← Product Code [FER](/submissions/GU/subpart-b%E2%80%94diagnostic-devices/FER) · K230439

# WellCare Anoscope (K230439)

_Well Care (Wuhan) Medical Technology Co., Ltd. · FER · Dec 22, 2023 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/GU/subpart-b%E2%80%94diagnostic-devices/FER/K230439

## Device Facts

- **Applicant:** Well Care (Wuhan) Medical Technology Co., Ltd.
- **Product Code:** [FER](/submissions/GU/subpart-b%E2%80%94diagnostic-devices/FER.md)
- **Decision Date:** Dec 22, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

WellCare Anoscope is intended for transanal use to provide physicians access to the anal sphincter, anus, and rectum and to perform various diagnostic and therapeutic procedures using additional accessories.

## Device Story

WellCare Anoscope is a single-use, disposable, sterile device used by physicians to access the anal canal for diagnostic and therapeutic procedures, including hemorrhoid surgery. The device consists of an anoscope, introducer, and sleeve, constructed from medical-grade polycarbonate. During operation, the device is inserted into the anus to open the anal passage; the sleeve fixes perianal tissue at a specific angle, providing a clear view of the operating field. It utilizes a separate disposable illuminator and approved light source for visualization. The device is used in clinical settings by physicians to facilitate access to the anal sphincter and rectum, aiding in procedures such as PPH and RPH hemorrhoid surgery. By providing stable access and visualization, it assists the clinician in performing surgical interventions, potentially improving procedural efficiency and patient outcomes.

## Clinical Evidence

No clinical data. Bench testing only, including physical strength, compatibility, chemical testing (reducing substances, pH, EO/ECH residues), and biocompatibility (cytotoxicity, sensitization, systemic toxicity, pyrogenicity, and irritation) per ISO 10993 standards.

## Technological Characteristics

Materials: Medical polymer PC. Components: Anoscope, introducer, sleeve. Energy: External light source. Connectivity: None. Sterilization: Ethylene Oxide. Form factor: Disposable, various sizes (diameters 15-28.4mm).

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

## Predicate Devices

- Family of disposable, sterile and not sterile THD Anoscope, Proctoscope, Rectoscope and Light-scope ([K121135](/device/K121135.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. Food & Drug Administration".

December 22, 2023

Well Care (Wuhan) Medical Technology Co., Ltd Zhou Huimin Ouality Manager No. 11# The No.1 Road Fozuling of East Lake High-Tech Development Zone Wuhan, Hubei 430205 China

Re: K230439

Trade/Device Name: WellCare Anoscope Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope And Accessories Regulatory Class: Class II Product Code: FER Dated: November 21, 2023 Received: November 21, 2023

#### Dear Zhou Huimin:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

{2}------------------------------------------------

Sincerely,

# Shanil P. Haugen -S

Shanil P. Haugen, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

# Indications for Use

510(k) Number (if known) K230439

Device Name WellCare Anoscope

Indications for Use (Describe)

WellCare Anoscope is intended for transanal use to provide physicians access to the anal sphincter, anus, and rectum and to perform various diagnostic and therapeutic procedures using additional accessories.

|--|

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

# Section 3 510(k) Summary

#### l 510(k) Submitter

| Device Submitter: | Well care (Wuhan) medical technology Co., Ltd..               |
|-------------------|---------------------------------------------------------------|
|                   | No. 11# The No. 1 road                                        |
|                   | Fozuling of East Lake High-Tech Development Zone Wuhan, Hubei |

Contact Person: ZhouHuimin Quality Manager Phone: +86-15626465068 E-mail: wec-qm@wellcare.onaliyun.com

# II Device

| Trade Name of Device: | WellCare Anoscope             |
|-----------------------|-------------------------------|
| Regulation Number:    | 21 CFR 876.1500               |
| Classification Name:  | Endoscope and acccessories    |
| Product Code:         | FER                           |
| Regulation Number:    | 21   CFR 876.1500             |
| Regulatory Class      | II                            |
| Review Panel          | GU - Gastroenterology-Urology |

## III Predicate Devices

| 510k Number           | K121135                                                                                               |
|-----------------------|-------------------------------------------------------------------------------------------------------|
| Trade Name of Device: | Family of disposable,sterile and not sterile THD<br>Anoscope,Proctoscope , Rectoscope and Light-scope |
| Regulation Number:    | 21 CFR 876.1500                                                                                       |
| Regulation Name:      | Endoscope and accessories                                                                             |
| Regulatory Class      | II                                                                                                    |
| Product Code:         | FER                                                                                                   |

## IV Device Description

WellCare Anoscope are intended for the examination of the anus,it also can be used as an auxiliary instrument for hemorrhoid surgery.WellCare Anoscope is composed of Anoscope, Introducer and Sleeve.The anoscopes are divided into the following models according to the way the components are configured and the differences in size : WellCare-ANO-Light Sterile . WellCare-ANO-Truncated Sterile.This is a single use, disposable device(s), provided sterile.

#### V Indications for use

WellCare Anoscope is intended for transanal use to provide physicians access to the anal sphincter, anus, and rectum and to perform various diagnostic and therapeutic procedures using additional accessories.

{5}------------------------------------------------

# VI Technological Characteristics Comparison

VI-1: Comparison of WellCare Anoscope

| Device<br>Characteristic | Subject Device                                                                                                                                                                                                                                                                                                                                                                                  | Predicate Device<br>(K121135)                                                                                                                                                                                                                                                                                                                                                   | Discussion                |        |                       |        |                           |
|--------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------|--------|-----------------------|--------|---------------------------|
| Indications<br>for Use   | WellCare Anoscope is intended for<br>transanal use to provide physicians<br>access to the anal sphincter, anus,<br>and rectum and to perform various<br>diagnostic and therapeutic<br>procedures using additional<br>accessories.                                                                                                                                                               | Family of disposable,<br>sterile and not sterile THD<br>Anoscope, Proctoscope,<br>Rectoscope and<br>Light-scope is a single-use<br>anoscope with integrated<br>LED lighting, suitable for<br>performing diagnostic and<br>operational anoscopes..                                                                                                                               | Identical                 |        |                       |        |                           |
| Operation<br>Principle   | WellCare Anoscope opens up the<br>anal passage and then exits the<br>anus, and the anal passage is<br>continuously visualized by the<br>speculum, and Sleeve fixes the<br>perianal tissue at a certain angle<br>during PPH and RPH hemorrhoid<br>surgery, leaving a space and angle<br>suitable for the operation. Thus, it<br>can be used as an auxiliary<br>instrument for hemorrhoid surgery | Gently press the The<br>ANOSCOPES into the<br>anus and slowly enter, then<br>pull out the core, adjust the<br>light, and slowly exit from<br>deep to shallow, The<br>compact and<br>asymmetricaldesign of the<br>THD Surgy MaxiLED<br>allows easy access to the<br>analcanal without sphincter<br>traumaensuring good<br>dilation and anexcellent<br>view of the operatingfield | Identical                 |        |                       |        |                           |
| Design                   | WellCare Anoscope is composed of<br>Anoscope, Introducer and Sleeve, a<br>disposable illuminator and approved<br>light source.                                                                                                                                                                                                                                                                  | The LightScope LED<br>anoscope combines ina<br>single device a light<br>anoscope and LED light<br>source.                                                                                                                                                                                                                                                                       | Different<br>Comment<br>1 |        |                       |        |                           |
| Anoscope<br>material     | medical polymer material PC                                                                                                                                                                                                                                                                                                                                                                     | medical polymer material<br>PC                                                                                                                                                                                                                                                                                                                                                  | Identical                 |        |                       |        |                           |
| Size<br>(Unit: mm)       | Anoscope                                                                                                                                                                                                                                                                                                                                                                                        | Introducer                                                                                                                                                                                                                                                                                                                                                                      | Sleeve                    |        |                       |        |                           |
|                          |                                                                                                                                                                                                                                                                                                                                                                                                 |                                                                                                                                                                                                                                                                                                                                                                                 | BEVELLED<br>diameter      | length | TRUNCATED<br>diameter | length | Different<br>Comment<br>1 |

{6}------------------------------------------------

|                         | diamet<br>er                                        | lengt<br>h | diamete<br>r | Heig<br>ht | diamet<br>er | er                                          | h  | er   | h  |                           |
|-------------------------|-----------------------------------------------------|------------|--------------|------------|--------------|---------------------------------------------|----|------|----|---------------------------|
|                         |                                                     | 41.1±1     |              |            |              | 25                                          | 70 | 25   | 70 |                           |
|                         | 28.4±1                                              |            | 28±1         | /          | /            | 21.7                                        | 91 | 21.7 | 91 |                           |
|                         | 28.4±1                                              | 41.1±1     | 28±1         | 114±2      | 26±1         | 15                                          | 65 | 15   | 65 |                           |
|                         | 28.4±1                                              | 41.1±1     | /            | 114±2      | 26±1         | 22                                          | 90 | 22   | 90 |                           |
|                         | 19.9±1                                              | 80.2±1     | 19.5±1       | /          | /            | 26                                          | 90 | 26   | 90 |                           |
| Anoscope<br>Packaging   | sealed bag, Individually packaged                   |            |              |            |              | sealed bag, Individually packaged           |    |      |    | Identical                 |
| Sterilization<br>method | Sterilized by Ethylene Oxide                        |            |              |            |              | Sterilized by Ethylene Oxide                |    |      |    | Identical                 |
| Light                   | a disposable illuminator and approved light source. |            |              |            |              | Self-light, with an integrated light source |    |      |    | Different<br>Comment<br>1 |
| Singe Use<br>Only       | disposable                                          |            |              |            |              | disposable                                  |    |      |    | Identical                 |
| Shelf Life              | 2 years                                             |            |              |            |              | 2 years                                     |    |      |    | Identical                 |

#### Comment 1

The Design of the two are only different ,For example, in terms of size, dimensions, and light source design,But the actual intended use is the same, which does not affect the safety and effectiveness of the product.

#### Comment 2

The light source of the two are different,But the function is the same Therefore, the safety and effectiveness of the product will not be affected.

## VII Summary of Non-clinical Testing (Bench)

The non-clinical testing for WellCare Anoscope was performed to demonstrate verification testing in conformance with the acceptance criteria of test methods and recognized consensus standards shown below.

Table VII-1: Performance testing was conducted on the subject device

{7}------------------------------------------------

| ID#      | Test                                       | Method                                                                                                                                                                                                                                                                                                         | Acceptance Criteria                                                                                                                                                                                                                            | Conclusion |
|----------|--------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| 1        | Appearance                                 | The shapeshouldbesmoothandsymmetrical, with nosharp edgesand no burrs in all parts.Themirror tube should be rounded, witha smooth inner walland a roundededgeatthewithouth eadend,burrs.                                                                                                                       | Smooth surface of all parts no sharp edges,no burrs.                                                                                                                                                                                           | Pass       |
| 2        | Physical Testing of Anoscope               |                                                                                                                                                                                                                                                                                                                |                                                                                                                                                                                                                                                |            |
| 2.1      | compatibility                              | Take 1 set of samples and test them manually                                                                                                                                                                                                                                                                   | The anal expander and the guide mirror opened and closed easily with the scope, without swing and jam. The end of the anal expander should be smooth and anastomosed with the endoscope, and the maximum gap should not be greater than 0.6mm. | Pass       |
| 2.2      | Material<br>strength                       | Take 1 set of samples, place the assembly horizontally and hang a 4 kg suspension in the middle of the assembly in accordance with the requirements                                                                                                                                                            | Any of the components (excluding the light source) should be able to withstand a 4 kg suspension without significant deformation                                                                                                               | Pass       |
| 3        | Chemical Testing of Anoscope               |                                                                                                                                                                                                                                                                                                                |                                                                                                                                                                                                                                                |            |
| 3.1      | Reducing<br>substances<br>(oxide<br>prone) | The sample was taken, configured at a ratio of 0.1g plus 1mL of water and then thermostated at 37°C ± 1°C for 8h, the sample was separated from the liquid and cooled to room temperature as the test solution. Take the same volume of water in a glass container and prepare a blank control solution by the | The difference in the volume of potassium permanganate solution [ c(1/5KMnO4) = 0.01 mol/L ] consumed by the test solution and the blank control solution should not exceed 2.0 mL                                                             | 0.4ml      |
|          |                                            |                                                                                                                                                                                                                                                                                                                |                                                                                                                                                                                                                                                |            |
| 3.2      | pH                                         | same method<br>Take the test solution of 3.1<br>and the blank control<br>solution and determine their<br>pH values respectively with<br>an acidity meter                                                                                                                                                       | the difference between the pH<br>of the test solution and that of<br>the blank control solution<br>should not exceed 1.5                                                                                                                       | 1.01       |
| 3.3      | EO/ECH<br>residues                         | ISO 10993-7:2008                                                                                                                                                                                                                                                                                               | average daily dose to patient<br>EO < 4 mg, ECH < 9mg                                                                                                                                                                                          | Pass       |
| 3.4      | Sterile                                    | ISO 11135:2014                                                                                                                                                                                                                                                                                                 | WellCare Anoscope should be<br>sterile                                                                                                                                                                                                         | Pass       |
| <b>4</b> | <b>Biocompatibility Testing</b>            |                                                                                                                                                                                                                                                                                                                |                                                                                                                                                                                                                                                |            |
| 4.1      | In Vitro<br>Cytotoxicity                   | ISO 10995-5:2009                                                                                                                                                                                                                                                                                               | Non-cytotoxic                                                                                                                                                                                                                                  | Pass       |
| 4.2      | Skin<br>Sensitization<br>Test              | ISO 10993-10:2021                                                                                                                                                                                                                                                                                              | Non-sensitizer                                                                                                                                                                                                                                 | Pass       |
| 4.3      | Acute<br>Systemic<br>Toxicity Test         | ISO 10993-11:2017                                                                                                                                                                                                                                                                                              | No systemic toxicity                                                                                                                                                                                                                           | Pass       |
| 4.4      | Pyrogen<br>Test                            | ISO 10993-11:2017                                                                                                                                                                                                                                                                                              | Non-pyrogen                                                                                                                                                                                                                                    | Pass       |
| 4.5      | Irritation Test                            | ISO 10993-23:2021                                                                                                                                                                                                                                                                                              | Non-Irritation                                                                                                                                                                                                                                 | Pass       |

{8}------------------------------------------------

## VIII Clinical Test Conclusion

No clinical study is included in this submission.

# IX Conclusion

The conclusions drawn from the nonclinical tests demonstrate that the WellCare Anoscope is as safe as effective, and performs as well as or better than the legally marketed device.

---

**Source:** [https://fda.innolitics.com/submissions/GU/subpart-b%E2%80%94diagnostic-devices/FER/K230439](https://fda.innolitics.com/submissions/GU/subpart-b%E2%80%94diagnostic-devices/FER/K230439)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
