# BRAVO PH MONITORING SYSTEM AND ACCESSORIES (K102543)

_Given Imaging , Ltd. · FFT · Dec 1, 2010 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/GU/subpart-b%E2%80%94diagnostic-devices/FFT/K102543

## Device Facts

- **Applicant:** Given Imaging , Ltd.
- **Product Code:** [FFT](/submissions/GU/subpart-b%E2%80%94diagnostic-devices/FFT.md)
- **Decision Date:** Dec 1, 2010
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1400
- **Device Class:** Class 1
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Pediatric

## Indications for Use

The Bravo pH Monitoring System is intended to be used for gastroesophageal pH measurement and monitoring of gastric reflux in adults and children from 4 years of age. The Bravo pH Capsule can be attached following either endoscopy or manometry. The RAPID pH software is intended to record, store, view, and analyze gastroesophageal pH data.

## Device Story

Ambulatory esophageal pH monitoring system; consists of antimony pH sensor capsule, delivery system, receiver, and analysis software. Capsule attaches to esophageal mucosal wall 6 cm above Z-line via delivery system following EGD or manometry. Capsule measures pH and transmits data via radiotelemetry to patient-worn receiver. Receiver stores data for later analysis. Physician uses RAPID pH and PolygramNet software to record, store, view, and analyze pH data to interpret study results. Procedure is minimally invasive, performed in outpatient setting. Output assists clinicians in diagnosing gastric reflux; enables monitoring of patient condition.

## Clinical Evidence

Clinical data summarized to demonstrate safety and effectiveness for the proposed indications for use.

## Technological Characteristics

Antimony pH sensor capsule; radiotelemetry transmission; receiver unit; RAPID pH and PolygramNet software. System includes calibration stand, vacuum pump, and pH buffer solutions.

## Regulatory Identification

A stomach pH electrode is a device used to measure intragastric and intraesophageal pH (hydrogen ion concentration). The pH electrode is at the end of a flexible lead which may be inserted into the esophagus or stomach through the patient's mouth. The device may include an integral gastrointestinal tube.

## Predicate Devices

- Bravo pH Monitoring System™ and Accessories ([K002028](/device/K002028.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

GI EN®
IMAGING

Given Imaging Limited New Industrial Park PO Box 258, Yoqneam 20692 Israel Voice 972 4 909 7777 Fax 972 4 959 2466

# 510(k) Summary

DEC 1 2010

This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of the Safe Medical Devices Act of 1990 and 21 CFR 807.92.

| Submitter Name and Address:        | Given Imaging Ltd.<br>Hermon Building<br>New Industrial Park<br>PO Box 258<br>Yoqneam 20692<br>Israel<br>Tel.: 011-972-4-9097730<br>Fax: 011-972-4-9938060 |
|------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person:                    | Tim Thomas<br>Vice President,<br>Regulatory Affairs and Quality Assurance<br>Email: tim.thomas@givenimaging.com                                            |
| Phone Number:                      | 770-662-0870 ext. 1006                                                                                                                                     |
| Fax Number:                        | 770-662-0510                                                                                                                                               |
| Establishment Registration Number: | 9710107                                                                                                                                                    |
| Date Prepared:                     | September 1, 2010                                                                                                                                          |
| Device Trade Name(s):              | Bravo pH Monitoring System™ and Accessories                                                                                                                |
| Device Common Name:                | Stomach pH electrode                                                                                                                                       |
| Classification:                    | Regulation No: 876.1400<br>Class: I<br>Panel: Gastroenterology<br>FFT - Stomach pH electrode                                                               |
| Predicate Device(s):               | Bravo pH Monitoring System™ and Accessories (K002028)                                                                                                      |
| General Device Description:        | The Bravo pH Monitoring System™ and Accessories is an<br>ambulatory esophageal pH testing device.                                                          |

Given Imaging Ltd. New Industrial Park, PO Box 258, Yoqneam 20692, Israel

{1}------------------------------------------------

Given Imaging Limited New Industrial Park PO Box 258, Yogneam 20692 Israel Voice 972 4 909 7777 Fax 972 4 959 2466

The Bravo system is a simple, minimally invasive outpatient procedure that involves the placement of a pH telemetry Capsule against the mucosal wall of the esophagus that simultaneously measures pH and transmits the data to a receiver attached to the patient's waist. The Bravo pH Monitoring System™ and Accessories consists of

### four main components:

#### Bravo pH Capsule with Delivery System

Bravo pH Capsule with Delivery System is an antimony pH sensor, in the form of a pH Capsule that is temporarily attached to the wall of the esophagus with a delivery system. The capsule is placed following an esophago- gastroduodenoscopy (EGD) at 6 cm above the Z-line. The capsule then transmits pH data via radiotelemetry to a receiver worn by the patient.

#### Bravo pH Receiver

The Bravo pH Receiver is lightweight and compact. It fits into a carrying case that comes with a strap and a belt clip allowing patients to carry it over the shoulder, attach it to a belt, or carry it in a pocket (without the carrying case). This flexibility allows the patient to carry the Receiver throughout the study period. The Receiver is designed with two separate interfaces: one for the patient and one for clinicians.

#### RAPID pH and PolygramNet software

The software is intended to record, store, view, and analyze gastroesophageal pH data, enabling physicians to interpret study results.

#### Bravo pH Delivery System Accessories:

Bravo pH Delivery System Accessories include calibration stand, vacuum pump, pH 1.07 and pH 7.01 calibration buffer solutions, Bravo DataLink, Bravo Adapter, and Bravo Pouch.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for Given Imaging. The logo consists of the word "GIVEN" in large, bold letters, with a stylized image of a chalice or goblet in between the "GI" and "EN". Below the word "GIVEN" is the word "IMAGING" in smaller, bold letters. The logo is black and white.

Given Imaging Limited New Industriai Park PO Box 258, Yoqneam 20692 Israel Voice 972 4 909 7777 Fax 972 4 959 2466

| Intended Use:                  | The Bravo pH Monitoring System is intended to be used for gastroesophageal pH measurement and monitoring of gastric reflux in adults and children from 4 years of age. The Bravo pH Capsule can be attached following either endoscopy or manometry.<br>The RAPID pH software is intended to record, store, view, and analyze gastroesophageal pH data. |
|--------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Technological Characteristics: | The technology characteristics are exactly the same as the predicate devices.                                                                                                                                                                                                                                                                           |
| Performance Data:              | Clinical data has been summarized to show safety and effectiveness for the proposed indications for use.                                                                                                                                                                                                                                                |
| Conclusion:                    | Based on the technological characteristics and clinical performance of the devices, Given Imaging Ltd. believes that The Bravo pH Monitoring System™ and Accessories and the predicate device selected is substantially equivalent and does not raise new issues of safety or effectiveness.                                                            |

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three stylized, curved lines.

Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002

Mr. Tim Thomas Vice President Regulatory Affairs and Quality Assurance Given Imaging Limited New Industrial Park P.O. Box 258, Yogneam 20692 ISRAEL

DEC 1 2010

Re: K102543

Trade/Device Name: Bravo pH Monitoring System™ and Accessories Regulation Number: 21 CFR §876.1400 Regulation Name: Stomach pH electrode Regulatory Class: I Dated: September 1, 2010 Received: September 3, 2010

Dear Mr. Thomas:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related

{4}------------------------------------------------

Page 2

adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Hubert Lemmer MD

Herbert P. Lerner, M.D., Director (Acting) Division of Reproductive, Gastro-Renal and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

## Enclosure

{5}------------------------------------------------

#### INDICATIONS FOR USE

#### K102543 510(k) Number (if known):

Device Name: Bravo pH Monitoring System™ and Accessories

DEC 1 2010

#### Indications for Use:

The Bravo pH Monitoring System is intended to be used for gastroesophageal pH measurement and monitoring of gastric reflux in adults and children from 4 years of age. The Bravo pH Capsule can be attached following either endoscopy or manometry. The RAPID pH software is intended to record, store, view, and analyze gastroesophageal pH data.

Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use _ (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off)

Division of Reproductive, Gastro-Rend
Urological Devices
510(k) Number ***_***_ K102543

page 1 of 1

---

**Source:** [https://fda.innolitics.com/submissions/GU/subpart-b%E2%80%94diagnostic-devices/FFT/K102543](https://fda.innolitics.com/submissions/GU/subpart-b%E2%80%94diagnostic-devices/FFT/K102543)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).