ACMI Circon's Semi-Rigid Ureteroscopes are intended for use by qualified physicians to provide access to the urinary tract to perform various diagnostic and therapeutic procedures.
Device Story
Semi-rigid ureteroscope provides urinary tract access for diagnostic and therapeutic procedures. Device incorporates fiber optical lenses and working channels within a semi-rigid shaft. Operates by insertion into the urinary tract; allows passage of accessory instrumentation through working channels. Connects to external light source for bladder surface illumination. Used by physicians in clinical settings. Output is direct visualization of the urinary tract to guide clinical decision-making and intervention.
Clinical Evidence
No clinical data; bench testing only.
Technological Characteristics
Semi-rigid shaft; fiber optical lenses; working channels for accessory instrumentation; external light source connectivity. Mechanical/optical design.
Indications for Use
Indicated for examination and operation of the urinary tract in patients requiring diagnostic or therapeutic procedures. For use by qualified physicians.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
K200369 — OES Elite Ureteroscopes and Accessories · Olympus Winter & Ibe GmbH · Aug 28, 2020
Submission Summary (Full Text)
{0}------------------------------------------------
APR - 9 2002
Pg 1 of r
11 012758
## 510(k) SUMMARY
# ACMI Circon' SEMI-RIGID URETEROSCOPES
Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared:
ACMI Circon Corporation 6500 Hollister Avenue Santa Barbara, CA 93117
Phone: 805-961-3290 Fax: 805-968-7385
Mr. Wayne B. Sterner Contact Person: Corporate Director Regulatory Affairs
Date Prepared: February 8, 2002
### Name of Device and Name/Address of Sponsor:
ACMI Circon's Semi-Rigid Ureteroscopes
ACMI Circon Corporation 6500 Hollister Avenue Santa Barbara, CA 93117
### Common or Usual Name:
Ureteroscopes
#### Classification Name and Classification Number:
Endoscope and accessories, 21 CFR Section 876.1500
### Predicate Device(s):
The predicate device is an ACMI Circon preamendment ureteroscope (circa, 1971).
{1}------------------------------------------------
K012758
pg 2 of 2
Intended Use:
ACMI Circon's Semi-Rigid Ureteroscopes are intended for use by qualified physicians to provide access to the urinary tract to perform various diagnostic and therapeutic procedures.
Technological Characteristics and Substantial Equivalence:
ACMI Circon's Semi-Rigid Ureteroscopes are substantially equivalent to other currently marketed ureteroscopes, referenced above. The Semi-Rigid Ureteroscopes and their predicate device(s) are all telescopes that incorporate fiber optical lenses and working channels within a semirigid shaft. The working channels of the Semi-Rigid Ureteroscopes and predicate device(s) allow the insertion of accessory instrumentation during urinary tract procedures. Both the Semi-Rigid Ureteroscopes and the predicate device(s) may be connected to an external light source to permit better illumination of the bladder surfaces.
The ACMI Circon's Semi-Rigid Ureteroscopes raise no new issues of safety or effectiveness.
7
{2}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image is a black and white circular seal. The seal contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the edge of the circle. In the center of the seal is an abstract image of an eagle.
#### Public Health Service
Food and Drug Administra 200 Cornorate Boulevard Rockville MD 20850
Mr. Wayne B. Sterner Director, Regulatory Affairs ACMI Circon Corporation 6500 Hollister Avenue SANTA BARBARA CA 93117-3019
## Re: K012758
Trade/Device Name: Semi-Rigid Ureteroscopes and Accessories Regulation Number: 21 CFR §876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: 78 FGB Dated: February 8, 2002 Received: February 11, 2002
Dear Mr. Sterner:
We have reviewed your Section 510(k) premarket notification of intent to market the device we nave rovious your betermined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for abounding to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 8xx.1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Nancy Cbrogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
CIRCON CORPORATION CONFIDENTIAL
K012758
CIRCON CORPORATION Semi-Rigid Ureteroscopes and Accessories FDA Premarket Notification 510(k)
# INDICATIONS FOR USE STATEMENT
510(k) Number (if known):
Semi-Rigid Ureteroscopes and Accessories Device Name:
The intended use for these Semi-Rigid ureteroscopes is for examination/operation of the urinary tract and, using additional accessories, to perform various diagnostic and therapeutic procedures.
## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CRDH, Office of Device Evaluation (ODE)
**Prescription Use**
(Per 21 CFR 801.109)
David A. Bergerson
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.