← Product Code [GCJ](/submissions/GU/subpart-b%E2%80%94diagnostic-devices/GCJ) · K022393

# MERDOC SYSTEM (K022393)

_Stryker Endoscopy · GCJ · Jun 5, 2003 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/GU/subpart-b%E2%80%94diagnostic-devices/GCJ/K022393

## Device Facts

- **Applicant:** Stryker Endoscopy
- **Product Code:** [GCJ](/submissions/GU/subpart-b%E2%80%94diagnostic-devices/GCJ.md)
- **Decision Date:** Jun 5, 2003
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

The Stryker Sidne™ System is indicated for use with compatible endoscopic and general surgery devices. The Stryker Sidne™ System can be used in general laparoscopy, nasopharyngoscopy, ear endoscopy, sinuscopy, and plastic surgery wherever a laparoscope fendoscope/arthroscope is indicated for use. A few examples of the more common endoscopic surgeries are laparoscopic cholecystectorny, laparoscopic hemia repair, laparoscopic appendectorny, laparoscopic pelvic lymph node dissection, laparoscopically assisted hysterectorny, laparoscopic & thorascopic anterior spinal fusion, anterior cruciate ligament reconstruction, knee arthroscopy, shoulder anthroscopy, small joint anthroscopy, decompression fixation, wedge resection, lung biopsy, pleural biopsy, dorsal sympathectorny, pleurodesis, internal mammary artery dissection for coronary artery bypass, coronary artery bypass grafting where endoscopic visualization is indicated and examination of the evacuated cardiac chamber during performance of valve replacement. The users of Sidne™ are general surgeons, gynecologists, cardiac surgeons, thoracic surgeons, plastic surgeons, orthopedic surgeons, ENT surgeons, and urologists.

## Device Story

Sidne™ System is an operating room control center designed for voice-controlled management of surgical equipment settings. It interfaces with compatible endoscopic and general surgery devices, eliminating manual adjustments or reliance on verbal communication between surgical staff. The system is operated by surgeons or OR personnel via voice commands to control the state, selection, and settings of networked medical equipment. By centralizing control, it aims to improve surgical workflow efficiency and reduce manual intervention during procedures.

## Clinical Evidence

No clinical data provided; substantial equivalence is based on technological characteristics and conformance to voluntary safety standards.

## Technological Characteristics

Medical electrical equipment conforming to EN 60601-1-1 (General Safety), EN 60601-1-1 (Collateral Standard for Medical Electrical Systems), EN 60601-1-2 (Electromagnetic Compatibility), and EN 60601-1-4 (Programmable Electrical Medical Systems). Functions as a networked control center for surgical equipment.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

## Predicate Devices

- HERMES OR Control Center ([K991444](/device/K991444.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health and Human Services. The logo features a stylized abstract symbol resembling a person embracing another person. The symbol is positioned to the right of the text, which reads "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA" in a circular arrangement around the symbol.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

Stryker Endoscopy Mr. Kiran Javadekar Senior Design Engineer 5900 Optical Court San Jose, CA 95138

JUL 2 7 2015

Re: K022393 Trade/Device Name: MerdocTM System Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: GCJ, HRX Dated (Date on orig SE ltr): March 31, 2003 Received (Date on orig SE ltr): April 1, 2003

Dear Mr. Javadekar,

This letter corrects our substantially equivalent letter of June 5, 2003.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be

{1}------------------------------------------------

Page 2 -

found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

## Benjamin R. Fisher -S

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

June 3, 2003

510(k) Number (if known): K022393

Device Name: Sidne ™

Indications for Use:

The Stryker Sidne™ System is indicated for use with compatible endoscopic and general surgery devices. The Stryker Sidne™ System can be used in general laparoscopy, nasopharyngoscopy, ear endoscopy, sinuscopy, and plastic surgery wherever a laparoscope fendoscope/arthroscope is indicated for use. A few examples of the more common endoscopic surgeries are laparoscopic cholecystectorny, laparoscopic hemia repair, laparoscopic appendectorny, laparoscopic pelvic lymph node dissection, laparoscopically assisted hysterectorny, laparoscopic & thorascopic anterior spinal fusion, anterior cruciate ligament reconstruction, knee arthroscopy, shoulder anthroscopy, small joint anthroscopy, decompression fixation, wedge resection, lung biopsy, pleural biopsy, dorsal sympathectorny, pleurodesis, internal mammary artery dissection for coronary artery bypass, coronary artery bypass grafting where endoscopic visualization is indicated and examination of the evacuated cardiac chamber during performance of valve replacement. The users of Sidne™ are general surgeons, gynecologists, cardiac surgeons, thoracic surgeons, plastic surgeons, orthopedic surgeons, ENT surgeons, and urologists.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)

iriam C. Provost

Division Sign-Off) (Division of General, Restorative

510(k) Number K022393

Prescription Use
(Per 21 CFR 801.109)

OR

Over-The-Counter-Use

p.3

Stryker Endoscopy Sidne™ System 510(k) submission

{3}------------------------------------------------

K022393

JUN - 5 2003

ax 1408) 754-250

## 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS

| Classification Name:   | Laparoscope, General and Plastic Surgery |
|------------------------|------------------------------------------|
| Common and Usual Name: | Sidne™                                   |
| Proprietary Name:      | Sidne™ System                            |

This 510(k) summary and effectiveness is being submitted in accordance with requirements of SMDA 1990

The Sidne™ System is substantially equivalent in safety and efficacy as the currently marketed HERMES OR Control Center. The HERMES OR Control Center was cleared under 510K# 991444

The Stryker Sidne™ System is a medical device that is designed to allow the surgeon to control the state, selection, and settings of the surgical equipment it Is networked with. Sidne™ is compatible with existing devices that are HERMES OR Control Center compatible.

The intent of the Sidne™ Control Center is to allow for voice control of medical device settings by the surgeon or operating room personnel, thereby eliminating the need for manual operation of those devices compatible with Sidne™ , or relying upon verbal communications between the surgeon and other personnel in the operating room in order to adjust the surgical equipment. The Sidne™ System is intended for use in general laparoscopy, nasopharyngoscopy, ear endoscopy, sinuscopy, and plastic surgery wherever a laparoscope lendoscope/arthroscope is indicated for use.

The Sidne™ System conforms to the following voluntary standards: EN 60601-1-1 Medical Electrical Equipment Part 1: General Requirements for Safety; EN 60601-1-1 Collateral Standard: Safety Requirements for Medical Electrical Systems; EN 60601-1-2 Collateral Standard: Electromagnetic Compatiblity - Requirements and Tesls; EN 60601-1-4 Collateral Standard: Programmable Electrical Medical Systems.

The technological differences between the Stryker Sidne™ System and the predicate HERMES device do not raise new issues of safety and efficacy of the predicate device. Therefore, the Stryker Sidne™ System is substantially equivalent to the currently marketed Computer Motion HERMES™ OR Control Center.

jivadikar

Kıran Javadekar Senior Design Engineer Stryker Endoscopy

Date: 6/03/2003

Stryker Endoscopy Sidne™ System 510(k) submission

---

**Source:** [https://fda.innolitics.com/submissions/GU/subpart-b%E2%80%94diagnostic-devices/GCJ/K022393](https://fda.innolitics.com/submissions/GU/subpart-b%E2%80%94diagnostic-devices/GCJ/K022393)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
