The Ureteral Access Sheath is intended to use in urologic endoscopic procedures to facilitate the passage of endoscopes.
Device Story
Disposable Ureteral Access Sheath facilitates urologic endoscopic procedures by providing a conduit for endoscopes and instruments into the urinary tract. Device consists of a sheath assembly and a dilator assembly with a hydrophilic coating to reduce tissue trauma. Physician advances dilator over a 0.038" guidewire to guide sheath into body orifice; placement confirmed via fluoroscopy. Dilator and sheath are radiopaque. Device accepts urological instruments with 10, 12, or 14 Fr outer diameters. Used in clinical settings by physicians to protect the ureter during device exchanges and instrument passage.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including biocompatibility (cytotoxicity, irritation, sensitization per ISO 10993-1:2009), tensile strength, dynamic friction force, bending resistance, and guidewire compatibility testing per EN 1618:1997 and internal requirements.
Indicated for use in urologic endoscopic procedures to facilitate the passage of endoscopes into the urinary tract for patients requiring such procedures.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:
1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the text "U.S. Food & Drug Administration" in blue.
April 30. 2021
Suzhou Beyo Medical Technology Co., Ltd. Aline Oin Supervisor of Quality Department No. 38 Beiguandu Road Suzhou, Jiangsu 215000 China
Re: K203165
> Trade/Device Name: Disposable Ureteral Access Sheath Regulation Number: 21 CFR§ 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: FED Dated: March 29, 2021 Received: April 5, 2021
Dear Aline Qin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Mark J. Antonino, M.S. Assistant Director DHT3B: Division of Reproductive. Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
### 510(k) Number (if known)
### K203165
### Device Name Disposable Ureteral Access Sheath
Indications for Use (Describe)
The Ureteral Access Sheath is intended to use in urologic endoscopic procedures to facilitate the passage of endoscopes.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (6/20)
{3}------------------------------------------------
011 Annex 1 510(k) Summary
This 510(k) Summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of Title 21, CFR Section 807.92.
The assigned 510(k) Number: K203165
#### SUBMITTER 1.
Suzhou Beyo Medical Technology Co., Ltd.
2nd Floor, 5th Building, No.38 Beiguandu Road, Wuzhong District, 215000 Suzhou,
P.R.China
Phone: +86-512-67269209
Contact Person: Aline Qin
Contact title: Supervisor of Quality Department
Mail: qinling@beyomed.com
Date Prepared: March 29, 2021
#### 2. DEVICE
Name of Device: Disposable Ureteral Access Sheath Common or Usual Name: Disposable Ureteral Access Sheath Classification Name: Endoscope and accessories Regulatory Class: II Product Code: FED Model Number: BY-AS-1025, BY-AS-1035, BY-AS-1040, BY-AS-1045, BY-AS-1055, BY-AS-1225, BY-AS-1235, BY-AS-1240, BY-AS-1245, BY-AS-1255, BY-AS-1425, BY-AS-1435, BY-AS-1440, BY-AS-1445, BY-AS-1455.
#### PREDICATE DEVICE 3.
Product Name: Well Lead Ureteral Access Sheath
510(k) Number: K151084
This predicate has not been subject to a design-related recall.
No reference devices were used in this submission.
#### DEVICE DESCRIPTION 4.
The Disposable Ureteral Access Sheath is one single use device, which is provided after EO sterilization. The shelf life of Disposable Ureteral Access Sheath is 3(three) years.
The Disposable Ureteral Access Sheath consists of two major components: sheath assembly
{4}------------------------------------------------
| Image: Beyomed logo | 011_Annex 1 510(k) Summary | Page: | 011-2 |
|---------------------|----------------------------|-------|-------|
|---------------------|----------------------------|-------|-------|
and dilator assembly. Sheath assembly consists of sheath hub, the dilator assembly consists of dilator, dilator hub and dilator clip. And there is a hydrophilic coating on the surface of the sheath and dilator.
The Disposable Ureteral Access Sheath is designed to provide the physician with reliable access to the urinary tract, the ability to inject fluids, and act as a conduit for device changes. Like all ureteral access sheath sets, Beyo's Disposable Ureteral Access Sheath also protects the ureter during device exchanges, thus helping reduce tissue trauma. Both the dilator and sheath are radiopaque and have a lubricous hydrophilic coating.
To guide the access sheath into the body orifice, the dilator is advanced over up to a 0.038" guidewire. The device can be visualized under X-ray(fluoroscopy) during placement to confirm location. The device can accept other urological instruments with OD's compatible with the sheath's OD of 10,12 and 14 Fr.
The device has two assemblies, sheath assembly and dilator assembly, the two assemblies are combined through dilator clip, the dilator can insert and withdrawn easily from sheath, The dilator assembly can be fixed on the sheath assembly.
#### 5. INDICATIONS FOR USE
The Ureteral Access Sheath is intended to use in urologic endoscopic procedures to facilitate the passage of endoscopes.
#### PERFORMANCE DATA 6.
The following performance data were provided in support of the substantial equivalence determination.
## Biocompatibility testing
According to ISO 10993-1:2009 Table A.1 -Evaluation tests for consideration and the intended use, the Disposable Ureteral Access Sheath is categorized as a surface device that contacts with mucosal membrane only, limited exposure. The biological safety tests recommended in ISO 10993-1:2009 for this category of devices are: cytotoxicity; irritation and sensitization.
The biological safety of the Disposable Ureteral Access Sheath has been evaluated in accordance with the requirements of ISO 10993-1:2009. The Disposable Ureteral Access Sheath contacts the human body component is sheath, dilator and hydrophilic coating and the material is summarized as below table 1.
Table 1 Patient Contact Material Identification
{5}------------------------------------------------
| Image: Beyomed logo | 011_Annex 1 510(k) Summary | Page: | 011-3 |
|---------------------|----------------------------|-------|-------|
|---------------------|----------------------------|-------|-------|
| NO. | Contact human<br>body<br>component<br>name | Material | Duration of<br>contact | Nature of<br>body<br>contact | |
|-----|--------------------------------------------|-------------------|-----------------------------------------------------------------|------------------------------|--------|
| 1 | Sheath | PEBAX+SUS+PTFE | <24hrs | Ureter | |
| 2 | Dilator | LDPE | <24hrs | Ureter | |
| 3 | Hydrophilic<br>coating | Primer<br>Coating | Polyurethane +<br>deionized water<br>mixture | <24hrs | Ureter |
| | | Top<br>Coating | Polyacrylamide +<br>isopropanol +<br>deionized water<br>mixture | | |
# Animal Study
This section is not applicable to the proposed device, because there is no animal testing of this device.
## Non-Clinical Test Conclusion
Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device.
| The test results demonstrated that the proposed device complies with the following standards: |
|-----------------------------------------------------------------------------------------------|
|-----------------------------------------------------------------------------------------------|
| No. | Standard Name | Title of Standard |
|-----|---------------|-------------------------------------------------------------------------------------|
| 1 | EN 1618:1997 | Catheters other than intravascular catheters-Test<br>methods for common properties. |
## Clinical Test Conclusion
No clinical study is included in this submission.
#### 7. Substantially Equivalent (SE) Comparison
{6}------------------------------------------------
| Image: Beyomed logo | 011_Annex 1 510(k) Summary | Page: | 011-4 |
|---------------------|----------------------------|-------|-------|
|---------------------|----------------------------|-------|-------|
| Item | Proposed Device | Predicate Device | Substantial Equivalence |
|-------------------------|--------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------|
| | Disposable Ureteral Access Sheath | Well Lead Ureteral Access Sheath (K151084) | |
| Product Code | FED | FED | Same |
| Regulation No. | 876.1500 | 876.1500 | Same |
| Class | II | II | Same |
| Supplied Sterile | Yes | Yes | Same |
| Sheath ID | 10Fr, 12 Fr, 14Fr | 10Fr, 12 Fr, 14Fr | Same |
| Sheath Effective Length | 25cm, 35cm, 40cm, 45cm, 55cm, | 13cm, 20cm, 28cm, 35cm, 45cm, 55cm | Similar |
| Indications for Use | The Ureteral Access Sheath is intended to use in<br>urologic endoscopic procedures to facilitate the<br>passage of endoscopes. | The Well Lead Ureteral Access Sheath is used to<br>establish a conduit during endoscopic urological<br>procedures facilitating the passage of endoscopes and<br>other instruments into the urinary tract. | Similar |
| Configuration | Sheath, sheath hub, dilator, dilator hub, dilator clip,<br>hydrophilic coating | sheath, dilator, connector,<br>hydrophilic coating | Similar |
| Single Use | Yes | Yes | Same |
| Packaging | Single-use EO sterilized pouch with one device per<br>pouch | Single-use EO sterilized pouch with one device per pouch | Same |
| Shelf Life | Three years (36 months) | Three years (36 months) | Same |
# Table 1 Comparison of Technology Characteristics
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the logo for Beyomed. The logo consists of an orange abstract shape on the left, followed by the text "Beyomed" in blue. Above the word "Beyomed" are three Chinese characters, also in blue. The logo is simple and modern, with a focus on the company name.
| NO. | Characteristics | Specifications | Test method | Standard | Comparison<br>Result |
|-----|--------------------------------------|-----------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------|----------------------|
| 1 | Dimensions | Dimension of ureteral access sheath shall<br>comply with the requirements of Table 3<br>Dimensions. | Using a universal gage. | Internal<br>requirement | SE |
| 2 | Dilator & Sheath<br>Compatibility | The dilator should be able to enter and<br>withdraw smoothly and unobstructed<br>within the sheath. | Pull the dilator out of the sheath, then insert it into<br>the sheath cavity, repeat three times, the results<br>should meet the requirements. | Internal<br>requirement | SE |
| 3 | Hub security | Dilator hub can lock onto the sheath hub,<br>dilator can be fixed on the sheath hub, no<br>relative slip. | The dilator hub is screwed off from the sheath hub<br>and then inserted into the sheath in the same<br>direction. The results should meet the<br>requirements. | Internal<br>requirement | SE |
| 4 | Sheath<br>Inner<br>Lumen Passability | The dilator should be able to pass<br>through the sheath with a minimum<br>bending radius of 55 mm. | The sheath is bent under the condition that the<br>marking length is 55 mm, the dilator pass through<br>the sheath inner cavity, the results should meet the<br>requirements. | Internal<br>Requirement | SE |
| 5 | Dynamic Friction<br>Force | The surface dynamic friction of the<br>sheath after water contact should be<br><0.5N. | Soaking the samples with the distilled water, use<br>the Automatic friction tester, set the grip force<br>200g force, speed 200mm/min, distance<br>15mm.Back and forth 10 times, the results should<br>meet the requirements. | Internal<br>requirement<br>and clinical<br>requirement. | SE |
# Table 2 Comparison of Performance testing
{8}------------------------------------------------
| Image: Beyomed logo | 011_Annex 1 510(k) Summary | Page: | 011-6 |
|---------------------|----------------------------|-------|-------|
|---------------------|----------------------------|-------|-------|
| 6 | Dilator to Hub<br>Tensile Strength | The Dilator to Hub<br>Strength≥15N. | Cut out a 5cm sample from the dilator (with hub<br>part) ; place the sample on the clamp of the<br>tension machine and clamp it. Set the standard<br>distance as 20mm and the tensile speed as<br>400mm/min. Run the tension machine to test and<br>record the maximum force value when the<br>product breaks. | EN1618:1997 | SE |
|---|------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------|----|
| 7 | Sheath to Hub<br>Tensile Strength | The Sheath to Hub<br>Strength≥15N. | Cut out a 5cm sample from the sheath (with hub<br>part); place the sample on the clamp of the tension<br>machine and clamp it. Set the standard distance as<br>20mm and the tensile speed as 400mm/min. Run<br>the tension machine to test and record the<br>maximum force value when the product breaks. | EN1618:1997 | SE |
| 8 | Guidewire<br>compatibility | 0.038" (0.97mm) guidewire can<br>smoothly enter and exit the inner cavity<br>of the dilator. | Inject liquid and simulate the medical guidewire<br>lumen to verify that it can pass smoothly. | Internal<br>requirement | SE |
| 9 | Sheath Flexibility | Sheath &Dilator should be flexible,<br>when the tip bending 90°, the sheath and<br>dilator can recover, when bending 10<br>times with a minimum bending radius of<br>55 mm, the sheath and dilator can not<br>kink or broken. | Fix the sheath and dilator tip, rotate the hub 90°,<br>the sheath and dilator should not permanent<br>deformation, after the sheath and dilator recover,<br>bending 10 times with a minimum bending radius<br>of 55 mm, the results should meet the<br>requirements. | Internal<br>Requirement | SE |
{9}------------------------------------------------
| Image: Beyomed logo | 011_Annex 1 510(k) Summary | Page: | 011-7 |
|---------------------|----------------------------|-------|-------|
|---------------------|----------------------------|-------|-------|
| 10 | Sheath<br>Bending<br>Resistance | The bending force should more than<br>2.1N, less than 7.6N. | Cutting 25cm samples from the distal end of<br>sheath, set the sample into the tensile machine<br>fixture with vertical direction, set the speed<br>10mm/min, fix the samples, compressed down<br>3cm. The results should meet the requirements | Internal<br>Requirement | SE |
|----|---------------------------------|-------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------|----|
|----|---------------------------------|-------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------|----|
{10}------------------------------------------------
#### Substantially Equivalent (SE) Conclusion 8.
Based on the comparison and analysis from the nonclinical tests, the proposed devices is safe and effective, the performance is as well as the predicate devices. It is determined to be Substantially Equivalent (SE) to the predicate devices.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.