OES Elite Ureteroscopes and Accessories

K200369 · Olympus Winter & Ibe GmbH · FGB · Aug 28, 2020 · Gastroenterology, Urology

Device Facts

Record IDK200369
Device NameOES Elite Ureteroscopes and Accessories
ApplicantOlympus Winter & Ibe GmbH
Product CodeFGB · Gastroenterology, Urology
Decision DateAug 28, 2020
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2
AttributesTherapeutic, Pediatric

Indications for Use

The Olympus ureteroscopes can be utilized for endoscopic observation and therapy in the ureters, urethra, and urinary bladder. Semi-rigid optical instrument for the visualization of the following diagnostic and therapeutic procedures: - Transurethral inspection of the ureters and renal pelvis for diagnosis - Transurethral insertion of catheters, guide wires, cannulas, forceps, electrodes, baskets, lithotripsy probes and laser fibres into the ureter - Transurethral treatment and removal of tissue, catheters, guide wires, debris and stones from urethra, urinary bladder and ureters - Transurethral treatment and removal of tissue and stones from the ureters and renal pelvis. The products are not intended for treatment of infants. For children (>2 years) and adults, refer to the particular constitution and anatomy of the patient.

Device Story

OES Elite Ureteroscopes are reusable, semi-rigid endoscopes for transurethral visualization and therapeutic intervention in the urinary tract. The device utilizes an internal image relay system and optical fiber bundle to transmit images and light from an external source to the distal tip. Physicians insert the device through the urethra to perform diagnostic inspections or guide surgical instruments (e.g., forceps, lithotripsy probes) through integrated instrument channels to treat stones or remove tissue. The device is used in clinical settings; it is delivered non-sterile and requires reprocessing between uses. Output is a direct optical image viewed by the clinician, facilitating real-time navigation and surgical decision-making. Benefits include minimally invasive access to the ureters and renal pelvis for stone management and diagnostic visualization.

Clinical Evidence

Bench testing only. No clinical data provided. Testing included electrical/thermal safety (ANSI AAMI ES60601-1, IEC 60601-2-18), optical/mechanical performance (ISO 8600 series), biocompatibility (ISO 10993-1), and reprocessing validation. Usability validation conducted per IEC 62366-1 and IEC 60601-1-6.

Technological Characteristics

Semi-rigid endoscope; optical fiber image relay and illumination system. Materials: autoclavable components. Dimensions: 330-430mm working length, 8.4-10.4 Fr distal tip. Connectivity: standalone optical device. Sterilization: reusable, requires validated reprocessing. Standards: ISO 8600-1/3/4, ANSI AAMI ES60601-1, IEC 60601-2-18, ISO 10993-1.

Indications for Use

Indicated for endoscopic observation and therapy in ureters, urethra, and urinary bladder in children (>2 years) and adults. Procedures include transurethral inspection, insertion of surgical instruments (catheters, wires, lithotripsy probes, etc.), and removal of tissue/stones. Contraindicated for infants.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls: 1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following: 1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures. 2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use. 3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use. 2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes. 3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches 4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error. 5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program. (6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms. (7) Labeling must include: (i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified; (ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA; (iii) Identification of compatible devices; (iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification; (v) Reprocessing instructions for reusable components; (vi) A shelf life for any sterile components; (vii) A description of the device-specific use training program; (viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and (ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance. (8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include: (i) Device motion accuracy and repeatability; (ii) System testing; (iii) Instrument reliability; (iv) Crosstalk; (v) Table motion control; (vi) Thermal effects on tissue; (vii) User-device interface performance; (viii) Workspace access testing; and (ix) Performance testing with compatible devices. (9) Software verification, validation, and hazard analysis must be performed. (10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed. (11) Performance data must demonstrate the sterility of all patient-contacting device components. (12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life. (13) Performance data must validate the reprocessing instructions for the reusable components of the device. (14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible. (15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic. (16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information: (i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and (ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue. August 28, 2020 Olympus Winter & Ibe GmbH % Christina Flores Regulatory Affairs Manager Olympus Surgical Technologies America 118 Turnpike Road Southborough, MA 01772-2104 Re: K200369 > Trade/Device Name: OES Elite Ureteroscopes and Accessories (WA2UR11A, WA2UR12A, WA2UR13A, WA2UR14A, WA2UR21A, WA2UR22A, WA2UR23A, WA2UR31A, WA2UR32A, WA2URADP, WA47778A) Regulation Number: 21 CFR& 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: II Product Code: FGB Dated: August 3, 2020 Received: August 4, 2020 Dear Christina Flores: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. Mark J. Antonino, M.S. Acting Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ### Indications for Use 510(k) Number (if known) #### Device Name OES Elite Ureteroscopes and Accessories (WA2UR12A, WA2UR13A, WA2UR14A, WA2UR21A, WA2UR22A, WA2UR23A, WA2UR31A, WA2UR32A, WA2URADP, WA47778A) #### Indications for Use (Describe) The Olympus ureteroscopes can be utilized for endoscopic observation and therapy in the ureters, urethra, and urinary bladder. Semi-rigid optical instrument for the visualization of the following diagnostic and therapeutic procedures: - Transurethral inspection of the ureters and renal pelvis for diagnosis - Transurethral insertion of catheters, guide wires, cannulas, forceps, electrodes, baskets, lithotripsy probes and laser fibres into the ureter - Transurethral treatment and removal of tissue, catheters, guide wires, debris and stones from urethra, urinary bladder and ureters - Transurethral treatment and removal of tissue and stones from the ureters and renal pelvis. The products are not intended for treatment of infants. For children (>2 years) and adults, refer to the particular constitution and anatomy of the patient. **Type of Use (Select one or both, as applicable)** X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ### 510(k) Summary August, 3d, 2020 ## 1. General Information - Olympus Winter and Ibe GmbH . Manufacturer/Holder Kuehnstr. 61 22045 Hamburg Germanv Establishment Registration No.: 9610773 - Official Correspondent Ms. Christina Flores, RAC . Olympus Surgical Technologies America 118 Turnpike Road Southborough, MA 01772 Phone: (508) 808-3341 Email: Christina.Flores@olympus.com Establishment Registration No .: 3003790304 ### 2. Device Identification - OES Elite Ureteroscopes and Accessories (WA2UR11A, Device Trade Name: . WA2UR12A, WA2UR13A, WA2UR14A, WA2UR21A, WA2UR22A, WA2UR23A, WA2UR31A, WA2UR32A, WA2URADP, WA47778A) - Common Name: Ureteroscopes - Classification Name: Endoscope and Accessories . - Regulation Number: 21 CFR 876.1500 ■ - Product Code: FGB ■ - Product Code Name: Ureteroscope and Accessories, Flexible/Rigid ■ - Regulatory Class: ## 3. Predicate Devices K052044, ACMI® MR-6A/MR-6LA Autoclavable Ureteroscopes - K011849, MRO-742A Ureteroscope No reference devices were used in this submission. {4}------------------------------------------------ # 4. Device Description The OES Elite Ureteroscopes are inserted directly through the natural orifice urethra and are used to visualize a wide range of therapeutic procedures or to support diagnosis. The OES Elite Ureteroscopes are reusable semi-rigid endoscopes, which consist of an image relay system inside the main body and an outer tube that guides an image fiber bundle to transmit the endoscopic image. For therapeutic procedures, the device is used in combination with surgical instruments which can be introduced through the instrument channels. The OES Elite Ureteroscopes and accessories are delivered in non-sterile condition. They are reusable and fully reprocessable. Before first and each subsequent use the device must be inspected and reprocessed according to defined and validated reprocessing methods in the instructions for use. The OES Elite Ureteroscopes are available in different working lengths, ocular directions, image sizes, directions of views and with one or two instrument channels. ## 5. Indications for Use The Olympus ureteroscopes can be utilized for endoscopic observation and therapy in the ureters, urethra, and urinary bladder. Semi-rigid optical instrument for the visualization of the following diagnostic and therapeutic procedures: - Transurethral inspection of the ureters and renal pelvis for diagnosis . - Transurethral insertion of catheters, quide wires, cannulas, forceps, baskets, lithotripsy ● probes and laser fibers into the ureter - Transurethral treatment and removal of tissue, catheters, guide wires, debris and stones from urethra, urinarv bladder and ureters - . Transurethral treatment and removal of tissue and stones from the ureters and renal pelvis. The products are not intended for treatment of infants. For children (>2 years) and adults, refer to the particular constitution and anatomy of the patient. The indications for use statement for the subject OES Elite Ureteroscopes is very similar to that of the predicate devices. A slightly different wording is chosen and specific indications are added, listing specific diagnostic and therapeutic procedures. Also, the patient population excludes the treatment of infants. . The differences do not alter the intended use of the device nor do they raise different questions of safety and effectiveness of the device relative to the predicate. {5}------------------------------------------------ # 6. Summary of Technological Characteristics The technological characteristics of the subject devices are considered equivalent to the predicate devices. Details can be found in the following table. | Characteristic | Subject Devices | Predicate Devices | |-----------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------| | General technology | A ureteroscope is a semi-rigid endoscope. An image relay system of image fibers transmits the endoscopic image. A bundle of optical fibers transmits light from a light source to illuminate the operating field | Same as subject device | | Distal end | Functional components: objective cover glass, instrument channel(s), light emission surface | Same functional component as subject device. | | Proximal end | All distal ends have the following functional components: irrigation connector, instrument port, eyepiece cup, finger rests and light guide connector.<br>Some models have incorporated stopcocks.<br>The subject device models differ in the positioning of these elements and the design of the finger rests. Some models are available with angled others with straight ocular. | Same functional components as subject device, exept for built-in stopcocks and finger rests. | | Insertion tube | Profile form distal: Triangular (dual-channel) and oval (single-channel). | Profile form distal: Triangular (dual-channel) and oval (single-channel), but minor differences exist. | | Irrigation inflow | 0.7 ml/min - 9.7ml/min with inserted wire | 0.2 ml/min - 17 ml/min with inserted wire | | Maximum working length | (330 mm - 430 mm) ± 1 mm | (330 mm - 430 mm) ± 5 mm | | Outer circumference on distal tip | (8.4 Fr - 10.4 Fr) ± 0.4 Fr. | 7 Fr - 8.4 Fr | | Maximum insertion portion width | 11.7 Fr. - 13.05 Fr. | 10.2 Fr - 11.2 Fr | | Minimum instrument channel width | 2.4 Fr. - 5.4 Fr. | 2.3 Fr - 5.4 Fr | | Direction of view | 5° ± 5° or 0° ± 5° | 5° ± 3° | | Field of view | (86° - 95°) ± 12° | 61° ± 3° | | Illumination | Minimum: (0.081 - 0.150) mlm/klx | (0.193 - 0.277) mlm/klx | | Resolution | ≥ 9.5 lp/mm @ best working distance | ≥ 5 lp/mm @ best working distance | | Distortion | Relative Distortion in air (max value): -40.0% | Relative Distortion in air (max value): -19.8 % | | Moiré filter | Included | Not included | | Total number of fibers/pixels | 30000 pixel or 50000 pixel | 30000 pixel | | Automated cleaning | yes | no | | Autoclavibility | yes | yes | {6}------------------------------------------------ As stated above, the subject and predicate devices have similar design characteristics and show comparable performance. As demonstrated in the nonclinical testing the different technological characteristics do not raise any new questions of safety and effectiveness when compared to the predicate device. # 7. Performance Data The following design verification and validation testing was performed on the subject devices to ensure that the subject devices function as intended and meet design specifications: - Electrical safety as per ANSI AAMI ES60601-1:2005/(R)2012 and A1:2012 and IEC . 60601-2-18 Edition 3.0 - . Thermal safety as per ANSI AAMI ES60601-1:2005/(R)2012 and A1:2012 and IEC 60601-2-18 Edition 3.0 - Mechanical and optical performance testing as per ISO 8600-1 Edition 4, ISO 8600-3 Edition 1 and ISO 8600-4 Edition 2 - . Transport and shipping testing as per ASTM D4169-16 - Tests related to the expected service life as per ANSI AAMI ES60601-1:2005/(R)2012 and A1:2012 and IEC 60601-2-18 Edition 3.0 - . Biocompatibility testing according ISO 10993-1 Ed. 5: 2018 - Reprocessing validation was conducted in accordance with the FDA guidance . document "Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling" issued March 17, 2015. - Usability validation was conducted according to IEC 62366-1 Ed. 1.0: 2015 and IEC . 60601-1-6 Ed. 3.1: 2013. Usability Validation of the Instructions for Use as per FDA quidance document "Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling" issued March 17, 2015. - Additional comparative testing: Visible image size, distortion, internal reflections, ● vignetting, transmission of the optical system, transmission of the illumination system, light distribution of illumination, remaining deformation after bending stress, flow rate in first and second instrument channel Risk analysis was carried out in accordance with established internal acceptance criteria based on ISO 14971:2007. ## 8. Conclusion The performance data support the safety of the device and demonstrate that the subject devices comply with the recognized standards as specified and support a substantial equivalence determination.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...