K043324 · KARL STORZ Endoscopy-America, Inc. · FET · Apr 4, 2005 · Gastroenterology, Urology
Device Facts
Record ID
K043324
Device Name
AIDA WITH DICOM AND HL7 INTERFACE
Applicant
KARL STORZ Endoscopy-America, Inc.
Product Code
FET · Gastroenterology, Urology
Decision Date
Apr 4, 2005
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.1500
Device Class
Class 2
Attributes
Software as a Medical Device
Intended Use
This software is intended for use by qualified personnel in the Operating Room and Nurses Station. The Advanced Image and Data Archiving System (AIDA) is a Windows® based still/video image capturing, archiving, documentation system and recording of audio sequences and patient data during a procedure. It allows capture and annotation of the surgical procedure for documentation purposes. Images captured and distributed by AIDA are for viewing and reference purposes and are not intended for primary diagnosis. It is also a Windows® based solution to communicate with other picture archival communication systems (PACS) using DICOM and with Hospital Information Systems (HIS) using the HL7 standard. Also as a part of the AIDA system the Storz Application Manager software (SAM) enables the selection and integrations of AIDA functions with various compatible applications, such as Karl Storz’s Storz Communication Bus (SCB) or other third party image capturing devices.
Device Story
AIDA is a Windows-based software system for capturing, archiving, and documenting surgical images, video, and audio. Used in ORs and nurses' stations by qualified personnel; integrates with hospital PACS via DICOM and HIS via HL7. Features Storz Application Manager (SAM) for integration with Storz Communication Bus (SCB) or third-party capture devices. Provides documentation and reference viewing; output is for clinical reference only, not primary diagnosis. Benefits include streamlined surgical documentation and data management.
Clinical Evidence
No clinical data. Safety and effectiveness established via risk analysis, software verification, and validation testing.
Technological Characteristics
Windows-based software; DICOM and HL7 connectivity; integrates with SCB and third-party capture devices. Includes built-in screen prompts, alarms, and error messages. Risk management via hazard analysis and software verification/validation.
Indications for Use
Indicated for use by qualified personnel in OR and Nurses Station for capturing, archiving, and documenting surgical still/video images, audio sequences, and patient data. Not for primary diagnosis.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
K052159 — AIDA COMPACT II SYSTEM · KARL STORZ Endoscopy-America, Inc. · Sep 7, 2005
K103308 — AIDA HD CONNECT WITH SMARTSCREEN / BLU-RAY DRIVE / SMARTSCREEN WITH BLU-RAY DRIVE · KARL STORZ Endoscopy-America, Inc. · Aug 29, 2011
K060777 — SMITH & NEPHEW MODEL 660HD IMAGE MANAGEMENT SYSTEM · Smith & Nephew, Inc. · Apr 13, 2006
K123426 — MEDICVOD, MEDICAST, MEDISHOW, MEDISEARCH, CENTRAL STORAGE SYSTEM - CSS · Orpheus Medical , Ltd. · Mar 1, 2013
Submission Summary (Full Text)
{0}------------------------------------------------
K043324
Page 1 of 2
APR - 4 2005
## 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of the Safe Medical Devices Act (SMDA) of 1990 and 21 CFR 807.92. All data included in this document is accurate and complete to the best of KSEA's knowledge.
| Applicant: | Karl Storz Endoscopy - America, Inc.<br>600 Corporate Pointe Drive<br>Culver City, CA 90230<br>(310) 338-8100 |
|------------|---------------------------------------------------------------------------------------------------------------|
|------------|---------------------------------------------------------------------------------------------------------------|
2020550 Establishment Number:
Paul S. Lee Contact: Senior Regulatory Affairs Specialist Telephone +1-310-410-2769 Telecopier +1-310-410-5519
### Device Identification / Classification:
- AIDA/DICOM/HL7 Trade Name: .
- Classification Name: Picture Archiving and Communications System .
- Classification Panel: Reproductive, Abdominal, Radiological Devices .
- 21CFR892.2050 CFR Section .
- Class II Device Class: .
- LIZ Product Code: .
#### Indication:
This software is intended for use by qualified personnel in the Operating Room and Nurses Station. The Advanced Image and Data Archiving System (AIDA) is a Windows® based still/video image capturing, archiving, documentation system and recording of audio sequences and patient data during a procedure. It allows capture and annotation of the
Confidential
{1}------------------------------------------------
KC43324
Page 2 of 2
surgical procedure for documentation purposes. Images captured and distributed by AIDA are for viewing and reference purposes and are not intended for primary diagnosis.
It is also a Windows® based solution to communicate with other picture archival communication systems (PACS) using DICOM and with Hospital Information Systems (HIS) using the HL7 standard. Also as a part of the AIDA system the Storz Application Manager software (SAM) enables the selection and integrations of AIDA functions with various compatible applications, such as Karl Storz's Storz Communication Bus (SCB) or other third party image capturing devices.
#### Device Description:
The Karl Storz AIDA/DICOM/HL7 is an image capturing device with DICOM viewer which communicates with the Hospital Information System via HL7 protocol.
#### Substantial Equivalence:
The Karl Storz AIDA/DICOM/HL7 is substantially equivalent to the predicate device (SIENET MagicWeb and Magic Link I: K973131) since the basic features and intended uses are the same. The minor differences between the Karl Storz AIDA/DICOM/HL7 and the predicate device raise no new issues of safety and effectiveness, as these differences have no effect on the performance, function or intended use of these devices.
#### General Safety and Effectiveness Concerns:
Built in screen prompts, alarms, error messages and warnings ensure safe and effective use of the AIDA/DICOM/HL7 software. Risk management is ensured via risk analyses, which are used to identify potential hazards. These hazards are controlled via software development, verification and validation testing. To minimize electrical, mechanical and radiation hazards, KSEA adheres to recognized and established industrial practice and standards.
Signed: ______________________________________________________________________________________________________________________________________________________________________
signature
Paul Lee Senior Regulatory Affairs Specialist
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/2 description: The image shows a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES" written around the edge. In the center of the seal is a stylized image of an eagle with its wings spread. The eagle is facing to the left and has three lines representing its body and wings.
APR - 4 2005
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. Paul S. Lee Senior Regulatory Affairs Specialist Karl Storz Endoscopy - America, Inc. 600 Corporate Pointe CULVER CITY CA 90230
Re: K043324
Trade/Device Name: AIDA with DICOM and HL7 interface Regulation Number: 21 CFR §876.1500 Regulation Name: Endoscope and accessories Product Codes: KOG and FET Regulation Number: 21 CFR §892.2050 Regulation Name: Picture archiving and communications system Product Code: LLZ Regulatory Class: II Dated: March 1, 2005 Received: March 16, 2005
#### Dear Mr. Lee:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) I mis leter with anow you to begin maneting of substantial equivalence of your device to a legally premarket notification. The PDA michig of oatsamments
marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific advice for your devices on the following numbers, based on the regulation number at the top of this letter:
| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|-----------------|----------------------------------|--------------|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Also, please note the regulation emilion on your responsibilities under the Act from the 807.97). I ou may obtain office general missional and Consumer Assistance at its toll-free number (800) Drvision of Sillan Manufacturers, Internet address http://www.fda.gov/cdrh/dsmadsmamain.html
Sincerely yours,
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indication for Use Form
# 510(k) Number (if known): Not yet-assigned- KO43324
Device Name: AIDA with DICOM and HL7 interface
Indications for Use:
This software is intended for use by qualified personnel in the Operating Room and Nurses This sollware to intented Image and Data Archiving System (AIDA) is a Windows® based still/video image capturing, archiving, documentation system and recording of audio sequences and patient data during a procedure. It allows capture and annotation of the surgical procedure for documentation purposes. Images captured and distributed by AIDA sure for viewing and reference purposes and are not intended for primary diagnosis.
It is also a Windows® based solution to communicate with other picture archival It is also a windows® based belated to being DICOM and with Hospital Information Systems (HIS) coinmunication systems (11108) as a part of the AIDA system the Storz Application Manager asing the HD7 otandard. the selection and integrations of AIDA functions with various Software (SAM) chaolos the sereotion and " " " Storz Communication Bus (SCB) or other third party image capturing devices.
Prescription Use: _/
(Per 21 CFR 801.Subpart D)
AND/OR Over-The-Counter Use: _ (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Concurrence of CDRH, Office of Device Evaluation (ODE)
Nancy Brogdon
(Division Sign-off)
Division of Reproductive, Abdominal. and Radiological Devices 510(k) Number .
Confidential
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches