GELPOINT PATH TRANSANAL ACCESS PLATFORM

K133393 · Applied Medical Resources Corp. · FER · Jan 2, 2014 · Gastroenterology, Urology

Device Facts

Record IDK133393
Device NameGELPOINT PATH TRANSANAL ACCESS PLATFORM
ApplicantApplied Medical Resources Corp.
Product CodeFER · Gastroenterology, Urology
Decision DateJan 2, 2014
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The GelPOINT Path Transanal Access Platform is indicated for multiple instrument or camera access through the anus to perform various diagnostic and therapeutic procedures by using additional accessories.

Device Story

GelPOINT Path is a transanal access platform inserted into the rectum to establish a port for transanal endoscopic microsurgery. Device consists of a 40mm diameter cannula and an airtight gel lid. Lid allows insertion of multiple trocars for laparoscopes and surgical instruments while maintaining insufflation. Used in clinical settings by physicians. Insufflation via CO2 pump enlarges operative space, facilitating diagnostic and therapeutic procedures. Provides greater access, operative space, and instrument articulation compared to traditional proctoscopes. Can be sutured to patient for retention. Detachable cap allows removal of large specimens.

Clinical Evidence

No clinical data. Substantial equivalence established via dimensional analysis and comparison of technological characteristics against predicate devices.

Technological Characteristics

Polymer-based construction; 40mm diameter cannula; 4.5cm length; three access ports for instruments up to 10mm; airtight gel lid for trocar placement; CO2 pump insufflation; disposable; manual insertion via dilator.

Indications for Use

Indicated for patients requiring diagnostic or therapeutic procedures via transanal access; allows multiple instrument or camera access through the anus.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls: 1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following: 1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures. 2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use. 3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use. 2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes. 3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches 4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error. 5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program. (6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms. (7) Labeling must include: (i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified; (ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA; (iii) Identification of compatible devices; (iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification; (v) Reprocessing instructions for reusable components; (vi) A shelf life for any sterile components; (vii) A description of the device-specific use training program; (viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and (ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance. (8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include: (i) Device motion accuracy and repeatability; (ii) System testing; (iii) Instrument reliability; (iv) Crosstalk; (v) Table motion control; (vi) Thermal effects on tissue; (vii) User-device interface performance; (viii) Workspace access testing; and (ix) Performance testing with compatible devices. (9) Software verification, validation, and hazard analysis must be performed. (10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed. (11) Performance data must demonstrate the sterility of all patient-contacting device components. (12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life. (13) Performance data must validate the reprocessing instructions for the reusable components of the device. (14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible. (15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic. (16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information: (i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and (ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K133393 Page 1 of 2 Image /page/0/Picture/1 description: The image shows the words "Applied Medical" in a bold, sans-serif font. The word "Applied" is in a larger font size than the word "Medical", and it is positioned above the word "Medical". The text is black and the background is white. Image /page/0/Picture/2 description: The image shows a logo with the text "JAN - 22" to the right of it. The logo is a stylized triangle shape with wavy lines inside. The triangle is filled with a pattern of small dots or stars, giving it a textured appearance. ## 510(k) SUMMARY Applied Medical Resources Corporation 510(k) OWNER: 22872 Avenida Empresa Rancho Santa Margarita, CA, 92688 (949) 713-8000 CONTACT PERSON: Frans VandenBroek Principal Specialist, Regulatory Affairs (949) 713-8369 (949) 713-8205 (FAX) DATE OF PREPARATION: November 1, 2013 TRADE NAME: GelPOINT Path Transanal Access Platform COMMON NAME: Transanal Endoscopic Microsurgery (TEM) device CLASSIFICATION NAME: Endoscope and Accessories, Gastroenterology and Urology Devices, 21CFR 876.1500. Product code FER, Anoscope and accessories. K070915, SAPIMED S.P.A. Disposable Sigmoidoscope/ PREDICATE DEVICE: proctoscope; product code KOG, Endoscope and accessories. K110792, Applied Medical Resources GelPOINT Path Transanal Access Platform; product code FER, Anoscope DEVICE DESCRIPTION: The subject device is inserted into the rectum to establish an access port into the lower sigmoid colon in preparation for transanal endoscopic microsurgery. The port lumen measures 40mm in diameter and may be capped off with an airtight lid. The lid allows the colon to be insufflated, thus enlarging the operative space. and accessories. The airtight lid is constructed of a gel material through which multiple trocars may be placed. These trocars establish multiple ports that allow passage of laparoscopes and laparoscopic instruments without loss of insufflation. The design has all the functionality of a simple, two-piece proctoscope but with significant additional versatility, capability and effectiveness. INTENDED USE: The subject device was previously cleared in K 110792 and its intended use is unchanged. It provides access to the rectal cavity to perform surgery, which is also the intended use of the Sapimed disposable proctoscope, cleared in K070915. The subject device has all the functionality of the Sapimed proctoscope with the additional benefit of {1}------------------------------------------------ delivering greater access and operative space. Therefore, the indications for the Sapimed device may be incorporated into the indications for the subject device as follows: The GelPOINT Path is indicated for multiple instrument or camera access through the anus to perform various diagnostic and/or therapeutic procedures by using additional accessories. SUMMARY OF TECHNOLOGICAL CHARACTERISTICS: The design of the subject device is unchanged since it was cleared in K110792. Similarities to the predicate Sapimed proctoscope are: - . Both dilate the anus to allow access to the rectal cavity - Both are inserted using a dilator ● - Both allow insertion of instruments including endoscopes . - Both are made of polymers . - Both are disposable . - � Both are capable of insufflation The predicate Sapimed proctoscope consists of two simple injection molded components, a 17mm cannula and an insertion dilator. By comparison, the subject device is more versatile. Differences include: - The cannula is larger in diameter (40mm vs. 17mm) and shorter (4.5cm vs 14cm) . - . Insufflates with a CO2 pump. The predicate device insufflates with a hand held bulb - May be sutured to the patient to assist retention . - Allows instrumentation to be articulated over a greater range of motion . - Has three access ports, each of which can accommodate standard laparoscopic . instruments 10mm and smaller - A detachable airtight cap allows removing specimens that are too large to be . removed through the proctoscope's lumen Safety and efficacy of the subject device and the Sapimed proctoscope were previously established in K110792 and K070915. In spite of the differences in design, the subject device functions-in essence-as a proctoscope with greater insufflation capability. Insufflation enlarges the operative field and therefore more effectively allows execution of medical procedures in the colon. DISCUSSION OF NONCLINICAL TESTS SUBMITTED: Subject and predicate device were both previously cleared for accessing the rectum. However, the subject device has numerous superior technological capabilities, not available in the Sapimed proctoscope. There are no recognized standard tests that can challenge and compare these technological differences. Therefore, substantial equivalence was established based on a dimensional analysis of each device. CONCLUSIONS DRAWN FROM TESTING: The subject device is substantially equivalent in performance to the Sapimed proctoscope but has numerous technological advantages that increase effectiveness. Therefore, the indications for use of the subject device may be undated by incorporating the indications of the Sapimed device. {2}------------------------------------------------ ## DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 January 2, 2014 Applied Medical Resources Corporation Frans VandenBroek Principal Specialist, Regulatory Affairs 22872 Avenida Empresa Rancho Santa Margarita, CA 92688 - Re: K133393 Trade/Device Name: GelPOINT Path Transanal Access Platform Regulation Number: 21 CFR8 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: FER Dated: November 1, 2013 Received: November 5, 2013 Dear Frans VandenBroek, We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. {3}------------------------------------------------ Page 2 - Frans VandenBroek You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/Centers0ffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Benjamin Fisher -S Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ## INDICATIONS FOR USE ## 510(k) Number (if known): K133393 Device Name: GelPOINT Path Transanal Access Platform Indications for Use: The GelPOINT Path Transanal Access Platform is indicated for multiple instrument or camera access through the anus to perform various diagnostic and therapeutic procedures by using additional accessories. x Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) Concurrence of CDRH, Office of Device Evaluation (ODE) Benjamin R=Eisher -S 2014.01.02 14.29 40 -05'00' Page 1 of Sec 1 - 1
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...