PRESSURIZED INFUSION PUMP

K122294 · Mui Scientific · FFX · Nov 1, 2012 · Gastroenterology, Urology

Device Facts

Record IDK122294
Device NamePRESSURIZED INFUSION PUMP
ApplicantMui Scientific
Product CodeFFX · Gastroenterology, Urology
Decision DateNov 1, 2012
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1725
Device ClassClass 2

Indications for Use

The Pressurized Infusion Pump (Electrically Powered Models #PIP-4-4SS, PIP-4-6SS, PIP-4-8SS, PIP-4-12SS, and Portable Models #PIP-6-4SS, PIP-6-6SS, PIP-6-8SS, PIP-6-12SS, PIP-6-16SS, PIP-6-22SS, PIP-6-24SS, PIP-6-36SS) is intended to provide a regulated and channeled perfusate of water to a motility catheter to perform manometric motility studies along the gastrointestinal tract. The Pressurized Infusion Pump is intended to be used with a separate legally marketed compatible motility catheter and separate computerized gastrointestinal monitoring system.

Device Story

Pressurized Infusion Pump provides regulated water perfusion to motility catheters for GI manometry. System components: electrical compressor motor, drying cylinder, pressure switches/gauges, flow restrictors, water reservoir. Operation: motor compresses air into drying cylinder; pressure switch maintains 18-42psi range; regulator reduces air to 17psi; regulated air drives water from reservoir through flow restrictors (0.60mL/min) into motility catheter. When GI muscles contract, water flow through catheter ports is constricted; pressure change transmits back through water column to pump-integrated pressure transducer; signal sent to external computerized monitoring system for display. Used in clinical settings by healthcare professionals. Benefits: enables standardized, constant-rate perfusion for accurate manometric pressure measurement during GI motility assessment.

Clinical Evidence

Bench testing only. Performance compared to predicate using pinch tests over 8 pressure ports on a motility catheter to simulate muscle contractions. Results demonstrated that pressure exerted and measured by the electrical/portable pumps was equal to the predicate pump.

Technological Characteristics

System includes electrical compressor motor, drying cylinder, pressure switches, gauges, and flow restrictors. Delivers water at constant 0.60mL/min rate. Regulates air pressure to 17psi. Connects to external motility catheter and computerized monitoring system. Non-sterile pump unit; uses sterilized irrigation water.

Indications for Use

Indicated for patients requiring manometric motility studies of the gastrointestinal tract. Used with compatible motility catheters and monitoring systems.

Regulatory Classification

Identification

A gastrointestinal motility monitoring system is a device used to measure peristalic activity or pressure in the stomach or esophagus by means of a probe with transducers that is introduced through the mouth into the gastrointestinal tract. The device may include signal conditioning, amplifying, and recording equipment. This generic type of device includes the esophageal motility monitor and tube, the gastrointestinal motility (electrical) system, and certain accessories, such as a pressure transducer, amplifier, and external recorder.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ Image /page/0/Picture/0 description: The image shows a black and white graphic. The graphic appears to be an abstract shape or design. The shape is irregular and has some jagged edges. The image is simple and minimalist. Ki22294 NOV 1 2012 PAGE 1 OF 2 Mivision of Mui Scientific Enterprises Inc 145 Traders Blvd. E., Unit #34, Mississauga, ON, Canada, L4Z 3L3 Tel: (905) 890-5525 Toll Free: 1-800-303-6611 Fax: (905) 890-3523 Website: www.muiscientific.com Email: mail@muiscientific.com October 31, 2012 ## 510(k) Summary #### RE: Electrical and Portable Pressurized Infusion Pump Summary prepared by: Contact Person: Tammy Mui Title: Operations Manager Manufaxturer: Mui Scientific, a Division of H&A Mui Enterprises Address: 145 Traders Blvd. East, Unit #33-34, Mississauga, Ontario, Canada L4Z 3L3 Phone: (905) 890-5525 Fax: (905) 890-3523 Email: tammy.mui@muiscientific.com Trade name: Pressurized Infusion Pump Common name: Water-perfused motility pump Classification name: Gastrointestinal motility system This 510(k) Summary is for the Pressurized Infusion Pump. Indications for Use: The Pressurized Infusion Pump (Electrically Powered Models #PIP-4-4SS, PIP-4-6SS, PIP-4-8SS, PIP-4-12SS, and Portable Models #PIP-6-4SS, PIP-6-6SS, PIP-6-8SS, PIP-6-12SS, PIP-6-16SS, PIP-6-22SS, PIP-6-24SS, PIP-6-36SS) is intended to provide a regulated and channeled perfusate of water to a motility catheter to perform manometric motility studies along the gastrointestinal tract. The Pressurized Infusion Pump is intended to be used with a separate legally marketed compatible motility catheter and separate computerized gastrointestinal monitoring system. These pressurized infusion pumps consist of an electrical compartment, a water chamber, a regulator and a set of gauges, and flow restrictors. The electrical compartment unit consists of a compressor motor, a pressure switch, a pressure gauge, and a drying cylinder. When turned on, the motor will compress air into the drying cylinder, which will remove the water byproduct from the compression. The pressure switch will turn the motor off at 42psi (as displayed on the pressure gauge), and will restart it when the pressure drops to 18psi. For the Portable Pressurized Infusion Pump, an air hose connects the electrical unit to the main pump housing to deliver the pressurized air to the pump. The pressurized air will flow through a high pressure gauge (to display the pressure of the supplied compressed air), to a regulator (which will regulate the pressurized air down to 17psi), then a low pressure gauge {2}------------------------------------------------ # 122294 AGE 2 OF 2 ### Division of Mui Scientific Enterprises Inc. 145 Traders Blvd. E., Unit #34, Mississauga, ON, Canada, L4Z 3L3 Tel: (905) 890-5525 Toll Free: 1-800-303-6611 Fax: (905) 890-3523 Website: www.muiscientific.com Email: mail@muiscientific.com (to display the pressure of the regulated air), to a water reservoir filled with sterilized irrigation water. The requlated air will push the water out of the water reservoir at a constant rate through resistors that will further reduce the water flow rate to 0.60mL/min before it travels through a pressure transducer to the motility catheter. At the proximal end of the motility catheter are female luers that connect to the top of the pressure transducers on the infusion The water will flow through the entire lenath of the motility catheter to the distal end pump. where the pressure ports are located. The motility catheter is then inserted into the patient and placed along the specific section of the gastrointestinal system being measured. With the requlated pressurized water flowing through the catheter at a constant rate, when the qut muscles contract, it will constrict the flow of water through the pressure ports, and the pressure change will be transmitted back through the water flow to the pressure transducer on the pressurized infusion pump, where the signal will be displayed onto the computer. We are claiming this system to be substantially equivalent to the following predicates: - Mui Scientific's Nitrogen gas model of the pressurized infusion pump - The Electrical and Portable pump versions are similar to this predicate in that they use pressure to push water from a water reservoir through resistors that further reduce the flow rate before passing through a motility catheter. The difference is the source of the pressure: compressed nitrogen gas in the predicate, and an air compressor motor in the Electrical and Portable pump models. Bench tests were conducted with the Electrical and Portable Pressurized Infusion Pumps, comparing performance data with that of its predicates. With the systems on and perfusing, a pinch test was performed over each of the 8 pressure ports along the motility catheter, to mimic a muscle contraction. The pressure exerted was shown to equal the pressure measured and transmitted on the predicate pump. {3}------------------------------------------------ ### DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling a stylized caduceus or a representation of human figures. #### Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002 Letter date: November 1, 2012 Mui Scientific Division of H&A Mui Enterprises Inc. % Ms. Tammy Mui Operations Manager 145 Traders Blvd. East, Unit #34 MISSISSAUGA ON CANADA L4Z 3L3 Re: K122294 Trade/Device Name: Pressurized Infusion Pump Regulation Number: 21 CFR§ 876.1725 Regulation Name: Gastrointestinal motility monitoring system Regulatory Class: II Product Code: FFX Dated: October 15, 2012 Received: October 19, 2012 Dear Ms. Mui: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {4}------------------------------------------------ Page 2 - Ms. Tammy Mui Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Benjamin R. Fisher -S Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ ## Indications for Use 510(k) Number (if known): K122294 Device Name: Pressurized Infusion Pum Indications For Use: The Pressurized Infusion Pump (Electrically Powered Models #PIP-4-4SS, PIP-4-8SS PIP-4-12SS, and Portable Models #PIP-6-4SS, PIP-6-6SS, PIP-6-8SS, PIP-6-12SS, PIP-6-16SS, PIP-6-22SS, PIP-6-24SS; PIP-6-36SS) is intended to provide a regulated and channeled perfusate of water to a motility catheter to perform manometric motility studies along the gastrointestinal tract. The Pressunzed Infusion Pump is intended to be used with a separate legally marketed compatible motility catheter and separate computerized gastrointestinal monitoring system. Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) oncurrence of CDRH, Office of Device Evaluation (ODE) Benjamin R. Fisher 1 16:46:29 -04'00 (Division Sign-Off) wision of Reproductive, Gastro-Renal, and Proiogical Devices SIU(K) NUr Page 1 of
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...