagle. The eagle is black, and the text is also black. The logo is simple and recognizable.

#### Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Applied Medical Resources Corporation % Underwriters Laboratories, Inc. Mr. Morten S. Christensen 455 E. Trimble Road San Jose, California 95131-1230

## FEB 1 8 2009

Re: K090275

Trade/Device Name: .. GelPort Single Incision Access .System Regulation Number: 21 CFR 876.1500 -Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: GCJ Dated: February 2, 2009 Received: Febriaru 4, 2009

Dear Mr. Christensen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration ..

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set

{5}------------------------------------------------

Page 2 - Mr. Morten S. Christensen

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarketsuracillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,
David Krone

Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and · Radiological Health

Enclosure

{6}------------------------------------------------

#### INDICATIONS FOR USE

# 510(k) Number (if known): Noteget assigned.

Device Name: GelPort Single Incision Access System.

Indications for Use:

For use in patients undergoing laparoscopic surgical procedures to obtain abdominal access by instrumentation. Trocars may be placed through the Gel Seal Cap to allow instrument access to the surgical site. The device may be used in procedures such as nephrectomy, colectomy and splenectomy in colorectal, wological and general surgery to access.the surgical site. 

Prescription Use *X*
(Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

Page 1 of

Concurrence of CDRH, Office of Device Evaluation (ODE)

1-1

Neil Klossner

(Division Sign-Off) Division of General, Restorative, and Neurological Devices

**510(k) Number** K090275

---

**Source:** [https://fda.innolitics.com/submissions/GU/subpart-b%E2%80%94diagnostic-devices/GCJ/K090275](https://fda.innolitics.com/submissions/GU/subpart-b%E2%80%94diagnostic-devices/GCJ/K090275)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
