PH ANALYSIS MODULE

K961346 · Synectics-Dantec · FFT · Jan 29, 1997 · Gastroenterology, Urology

Device Facts

Record IDK961346
Device NamePH ANALYSIS MODULE
ApplicantSynectics-Dantec
Product CodeFFT · Gastroenterology, Urology
Decision DateJan 29, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1400
Device ClassClass 1
AttributesSoftware as a Medical Device, Pediatric

Indications for Use

The pH Analysis Module is a software program that has been designed to analyze intragastroesophageal pH data recorded from the gastrointestinal tract in pediatric and adult populations. The program is to be used on a personal computer analyzing patient data in the hospital environment under supervision of a trained physician. The analyzed data can be viewed on the computer screen or printed out on a separate paper.

Device Story

pH Analysis Module; software add-on for Polygram Software for Windows. Inputs: recorded intragastroesophageal pH data. Processing: automated analysis of pH tracings; calculation of parameters including duration of recording, number of periods of interest, time below pH threshold, symptom correlation, and scoring. Output: analysis report containing patient demographics, procedure summary, pH tracing, period table, and physician signature section. Used in hospital environment; operated by trained physician. Benefit: assists physician in diagnosis and post-treatment evaluation of gastrointestinal pH data.

Clinical Evidence

No clinical data provided; bench testing only.

Technological Characteristics

Software-based analysis module; runs on personal computer; Windows-based environment; integrates with Polygram Software for Windows platform.

Indications for Use

Indicated for analysis of intragastroesophageal pH data recorded from the gastrointestinal tract in pediatric and adult populations.

Regulatory Classification

Identification

A stomach pH electrode is a device used to measure intragastric and intraesophageal pH (hydrogen ion concentration). The pH electrode is at the end of a flexible lead which may be inserted into the esophagus or stomach through the patient's mouth. The device may include an integral gastrointestinal tube.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{1} JAN 29 1997 K961346 # 9. SMDA 510(k) Summary or Statement ## DRAERD 510(k) Summary of Safety and Effectiveness Information | Submitter's name | Synectics Medical AB Renstiernas gata 12 S-116 28 Stockholm Sweden | | --- | --- | | | Contact: Anna Pettersson | | Name of device | pH Analysis Module | | Name of equivalent device | EsopHogram Software (DOS environment) (included in K859361 - Synectics pH System Comprising Digitrapper Mark II, EsopHogram Software & Monocrystant Catheter) | | Description of device | The Polygram Software for Windows (K946322) was designed to record and handle/store physiological parameters. After a recording, the user is able to review the tracings on the computer screen and print the signal tracings or just parts of them. When reviewing the data on the computer screen, the user is able to mark certain segments and calculate certain parameters from the selected signal segments. These parameters include min and max values, length of selection. By adding the pH Analysis Module to the Polygram Software for Windows, the user can also have all the intragastroesophageal pH data analyzed (data recorded from the gastrointestinal tract in pediatric and adult population). The analysis report includes sections such as patient demographics; interpretation and comment (for user to insert); procedure summary; pH tracing; period table including number such as duration of recording, number of periods of interest, time where pH was below certain threshold value, symptom correlation, scoring; physician signature section. The analysis report can then serve as a part of the physician’s diagnosis and post treatment evaluation. | | Statement of intended use/indication for use | The pH Analysis Module is a software program that has been designed to analyze intragastroesophageal pH data recorded from the gastrointestinal tract in pediatric and adult populations. The program is to be used on a personal computer analyzing patient data in the hospital environment under supervision of a trained physician. The analyzed data can be viewed on the computer screen or printed out on a separate paper. | 13-Jan-97
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...