Veress Needle

K172120 · Wickimed (Huizhou) Medical Equipment Manufacturing Co.,Ltd · FHO · Sep 14, 2017 · Gastroenterology, Urology

Device Facts

Record IDK172120
Device NameVeress Needle
ApplicantWickimed (Huizhou) Medical Equipment Manufacturing Co.,Ltd
Product CodeFHO · Gastroenterology, Urology
Decision DateSep 14, 2017
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2

Indications for Use

The Veress Needle is intended for percutaneous insertion into the peritoneal cavity for the purpose of insufflation with carbon dioxide to establish pneumoperitoneum prior to the placement of trocars during laparoscopic procedures.

Device Story

Sterile, single-use Veress needle; 120 mm and 150 mm lengths; incorporates spring-loaded blunt stylet mechanism. Used by surgeons in laparoscopic procedures to access the peritoneal cavity. Device connects to an external insufflator via tubing; carbon dioxide is delivered through the needle to create pneumoperitoneum, facilitating subsequent trocar and cannula insertion. Benefits include safe, controlled access to the abdominal cavity for minimally invasive surgery.

Clinical Evidence

Bench testing only. Biocompatibility testing performed per ISO 10993-1, -5, -7, -10, -11, and -12 (cytotoxicity, sensitization, intracutaneous reactivity, acute systemic toxicity, material-mediated pyrogenicity). Sterilization validation per ISO 11135:2014. Performance testing included tip pull, switch operation, spring obturator operation, and needle puncture force tests. All results met pre-defined acceptance criteria.

Technological Characteristics

Materials: SUS 304 (needle), PC (handle). Spring-loaded blunt stylet mechanism. Dimensions: 120 mm and 150 mm lengths. Energy source: None (mechanical). Sterilization: Ethylene Oxide (EO). Single-use.

Indications for Use

Indicated for patients undergoing laparoscopic procedures, including gynecological laparoscopy, requiring percutaneous insertion into the peritoneal cavity for carbon dioxide insufflation to establish pneumoperitoneum.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing right, with flowing lines representing hair or movement. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 September 14, 2017 WickiMed (Huizhou) Medical Equipment Manufacturing Co.,Ltd. Haobin Li General Manager Tang Jiao Wang Street, Lilin Town, Zhongkai Hi-Tech Zone Huizhou, Guangdong 516000 China Re: K172120 Trade/Device Name: Veress Needle, models: WVN0112T, WVN0115T Regulation Number: 21 CFR§ 884.1730 Regulation Name: Laparoscopic insufflator Regulatory Class: II Product Code: HIF, FHO Dated: July 7, 2017 Received: July 13, 2017 Dear Haobin Li: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely, # Benjamin R. Fisher -S Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ### Indications for Use 510(k) Number (if known) K172120 Device Name Veress Needle, models :WVN0112T,WVN0115T Indications for Use (Describe) The Veress Needle is intended for percutaneous insertion into the peritoneal cavity for the purpose of insufflation with carbon dioxide to establish pneumoperitoneum prior to the placement of trocars during laparoscopic procedures. Type of Use (Select one or both, as applicable) | <span style="text-decoration: underline;">Prescription Use (Part 21 CFR 801 Subpart D)</span> | <span style="text-decoration: underline;">Over The Counter Use (21 CFR 801 Subpart C)</span> | |-----------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------| | <div style="display:inline-block; vertical-align:middle;">☑</div> | <div style="display:inline-block; vertical-align:middle;">☐</div> | |× | Prescription Use (Part 21 CFR 801 Subpart D) __ Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW * The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ### 510(K) SUMMARY This summary of 510(K) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR §807.92. Type of submission :Traditional The assigned 510(K) number is: K172120 The date the summary was prepared: September 11, 2017 ## 1. Submitter information: Manufacturer Name: WickiMed(Huizhou)Medical Equipment Manufacturing Co.,Ltd. Address: TangJiao XingWang Street, LiLin Town, ZhongKai Hi-Tech Zone, HuiZhou,GuangDong, China. Tel : 0086-0752-3860807 Fax : 0086-0752-3863017 Establishment Registration Number:3010601992 ### 2. Contact person: Haobin Li (General Manager) WickiMed(Huizhou)Medical Equipment Manufacturing Co.,Ltd TangJiao XingWang Street, LiLin Town, ZhongKai Hi-Tech Zone, HuiZhou,GuangDong, China. Tel : 0086-0752-3860807 Fax : 0086-0752-3863017 E-mail: mac lai@wickimed.com ### 3. Identification of the Device : Trade Name: Veress Needle Model: WVN0112T,WVN0115T Common Name: Veress Needle {4}------------------------------------------------ | Classification Name | Product<br>Code | Regulation<br>Number | Regulatory<br>Class | Review Panel | |------------------------------|-----------------|----------------------|---------------------|---------------------------| | Insufflator,<br>Laparoscopic | HIF<br>FHO | 21CFR<br>884.1730 | II | Obstetrics/Gynec<br>ology | #### 4. Identification of the Predicative Device | Device<br>Name | Common<br>Name | Manufacturer | Classification<br>and Code | Classification<br>regulation | 510(k)<br>number | |------------------------------|------------------|--------------------------------------------------------|----------------------------|------------------------------|------------------| | Unimicro<br>Veress<br>Needle | Veress<br>Needle | Unimicro Medical<br>Systems<br>(ShenZhen) Co.,<br>Ltd. | Class II,<br>HIF | 21CFR<br>884.1730 | K150068 | | | | | Class II,<br>FHO | 21CFR<br>876.1500 | | There are no design-related recall records with Predicative Device K150068. #### 5. Intended Use and Indications for Use of the subject device The Veress Needle is intended for percutaneous insertion into the peritoneal cavity for the purpose of insufflation with carbon dioxide to establish pneumoperitoneum prior to the placement of trocars during laparoscopic procedures. #### 6. Device Description The Veress Needle is a sterile and single-use product. It incorporated a spring loaded blunt stylet mechanism similar to the needle. It is used to establish peritoneum prior to trocar and cannula insertion in laparoscopic procedures. The Veress Needle, available in 120 mm and 150 mm lengths, has applications in gynecological laparoscopy and other laparoscopic procedures. #### 7. Substantial Equivalence Determination The Veress Needle submitted in this 510(k) file is substantially equivalent to the cleared Unimicro Veress Needle which is the subject of K150068. {5}------------------------------------------------ The comparison to the predicate device is provided below in Table 2. | Item | Predicate Device | Proposed Device | |------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Trade Name | Unimicro Veress Needle | Veress Needle | | 510(K) Submitter | Unimicro Medical Systems<br>(ShenZhen) Co.,Ltd. | WickiMed(Huizhou)Medical<br>Equipment Manufacturing<br>Co.,Ltd | | 510(K) Number | K150068 | - | | Classification<br>regulation | 21CFR 884.1730 | 21CFR 884.1730 | | Classification and<br>Product Code | Class II,<br>HIF,FHO | ClassII,<br>HIF,FHO | | Common name | Veress Needle | Veress Needle | | Intended Use | The Unimicro Veress Needle<br>is intended for percutaneous<br>insertion into the peritoneal<br>cavity for the purpose of<br>insufflation with carbon<br>dioxide to establish<br>pneumoperitoneum prior to<br>the placement of trocars<br>during laparoscopic<br>procedures. | The Veress Needle is intended<br>for percutaneous insertion<br>into the peritoneal cavity for<br>the purpose of insufflation<br>with carbon dioxide to<br>establish pneumoperitoneum<br>prior to the placement of<br>trocars during laparoscopic<br>procedures. | | Components | Veress Needle<br>Obturator | Veress Needle<br>Obturator | | Models | MND 11200,MND 11500 | WVN0112T,WVN0115T | | Length | 120mm,150mm | 120mm,150mm | | Sterilization | EO Sterilized | EO Sterilized | | Disposable | Yes | Yes | | Materials | Patient-Contacting Material<br>SUS 420 | Patient-Contacting Material<br>SUS 304 | #### Table 2 : Comparison to Predicate Device {6}------------------------------------------------ | | ( Outer tube<br>inner tube ) | | ( Outer tube<br>inner tube ) | | |----------------------------|---------------------------------------------------------------------------------------------------------------------------------------------|-----|---------------------------------------------------------------------------------------------------------------------------------------------|----| | | Handle<br>Material | ABS | Handle<br>Material | PC | | Principles of<br>operation | Connect the device to the<br>insufflators with insufflation<br>tubing, insufflating with<br>carbon dioxide to establish<br>pneumoperitoneum | | Connect the device to the<br>insufflators with insufflation<br>tubing, insufflating with<br>carbon dioxide to establish<br>pneumoperitoneum | | The subject and predicate device have the same intended use. The subject and predicate device designs are nearly identical. Both are single-use devices and the lengths of the devices are identical. The differences in technological characteristics between the subject and predicate devices (i.e., different materials, and different needle inner diameter) do not raise different questions of safety and effectiveness. #### 8. Non-clinical Testing A series of tests were performed to assess the safety and effectiveness of the Veress Needle. Biocompatibility tests were conducted in accordance with ISO 10993-1:2009, ISO 10993-5:2009, ISO 10993-7:2008, ISO 10993-10:2010, ISO 10993-11:2006, and ISO 10993-12:2012. The biocompatibility tests included Cytotoxicity, Sensitization, Intracutaneous reactivity, Acute systemic toxicity and Material-Mediated Pyrogenicity. Sterilization validation was performed per ISO 11135:2014. The tests listed below evaluated the performance of the subject device. - Tip Pull Test - Switch Operation - Spring Obturator Operation - Needle Puncture Force Test {7}------------------------------------------------ All the test results demonstrate the performance of Veress Needle meets the requirements of its pre-defined acceptance criteria and intended uses. The results of the non-clinical testing demonstrate that the Veress Needle is as safe and effective as the predicate device. #### 9. Conclusion Based on the results of the above described performance testing data, it can be concluded that Veress Needle is as safe and effective as the predicate device.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...