Medical Video Endoscope [Standard Deflection] (URS3005); Medical Video Endoscope [Reverse Deflection] (URS3005E); Medical Video Endoscope [Standard Deflection] (URS3006); Medical Video Endoscope [Reverse Deflection] (URS3006E); Medical Video Endoscope [Standard Deflection] (URS3016); Medical Video Endoscope [Reverse Deflection] (URS3016E); Image processor (DIS8000)
Medical Video Endoscope [Standard Deflection] (URS3005); Medical Video Endoscope [Reverse Deflection] (URS3005E); Medical Video Endoscope [Standard Deflection] (URS3006); Medical Video Endoscope [Reverse Deflection] (URS3006E); Medical Video Endoscope [Standard Deflection] (URS3016); Medical Video Endoscope [Reverse Deflection] (URS3016E); Image processor (DIS8000)
Applicant
Seplou (Zhuhai) Co., Ltd.
Product Code
FGB · Gastroenterology, Urology
Decision Date
Aug 26, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.1500
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Medical Video Endoscope is designed to be used with endo-therapy accessories such as a biopsy forceps and other ancillary equipment for endoscopy and endoscopic surgery within urinary tract and interior of the kidney. The image processor is used with the specified endoscope designed by SEPLOU during minimally invasive surgery. The image processor provides power and processes the images for medical electronic endoscope.
Device Story
System comprises flexible, single-use video endoscopes (standard or reverse deflection) and reusable image processor (DIS8000). Endoscope features on-tip CMOS camera and LED illumination; control body includes articulation controls and accessory ports. Processor provides power to endoscope, processes video signals, and outputs to display. Used in hospital settings by physicians for minimally invasive urinary tract and kidney procedures. Enables visualization for diagnostic or therapeutic interventions using ancillary tools like biopsy forceps or laser fibers. Benefits include real-time visualization for precise surgical navigation and reduced risk of cross-contamination due to single-use scope design.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench performance testing, including optical performance (ISO 8600), mechanical testing, electrical safety (IEC 60601-1, 60601-2-18), EMC (IEC 60601-1-2), and biocompatibility (ISO 10993).
Technological Characteristics
Flexible ureteroscope with CMOS sensor and LED illumination. Materials biocompatibility per ISO 10993. Sterilization via EO (ISO 11135). Shelf life 3 years (ASTM F1980). Compliance with ISO 8600-1, 3, 4 for optical performance. Electrical safety per IEC 60601-1/2-18; EMC per IEC 60601-1-2. System includes dedicated image processor (DIS8000).
Indications for Use
Indicated for adult patients undergoing endoscopy and endoscopic surgery within the urinary tract and interior of the kidney, using endo-therapy accessories.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
Build Correspondence
Convert to PDF
August 26, 2025
Seplou (ZHUHAI) Co., Ltd.
% Jie Yang
Consultant
Chonconn Consulting Co., Ltd.
Room 504, Block C, No. 1029 Nanhai Avenue, Nanshan District
Shenzhen, Guangdong
CHINA
Re: K250049
Trade/Device Name: Medical Video Endoscope [Standard Deflection] (URS3005);
Medical Video Endoscope [Reverse Deflection] (URS3005E);
Medical Video Endoscope [Standard Deflection] (URS3006);
Medical Video Endoscope [Reverse Deflection] (URS3006E);
Medical Video Endoscope [Standard Deflection] (URS3016);
Medical Video Endoscope [Reverse Deflection] (URS3016E);
Image processor (DIS8000)
Regulation Number: 21 CFR 876.1500
Regulation Name: Endoscope and accessories
Regulatory Class: II
Product Code: FGB
Dated: July 21, 2025
Received: July 21, 2025
Dear Jie Yang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K250049 - Jie Yang
Page 2
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-
{2}
K250049 - Jie Yang
Page 3
devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
MARK J. ANTONINO -S
Mark J. Antonino, M.S.
Assistant Director
DHT3B: Division of Reproductive, Gynecology, and Urology Devices
OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Submission Number (if known)
K250049
Device Name
Medical Video Endoscope [Standard Deflection] (URS3005);
Medical Video Endoscope [Reverse Deflection] (URS3005E);
Medical Video Endoscope [Standard Deflection] (URS3006);
Medical Video Endoscope [Reverse Deflection] (URS3006E);
Medical Video Endoscope [Standard Deflection] (URS3016);
Medical Video Endoscope [Reverse Deflection] (URS3016E);
Image processor (DIS8000)
Indications for Use (Describe)
The Medical Video Endoscope is designed to be used with endo-therapy accessories such as a biopsy forceps and other ancillary equipment for endoscopy and endoscopic surgery within urinary tract and interior of the kidney.
The image processor is used with the specified endoscope designed by SEPLOU during minimally invasive surgery.
The image processor provides power and processes the images for medical electronic endoscope.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
“An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number.”
{4}
K250049
Page 1 of 5
# 510(K) Summary
Prepared in accordance with the requirements of 21 CFR Part 807.92 on August 18, 2025
## 1. Submission sponsor
Name: SEPLOU (ZHUHAI) CO., LTD.
Address: 5th Floor, Building 1, No. 8, Keji 10th Road, Tangjiawan Town, High-tech Zone, 519085, Zhuhai, Guangdong, P.R. China
Contact person: Zeng QingFu
Title: Regulatory Affairs Manager
E-mail: Vince@seplou.com
Tel: 86-0756-8858005
## 2. Submission correspondent
Name: Chonconn Consulting Co., Ltd.
Address: Room 504, Block C, No. 1029 Nanhai Avenue, Nanshan District, Shenzhen, Guangdong, P.R. China
Contact person: Yang Jie
E-mail: yangjie@chonconn.com
Tel: +86-755 33941160
## 3. Subject Device Information
| Trade/Device Name | Medical Video Endoscope [Standard Deflection] (URS3005); Medical Video Endoscope [Reverse Deflection] (URS3005E); Medical Video Endoscope [Standard Deflection] (URS3006); Medical Video Endoscope [Reverse Deflection] (URS3006E); Medical Video Endoscope [Standard Deflection] (URS3016); Medical Video Endoscope [Reverse Deflection] (URS3016E); Image processor (DIS8000) |
| --- | --- |
| Common Name | Endoscope and accessories |
| Regulatory Class | Class II |
| Classification | 21CFR 876.1500 / Ureteroscope And Accessories, Flexible/Rigid / FGB |
| Submission type | Traditional 510(K) |
## 4. Predicate Device
510(k) number: K172098
Product name: Medical Video Endoscope System
Product code: FGB
1 / 5
{5}
K250049
Page 2 of 5
# 5. Device Description
The Medical Video Endoscope system is designed to provide image solution for endoscopy and endoscopic surgery, and perform procedures in the urinary tract and interior of the kidney using appropriate accessory devices (e.g. laser fibers, forceps baskets).
The Medical Video Endoscope is a sterile Medical Video Endoscope. The Image Processor for Endoscopy is a reusable monitor.
The Medical Video Endoscope is comprised of a control body with articulation controls and accessory access ports, and a flexible insertion tube with an on-tip camera module and LED lighting source. The Image Processor processes the images from the Endoscope and outputs video signals to a display.
# 6. Indications for use
The Medical Video Endoscope is designed to be used with endo-therapy accessories such as a biopsy forceps and other ancillary equipment for endoscopy and endoscopic surgery within urinary tract and interior of the kidney.
The image processor is used with the specified endoscope designed by SEPLOU during minimally invasive surgery.
The image processor provides power and processes the images for medical electronic endoscope.
# 7. Comparison to the Predicate Device
| Features | Subject Device
Medical Video Endoscope:
[Standard Deflection]:
URS3005, URS3006, URS3016
[Reverse Deflection]:
URS3005E, URS3006E, URS3016E
Image Processor: DIS8000 | Predicate Device
Medical Video Endoscope System | Comparison |
| --- | --- | --- | --- |
| K number | K250049 | K172098 | / |
| Product Code | FGB | FGB | Same |
| Regulation No. | 876.1500 | 876.1500 | Same |
| Device trade name | Medical Video Endoscope
Image Processor | Medical Video Endoscope System | / |
| Indication for use | The Medical Video Endoscope is designed to be used with | This instrument has been | Same |
{6}
K250049
Page 3 of 5
| | endo-therapy accessories such as a biopsy forceps and other ancillary equipment for endoscopy and endoscopic surgery within urinary tract and interior of the kidney. The image processor is used with the specified endoscope designed by SEPLOU during minimally invasive surgery. The image processor provides power and processes the images for medical electronic endoscope. | designed to be used with endo-therapy accessories such as a biopsy forceps and other ancillary equipment for endoscopy and endoscopic surgery within urinary tract and interior of the kidney. | |
| --- | --- | --- | --- |
| Anatomical Site | Urinary tract and interior of the kidney | Urinary tract and interior of the kidney | Same |
| Target population | Adults | Adults | Same |
| Where used | Hospital | Hospital | Same |
| **Medical Video Endoscope** | | | |
| Scope type | Flexible | Flexible | Same |
| Single use/ Reuse | Single use | Single use | Same |
| Sterile | Yes | Yes | Same |
| Sterilization Method | EO sterilization | EO sterilization | Same |
| SAL | 10^{-6} | 10^{-6} | Same |
| Field of view | 120° | 120° | Same |
| Direction of view | 0° | 0° | Same |
| Sensor type | CMOS | CMOS | Same |
| Illumination source | LED | LED | Same |
| Max. outer diameter of insertion section | 3.0mm
2.5mm
2.2mm | 3.2mm | Similar |
| Up/down deflection | Up:275° Down: 275° | Up:270° Down: 270° | Similar |
| Work length | 650mm | 650mm | Same |
| Minimum instrument channel width | 1.2mm | 1.0mm | Similar |
3 / 5
{7}
K250049
Page 4 of 5
| Product Performance | Comply with ISO 8600-1, ISO 8600-3, ISO 8600-4 | Comply with ISO 8600-1, ISO 8600-3, ISO 8600-4 | Same |
| --- | --- | --- | --- |
| Biocompatibility | Cytotoxicity: ISO 10993-5
Sensitization: ISO 10993-10
Irritation: ISO 10993-10
Acute Systemic Toxicity: ISO 10993-11
Material-mediated Pyrogens: ISO 10993-11 | Cytotoxicity: ISO 10993-5
Sensitization: ISO 10993-10
Irritation: ISO 10993-10 | Different |
| Image system | Image Processor DIS8000 | UTV100 as the image system | Same |
## 8. Non-clinical Data
The following performance data were provided in support of the substantial equivalence determination.
### Biocompatibility testing
Biocompatibility of the Medical Video Endoscope was evaluated in accordance with the FDA guidance “Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" and International Standard ISO 10993-1 “Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing Within a Risk Management Process,” as recognized by FDA.
### Sterilization and shelf life testing
The Medical Video Endoscope is provided sterile and its shelf-life is 3 years.
Sterilization Process has been validated accordance with ISO 11135:2014.
EO/ECH residual test was performed according to ISO 10993-7:2008.
The shelf life is determined based on optical testing and product performance testing after accelerated aging test according to ASTM F1980-21 Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices.
Package validation was conducted according to ISO 11607-1:2019 and ISO 11607-2:2019, and ASTM F1886/F1886M-16, ASTM F88/F88M-23, ASTM F1929-23.
### Electrical safety and electromagnetic compatibility (EMC)
Electrical safety and EMC testing were conducted on the Medical Video Endoscope system. The system complies with the IEC 60601-1 and IEC60601-2-18 for safety and the IEC 60601-1-2 for EMC.
{8}
K250049
Page 5 of 5
## Software Verification and Validation Testing
Software verification and validation testing were conducted, and documentation was provided as recommended by FDA’s Guidance for Industry and FDA Staff, “Content of Premarket Submissions for Device Software Functions”.
## Bench performance testing
The following bench tests were performed:
1. Optical performance testing according to ISO 8600 series.
2. Color performance (color reproduction), geometric distortion, optical performance (resolution, depth of field and image intensity uniformity), Noise and dynamic range compared with the predicate device.
3. Mechanical testing including use-life of bending section and control knob, connection strength, peak tensile force and corrosion resistance etc.
## 9. Clinical study
Not applicable.
## 10. Conclusion
Performance testing and compliance with voluntary standards demonstrate that the proposed Medical Video Endoscope system is substantially equivalent to the predicate device.