Single Use Suction Access Ureteroscope (PU400A, PU411A), Single Use Flexible Video Ureteroscope (PU3033H, PU3033AH)
Applicant
Zhuhai Pusen Medical Technology Co., Ltd.
Product Code
FGB · Gastroenterology, Urology
Decision Date
Jul 16, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.1500
Device Class
Class 2
Indications for Use
This instrument has been designed to be used with Endo-Therapy accessories such as a biopsy forceps and other ancillary equipment for endoscopy and endoscopic surgery within urinary tract and interior of the kidney. This instrument is also used as a suction catheter that establishes a conduit used for irrigation of kidney stones and stone dust during ureteral lithotripsy.
Device Story
Single-use flexible video ureteroscope; provides visualization and suction access for urological procedures. Inputs: optical signals captured by distal CMOS sensor; LED illumination. Device connects to external HD Medical Video Endoscope Image Processor (PV300) for image processing and power. Output: real-time video feed displayed to physician. Used in hospitals/medical institutions by trained clinicians. Facilitates stone management via irrigation/aspiration conduit and accessory channel for biopsy forceps. Benefits: single-use design eliminates reprocessing requirements; provides visualization and stone removal capability during lithotripsy.
Clinical Evidence
No clinical or animal studies were performed. Substantial equivalence is supported by bench testing, including biocompatibility (ISO 10993-1), electrical safety (IEC 60601-1, IEC 60601-2-18), EMC (IEC 60601-1-2), mechanical/optical performance (ISO 8600), image quality comparison, photobiological safety (IEC 62471), and luer taper testing (ISO 80369-7).
Technological Characteristics
Flexible, single-use ureteroscope. CMOS imaging sensor; LED illumination. Powered by external processor (PV300). Dimensions: 2.7-3.2 mm insertion width; 650-680 mm working length; >=1.2 mm or >=5.1 Fr working channel. Deflection: 270° up/down. Sterilization: Ethylene Oxide (EO) to SAL 10^-6. Connectivity: Wired connection to image processor.
Indications for Use
Indicated for use in endoscopic surgery and endoscopy within the urinary tract and kidney interior, including irrigation and aspiration of kidney stones and stone dust during ureteral lithotripsy, using Endo-Therapy accessories.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
K241956 — Single-Use Digital Flexible Ureteroscope (PUR12-1, PUR12-2, PUR12-3, PUR12-4, PUR12-5, PUR12-6, PUR12-S, PUR12-S1, PUR12-S2) · Hunan Endoso Life Technology Co., Ltd. · Jan 22, 2025
K230200 — Single-Use Flexible Ureteroscope · Hunan Vathin Medical Instrument Co., Ltd. · May 18, 2023
K250785 — Single-use Flexible Ureteroscope · Hunan Vathin Medical Instrument Co., Ltd. · May 6, 2025
K233275 — RenaFlex Single-use Flexible Ureteroscope (SUURF-V Standard); RenaFlex Single-use Flexible Ureteroscope (SUURF-VR Reverse); Video System Center for Single-use Endoscopes (CV-S1) · Gyrus Acmi, Inc. · Mar 25, 2024
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
July 16, 2024
Zhuhai Pusen Medical Technology Co., Ltd. Changshen Wang Regulatory Affairs Director 5/F, Building 1, No. 33, Ke Ji San Road, High-tech zone, TangJiaWan Town Zhuhai, Guangdong 519000 CHINA
Re: K233778
Trade/Device Name: Single Use Suction Access Ureteroscope (PU400A, PU411A), Single Use Flexible Video Ureteroscope (PU3033H, PU3033AH) Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: FGB, FED Received: June 18, 2024
Dear Changshen Wang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the
{2}------------------------------------------------
Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Mark R. Kreitz -S
for Mark J. Antonino, M.S. Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
### Indications for Use
510(k) Number (if known) K233778
### Device Name
Single Use Suction Access Ureteroscope (PU400A, PU411A); Single Use Flexible Video Ureteroscope (PU3033H, PU3033AH)
### Indications for Use (Describe)
This instrument has been designed to be used with Endo-Therapy accessories such as a biopsy forceps and other ancillary equipment for endoscopy and endoscopic surgery within urinary tract and interior of the kidney. This instrument is also used as a suction catheter that establishes a conduit used for irrigation of kidney stones and stone dust during ureteral lithotripsy.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|
| <span style="font-family: DejaVu Sans, sans-serif">✘</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: DejaVu Sans, sans-serif">□</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary
# I. SUBMITTER
Submitter's name: Zhuhai Pusen Medical Technology Co. Ltd.
Submitter's address: 5/F, Building 1, No 33, Keji San Road, High-tech Zone, Tangjiawan Town, 519085 Zhuhai, Guangdong, People's Republic of China.
Phone: +86 756 688 0865
Contact Person: Ellen Wang
Date Prepared: July 16, 2024
# II. DEVICE
Name of Device: Pusen Single Use Flexible Video Ureteroscope (Model: PU3033H, PU3033AH) and Pusen Single Use Suction Access Ureteroscope (Model: PU400A, PU411A) Common Name: Ureteroscope and Accessories 510(k) number: K233778 Classification Name: Ureteroscope and Accessories, Flexible/Rigid (21 CFR 876.1500) Regulatory Class: Class II Product Code: FGB
### III. PREDICATE DEVICE
### Predicate device
Name of Device: Medical Video Endoscope system
510(k) number: K172098
Classification regulation: Ureteroscope and Accessories, Flexible/rigid
Product code: FGB
This predicate device has not been subject to a design-related recall.
{5}------------------------------------------------
### Reference device
Name of Device: ClearPetra Suction-Evacuation Sheath 510(k) number: K203119 Classification regulation: Endoscope and Accessories Product code: FED, FAJ, FGA This reference device has not been subject to a design-related recall.
## IV. DEVICE DESCRIPTION
The Pusen Single Use Flexible Video Ureteroscope (Model: PU3033H, PU3033AH) and Pusen Single Use Suction Access Ureteroscope (Model: PU400A, PU411A) have been designed to be used with Endo-Therapy accessories such as a biopsy forceps and other ancillary equipment for endoscopy and endoscopic surgery within urinary tract and interior of the kidney. This instrument is also used as a suction catheter that establishes a conduit used for irrigation and aspiration of kidney stones and stone dust during ureteral lithotripsy. This product is for use in a hospital or qualified medical institution.
- Pusen Single Use Flexible Video Ureteroscope and Pusen Single Use Suction Access Ureteroscope are provided sterile with the following model differences:
| Product name | Single Use<br>Ureteroscope | Suction Access | Single Use<br>Ureteroscope | Flexible Video |
|----------------|----------------------------|----------------|----------------------------|----------------|
| Model | PU400A | PU411A | PU3033H | PU3033AH |
| Suction access | Yes | Yes | Yes | Yes |
| Buttons | Yes | Yes | No | Yes |
- . The Pusen Single Use Flexible Video Ureteroscope and Pusen Single Use Suction Access Ureteroscope have two LEDs and CMOS imaging sensor at its distal tip. The Pusen Single Use Flexible Video Ureteroscope and Pusen Single Use Suction Access Ureteroscope need to be connected with the HD Medical Video Endoscope Image Processor PV300 as a system. The Pusen Single Use Flexible Video Ureteroscope and 510(k) summary
{6}------------------------------------------------
Pusen Single Use Suction Access Ureteroscope are powered by the HD Medical Video Endoscope Image Processor PV300.
- The HD Medical Video Endoscope Image Processor PV300 was cleared under 510(k) number K222602.
### V. INDICATIONS FOR USE
This instrument has been designed to be used with Endo-Therapy accessories such as a biopsy forceps and other ancillary equipment for endoscopy and endoscopic surgery within urinary tract and interior of the kidney. This instrument is also used as a suction catheter that establishes a conduit used for irrigation and aspiration of kidney stones and stone dust during ureteral lithotripsy.
### VI. COMPARISON WITH THE PREDICATE DEVICE
### a. INDICATIONS FOR USE
| Item | Subject device | Predicate device | Reference device |
|------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade name | Single Use Flexible Video<br>Ureteroscope<br>Single Use Suction Access<br>Ureteroscope | Medical Video Endoscope<br>System | ClearPetra Suction-<br>Evacuation Sheath |
| 510(K) number | K233778 | K172098 | K203119 |
| Indications for<br>use | This instrument has been<br>designed to be used with<br>Endo-Therapy accessories<br>such as a biopsy forceps<br>and<br>other<br>ancillary<br>equipment for endoscopy<br>and<br>endoscopic surgery<br>within urinary tract and<br>interior of the kidney. This<br>instrument is also used as a<br>suction<br>catheter<br>that<br>establishes a conduit used | This instrument has been<br>designed to be used with<br>endo-therapy accessories<br>such as a biopsy forceps<br>and<br>other<br>ancillary<br>equipment for endoscopy<br>and<br>endoscopic surgery<br>within urinary tract and<br>interior of the kidney. | The ClearPetra Suction-<br>Evacuation Sheath is<br>used to establish a<br>conduit<br>during<br>endoscopic<br>urological<br>procedures facilitating<br>the<br>passage<br>of<br>endoscopes and other<br>instruments into the<br>urinary tract. It is<br>designed to establish a<br>conduit for the treatment |
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image displays the logo for Zhuhai Pusen Medical Technology Co., Ltd. The logo features the company name in a clean, sans-serif font. The word "PUSEN" is prominently displayed in a light blue color, while the rest of the company name is in a smaller, gray font.
| for irrigation and aspiration<br>of kidney stones and stone<br>dust during ureteral<br>lithotripsy | of urinary stones or other<br>urinary diseases during<br>endoscopic procedures. |
|----------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------|
|----------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------|
The subject device's indications for use is similar to that of the predicate device, the difference is only supplementary descriptions to the suction access functionality which aligns with that of the reference device, and this difference does not change the intended use of the subject device.
### TECHNOLOGICAL CHARACTERISTICS b.
| Item | Subject device | Predicate device |
|---------------------------------------|----------------------------------------------------------------------------------|-------------------------------------------------|
| Trade name | Single Use Flexible Video Ureteroscope<br>Single Use Suction Access Ureteroscope | Medical Video Endoscope System |
| 510(K) number | K233778 | K172098 |
| Scope type | Flexible | Flexible |
| Scope reusability | Single-use | Single-use |
| Energy used | Powered by chargeable battery or line<br>power. | Powered by chargeable battery or line<br>power. |
| Digital<br>video<br>technology | CMOS | CMOS |
| Illumination<br>source | LED | Fiber |
| Field of view | 120 ° | 120 ° |
| Direction of view | 0° | 0° |
| Depth of field | 3~50 mm | 3~50 mm |
| Maximum<br>insertion portion<br>width | PU3033AH, PU3033H: 2.7 mm<br>PU400A, PU411A: 3.2 | 3.2 mm |
| Working length | PU3033AH, PU3033H: 650 mm | 650 mm |
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image contains the logo for Zhuhai Pusen Medical Technology Co., Ltd. The logo consists of the word "PUSEN" in a stylized, sans-serif font, with each letter rendered in a light blue color. Below the logo, the full company name, "Zhuhai Pusen Medical Technology Co., Ltd.," is written in a smaller, gray font.
| | PU400A: 650 mm<br>PU411A: 680 mm | |
|-------------------------|-----------------------------------------------------------------|----------------------------------|
| Working channel<br>size | PU3033AH, PU3033H: $\ge$ 1.2 mm<br>PU400A, PU411A: $\ge$ 5.1 Fr | 1.0 mm |
| Up/down<br>deflection | Up: 270°<br>Down: 270° | Up: 270°<br>Down: 270° |
| Irrigation | Provided | Provided |
| Suction | Provided | Not Provided |
| Sterility | Ethylene Oxide (EO)<br>SAL: 10-6 | Ethylene Oxide (EO)<br>SAL: 10-6 |
The subject and the predicate device have different dimensions (e.g., working length, channel size) and illumination source. However, these differences do not raise different questions of safety and effectiveness.
{9}------------------------------------------------
### VII. PERFORMANCE DATA
The subject device has been verified for the safety and effectiveness based on the following performance data.
### Biocompatibility testing
The biocompatibility evaluation for the Pusen Single Video Ureteroscope and Pusen Single Use Suction Access Ureteroscope was conducted in accordance with the FDA final guidance "Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" (issued on September 4, 2020) and international standard ISO 10993-1:2018 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA. To support the contact classification as breached/compromised or compromised surfaces, contacting tissue for a limited duration (< 24 hours), the following endpoints were evaluated:
- . Cytotoxicity
- Intracutaneous Irritation .
- Sensitization ●
- . Material-Mediated Pyrogenicity
- . Acute Systemic Toxicity
The subject device passed the relevant tests and complies with ISO 10993-1.
### Electrical safety and electromagnetic compatibility (EMC)
Electrical safety and EMC testing were conducted on the subject device system, and the system complies with the IEC 60601-1 and IEC 60601-2-18 standards for safety and the IEC 60601-1-2 standard and IEC TS 60601-4-2 for EMC.
### Software Verification and Validation Testing
Software verification and validation testing were conducted, and relevant software documentations were provided as recommended by FDA's Guidance for Industry and FDA
{10}------------------------------------------------
Image /page/10/Picture/1 description: The image shows the word "PUSEN" in a blue sans-serif font. Below the word "PUSEN" is the text "Zhuhai Pusen Medical Technology Co.," also in a smaller, darker font. The text appears to be a company logo.
staff, "Guidance for the Content of Premarket Submissions for Device Software Function" (issued on June 14, 2023).
### Bench Performance Testing
### Mechanical and Optical Performance
The subject device was designed to comply with applicable parts of ISO 8600. Optical measurements and physical performances tests were performed according to applicable part of ISO 8600 standard, include appearance, working length, working channel diameter, maximum insertion width, angle of deflection, field of view and direction of view. Mechanical characteristics were tested and include flow rate of water, suction rate and tensile strength testing.
### lmage quality
Comparative testing related to image quality performances were performed for the subject device and the predicate device to support substantial equivalence. The tests included color reproduction performance, optical performance (such as resolution, depth of field, image intensity uniformity, signal-to-noise ratio (SNR), dynamic range and distortion).
### Photobiological safety
The LEDs in the subject device were tested according to the following FDA recognized standards:
• IEC 62471:2006 Medical electrical equipment, Photobiological safety of lamps and lamp systems.
### Luer taper
Luer taper in the subject device was tested according to ISO 80369-7: 2021, Small-bore connectors for liquids and gases in healthcare applications - Part 7: Connectors for intravascular or hypodermic applications.
### Sterilization and shelf life
- A shelf life of 3 years for the Pusen Single Use Flexible Video Ureteroscope and Pusen Single Use Suction Access Ureteroscope was supported with accelerated aging of final
{11}------------------------------------------------
Image /page/11/Picture/1 description: The image contains the word "PUSEN" in a large, sans-serif font, with each letter in a light blue color. Below the word "PUSEN", there is the text "Zhuhai Pusen Medical Technology Co.," in a smaller, gray font. The text appears to be the name of a company.
devices, followed by simulated shipping distribution, package integrity testing, and product performance testing.
- The EO sterilization method has been validated to ISO 11135:2014.
### Animal Study and Clinical study
No animal study or clinical study is included in this submission.
### Summary of performance data
All tests were passed and all evaluation acceptance criteria were met.
### VIII. CONCLUSIONS
The performance data described above demonstrate that the Pusen Single Use Flexible Video Ureteroscope and Pusen Single Use Suction Access Ureteroscope is as safe and effective as the predicate device and supports a substantial equivalence determination.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.