K233472 · Calyxo, Inc. · FED · Feb 2, 2024 · Gastroenterology, Urology
Device Facts
Record ID
K233472
Device Name
CVAC Aspiration System + CVAC Image Processor
Applicant
Calyxo, Inc.
Product Code
FED · Gastroenterology, Urology
Decision Date
Feb 2, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.1500
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The CVAC Aspiration System and CVAC Image Processor consists of a sterile single use, steerable ureteral catheter and a reusable software-controlled image processor. It is intended to establish a conduit during endoscopic urological procedures for the treatment and removal of urinary stones (kidney stones, fragments, and dust). It employs flexible ureteroscopy within the urinary tract for endoscopic examination of the urinary tract and the interior of the kidney. The steerable ureteral catheter is used for irrigation and aspiration of kidney stones and stone dust with dedicated irrigation and vacuum channels under ureteroscopy. The CVAC Aspiration System and CVAC Image Processor can be used with additional accessories to perform various diagnostic and therapeutic procedures.
Device Story
CVAC Aspiration System and CVAC Image Processor facilitate kidney stone removal during endoscopic urological procedures. System comprises sterile, single-use steerable ureteral catheter and reusable, software-controlled image processor. Catheter features dedicated irrigation and vacuum channels; includes introducer, laser bridge, trigger tie, 3-way stopcock, and vacuum port cap. Processor connects to catheter via video cable; houses LED control, display interface, and System-on-Module (SOM) to generate video for standard hospital monitors. Used in surgical suites by physicians. Catheter provides direct visualization via distal CMOS chip and dual LEDs; allows insertion of laser fibers and stone baskets. Vacuum (150-200 mmHg) enables aspiration of stones/dust. Hydrophilic coating aids catheter placement. System streamlines stone removal by combining visualization with aspiration capabilities.
Clinical Evidence
Bench testing only. Performance verification/validation confirmed aspiration/irrigation functionality and design specifications. Usability assessment confirmed intended users safely performed critical tasks. Biocompatibility testing (ISO 10993-1) passed. Electrical safety/EMC (IEC 60601 series) and optical testing (ISO 8600 series) passed. Sterilization (EtO) validated to SAL 10^-6 per ISO 11135. Packaging/shelf-life testing supported 12-month shelf life.
Technological Characteristics
Single-use steerable ureteral catheter (PEBAX tip, hydrophilic coating) and reusable electrical image processor. CMOS camera chip (200x200 resolution), LED illumination (6000 lux). Irrigation/vacuum channels. Connectivity: HDMI/DVI output to hospital monitors. Power: 110V AC. Sterilization: EtO (ISO 11135). Software: IEC 62304 compliant.
Indications for Use
Indicated for endoscopic urological procedures to establish a conduit for treatment and removal of urinary stones (kidney stones, fragments, and dust) and for endoscopic examination of the urinary tract and kidney interior.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:
1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.
Axis Single-Use Digital Flexible Ureteroscope (K180367)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the text "FDA U.S. FOOD & DRUG ADMINISTRATION" on the right. The symbol on the left is a stylized image of a human figure, and the text on the right is in blue. The text is arranged in three lines, with "FDA" on the first line, "U.S. FOOD & DRUG" on the second line, and "ADMINISTRATION" on the third line.
February 2, 2024
Calyxo, Inc. Diane Mandell Horwitz, Ph.D. Regulatory Consultant 4473 Willow Road. Suite 100 Pleasanton, CA 94588
Re: K233472
Trade/Device Name: CVAC Aspiration System and CVAC Image Processor Regulation Number: 21 CFR§ 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: II Product Code: FED, FGB Dated: January 4, 2024 Received: January 4, 2024
Dear Diane Mandell Horwitz:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
## Mark R. Kreitz -S
for Mark J. Antonino, M.S. Assistant Director DHT3B: Division of Reproductive. Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K233472
Device Name
CVAC Aspiration System and CVAC Image Processor
## Indications for Use (Describe)
The CVAC Aspiration System and CVAC Image Processor consists of a sterile single use, steerable ureteral catheter and a reusable software-controlled image processor. It is intended to establish a conduit during endoscopic urological procedures for the treatment and removal of urinary stones (kidney stones, fragments, and dust). It employs flexible ureteroscopy within the urinary tract for endoscopic examination of the urinary tract and the interior of the kidney.
The steerable ureteral catheter is used for irrigation of kidney stones and stone dust with dedicated irrigation and vacuum channels under ureteroscopy. The CVAC Aspiration System and CVAC Image Processor can be used with additional accessories to perform various diagnostic and therapeutic procedures.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Calyxo. The logo is blue and features the word "Calyxo" in a stylized font. The "C" in Calyxo is large and bean-shaped. The rest of the letters are smaller and connected.
## 510(k) SUMMARY
#### GENERAL INFORMATION 1.
#### Submitter and 510(k) Owner 1.1
Calyxo, Inc. 4473 Willow Road, Suite 100 Pleasanton, CA 94588
#### 1.2 Official Correspondent
Diane Horwitz, Ph.D. Regulatory Consultant 5 Lake Como Ct. Greenville SC 29609 Telephone: 703.307.2921 Email: diane.horwitz(@calyxoinc.com
#### Date of Preparation 1.3
January 11, 2024
#### 2. NAME OF THE DEVICE
#### 2.1.1 Trade/Proprietary Name
CVAC Aspiration System and CVAC Image Processor
#### 2.1.2 Common/Usual Name
Endoscopic Access Overtube, Gastroenterology-Urology Ureteroscope and Accessories, Flexible/Rigid
#### 2.1.3 Classification Information
| Classification Name: | Endoscope and Accessories |
|----------------------------|----------------------------------------------------------------------------------------------------------------|
| Classification Regulation: | 21 CFR 876.1500 |
| Class: | II |
| Product Code: | FED, Endoscopic Access Overtube, Gastroenterology-Urology<br>FGB, Ureteroscope and Accessories, Flexible/Rigid |
| Panel: | Urology |
#### PREDICATE DEVICES 3.
The primary predicate device is the CVAC Aspiration System (formerly K-VAC Aspiration Catheter, Calyxo, Inc, K200149). The secondary predicate is Axis Single-Use Digital Flexible Ureteroscope (K180367, Shanghai AnQing Medical Instrument CO., Ltd.).
l
{4}------------------------------------------------
#### 4. DESCRIPTION OF THE DEVICE
The single use CVAC Aspiration System and the reusable CVAC Image Processor are specialized devices designed to allow physicians to locate kidney stones and to establish a conduit during endoscopic urological procedures for the treatment and removal of urinary stones (kidney stones, fragments, and dust). The devices are used in a surgical suite and are compatible with standard hospital monitors, vacuum sources, and irrigation sources.
The CVAC Aspiration System is a sterile single-use steerable ureteral catheter employing ureteroscopy with built-in irrigation and vacuum functionality. Accessories provided with the CVAC Aspiration System include an introducer, a laser bridge, a trigger tie, a 3-way stopcock, and a vacuum port cap. The CVAC Aspiration System catheter is compatible with commercially available accessories for stone removal such as appropriately sized single-use laser fibers and stone baskets, which are inserted and used in the vacuum channel. The CVAC Aspiration System connects to the CVAC Image Processor via an attached video cable.
The CVAC Image Processor is a software-controlled, reusable and electrical image processor. The CVAC Image Processor is connected to a standard 110V hospital electrical outlet via a provided power cord. The internal carrier board inside the CVAC Image Processor houses the LED control circuit, the display interface, and supports electrical interconnect between the image signal processing module and the System-on-Module (SOM). This allows the CVAC Image Processor to generate a video signal compatible with standard hospital grade external monitors. Accessories provided with the CVAC Image Processor include a HDMI-to-DVI adapter, a HDMI cable, and a power cable.
#### ട്. INTENDED USE
The intended use / indications for the CVAC Aspiration System and CVAC Image Processor is as follows:
The CVAC Aspiration System and CVAC Image Processor consists of a sterile single use, steerable ureteral catheter and a reusable software-controlled image processor. It is intended to establish a conduit during endoscopic urological procedures for the treatment and removal of urinary stones (kidney stones, fragments, and dust). It employs flexible ureteroscopy within the urinary tract for endoscopic examination of the urinary tract and the interior of the kidney.
The steerable ureteral catheter is used for irrigation and aspiration of kidney stones and stone dust with dedicated irrigation and vacuum channels under ureteroscopy. The CVAC Aspiration System and CVAC Image Processor can be used with additional accessories to perform various diagnostic and therapeutic procedures.
#### INTENDED USE COMPARED TO THE PREDICATE 6.
The intended use of the CVAC Aspiration System and CVAC Image Processor includes the combined intended uses of both predicate devices. The mode of operation is the same as the combined modes of action of both predicate devices but is based on the mode of action of the primary predicate. Placement procedures for the subject device and primary predicate are the same. The subject device provides visualization during the procedure, similar to the secondary predicate. The combination of the two capabilities provides a streamlined ability to remove kidney stones and stone dust.
{5}------------------------------------------------
#### 7. TECHNOLOGY CHARACTERISTICS COMPARED TO THE PREDICATE
There are similarities between the technological characteristics of CVAC Aspiration System and CVAC Image Processor and the primary predicate, and between the subject devices and the secondary predicate. Similarities to the primary predicate include the mode of use of the device, the working length, OD, ID, external catheter profile, the dedicated irrigation channel and aspiration channel and the functionality for irrigation of small kidney stones and stone dust.
Differences compared to the primary predicate include the built-in digital imaging in the subject device, the software controls and electronics, while the primary predicate relies on fluoroscopy for visualization. Additionally, the subject device has a hydrophilic coating to aid in placement.
The CVAC Aspiration System and CVAC Image Processor and the secondary predicate are similar in that both devices are flexible single-use sterile ureteroscopes designed to be used with one companion image processor. Both have visualization via a camera chip at the catheter tip, both are software-controlled and electronic. Table 1 below reviews the technological comparison between the subject device and both predicates.
The different technological characteristics between the subject device and the predicates have been evaluated and it has been determined the subject device does not raise new questions of safety and effectiveness. The sum of the technology of the primary predicate and the secondary predicate supports the substantial equivalence determination.
| | Subject Device<br>CVAC Aspiration System<br>and CVAC Image Processor | Primary Predicate<br>CVAC Aspiration System<br>K200419 | Secondary Predicate<br>Axis Disposable<br>Ureteroscope<br>K180367 |
|---------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------|
| Working<br>Length | 70 cm | 70 cm | 65 cm |
| Total Length | 100 cm | 90 cm | 92 cm |
| Shaft OD | <12 Fr (4 mm) | <12 Fr (4 mm) | 9.3 Fr (3.1 mm) |
| Working<br>Channel ID<br>(also referred<br>to as Vacuum<br>Channel) | 7 Fr (2.3 mm) | 7.5 Fr (2.5 mm) | 3.6 Fr (1.2 mm) |
| External<br>Catheter<br>Profile | Diameter and length<br>sufficient to reach kidney,<br>low durometer PEBAX outer<br>material at tip, smooth outer<br>surface | Diameter and length sufficient<br>to reach kidney, low durometer<br>PEBAX outer material at tip,<br>smooth outer surface | Diameter and length<br>sufficient to reach kidney |
| Navigation | Ureteroscopic navigation. | Initial placement and<br>navigation are done under<br>fluoroscopy. | Ureteroscopic navigation. |
| Coating | Hydrophilic Coating on distal<br>~45 cm of the catheter | No coating | No coating |
| Distal Tip<br>Configuration<br>(Camera) | Recessed CMOS chip and<br>dual LEDs for direct vision<br>capability | N/A | Recessed CMOS chip and<br>dual LEDs for direct vision<br>capability |
| Single or<br>multiple use | Single patient use | Single patient use | Single patient use |
| | Subject Device<br>CVAC Aspiration System<br>and CVAC Image Processor | Primary Predicate<br>CVAC Aspiration System<br>K200419 | Secondary Predicate<br>Axis Disposable<br>Ureteroscope<br>K180367 |
| IRRIGATION / ASPIRATION | | | |
| Irrigation | Via the irrigation port and<br>irrigation channel at catheter<br>tip. | Via the irrigation port and<br>irrigation channel at catheter<br>tip. | Via access port through the<br>working channel |
| | Irrigation channel is separate<br>and distinct from working<br>channel. | Irrigation channel is separate<br>and distinct from working<br>channel. | |
| Aspiration | Use hospital vacuum, setting<br>of 150 to 200 mmHg | Use hospital vacuum, setting of<br>150 to 200 mmHg | No vacuum capabilities |
| Stone<br>Collector | Integrated into handle | In-line stone collector separate<br>from device | No stone collector |
| | ACCESSORIES AND COMPATIBILITY WITH OTHER DEVICES | | |
| Compatible<br>medical<br>devices | Compatible with standard<br>guidewires and 200-series<br>laser fibers. | Compatible with standard<br>guidewires. | "Additional accessories" may<br>be used per Axis IFU. |
| | Compatible with standard<br>stone baskets size 1.9Fr | | This includes laser fibers that<br>fit inside the working channel<br>(including 200-series laser<br>fibers). Compatible with stone<br>baskets. |
| Introducer<br>Provided? | Yes (provided in sterile tray) | Yes (provided in sterile tray) | No |
| Compatible<br>Ureteral<br>Access Sheath<br>sizes | If clinician desires, can be<br>used with 12/14 Fr ureteral<br>sheath | If clinician desires, can be used<br>with 12/14 Fr ureteral sheath | If clinician desires, can be<br>used with 11/13 Fr or 12/14<br>Fr ureteral sheath |
| IMAGE PROCESSOR | | | |
| Image System | The CVAC Image Processor<br>is connected to the CVAC<br>Aspiration System and<br>provides power to the<br>ureteroscope | None | The Video Processor is<br>connected to the flexible<br>ureteroscope and provides<br>power to the ureteroscope |
| Compatible<br>Display | Hospital-grade High-<br>Definition Monitor with<br>HDMI or DVI input | N/A | Hospital-grade High-<br>Definition Monitor with<br>HDMI or DVI input |
| Resolution<br>(pixels) | 200 x 200 | N/A | 400 x 400 |
| Digital Video<br>Technology | Color CMOS | N/A | Color CMOS |
| Illumination | LED | N/A | LED |
| Field of View<br>(°) | 100° | N/A | 110° |
| Direction of<br>View (°) | 0° | N/A | 0° |
| Light (lux) | 6000 lux | N/A | >5000 lux |
Table 1. Technology Comparison of CVAC Aspiration System and CVAC Image Processor and Predicate Devices
{6}------------------------------------------------
# Table 1. Technology Comparison of CVAC Aspiration System and CVAC Image Processor
and Predicate Devices
{7}------------------------------------------------
### 8. PERFORMANCE TESTING
The 510(k) submission provided performance data to establish the substantial equivalence of the CVAC Aspiration System and CVAC Image Processor (the subject device) to the predicate devices. A summary of these performance tests is provided below.
Performance Test, Bench: Results of extensive bench performance verification and validation testing confirm that the aspiration and irrigation and other device characteristics meet product specifications, and the device meets design input requirements.
Usability Assessment: A usability assessment was conducted to evaluate the User's understanding of the User Manuals and Quick Reference Guide when performing and asked about critical tasks during use of the CVAC Aspiration System and CVAC Image Processor. Results confirm that the acceptance criteria for usability have been met by intended users in the intended use environments. All participants were observed to safely perform all primary operating functions and critical tasks. The risk profile was acceptable, user training and User Manuals and Quick Reference Guide were interpreted correctly.
Biocompatibility Data: Biocompatibility testing was conducted for the CVAC Aspiration System and accessories as an external communicating device, tissue/bone/dentin contacting, limited duration exposure. Cytotoxicity (MEM Elution Method), Sensitization (Guinea Pig Maximization Test, Magnusson-Kligman Method), Irritation (Intracutaneous test), Acute Systemic Toxicity, Material Mediated Pyrogen Test (LAL Test) testing all resulted in passing test results. This conclusion is that the CVAC Aspiration is biocompatible and meets the requirements of ISO 10993-1.
Electrical Safety and Electromagnetic Compatibility: Electrical safety and electromagnetic compatibility (EMC) was successfully performed to applicable requirements of IEC 60601-1, IEC 60601-1-2, and IEC 60601-4-2.
Optical Testing: Optical testing was successfully performed to appliable requirements of ISO 8600-1, ISO 8600-3, ISO 8600-4 and ISO 8600-6.
Software: Software verification and validation testing were conducted, and documentation was provided as recommended by FDA guidance documents, "Content of Premarket Submissions for Device Software Functions," issued on June 14, 2023, and "Off-The-Shelf Software Use in Medical Devices," issued on August 11, 2023. Cybersecurity risk assessment was conducted in accordance with FDA guidance document, "Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions," issued on September 27, 2023. The software complies with IEC 62304.
Sterilization: Sterilization validation studies using ethylene oxide (EtO) confirmed sterilization to a sterility assurance level (SAL) of 10°. The sterilization process was validated in accordance with ANSI AAMI ISO 11135: 2014, Over-Kill Approach.
Packaging and Shelf Life: Packaging validation testing, including transportation and distribution testing, demonstrated the product packaging protects and maintains device integrity and functionality for the CVAC Aspiration System and CVAC Image Processor. It also confirmed that the product packaging maintained product sterility and supports a 12-month shelf life for the CVAC Aspiration System.
#### CONCLUSIONS 9.
This 510(k) submission demonstrates that the CVAC Aspiration System and CVAC Image Processor is substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.