ZEUS MICROWRIST ROBOTIC SURGICAL SYSTEM

K013551 · Computer Motion, Inc. · GCJ · Mar 28, 2002 · Gastroenterology, Urology

Device Facts

Record IDK013551
Device NameZEUS MICROWRIST ROBOTIC SURGICAL SYSTEM
ApplicantComputer Motion, Inc.
Product CodeGCJ · Gastroenterology, Urology
Decision DateMar 28, 2002
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The ZEUS® MicroWrist™ Robotic Surgical System is intended to assist in the accurate control of blunt dissectors, retractors, atraumatic graspers and stabilizers during laparoscopic and thoracoscopic surgery. The users of the ZEUS® MicroWrist™ Robotic Surgical System are surgeons trained in minimally invasive surgery. Assistant surgeons can also use this system while under the direction of the primary surgeon in accordance with hospital's customary practice.

Device Story

Robotic surgical system; assists surgeons in controlling laparoscopic/thoracoscopic instruments (blunt dissectors, retractors, atraumatic graspers, stabilizers). System provides accurate control during minimally invasive procedures; operated by trained surgeons or assistant surgeons under primary surgeon direction. Enhances precision in instrument manipulation; facilitates complex maneuvers within confined surgical fields. Used in OR settings; provides mechanical assistance to improve surgical outcomes.

Clinical Evidence

No clinical data provided; substantial equivalence based on bench testing and adherence to voluntary standards including EN60601-1, EN60601-1-1, EN60601-2-18, UL 2601-1, and electromagnetic compatibility standards (EN55011, EN61000 series).

Technological Characteristics

Robotic surgical system for laparoscopic/thoracoscopic use. Complies with EN60601-1 (medical electrical equipment), EN60601-1-1, EN60601-2-18, UL 2601-1, and CAN/CSA-C22.2 standards. Features electromagnetic compatibility per EN55011/CISPR 11 and EN61000 series immunity tests. System architecture designed for tele-manipulation of surgical instruments.

Indications for Use

Indicated for use by surgeons trained in minimally invasive surgery to assist in the accurate control of blunt dissectors, retractors, atraumatic graspers, and stabilizers during laparoscopic and thoracoscopic surgery.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ 01355/ ## Premarket Notification 510(k) Safety and Effectiveness Summary 11. ## ZEUS® MicroWrist™ Robotic Surgical System 510(k) Summary In accordance with 21 CFR section 807.92 Computer Motion is submitting the following safety and effectiveness summary. 1) Submitter Information Computer Motion. Inc. 130-B Cremona Drive Goleta, CA 93117 Contact: David U. Thomas Prepared: October 23, 2001 2) Name of Device: Proprietary Name: ZEUS® MicroWrist™ Robotic Surgical System Common Name is ZEUS MicroWrist System Classification Name: Laparoscope for Use in General and Plastic Surgery, Regulation Number 876.1500, Class II. 3) Substantially equivalent to ZEUS® Robotic Surgical System K003431 - 4) The ZEUS® MicroWrist™ Robotic Surgical System is intended to assist in the accurate control of blunt dissectors, retractors, atraumatic graspers and stabilizers during laparoscopic and thoracoscopic surgery. The users of the ZEUS® MicroWrist™ Robotic Surgical System are surgeons trained in minimally invasive surgery. Assistant surgeons can also use this system while under the direction of the primary surgeon in accordance with hospital's customary practice. - 5) The ZEUS MicroWrist System is designed and tested to the following Computer Motion and voluntary standards. EN60601--1 Second Edition 1994 International Standard for Medical Electrical Equipment EN60601-1-1 1994 International Standard for Medical Electrical Equipment EN60601-2-18 First Edition 1997 International Standard for Medical Electrical Equipment UL 2601-1 Conducted & Radiated Emission EN55011/A1: 1998; CISPR 11:1997 Immunity Tests EN61000-4-2: 1995; EN61000-4-3: 1996; EN50204:1996; EN61000-4-4:1995; EN61000-4-5:1995; EN61000-4-6:1996. CAN/CSA-C22.2 NO. 601.1-M90 & NO. 601.2.18-92 {1}------------------------------------------------ Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 MAR 2 8 2002 Computer Motion, Inc. Mr. Douglas Bueschel Director, Regulatory Affairs and Quality Assurance 130 B Cremona Drive, Suite B Goleta. California 93117 Re: K013551 Trade Name: Zeus Microwrist Robotic Surgical System Regulation Number: 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: GCJ Dated: December 28, 2001 Received: January 3, 2002 Dear Mr. Bueschel: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {2}------------------------------------------------ ## Page 2 - Mr. Douglas Bueschel This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally presidence modicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, «Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours, uriam C. Provost Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ ## 510(k) Number (if known): K013551 Device Name: Microwrist™ ZEUS® Robotic Surgical System Indication For Use: The ZEUS® Microwrist™ Robotic Surgical System is intended to assist in the accurate control of blunt dissectors, retractors, atraumatic graspers and stabilizers during laparoscopic and thoracoscopic surgery. The users of the ZEUS® Microwrist™ Robotic Surgical System are surgeons trained in minimally invasive surgery. Assistant surgeons can also use this system while under the direction of the primary surgeon in accordance with hospital's customary practice. (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE, IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) OR Prescription Use (Per 21 CFR 801.109) Over-The-Counter Use (Optional Format) Miriam C. Provost (Division Sign-Off) Division of General, Restorative and Neurological Devices 510(k) Number K013551
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...