← Product Code [FEQ](/submissions/GU/subpart-b%E2%80%94diagnostic-devices/FEQ) · K132340

# ERBEFLO CLEVERCAP HYBRID TUBING/CAP SET FOR PENTAX SCOPES, ERBEFLO CLEVERCAP HYBRID CO2 TUBING/CAP SET FOR OLYMPUS 160/1 (K132340)

_Erbe USA, Inc. · FEQ · Dec 17, 2013 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/GU/subpart-b%E2%80%94diagnostic-devices/FEQ/K132340

## Device Facts

- **Applicant:** Erbe USA, Inc.
- **Product Code:** [FEQ](/submissions/GU/subpart-b%E2%80%94diagnostic-devices/FEQ.md)
- **Decision Date:** Dec 17, 2013
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** 3rd-Party Reviewed

## Intended Use

The ERBEFLO CleverCap® Hybrid CO2 Tubing/Cap Sets and CO2 Connector Tube provide sterile water and CO2 or air (if CO2 is not used) to an endoscope for endoscopic procedures.

## Device Story

Device consists of sterile, disposable Hybrid CO2 Tubing/Cap Sets and non-sterile, disposable CO2 Connector Tubes. Sets interface with specific endoscopes (Pentax, Olympus, Fujinon) and irrigation pumps. Cap attaches to sterile water bottle; irrigation tubing delivers water for lavage; air/water tubing delivers air/CO2 for insufflation and lens cleaning. CO2 Connector Tube includes hydrophobic air/gas filter to prevent fluid backflow into CO2 source. Used in clinical endoscopic settings by physicians. Provides conduit for irrigation and insufflation; enhances visualization and procedural utility. Benefits include sterile fluid delivery and controlled insufflation during endoscopy.

## Clinical Evidence

Bench testing only. Evaluations included biocompatibility, flow rate performance, pressure testing, 24-hour durability, 2X sterilization functional testing, and packaging integrity. Results demonstrated performance comparable to or better than predicate devices.

## Technological Characteristics

Medical grade materials including plastics, nickel-plated brass, nitrile rubber, acrylic, nylon, ink, solvent, and adhesive. Components include tubing segments, caps, luer connectors, and hydrophobic air/gas filters. Disposable, ethylene oxide sterilized. No software or electronic components.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

## Predicate Devices

- Universal Irrigation Solution Hybrid™ ([K102855](/device/K102855.md))
- ERBEFLO CleverCap® Hybrid Tubing/Cap Sets ([K103696](/device/K103696.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K132340

page 1 of 4

ERBE USA. Inc. Traditional 510(k)/Third Party Review: ERBEFLO CleverCap® Hybrid CO2 Tubing/Cap Sets and CO2 Connector Tube

DEC 17 2013

| 510(k) SUMMARY                        |                                                                                                                                         |
|---------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------|
| Submitted By:                         | ERBE USA, Inc.<br>2225 Northwest Parkway<br>Marietta, GA 30067<br>Tel: 770-955-4400<br>Fax: 770-955-2577                                |
| Contact Person:                       | John Tartal<br>QA/RA Director                                                                                                           |
| Date Prepared:                        | May 22, 2013                                                                                                                            |
| Common Name:                          | Endoscopic Irrigation and CO2 or Air Tubing/Cap Sets<br>and CO2 Connector Tube                                                          |
| Trade/Proprietary Name:               | ERBEFLO CleverCap® Hybrid CO2 Tubing/Cap Sets and<br>CO2 Connector Tube                                                                 |
| Classification Name:                  | Endoscopes and Accessories (21 CFR Part 876.1500)                                                                                       |
| Product Code:                         | FEQ                                                                                                                                     |
| Legally Marketed<br>Predicate Device: | Universal Irrigation Solution Hybrid™, 510(k) Number<br>K102855 and ERBEFLO CleverCap® Hybrid Tubina/Cap<br>Sets, 510(k) Number K103696 |

Device Description:

In general; the Hybrid CO2 Tubing/Cap Sets and CO2 Connector Tubes will be manufactured with medical grade materials or agents used in the medical device industry such as plastics, nickel plated brass, nitrile rubber, acrylic, nylon, ink. solvent, adhesive, etc. The devices provide as a conduit for water for endoscopic irrigation and lens cleaning as well as air or CO2 for insufflation.

There are three (3) types of ERBEFLO CleverCap Hybrid CO>Tubing/Cap Sets for which each respectively interfaces with a specified brand of scope (i.e., Pentax, Olympus, and Fujinon) along with one (1) CO2 Connector Tube to attach to a CO2 source. The Sets consist of multiple tubing segments and a cap. The cap of a Set attaches with an air tight seal to a water source (i.e., a sterile water bottle). Then from the water bottle cap, irrigation tubing of a Set interfaces with a designated pump and via ERBEFLO 2 connector accessories to the specified endoscope for endoscopic lavage. The next segment, the air/water tubing (also coming from the same water bottle cap), connects to an air/water port of a specified scope for air insufflation as well as lens cleaning [Note: The air/water tubing is a tube within a

{1}------------------------------------------------

#### ERBE USA. Inc. Traditional 510(k)/Third Party Review: ERBEFLO CleverCap® Hybrid CO2 Tubing/Cap Sets and CO2 Connector Tube

tube in which the endoscope's processor or CO2 Source (if used) is used to pressurize the bottle for functionality (air and water to the endoscope). Also, for the Pentax Set there is an additional air inlet tube that directs air for endoscope functional use. Or if CO2 is used, pressurization (air and water function) occurs via a CO2 Source through a CO2 Connector Tube and a Pentax CO2 Adaptor (Pentax part #OF-G11).]. The third and final seqment for both the Olympus and Fujinon Sets is for connecting to the CO2 Connector Tube for CO2 insufflation. For each Set, both the irrigation as well as the air/water tubing segments have a backflow valve and a clamp to close the tubing while not in use. Additionally, each Set is designed for use with designated irrigation pumps and has an air/water connector(s) for its specified The Hybrid CO2Tubing/Cap Sets are provided sterile and are endoscope. disposable.

The CO₂ Connector Tube has a standard female luer connector for accessing a CO₂ Source. The other end of the Tube has a hydrophobic air/gas filter which filters particulates from the CO2 Source and keeps fluid from flowing into the CQ2 Source. The filtered end of the CO2 Connector Tube attaches to the CO2 segments of the Sets or in the case of the Pentax Set, the Pentax CO2 Adaptor that adjoins the Set to the Endoscope. The CO2 Connector Tube is provided non-sterile and is disposable.

## Intended Use:

The ERBEFLO CleverCap Hybrid CO2 Tubing/Cap Sets and CO2 Connector Tube provide sterile water and CO2 or air (if CO2 is not used) to an endoscope for endoscopic procedures.

Similarities and Differences of the Proposed Device to the Current Device (Predicate Comparison/Substantial Equivalence):

## Similarities

The ERBEFLO CleverCap Hybrid CO2 Tubing/Cap Sets have the same intended use as the Universal Irrigation Solution Hybrid™ Sets and a similar intended use to the ERBEFLO CleverCap Hybrid Tubing/Cap Sets. The proposed Sets have the same thread connections, cap, tubing seqments, and valve placements in comparison to one or both of the predicates. The tubing of the Hybrid CO2 Sets has the same or comparable durometer, Inner Diameters (I.D.s), Outer Diameters (O.D.s), and lengths as the predicate devices. The proposed and predicate devices both have the same locking clamp. Also, the Hybrid Tubing/Cap Sets have the same duration of use as the predicate devices (24 hour use). The proposed Sets and predicate devices use the same types of water bottles, pumps, endoscope connection accessories, endoscopes, and CO2 sources. Finally, the proposed and predicate devices are sterilized via Ethylene Oxide and disposable.

The ERBEFLO CleverCap CO2 Connector Tube compared to Universal Irrigation Solution Hybrid™ CO2 Source Tubing in that both are manufactured with similar materials, are provided non sterile, have the same luer fittings and a hydrophobic

{2}------------------------------------------------

K132340 pg. 3 of 4

## ERBE USA. Inc. Traditional 510(k)/Third Party Review: ERBEFLO CleverCap® Hybrid CO2 Tubing/Cap Sets and CO2 Connector Tube

air/gas filter, and are used in the same manner for the same duration. Differences

For the Sets, there is one difference not covered by either predicate. The ERBEFLO CleverCap Hybrid CO2 Tubing/Cap Set will have a luer actuated valve on the CO2 access tube as compared to the predicate device which has a backflow valve. Evaluations and testing as described below demonstrated the safety and efficacy of this Set.

The ERBEFLO CleverCap CO2 Connector Tube is different than the Universal Irrigation Solution Hybrid ™ CO2 Source Tubing in that it is one piece (i.e., it is not in seaments), it is shorter in length, and its filter is not as prominent (i.e., a smaller footbrint). None of these differences impacted the safety or efficacy of the ERBEFLO CleverCap CO2 Connector Tube.

#### Evaluations and Testing:

The following evaluations and tests were performed on the ERBEFLO CleverCap Hybrid CO2 Tubing/Cap Sets and CO2 Connector Tube to demonstrate the safety and efficacy.

Biological Evaluation

The evaluation was performed per the current recognized standard and demonstrated that there were no biocompatibility issues with the materials used for the proposed products.

Performance Feasibility Testing

The feasibility testing showed that the flow performance of the proposed products was as good or better than the predicate devices.

Flow Rate Testing

Flow rates for irrigation and lens cleaning were found to be comparable to or better than the predicate devices and/or commercially available products. Pressure Testing

The pressure testing demonstrated that the proposed devices would withstand maximum internal pressures that could be encountered during normal use.

Durability Testing

The products were tested and proved durable for 24 hour use.

2X Sterilization Functional Testing

The testing demonstrated that the proposed products upon 2X sterilization met established performance specifications.

Packaging Evaluation

The evaluation demonstrated the adequacy and integrity of the packaging for the proposed products.

Sterilization Evaluation

The evaluation was performed using current recognized standards to demonstrate product sterility.

Conclusion:

{3}------------------------------------------------

## ERBE USA, Inc. Traditional 510(k)/Third Party Review: ERBEFLO CleverCap® Hybrid CO2 Tubing/Cap Sets and CO2 Connector Tube

The ERBEFLO CleverCap Hybrid CO2 Tubing/Cap Sets and CO2 Connector Tube's intended use is the same, although written differently, as the Universal Irrigation Solution Hybrid™ predicate. Also, the proposed device's intended use is similar to the ERBEFLO CleverCap Hybrid Tubing/Cap Sets with exception of the CO2 access/use addition. The proposed Hybrid CO2 Tubing/Cap Sets and CO2 Connector Tube have the same principles of operation and technological characteristics as the predicate devices. The duration of use for the proposed and predicates is the same. As compared to the predicates, the proposed Sets are constructed with the same type of materials as well as have comparable performance characteristics. In conclusion, the ERBEFLO CleverCap Hybrid CO2 Tubing/Cap Sets and CO2 Connector Tube did not adversely affect safety or effectiveness.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the text "DEPARTMENT OF HEALTH & HUMAN SERVICES" in all capital letters. The text is in a bold, sans-serif font and is arranged on a single line. The words are evenly spaced and the overall appearance is clean and professional.

Image /page/4/Picture/1 description: The image shows a logo for the Department of Health & Human Services - USA. The logo features a stylized eagle-like symbol with three curved lines forming the body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the symbol. The text is in all capital letters and appears to be in a sans-serif font.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

December 17, 2013

ERBE USA, Inc. % Mark Job Responsible Third Party Regulatory Technology Services, LLC 1394 25d Street NW Buffalo, MN 55313

K132340 Re:

Trade/Device Name: ERBEFLO CleverCap® Hybrid CO2 Tubing/Cap Sets and CO2 Connector Tube Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscopes and Accessories Regulatory Class: Class II Product Code: FEQ Dated: November 27, 2013 Received: December 2, 2013

Dear Mark Job: .

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be may of subject to additions, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{5}------------------------------------------------

Page 2 - Mark Job

, .. . . . . . . . .

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# Herbert P. Lerner -S

for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

## Indications for Use

510(k) Number (if known): __K132340

Device Name: ERBE USA, Inc.'s ERBEFLO CleverCap® Hybrid CO2 Tubing/Cap Sets and CO2 Connector Tube

Indications For Use:

The ERBEFLO CleverCap® Hybrid CO2 Tubing/Cap Sets and CO2 Connector Tube provide sterile water and CO2 or air (if CO2 is not used) to an endoscope for endoscopic procedures.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Herbert P. Lerner -S 2013.12.17 16:18:08 -05'00'

Page 1 of 1

---

**Source:** [https://fda.innolitics.com/submissions/GU/subpart-b%E2%80%94diagnostic-devices/FEQ/K132340](https://fda.innolitics.com/submissions/GU/subpart-b%E2%80%94diagnostic-devices/FEQ/K132340)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
