K181760 · Medifactia AB · FFX · Aug 8, 2019 · Gastroenterology, Urology
Device Facts
Record ID
K181760
Device Name
Transit-Pellets
Applicant
Medifactia AB
Product Code
FFX · Gastroenterology, Urology
Decision Date
Aug 8, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.1725
Device Class
Class 2
Indications for Use
For evaluation of colonic transit in adult patients with chronic constipation and used to aid in differentiating slow and normal constipation.
Device Story
Transit-Pellets are radiopaque markers used to evaluate colonic transit time. Patient swallows fixed dose of markers (10/day) for 6 days; on day 7, abdominal radiograph is taken. Markers consist of barium sulphate and silicone rubber in HPMC capsules. Clinician counts retained markers and assesses position in colon from radiograph to calculate transit time; results compared to reference values. Used in hospitals/clinics to differentiate slow vs. normal transit constipation. Device aids clinical decision-making by quantifying transit dysfunction; benefits include standardized assessment of chronic constipation.
Clinical Evidence
No clinical data was required. Evidence consists of bench testing, including simulated capsule digestion, dimensional/mass analysis, radiopacity testing per ASTM F640-12, and biocompatibility testing per ISO 10993-5, -10, and -11. 14-day repeat IV and intraperitoneal toxicity studies in rats showed no systemic toxicity.
Technological Characteristics
Markers: 22% Barium Sulphate, 78% Elastosil R 401/60 Silicone rubber. Capsules: HPMC (Hydroxypropyl-methylcellulose). Dimensions: 2x4.5mm (ring) and 6x2mm (tube). Radiopacity per ASTM F640-12. Biocompatibility per ISO 10993-5, -10, -11. Non-sterile. Shelf life 2.5 years.
Indications for Use
Indicated for adult patients with chronic constipation to aid in differentiating slow and normal transit constipation.
Regulatory Classification
Identification
A gastrointestinal motility monitoring system is a device used to measure peristalic activity or pressure in the stomach or esophagus by means of a probe with transducers that is introduced through the mouth into the gastrointestinal tract. The device may include signal conditioning, amplifying, and recording equipment. This generic type of device includes the esophageal motility monitor and tube, the gastrointestinal motility (electrical) system, and certain accessories, such as a pressure transducer, amplifier, and external recorder.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the seal of the Department of Health & Human Services. To the right of the seal is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in a smaller font size below.
August 8, 2019
Medifactia AB % Connie Qiu Regulatory Consultant III M Squared Associates, Inc 575 8th Ave Suite 1212 New York, NY 10018
Re: K181760
Trade/Device Name: Transit-Pellets Regulation Number: 21 CFR 876.1725 Regulation Name: Gastrointestinal Motility Monitoring System Regulatory Class: Class II Product Code: FFX Dated: December 27, 2018 Received: July 12, 2019
Dear Connie Qiu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for
Martha W. Betz, PhD Acting Assistant Division Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K181760
Device Name Transit-Pellets
Indications for Use (Describe)
For evaluation of colonic transit in adult patients with chronic constipation and used to aid in differentiating slow and normal constipation.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) number: K181760
In accordance with 21 CFR 807.92, the following information constitutes the Medifactia AB summary for the Transit-Pellets.
### Submission Date l.
July 11, 2019
### II. Submitter Medifactia AB
Lumaparksvagen 7 Stockholm SE 120 30 Sweden Contact person: Diana Nyström Phone number: +46 (0) 31-787 70 77 E-mail: diana.nystrom@medifactia.com
Official Correspondent: Connie Qui Regulatory Consultant M Squared Associates, Inc. 575 8th Avenue, Rm 1212 New York, New York 10018 Ph: 347-954-0411 Fax: 702-562-9797 E-mail: CQui@Msquaredassociates.com
#### III. Device
Name of device: Transit-Pellets Common or usual name: Radiopaque markers Classification name: Gastrointestinal motility monitoring system Regulatory number: 21 CFR 876.1725 Regulation Class: II Product code: FFX Panel: Gastroenterology/Urology
#### IV. Predicate device
| 510(k) | Company | Device |
|---------|--------------------------|-----------|
| K881609 | Lafayette Pharmacal, Inc | Sitzmarks |
{4}------------------------------------------------
#### V. Device Description
Transit-Pellets is intended to be used for evaluation of colonic transit time in patients with chronic constipation and used to aid in differentiating slow and normal transit constipation. Ten (10) radiopaque markers per day are swallowed for six consecutive days. On day seven an abdominal radiograph is taken. Based on the number of retained markers and the position in colon a colonic transit time is calculated and compared to reference values. Both total transit and segmental transit dysfunction in the colon can be evaluated with the device. By dividing the particle dose on day six by taking five (5) markers in the morning and five (5) markers in the evening, the whole range of transit times (slow, normal, rapid) transit can be measured from the radiograph.
Transit-Pellets is a convenience package of radiopaque markers (22% Barium Sulphate, 78% Elastosil® R 401/60 Silicone rubber) placed in veqetarian capsules from cellulose (HPMC, Hydroxypropyl-methylcellulose), intended for single patient use. The dimension of the markers is 2x4.5mm (ring-formed markers) and 6x2mm (tube-formed markers). The capsules are packed in a blister pack and the blister pack is placed inside a folding box.
#### VI. Intended use
For evaluation of colonic transit in adult patients with chronic constipation and used to aid in differentiating slow and normal constipation.
| Device name | Company Device | Predicate Device | SE |
|------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----|
| Applicant | Medifactia AB<br>Transit-Pellets | Lafayette Pharmacal, Inc<br>Sitzmarks | |
| K# | -- | K881609 | |
| Indication for use<br>Intended use | For evaluation of colonic<br>transit in patients with<br>chronic constipation and<br>used to aid in differentiating<br>slow and normal<br>constipation. | For evaluation of colonic<br>transit in patients with<br>chronic constipation and<br>used to aid in<br>differentiating slow and<br>normal constipation. | X |
| Application | Measuring and using<br>colonic transit time to<br>evaluate patients with<br>chronic (severe)<br>constipation. | Measuring and using<br>colonic transit time to<br>evaluate patients with<br>chronic (severe)<br>constipation. | X |
| Target population | Adult patients with chronic<br>(severe) constipation. | Adult patients with chronic<br>(severe) constipation. | X |
| Anatomical site | Gastrointestinal tract/colon | Gastrointestinal tract/colon | X |
| Where used | Hospitals, clinics | Hospitals, clinics | X |
| Device name | Company Device | Predicate Device | SE |
| Applicant | Medifactia AB | Lafayette Pharmacal, Inc | |
| | Transit-Pellets | Sitzmarks | |
| Design | Ring-formed and tube-<br>formed markers. | Three different shapes of<br>markers: O Rings, Double<br>D and Tri-Chamber. | X |
| Dimensions | 2 x 4.5mm, 6 x 2mm | 1mm x 4.5mm | X |
| Materials | Markers: Barium sulphate,<br>Elastosil® R 401/60<br>Silicone rubber<br>Capsules: HPMC,<br>Hydroxypropyl-<br>methylcellulose | Copied from<br>www.sitzmarks.com:<br>Markers: Rings-<br>Polyvinylchloride, Barium<br>sulphate, DEHP, soya oil,<br>calcium, zinc, phosphate,<br>ultramarine tinting agent,<br>mono & diglycerides, DOP<br>plasticizer<br>Capsules: Hypromellosa<br>Copied from 510(k);<br>Markers: Radiopaque<br>polyvinyl chloride<br>Gelatin capsules | X |
| Principle of<br>operation | Patient swallows a fixed<br>dose radiopaque marker for<br>a number of days. On day<br>7 a single abdominal<br>radiograph or fluoroscopy<br>is taken. Based on the<br>number of retained markers<br>on abdominal film and their<br>position in colon a colonic<br>transit time is calculated<br>and compared to reference<br>value. | Patient swallows a fixed<br>dose radiopaque marker<br>for a number of days.<br>Several abdominal<br>radiographs may be<br>necessary. Based on the<br>number of retained<br>markers on abdominal<br>film(s) and their position in<br>colon a colonic transit time<br>is calculated and<br>compared to reference<br>value. | X |
| Performance | Colonic Transit Time<br>(CTT/OATT) numerical<br>values reported in days.<br>ASTM F640-12 RO<br>confirmed. | Colonic Transit Time<br>(CTT/OATT) numerical<br>values reported in hours. | X |
| Biocompatibility | Yes, safe for intended use<br>ISO 10993-5<br>ISO10993-10<br>ISO 10993-11 | No information | |
| Packaging | Blister and folding box | Blister and folding box | X |
| Sterile | No | No | X |
| Shelf Life | 2.5 years | unknown | X |
| Rx Only | Yes | Yes | X |
#### VII. Comparison of the technology characteristics with the predicate devices
{5}------------------------------------------------
{6}------------------------------------------------
#### VIII. Performance testing
Transit-Pellets have a size <8mm and are made tube-formed and ring-formed. To test transit characteristics of various types of markers, five types of distinguishable markers in the specific gravity range 1.2-1.6 were examined. The size of the particles was chosen in the range 2-7 mm so that emptying from stomach was likely to occur also with a meal. In studies with markers of increasing sizes it was observed that markers with a size >8 mm will have a slower emptying rate from the stomach and thus the transit time from the stomach to the colon will be delayed and can lead to a falsely prolonged value of colonic transit time. Studies with markers <8 mm but of different shape showed that the transit through colon was identical with ring-formed, tube-formed and cube-formed markers. For these reasons, the Transit-Pellets were designed for their size and configurations and are safe and effective for their intended use when compared to the predicate device. Simulated capsule digestion testing in a simulated gastric fluid, dimensional analysis, and mass analysis, as well as the clinical literature provided in the 510k submission, confirms the device performs as intended for the proposed indication for use, i.e., is safe and effective for evaluation of colonic transit in patients with chronic constipation and used to aid in differentiating slow and normal constipation. The Transit-Pellets is substantially equivalent to Sitzmarks regarding material, technological characteristics, and indications for use.
Biocompatibility testing to ISO 10993-5 and -10 of the Transit-Pellets confirm the device is non-cytotoxic, non- irritant, and non-sensitizing. Testing to ISO 10993-11 confirmed that the Transit-Pellet extracts did not induce a significantly greater biologic reaction than the control following systemic injection and produced a non-pyrogenic response. Additionally, 14 Day Repeat IV and Intraperitoneal Toxicity in Rats did not demonstrate systemic signs of toxicity over 14 days.
Radiopacity Testing in accordance with ASTM F640-12 confirms visibility of the Transit-Pellets for the duration and environment of their intended use during colonic transit.
#### IX. Clinical Data
Clinical data was not required to establish the substantial equivalence of the subject and predicate devices.
#### X. Conclusion
The two devices share the indication for use of measuring and using colonic transit time to evaluate patients with chronic constipations of our device cover the application of the predicate device and thus the difference will not affect safety and efficiency.
The information provided in this510(k)support that the Transit-Pellets is substantially equivalent in function, composition, and intended use to the predicate devices. The proposed device raises no new issues of safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.