SILS PORT, MODELS SILSPT5, SILSPT12

K082619 · Covidien Lp, Formerly Registered AS United States · GCJ · Nov 7, 2008 · Gastroenterology, Urology

Device Facts

Record IDK082619
Device NameSILS PORT, MODELS SILSPT5, SILSPT12
ApplicantCovidien Lp, Formerly Registered AS United States
Product CodeGCJ · Gastroenterology, Urology
Decision DateNov 7, 2008
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2

Indications for Use

For multiple instrument or camera access to the abdominal cavity through a single incision for performing minimally invasive laparoscopic procedures

Device Story

SILS Port is a surgical access device designed for single-incision laparoscopic surgery. It consists of three laparoscopic trocars and an insufflation tube integrated into a flexible port. The device is inserted into a single abdominal incision to provide multiple access points for instruments or cameras while maintaining pneumoperitoneum. It is used by surgeons in an operating room setting. By consolidating access, the device facilitates minimally invasive procedures through a single entry point, potentially reducing the number of incisions required for the surgery.

Clinical Evidence

Bench testing only. In-vitro and in-vivo testing performed to support the intended use of the device.

Technological Characteristics

Flexible port housing three laparoscopic trocars and an insufflation tube. Materials evaluated per ISO 10993-1:2003. Mechanical access device; no energy source or software components.

Indications for Use

Indicated for patients undergoing minimally invasive laparoscopic procedures requiring multiple instrument or camera access to the abdominal cavity through a single incision.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Kob2619 ## 510(K) SUMMARY: 510(k) Summary of Safety and Effectiveness: SUBMITTER: Surgical Devices, a global business unit of Tyco Healthcare Group LP (d/b/a Covidien) 60 Middletown Avenue North Haven, CT 06473 CONTACT PERSON: Robert Zott Program Director, Regulatory Affairs Phone: (203) 492-6013 Fax: (203) 492-5029 DATE PREPARED: TRADE/PROPRIETARY NAME: COMMON/USUAL NAME: CLASSIFICATION NAME: PREDICATE DEVICE(S): September 8, 2008 SILSTM Port Surgical Trocar Endoscope and/or accessories K981941: Dexide™ Multaport Cannula Reducer and Accessories K012539: Versa Step™ Reposable System K954108: Modified Versaport™ Trocar K072814: Convenience Kit for "Single-Incision Laparoscopic Surgery and other advanced laparoscopic procedures." K945457: Auto Suture™ Modified Grip K073170: R-Port II Laparoscopic Access Device Three laparoscopic trocars and an insufflation tube bound by a flexible port. #### DEVICE DESCRIPTION: {1}------------------------------------------------ 9 2619 #### INTENDED USE: ## TECHNOLOGICAL CHARACTERISTICS: #### MATERIALS: For multiple instrument or camera access to the abdominal cavity through a single incision for performing minimally invasive laparoscopic procedures The SILSTM Port provides the ability to use three conventional laparoscopic trocars and an insufflation port through a single incision while providing the ability to maintain pneumoperitoneum. All components of the SILSTM Port are comprised of materials that have been evaluated in accordance with ISO 10993-1: 2003, Biological Evaluation of medical devices --Part 1: Evaluation and Testing. ### PERFORMANCE DATA: In-vitro and in-vivo testing has been performed in support of the intended use of this device. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States of America. The seal features a stylized eagle with its wings spread, symbolizing the federal government. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES" are arranged in a circular pattern around the left side of the eagle, while "USA" is positioned on the right side. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Covidien % Mr. Robert Zott Regulatory Affairs Program Director 60 Middletown Avenue North Haven, Connecticut 06473 NOV - 7 2008 Re: K082619 Trade/Device Name: SILS™ Port Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: GCJ Dated: October 27, 2008 Received: October 28, 2008 Dear Mr. Zott: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {3}------------------------------------------------ Page 2 - Mr. Robert Zott This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket, Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours. Mark M Millman Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ Ko82619 # Indications for Use 510(k) Number (if known): K082619 Device Name: SILS ™ Port Indications For Use: For multiple instrument or camera access to the abdominal cavity through a single incision for performing minimally invasive laparoscopic procedures Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-On) Division of General. Restorative, and Neurological Devices K082619 510(k) Number_ Page 1 of
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...